SALTERLABS 8960 SERIES NEBULIZER (8960/8961/8966/8984)
Applicant
Salter Labs
Product Code
CAF · Anesthesiology
Decision Date
Jul 16, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The nebulizer is a disposable device which is used where liquids are to be delivered to a patient in an aerosol form. The device is not life-supporting or life-sustaining, nor is it implanted. It does not use software, nor is it sterile. The device is for single patient use. It is intended to be used with accessories, such as masks and tubing which are industry standard devices used with nebulizers.
Device Story
Pneumatic-powered disposable nebulizer; consists of nebulizer top and cup. Input: pressurized nebulizing gas and liquid medication. Operation: gas flow nebulizes liquid into aerosol for patient inhalation. Features internal aerosol reservoir, inspiration valve in top, and expiration valve in mouthpiece. Used in clinical or home settings; operated by patient or caregiver. Output: aerosolized medication. Benefits: facilitates respiratory drug delivery.
Indicated for patients requiring delivery of liquid medication in aerosol form via inhalation.
Regulatory Classification
Identification
A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.
Predicate Devices
Salter Labs Hand Held Pneumatic Powered (disposable) Nebulizer Model 8900 (K870027)
Related Devices
K962879 — SALTERLABS 8660 SERIES NEBULIZER · Salter Labs · Oct 22, 1996
K962085 — SALTER LABS 8990 SERIES NEBULIZER · Salter Labs · Aug 26, 1996
K041327 — MED2000 SPA NEBULIZER ANDYFLOW, MODEL A1/C WITH ACCESSORIES · Med2000 S.R.L. · Jul 20, 2004
Submission Summary (Full Text)
{0}
K961476
JUL 16 1996
# 15. 510(k) Summary
## 15.1 Purpose:
This summary of 510(k) Safety and Effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
## 15.2 Submitter Identification:
Salter Labs
100 W. Sycamore
Arvin, CA 93203
Telephone: (805) 854-3166
Contact Person: James N. Curti
## 15.3 Date of Summary Preparation:
April 15, 1996
## 15.4 Name of Device:
### 15.4.1 Classification Name:
Nebulizer (direct patient interface)
### 15.4.2 Trade Name:
Nebulizer
### 15.4.3 Proprietary Name:
Salter Labs Ultramist Nebulizer, 8960 Series
## 15.5 Identification of Predicate Device:
Salter Labs Hand Held Pneumatic Powered (disposable) Nebulizer Model 8900, which was cleared for marketing by 510(k) No, K870027
## 15.6 Description of the Device:
Each of the Salter Labs 8960 Series Nebulizers consists of a Nebulizer Top which is screwed onto a Nebulizer Cup. The bottom of the cup has a fitting to accept a source of nebulizing gas.
Liquids to be nebulized are placed in the nebulizer cup. The top of the nebulizer housing is then applied, and the supply line for the nebulizing gas source is then connected to the bottom of the nebulizer cup.
Salter Labs JNC 96022901
100 W. SYCAMORE ROAD · ARVIN, CALIFORNIA 93203-2300
PHONE: 805-854-3166
FAX: 805-854-3850
{1}
After the pressurized nebulizing gas is applied to the nebulizer, the solution in the cup is nebulized and pushed from the unit in aerosol form that can be inhaled by the patient.
## 15.7 Intended Use:
The Model 8960 Series Nebulizers have the same intended use as their predicate device, The Salter Model 8900 Nebulizer, which was cleared for marketing by 510(k) No. K870027. The nebulizer is a disposable device which is used where liquids are to be delivered to a patient in an aerosol form. The device is not life-supporting or life-sustaining, nor is it implanted. It does not use software, nor is it sterile. The device is for single patient use. It is intended to be used with accessories, such as masks and tubing which are industry standard devices used with nebulizers.
## 15.8 Difference(s) Between This Device and its Predicate Device:
The Model 8960 Series Nebulizers are the same as their predicate device except:
Each includes an internal reservoir for the aerosol, which eliminates the need for the external tee adapter and reservoir tube
Each includes an inspiration valve in the nebulizer top.
The Model 8960 includes an expiration valve in its mouthpiece.
None of these differences has a significant effect on the safety or effectiveness of the device.
## 15.9 Signature of contact person:

James N. Curti
End of 510(k) Summary
Salter Labs JNC 96022901
100 W. SYCAMORE ROAD • ARVIN, CALIFORNIA 93203-2300
PHONE: 805-854-3166
FAX: 805-854-3850
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.