Hudson RCI Variable concentration Large Volume Nebulizer (1770)
K220955 · Medline Industires, Inc. · CAF · Jun 23, 2023 · Anesthesiology
Device Facts
Record ID
K220955
Device Name
Hudson RCI Variable concentration Large Volume Nebulizer (1770)
Applicant
Medline Industires, Inc.
Product Code
CAF · Anesthesiology
Decision Date
Jun 23, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5630
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The non-prefilled nebulizer is indicated for use when humidity needs to the air flow going to a patient in the form of an aerosol. The large volume nebulizer also allows for the provision of humidified oxygen at select oxygen concentrations from 28% to 98%. This product is for single patient use and is designed to be used in hospital, nursing homes, extended care facilities, outpatient clinics as prescribed by a healthcare professional.
Device Story
Non-prefilled reservoir nebulizer; supplies humidity for inhalation therapy. Input: medical gas source (flow-metered). Operation: internal venturi nozzle draws solution from 500mL jar via pickup tube into gas stream; rotating collar adjusts air entrainment for oxygen concentration (28-98%); aerosol delivered via 22mm tubing. Used in hospitals, nursing homes, extended care facilities, outpatient clinics; operated by healthcare professionals. Output: humidified aerosol mist. Benefit: provides humidity and oxygen to patient airways; supports respiratory therapy.
Clinical Evidence
Bench testing only. Performance evaluation included biocompatibility (ISO 10993-1, -5, -10, -11, -18; ISO 18562-1, -2, -3), packaging, environmental conditioning, aging, oxygen entrainment, lift testing, humidity output, useful life, and cleaning process validation.
Technological Characteristics
Jet nebulizer; 500mL reservoir; polypropylene, polyethylene, LDPE, polystyrene, acrylic construction. Adjustable venturi nozzle for 28-98% FIO2. 22mm bull-nose output connector. Non-sterile. Standards: ISO 10993 (biocompatibility), ISO 18562 (gas pathway).
Indications for Use
Indicated for pediatric (ages 2+) and adult patients requiring humidified air or oxygen (28-98% concentration) via aerosol therapy in clinical or home settings.
Regulatory Classification
Identification
A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.
Predicate Devices
AirLife Misty Finity Large Volume Continuous Nebulizer (K041418)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 23, 2023
Medline Industires, Inc. % Joy Gutermuth Senior Specialist (Consultant) Ram+ 2790 Mosside Blvd. Suite 800 Monroeville, Pennsylvania 15146
Re: K220955
Trade/Device Name: Hudson RCI Variable concentration Large Volume Nebulizer (1770) Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: Class II Product Code: CAF Dated: May 26, 2023 Received: May 26, 2023
Dear Joy Gutermuth:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Ting Song -S
# For
Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220955
Device Name Hudson RCI Large Volume Nebulizer
#### Indications for Use (Describe)
The non-prefilled nebulizer is indicated for use when humidity needs to the air flow going to a patient in the form of an aerosol. The large volume nebulizer also allows for the provision of humidified oxygen at select oxygen concentrations from 28% to 98%. This product is for single patient use and is designed to be used in hospital, nursing homes, extended care facilities, outpatient clinics as prescribed by a healthcare professional.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|-----------------------------------------------|
| <div> </div> | |
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
## DATE PREPARED
June 23, 2023
# MANUFACTURER AND 510(k) OWNER
Medline Industries, Inc. 1 Three Lake Dr. Northfield, IL 60093 Phone: 724-640-9680
Contact Person: Nicole Schaffer, Manager Regulatory Affairs, Respiratory
# REPRESENTATIVE/CONSULTANT
Joy Gutermuth, Regulatory Consultant RQM+ Telephone: +1 (412) 810-8172 Email: jgutermuth@rqmplus.com
## DEVICE INFORMATION
| Proprietary Name/Trade Name: | Hudson RCI Variable Concentration Large Volume<br>Nebulizer |
|------------------------------|-------------------------------------------------------------|
| Common Name: | Large Volume Nebulizer |
| Regulation Number: | §868.5630 |
| Class: | II |
| Product Code: | CAF |
# PREDICATE DEVICE IDENTIFICATION
The Hudson RCI Variable Concentration Large Volume Nebulizer is substantially equivalent to the following predicates:
| 510(k) Number | Predicate Device Name / Manufacturer | Primary<br>Predicate |
|---------------|-----------------------------------------------------------|----------------------|
| K041418 | AirLife Misty Finity Large Volume Continuous<br>Neublizer | ✓ |
| K141214 | Hudson RCI® AquaPak® Prefilled Nebulizer | Reference Device |
The predicate devices have not been subject to a design related recall.
{4}------------------------------------------------
# DEVICE DESCRIPTION
The large volume nebulizer is a non-prefilled reservoir nebulizer for supplying humidity for inhalation therapy. The device features a wing nut style connector that fits standard flow metered medical gas sources and includes a 500mL capacity jar with minimum and maximum fill lines. Large volume nebulize an internal venturi nozzle to draw the solution up from the jar through a small plastic pickup tube and into the gas stream to be aerosolized. A rotating collar sets the delivered oxygen concentration by controlling the size of the room air opening around the venturi. A bull-nose style output connector is used to connect 22mm aerosol tubing for delivery to the patient.
## INDICATIONS FOR USE
The non-prefilled nebulizer is indicated for use when humidity needs to the air flow going to a patient in the form of an aerosol. The large volume nebulizer also allows for the provision of humidified oxygen at select oxygen concentrations from 28% to 98%. This product is for single patient use and is designed to be used in hospital, nursing homes, extended care facilities, outpatient clinics as prescribed by a healthcare professional.
## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
Medline Industries believes that the large volume nebulizer is substantially equivalent to the predicate devices based on the information summarized here:
The subject device has a similar design and dimensions, and uses similar or identical materials as the devices cleared in K041418 and K141214. The subject device has the same intended use and similar technological characteristics to the devices cleared in both K041418 and K141214. The device has similar instrumentation to the device cleared in K041418 and K141214. These technological characteristics have undergone testing to ensure the device is as safe and effective as the predicates.
| Product Features | Proposed<br>Hudson RCI<br>Variable<br>Concentration<br>Large Volume<br>Nebulizer | Predicate<br>AirLife Misty Finity<br>Large Volume<br>Continuous<br>Nebulizer<br>(K041418) | Reference Device<br>Hudson RCI®<br>AquaPak® Prefilled<br>Nebulizer (K141214) |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification | Class II | Class II | Class II |
| Product Code | CAF | CAF | CAF |
| Regulation<br>Number | §868.5630 | §868.5630 | §868.5630 |
| Regulation Name | Nebulizer | Nebulizer | Nebulizer |
| Product Features | Proposed<br>Hudson RCI<br>Variable<br>Concentration<br>Large Volume<br>Nebulizer | Predicate<br>AirLife Misty Finity<br>Large Volume<br>Continuous<br>Nebulizer<br>(K041418) | Reference Device<br>Hudson RCI®<br>AquaPak® Prefilled<br>Nebulizer (K141214) |
| Intended Use | To add humidity in aerosol form to a patient's breathing gases. | This device is intended to be used to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. | To add humidity in aerosol form to a patient's breathing gases. |
| Indications for Use | The non-prefilled nebulizer is indicated for use when humidity needs to be added to the air flow going to a patient in the form of an aerosol. The large volume nebulizer also allows for the provision of humidified oxygen at select oxygen concentrations from 28% to 98%. This product is for single patient use and is designed to be used in hospital, nursing homes, extended care facilities, outpatient clinics as prescribed by a healthcare professional. | This device is intended to be used to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Its use is indicated whenever a physician or healthcare professional administers or prescribes medical aerosol products to a patient using a Continuous Large Volume Nebulizer. This product is a single patient use, non-sterile prescriptive device and is designed to be used under medical supervision in hospitals, nursing homes, extended care facilities or outpatient clinics. | The Hudson RCI® AquaPak® Prefilled Nebulizer adds sterile water or saline solution in aerosol form to a patient's breathing gases. |
| Product Features | Proposed<br>Hudson RCI<br>Variable<br>Concentration<br>Large Volume<br>Nebulizer | Predicate<br>AirLife Misty Finity<br>Large Volume<br>Continuous<br>Nebulizer<br>(K041418) | Reference Device<br>Hudson RCI®<br>AquaPak® Prefilled<br>Nebulizer (K141214) |
| Representative<br>Image | Image: Proposed Hudson RCI Variable Concentration Large Volume Nebulizer | Image: Predicate AirLife Misty Finity Large Volume Continuous Nebulizer (K041418) | Image: Reference Device Hudson RCI® AquaPak® Prefilled Nebulizer (K141214) |
| Product Features | Proposed<br>Hudson RCI<br>Variable<br>Concentration<br>Large Volume<br>Nebulizer | Predicate<br>AirLife Misty Finity<br>Large Volume<br>Continuous<br>Nebulizer<br>(K041418) | Reference Device<br>Hudson RCI®<br>AquaPak® Prefilled<br>Nebulizer (K141214) |
| Device | The large volume | The nebulizer is a | The Hudson RCI® |
| Description | nebulizer is a non-<br>prefilled reservoir | single patient use | AquaPak® Prefilled |
| | nebulizer for | device, which is filled<br>with a fluid, typically | Nebulizers provide<br>sterile water or sterile |
| | supplying humidity | respiratory | saline for inhalation |
| | for inhalation | medication and | therapy. Nebulizers |
| | therapy. The device | connected to an air | generate aerosol, a fine |
| | features a wing nut | source via flexible | mist of liquid water (or |
| | style connector that | tubing. The nebulizer | sodium chloride |
| | fits standard flow<br>metered medical gas | works by having the<br>fluid come into | solution) that is<br>suspended in the gas to |
| | sources and includes | contact with the | be inhaled by the |
| | a 500mL capacity jar | steam of gas. The gas | patient. The Hudson |
| | with minimum and | shatters the liquid | RCI® AquaPak® |
| | maximum fill lines. | into small particles. | Prefilled Nebulizers are |
| | Large volume | These particles then | designed to aerosolize |
| | nebulizers utilize an | impact a baffle that | and provide a fine mist |
| | internal venturi | further reduces the | of sterile water or |
| | nozzle to draw the | size of the particles. | saline solution to |
| | solution up from the<br>jar through a small | The majority of the<br>larger particles settle | inspired gas during<br>aerosol therapy. |
| | plastic pickup tube | inside the nebulizer | Prefilled sterile |
| | and into the gas | as a result of gravity | reservoirs for |
| | stream to be | and inertia, returning | AquaPak® Nebulizers |
| | aerosolized. A | the mist to liquid to | come in three sizes: |
| | rotating collar sets | repeat the | 440ml, 760ml, and |
| | the delivered oxygen | nebulization process. | 1070ml. Each reservoir |
| | concentration by | The smaller particles | must be used with a |
| | controlling the size | are then administered | suitable adaptor |
| | of the room air<br>opening around the | as the patient inhales.<br>The treatment is | component, which<br>connects the system to |
| | venturi. A bull-nose | completed when the | a flow-metered gas |
| | style output | majority of fluid is | source and provides |
| | connector is used to | nebulized. | nebulizer functionality. |
| | connect 22mm | | |
| | aerosol tubing for | | |
| | delivery to the | | |
| | patient. | | |
| Product Features | Proposed<br>Hudson RCI<br>Variable<br>Concentration<br>Large Volume<br>Nebulizer | Predicate<br>AirLife Misty Finity<br>Large Volume<br>Continuous<br>Nebulizer<br>(K041418) | Reference Device<br>Hudson RCI®<br>AquaPak® Prefilled<br>Nebulizer (K141214) |
| Accessories | Masks (not<br>included), tubing<br>(not included) | Masks (not included),<br>tubing (not included) | Nebulizer Adaptors,<br>masks (no included),<br>tubing (not included) |
| Environment | Hospital or home<br>care setting. | Hospitals, nursing<br>homes, extended care<br>facilities or outpatient<br>clinics. | Hospital or home care<br>setting. |
| Patient<br>Population | Pediatric (ages 2<br>years and above) and<br>adults. | Infant, pediatric, and<br>adult | Pediatric (ages 2<br>years and above) and<br>adults. |
| Principle of<br>Operation | Jet nebulizer with<br>adjustable air<br>entrainment | Jet nebulizer with<br>adjustable air<br>entrainment | Jet nebulizer with<br>adjustable air<br>entrainment |
| Sterilization<br>Method | Non-sterile | Non-sterile | Sterile |
| Patient<br>Contacting<br>Materials | Polypropylene,<br>polyethylene, LDPE,<br>polystyrene, acrylic,<br>dye | Thermoplastics | Thermoplastics |
| Single Use | Single patient use,<br>disposable | Single patient use,<br>disposable | Single patient use,<br>disposable |
| Adjustable<br>Oxygen Settings | FIO2 28-98% | FIO2 28-98% | FIO2 28-98% |
| Biocompatibility | ● External<br>communicating,<br>prolonged/perman<br>ent contact device<br>that indirectly<br>contacts<br>tissue/bone/dentin<br>● Indirect gas<br>pathway | ● Unknown from<br>publicly available<br>summary what<br>biocompatibility<br>endpoints were<br>assessed. | ● External<br>communicating,<br>prolonged/permanent<br>contact device that<br>indirectly contacts<br>tissue/bone/dentin.<br>● Indirect gas pathway |
| Product Features | Proposed<br>Hudson RCI<br>Variable<br>Concentration<br>Large Volume<br>Nebulizer | Predicate<br>AirLife Misty Finity<br>Large Volume<br>Continuous<br>Nebulizer<br>(K041418) | Reference Device<br>Hudson RCI®<br>AquaPak® Prefilled<br>Nebulizer (K141214) |
| Standards<br>Utilized | ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10<br>ISO 10993-11<br>ISO 10993-18<br>ISO 18562-1<br>ISO 18562-2<br>ISO 18562-3 | ISO 10993-1 | ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10<br>ISO 10993-11<br>ISO 10993-18 |
| Non-Clinical<br>Testing | • Packaging<br>• Environmental<br>Conditioning<br>(high and low<br>humidity)<br>• Aging<br>• Oxygen<br>entrainment<br>• Lift testing<br>• Humidity output<br>• Useful life testing<br>• Cleaning process | "Performance<br>evaluation of the<br>proposed and<br>predicated devices<br>consisted of cascade<br>impaction and output<br>rate testing." | • Packaging<br>• Environmental<br>Conditioning (high<br>and low humidity)<br>• Aging |
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
# SUMMARY OF NON-CLINICAL TESTING
No FDA performance standards have been established for the Hudson RCI Variable Concentration Large Volume Nebulizer. The following tests were performed to demonstrate safety based on current industry standards:
- Biocompatibility
- . Packaging
- Environmental Conditioning (high and low humidity) ●
- Aging
- Oxygen entrainment ●
- Lift testing .
- Humidity output ●
- Useful life testing ●
- Cleaning process .
{10}------------------------------------------------
The results of these tests indicate that the Hudson RCI Variable Concentration Large Volume Nebulizer is substantially equivalent to the predicate devices.
## CONCLUSION
Based on the testing performed, including additional gas pathway biocompatibility according to ISO 18562, it can be concluded that the subject device does not raise…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.