DEVILBISS NEBULIZER MODEL 800

K993492 · Sunrise Medical Hhg, Inc. · CAF · Apr 21, 2000 · Anesthesiology

Device Facts

Record IDK993492
Device NameDEVILBISS NEBULIZER MODEL 800
ApplicantSunrise Medical Hhg, Inc.
Product CodeCAF · Anesthesiology
Decision DateApr 21, 2000
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5630
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The Sunrise Medical Model 800 Series iet nebulizer is designed for use in conjunction with a compressed gas source such as the Sunrise Medical Pulmo-Aide Model 5650 Compressor (FDA 510(k) Registration # K923888), Sunrise Medical Pulmo-Mate Model 4650 Compressor (FDA 510(k) Registration # K931015) or Sunrise Medical Pulmo-Aide LT Model 3650 Compressor (FDA 510(k) Registration # K970289). The system is used on the order of a physician to treat various respiratory problems. The nebulizer converts the liquid medication into a fine aerosol mist, which can be inhaled for home respiratory therapy. This allows the medication to be deposited at the affected areas of the patient's airway. The Sunrise Medical Model 800 Series jet nebulizer is designed to offer high respirable aerosol output for the patient. The nebulizer is intended as a long term usage nebulizer (approximately 1 year) similar to other reusable nebulizers that are currently being marketed by competitors of DeVilbiss such as the Pari LC Plus and Pari Jet permanent nebulizers manufactured by Pari of Germany and distributed here in the U.S. The target patient population for this device is comprised of both children and adults suffering from asthma, bronchitis, emphysema, and other respiratory disorders. The device is to be used by the patient only on the order of a physician in conjunction with prescribed medication(s). The intended environment of use is in the home with AC powered compressors or for portable use with DC/battery powered compressors.

Device Story

Reusable small volume jet nebulizer; operates via connection to external compressed gas source (e.g., Pulmo-Aide/Pulmo-Mate compressors). Converts liquid medication into fine aerosol mist for inhalation; deposits medication into patient airways. Includes optional triggering system to conserve medication between breaths. Designed for home use by patients; long-term usage (approx. 1 year). Physician prescribes medication and therapy; patient self-administers. Benefits include targeted delivery of respiratory medication to affected airway areas.

Clinical Evidence

Bench testing only. Evaluated aerosol output and particle size to confirm performance equivalence to predicate devices.

Technological Characteristics

Reusable small volume jet nebulizer; pneumatic operation via external compressed gas source. Includes optional airflow triggering system. Designed for approximately 1-year lifespan. No software or electronic components.

Indications for Use

Indicated for children and adults with asthma, bronchitis, emphysema, and other respiratory disorders requiring home aerosolized medication therapy as prescribed by a physician.

Regulatory Classification

Identification

A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ KG93492 ## 510(k) Summary | Submitter's Name: | Sunrise Medical HHG, Inc.<br>100 DeVilbiss Drive<br>Somerset PA 15501<br>Frank Clementi<br>814-443-7474 | |---------------------------------------------------------|---------------------------------------------------------------------------------------------------------| | Date Prepared: | October 7, 1999 | | Device Name: | Nebulizer FDA Classification CAF | | Common or Usual Name: | Jet Nebulizer | | DeVilbiss Model Number: | Nebulizer Model 800 | | Trade Proprietary Name: | N/A Model number only. | | Established Registration<br>Number: | 2515872 | | FDA Classification: | Class II | | Equivalent Legally Marketed Predicate Devices: | | | Legally Marketed Predicate Device 510(k) Registration # | | | Sunrise Medical Model 700 | K952249 | | Pari LC Plus | K894555 | Pari PariJet ## Description of Device: The Sunrise Medical Model 800 Series jet nebulizer is a reusable hand held, small volume jet nebulizer. The nebulizer is designed for use in conjunction with a compressed gas source such as the Sunrise Medical Pulmo-Aide Model 5650 Compressor (FDA 510(k) Registration # K923888), Sunrise Medical Pulmo-Mate Model 4650 Compressor (FDA 510(k) Registration # K931015) or Sunrise Medical Pulmo-Aide LT Model 3650 Compressor (FDA 510(k) Registration # K970289). The nebulizer & compressor system is used on the order of a physician to treat various respiratory problems. The nebulizer converts the liquid medication into a fine aerosol mist, which can be inhaled for home respiratory therapy. This allows the medication to be deposited at the affected areas of the patient's airway. K896984 {1}------------------------------------------------ The Sunrise Medical Model 800 Series iet nebulizer is designed to offer high respirable aerosol output for the patient. The nebulizer is intended as a long term usage nebulizer (approximately 1 year) similar to other reusable nebulizers that are currently being marketed by competitors of Sunrise Medical such as the Pari LC Plus and Pari Jet permanent nebulizers manufactured by Pari of Germany and distributed here in the U.S. Accessories include a triggering system to shut off airflow to the nebulizer to conserve medication between breaths. ## Statement of Intended Use: The Sunrise Medical Model 800 Series iet nebulizer is designed for use in conjunction with a compressed gas source such as the Sunrise Medical Pulmo-Aide Model 5650 Compressor (FDA 510(k) Registration # K923888), Sunrise Medical Pulmo-Mate Model 4650 Compressor (FDA 510(k) Registration # K931015) or Sunrise Medical Pulmo-Aide LT Model 3650 Compressor (FDA 510(k) Registration # K970289). The system is used on the order of a physician to treat various respiratory problems. The nebulizer converts the liquid medication into a fine aerosol mist, which can be inhaled for home respiratory therapy. This allows the medication to be deposited at the affected areas of the patient's airway. The Sunrise Medical Model 800 Series jet nebulizer is designed to offer high respirable aerosol output for the patient. The nebulizer is intended as a long term usage nebulizer (approximately 1 year) similar to other reusable nebulizers that are currently being marketed by competitors of DeVilbiss such as the Pari LC Plus and Pari Jet permanent nebulizers manufactured by Pari of Germany and distributed here in the U.S. The target patient population for this device is comprised of both children and adults suffering from asthma, bronchitis, emphysema, and other respiratory disorders. The device is to be used by the patient only on the order of a physician in conjunction with prescribed medication(s). The intended environment of use is in the home with AC powered compressors or for portable use with DC/battery powered compressors. ## Technological Characteristics: The Sunrise Medical Model 800 Nebulizer is equivalent in functional characteristics to the existing legally marketed predicate devices. The devices all utilize a compressed air source for operation. Testing performed on the aerosol output and particle size show that the new Sunrise Medical Model 800 Nebulizer is substantially equivalent to the existing legally marketed predicate devices and that all of these devices will produce a similar aerosol treatment. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an emblem featuring a stylized caduceus, a symbol often associated with medicine and healthcare. Public Health Service APR 2 1 2000 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Frank Clementi Sunrise Medical HHG Inc. Respiratory Products Division Dba DeVilbiss Health Care 100 DeVilbiss Drive P.O. Box 635 Somerset, PA 15501-0635 K993492 Re: Sunrise Medical Model 800 Nebulizer Regulatory Class: II (two) Product Code: CAF Dated: January 21, 2000 Received: January 24, 2000 Dear Mr. Clementi: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such Failure to comply with the GMP regulation may result in assumptions. regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act {3}------------------------------------------------ Page 2 - Mr. Frank Clementi for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Joanna Westerhausen for James E. Dillard III Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ | Page | 1 of 1 | |------|--------| |------|--------| 510(k) Number: (if known): Sunrise Medical Model 800 Nebulizer Device Name: Indications For Use: The system is used on the order of a physician for the treatment of asthma, bronchitis and other respiratory diseases. The nebulizer converts the liquid medication into a fine aerosol mist which can be inhaled for home respiratory therapy. This allows the medication to be deposited at the affected areas of the patients airway. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurence of CDRH, Office of Device Evaulation (ODE) f. Allertson
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...