LIFECHOICE OXYGEN CONCENTRATOR

K113317 · Inova Labs, Inc. · CAW · Dec 28, 2011 · Anesthesiology

Device Facts

Record IDK113317
Device NameLIFECHOICE OXYGEN CONCENTRATOR
ApplicantInova Labs, Inc.
Product CodeCAW · Anesthesiology
Decision DateDec 28, 2011
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 868.5440
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The LifeChoice Oxygen Concentrator is used on a prescriptive basis by patients who are diagnosed as requiring supplemental oxygen concentrator will provide supplemental, high concentration oxygen to these patients. It is not life supporting nor life sustaining. It may be used continuously in a home, institution or travel environment. The LifeChoice is also portable.

Device Story

LifeChoice Oxygen Concentrator (Model OXY1000) is a portable device providing supplemental oxygen to patients. It utilizes molecular sieve pressure swing adsorption technology to concentrate room air into ~90% oxygen. The device operates on a demand flow basis, delivering oxygen at rates from 1.0 to 5.0 liters per minute in 1.0 L/min increments. It is intended for use in home, institutional, or travel settings. The device is powered by AC, DC, or rechargeable batteries. It is a prescription-only device; it is not life-supporting or life-sustaining. Healthcare providers prescribe the device to manage oxygen therapy needs, and patients use it to receive supplemental oxygen in various environments.

Clinical Evidence

Bench testing only.

Technological Characteristics

Molecular sieve pressure swing adsorption technology; demand flow delivery system; portable form factor; power sources include AC, DC, and rechargeable batteries; delivers ~90% oxygen concentration.

Indications for Use

Indicated for patients diagnosed as requiring supplemental oxygen. Not life-supporting or life-sustaining. Suitable for continuous use in home, institution, or travel environments.

Regulatory Classification

Identification

A portable oxygen generator is a device that is intended to release oxygen for respiratory therapy by means of either a chemical reaction or physical means (e.g., a molecular sieve).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for LifeChoice by InovaLabs. The logo is black and white and features the words "LifeChoice" in a bold, sans-serif font. Below the word "LifeChoice" is the phrase "by InovaLabs" in a smaller font. To the right of the logo is a black and white image of a medical device. K113317 DEC 2 8 2011 Special 510(k) Summary (reference 510(k) #K072688) Non-confidential Summary of Safety and Effectiveness Submitter Information: Inova Labs, Inc. 3500 Comsouth Dr. Suite 100 Austin, TX 78744 Contact: H. David Shockley V.P. 512-617-1700 off. 512-617-1692 fax Date: Submitted: November, 2011 Device Classification Name: Generator, Oxygen, Portable Product Code: CAW Classification: Class II Regulation No .: 21 CFR 868.5440 Panel: Anesthesiology Specific Device Name: LifeChoice Oxygen Concentrator Model: OXY1000 Description of the Device: The LifeChoice Oxygen Concentrator is a prescription use device for patients needing supplemental high concentration oxygen. The LifeChoice is not intended to be life supporting. The device provides approximately 90% oxygen to patients on a demand flow basis at an equivalent rate of 1.0 liter per minute to 5.0 liters per minute in increments of 1.0 liter per minute. The LifeChoice is a portable device which can be used in a home, institution or travel environment. The device uses molecular sieve pressure swing adsorption technology. This is a proven technology to concentrate room air into high concentration oxygen. Indications for Use: The LifeChoice Oxygen Concentrator is used on a prescriptive basis by patients who are diagnosed as requiring supplemental oxygen concentrator will provide supplemental, high concentration oxygen to these patients. It is not life supporting nor life sustaining. It may be used continuously in a home, institution or travel environment. The LifeChoice is also portable. Technological Characteristics: The LifeChoice Oxygen Concentrator utilizes well established technologies. Molecular sieve pressure swing adsorption technology has been used for many years to produce high concentration oxygen. Also, demand flow delivery systems have been in use on portable oxygen devices for many years. The capability to power the device with AC, DC and/or rechargeable batteries has also been in use for many years. The technologies used in the modifications to the LifeChoice Oxygen Concentrator do not create any new questions of safety and effectiveness. Benchtop performance testing and comparison of performance characteristics have demonstrated that the modified LifeChoice is substantially equivalent to the previously marketed version. Inova Labs Inc. | 3500 Comsouth Road, Suite 100 | Austin, Texas 78744 P 512-326-3237 | F 512-326-3299 LifeChoiceOxvgen.com {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The eagle is facing to the the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle. ## Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 DEC 2 8 2011 Mr. H. David Shockley Vice President Inova Labs, Incorporated 3500 Comsouth Drive, Suite 100 Austin, Texas 78744 Re: K113317 Trade/Device Name: LifeChoice Concentrator, Model OXY 1000 Regulation Number: 21 CFR 868.5440 Regulation Name: Portable Oxygen Generator Regulatory Class: II Product Code: CAW Dated: December 16, 2011 Received: December 20, 2011 Dear Mr. Shockley: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ Page 2 - Mr. Shockley Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, h for Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health {3}------------------------------------------------ ## Indications for Use 510(k) Number (if known): ____K113317_________________________________________________________________________________________________________________________________________ Device Name: Inova Labs LifeChoice Oxygen Concentrator Indications for Use: The LifeChoice Oxygen Concentrator is used on a prescriptive basis by patients who are diagnosed as requiring supplemental oxygen concentrator will provide supplemental, high concentration oxygen to these patients. It is not life supporting nor life sustaining. It may be used continuously in a home, institution or travel environment. The LifeChoice is also portable. X Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) | Concurrence of CDRH, Office of Device Evaluation (ODE) | | | |--------------------------------------------------------|--|--| | | | | | | | | (Division Sign=Off) (Division Sign-On) Division of Anesthesiology, General Hospital Production of Anesthesion And Bayanas Division of Antrol, Dental Devices | 510(k) Number: | <i>K113317</i> | |----------------|----------------| |----------------|----------------| Page 1 of 1
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...