SALTER LABS 8990 SERIES NEBULIZER

K962085 · Salter Labs · CAF · Aug 26, 1996 · Anesthesiology

Device Facts

Record IDK962085
Device NameSALTER LABS 8990 SERIES NEBULIZER
ApplicantSalter Labs
Product CodeCAF · Anesthesiology
Decision DateAug 26, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5630
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The nebulizer is a disposable device which is used where liquids are to be delivered to a patient in an aerosol form. The device is not life-supporting or life-sustaining, nor is it implanted. It does not use software, nor is it sterile. The device is for single patient use. It is intended to be used with accessories, such as tubing which are industry standard devices used with nebulizers.

Device Story

Disposable pneumatic nebulizer; consists of nebulizer cup and top. Input: pressurized nebulizing gas and liquid medication. Operation: gas flow nebulizes liquid into aerosol for patient inhalation. Output: aerosolized medication. Used in clinical or home settings; operated by patient or caregiver. Larger internal reservoir design eliminates need for external tee adapter and reservoir tube compared to predicate. Non-sterile; single-patient use.

Clinical Evidence

No clinical data; bench testing only.

Technological Characteristics

Pneumatic-powered disposable nebulizer. Components: nebulizer cup and top. Design features larger internal reservoir. Non-sterile. No software, electronics, or energy source other than pneumatic gas supply.

Indications for Use

Indicated for patients requiring delivery of liquid medication in aerosol form via inhalation.

Regulatory Classification

Identification

A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.

Predicate Devices

Submission Summary (Full Text)

{0} K962085 AUG 26 1996 # 15. 510(k) Summary ## 15.1 Purpose: This summary of 510(k) Safety and Effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. ## 15.2 Submitter Identification: Salter Labs 100 W. Sycamore Arvin, CA 93203 Telephone: (805) 854-3166 Contact Person: James N. Curti ## 15.3 Date of Summary Preparation: April 22, 1996 ## 15.4 Name of Device: ### 15.4.1 Classification Name: Nebulizer (direct patient interface) ### 15.4.2 Trade Name: Nebulizer ### 15.4.3 Proprietary Name: Salter Labs Ultramist Nebulizer, 8990 Series ## 15.5 Identification of Predicate Device: Salter Labs Hand Held Pneumatic Powered (disposable) Nebulizer Model 8900, which was cleared for marketing by 510(k) No, K870027. ## 15.6 Description of the Device: Each of the Salter Labs 8990 Series Nebulizers consists of a Nebulizer Top which is screwed onto a Nebulizer Cup. The bottom of the cup has a fitting to accept a source of nebulizing gas. Liquids to be nebulized are placed in the nebulizer cup. The top of the nebulizer housing is then Salter Labs JNC 96032001 100 W. SYCAMORE ROAD • ARVIN, CALIFORNIA 93203-2300 PHONE: 805-854-3166 FAX: 805-854-3850 {1} applied, and the supply line for the nebulizing gas source is then connected to the bottom of the nebulizer cup. After the pressurized nebulizing gas is applied to the nebulizer, the solution in the cup is nebulized and pushed from the unit in aerosol form that can be inhaled by the patient. ## 15.7 Intended Use: The Model 8990 Series Nebulizers have the same intended use as their predicate device, The Salter Model 8900 Nebulizer, which was cleared for marketing by 510(k) No. K870027. The nebulizer is a disposable device which is used where liquids are to be delivered to a patient in an aerosol form. The device is not life-supporting or life-sustaining, nor is it implanted. It does not use software, nor is it sterile. The device is for single patient use. It is intended to be used with accessories, such as tubing which are industry standard devices used with nebulizers. ## 15.8 Difference(s) Between This Device and its Predicate Device: The Model 8990 Series Nebulizers are the same as their predicate device except each includes a larger internal reservoir for the aerosol, which eliminates the need for the external tee adapter and reservoir tube This difference has no significant effect on the safety or effectiveness of the device. ## 15.9 Signature of contact person: ![img-0.jpeg](img-0.jpeg) End of 510(k) Summary Salter Labs JNC 96032001 100 W. SYCAMORE ROAD · ARVIN, CALIFORNIA 93203-2300 PHONE: 805-854-3166 FAX: 805-854-3850
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...