← Product Code [CAF](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/CAF) · K203400

# Corrugated tube with mouthpiece accessory (K203400)

_Trudell Medical International · CAF · Jun 14, 2021 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/CAF/K203400

## Device Facts

- **Applicant:** Trudell Medical International
- **Product Code:** [CAF](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/CAF.md)
- **Decision Date:** Jun 14, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5630
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The corrugated tube is an accessory to the MC 300 ® Nebulizer. It is designed to work in conjunction with the MC 300 ® Nebulizer to extend the patient interface away from the nebulizer. The accessory is a single patient is intended to be used with either mouthpiece or mask for pediatric (ages 2 years and above) and adult patients, who are under the care of a licensed healthcare provider or physician.

## Device Story

Corrugated tube accessory extends patient interface away from MC 300 Nebulizer; used with mouthpiece or mask. Device is disposable, single-patient use, non-sterile. Operates via pneumatic jet nebulization (inherited from predicate). Used in hospital, clinic, or home environments by patients under physician care. Accessory supplements parent device performance without altering principle of operation. Benefits include increased flexibility in patient positioning during aerosolized medication delivery.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including aerosol characterization (Total Mass, Fine Particle Fraction, MMAD, GSD) across multiple medications (Albuterol Sulfate, Budesonide, Ipratropium Bromide) and biocompatibility testing (ISO 10993-1, 3, 5, 10, 11, 12, 17, 18).

## Technological Characteristics

Plastic molded corrugated tube accessory. Pneumatic jet nebulization principle. Non-sterile, disposable, single-patient use. Biocompatibility per ISO 10993. Dry gas pathway VOC and PM2.5 assessment performed.

## Regulatory Identification

A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.

## Predicate Devices

- MC300 Nebulizer ([K173367](/device/K173367.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Trudell Medical International Marianne Tanton Director, Quality and Regulatory Affairs 725 Baransway Drive London, Ontario N5V 5G4 Canada

### Re: K203400

Trade/Device Name: Corrugated tube with mouthpiece accessory Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: Class II Product Code: CAF Dated: May 20, 2021 Received: May 21, 2021

### Dear Marianne Tanton:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Brandon Blakely, PhD Acting Assistant Director DHT1C: Division of ENT. Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K203400

#### Device Name

Corrugated tube with mouthpiece accessory

#### Indications for Use (Describe)

The corrugated tube is an accessory to the MC 300 ® Nebulizer. It is designed to work in conjunction with the MC 300 ® Nebulizer to extend the patient interface away from the nebulizer. The accessory is a single patient is intended to be used with either mouthpiece or mask for pediatric (ages 2 years and above) and adult patients, who are under the care of a licensed healthcare provider or physician.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D) | <div></div> |
|----------------------------------------------|-------------|
| Over-The-Counter Use (21 CFR 801 Subpart C)  | <div></div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Prepared: 14 June 2021

### 1. Submitter

Trudell Medical International 725 Baransway Drive London, Ontario N5V 5G4, Canada

Contact: Marianne Tanton Director, Quality and Regulatory Affairs Phone: 1-519-455-7060 Email: mtanton@trudellmed.com

### 2. Device Name

| Trade Name:          | None                                      |
|----------------------|-------------------------------------------|
| Common Name:         | Corrugated tube with mouthpiece accessory |
| Classification Name: | Nebulizer 21 CFR 868.5630                 |
| Regulatory Class:    | II                                        |
| Product Code:        | CAF                                       |

### 3. Predicate Device

MC300* Nebulizer - K173367 Trudell Medical International

The predicate device has not been subject to a recall.

### 4. Device Description and Principle of Operation

The corrugated tube with mouthpiece accessory is designed to work in conjunction with the cleared predicate device, MC300 * Nebulizer (K173367) to extend the patient interface away from the nebulizer.

In line with FDA's definition of a "medical device accessory", the corrugated tube with mouthpiece accessory is intended to supplement the performance of the parent device, MC300* Nebulizer (K173367).

The principle of operation of MC300* Nebulizer remains the same as cleared under K173367.

### 5. Indications for Use

The corrugated tube is an accessory to the MC 300 ® Nebulizer. It is designed to work in conjunction with the MC 300 ® Nebulizer to extent interface away from the nebulizer. The accessory is a single patient use device that is intended to be used with either

<sup>*</sup>Trademarks and registered trademarks of Trudell Medical International

{4}------------------------------------------------

mouthpiece or mask for pediatric (ages 2 years and above) and adult patients, who are under the care of a licensed healthcare provider or physician.

The MC300* Nebulizer indications for use remains the same as cleared under K173367.

### 6. Comparison to predicate device

The predicate device, MC300* Nebulizer (K173667) and the subject device (corrugated tube with mouthpiece) used with MC300* Nebulizer, are identical in purpose, function, core technology and method of operation.

The purpose of this submission is the addition of corrugated tube with mouthpiece accessory for use with the predicate device, MC300* Nebulizer (K173367). Table 1 provides a comparison of the subject and predicate devices.

| Element of<br>Comparison  | Corrugated tube with<br>mouthpiece to be used<br>with MC300* Nebulizer<br>(Subject Device)                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | MC300* Nebulizer<br>K173367<br>(Predicate Device)                                                                                                                                                                                                                                                                                                                                                        | Comparison |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Indications for<br>Use    | The nebulizer indications<br>for use remains the same<br>as cleared under K173367.<br>The corrugated tube is an<br>accessory to the MC 300®<br>Nebulizer. It is designed to<br>work in conjunction with the<br>MC 300® Nebulizer to<br>extend the patient interface<br>away from the nebulizer.<br>The accessory is a single<br>patient use device that is<br>intended to be used with<br>either mouthpiece or mask<br>for pediatric (ages 2 years<br>and above) and adult<br>patients, who are under the<br>care of a licensed<br>healthcare provider or<br>physician. | The nebulizer is intended to<br>be used with pediatric<br>(ages 2 years and above)<br>and adult patients, who are<br>under the care of a<br>licensed healthcare<br>provider or physician. The<br>device is designed to<br>aerosolize prescribed<br>medication for inhalation by<br>a patient in the hospital,<br>clinic or home care<br>environment. The nebulizer<br>is a single patient use<br>device. | Supplement |
| Principle of<br>Operation | The nebulizer's principle of<br>operation remains the<br>same as cleared under<br>K173367.                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Pneumatic Jet Nebulizer                                                                                                                                                                                                                                                                                                                                                                                  | Supplement |

### Table 1: Comparison to Predicate Device

<sup>*</sup>Trademarks and registered trademarks of Trudell Medical International

{5}------------------------------------------------

| Element of<br>Comparison | Corrugated tube with<br>mouthpiece to be used<br>with MC300* Nebulizer<br>(Subject Device)                                                                                                                                 | MC300* Nebulizer<br>K173367<br>(Predicate Device) | Comparison |
|--------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------|------------|
|                          | The corrugated tube with<br>mouthpiece is an accessory<br>to the MC300* Nebulizer. It<br>is designed to work in<br>conjunction with the<br>MC300* Nebulizer to<br>extend the patient interface<br>away from the nebulizer. |                                                   |            |
| Environment of<br>use    | Hospital, Clinic or Home                                                                                                                                                                                                   |                                                   | Similar    |
| Patient<br>population    | Adult and pediatric patients<br>(ages 2 years and above)                                                                                                                                                                   |                                                   | Similar    |
| Single Patient<br>Use    | Yes                                                                                                                                                                                                                        |                                                   | Similar    |
| Aerosolization           | Continuous during inhalation and exhalation                                                                                                                                                                                |                                                   | Similar    |
| Type of device           | Disposable, single patient use, prescription only, non-<br>sterile                                                                                                                                                         |                                                   | Similar    |
| Manufacturing<br>process | Plastic molding                                                                                                                                                                                                            |                                                   | Similar    |

Note: Type of gas source, flow rates and maximum fill volume are nebulizer characteristics and are not listed in the comparison table.

## 7. Performance Data

### 7.1. Aerosol Characterization

Aerosol characterization testing was performed in accordance with relevant sections of the CDRH Guidance Document "Reviewer Guidance for Nebulizers, Metered Dose Inhalers, Spacers and Actuators" (FDA/CDRH - 1993). The test results demonstrated substantially equivalent in-vitro performance between the subject device (corrugated tube with mouthpiece or mask) used with MC300* Nebulizer, and the predicate device, MC300* Nebulizer (K173367). The results of the testing are summarized in Tables 2 to 4 below.

{6}------------------------------------------------

#### Table 2. Summary of Aerosol Characterization Test Results for Corrugated Tube with Mouthpiece (Subject Device) Used with MC300* Nebulizer and MC300* Nebulizer (K173367) (Predicate Device) at 8 L/min

| Metric                                                                    | Albuterol Sulfate |                     | Budesonide        |                     | Ipratropium Bromide |                     |
|---------------------------------------------------------------------------|-------------------|---------------------|-------------------|---------------------|---------------------|---------------------|
|                                                                           | Subject Device    | Predicate<br>Device | Subject<br>Device | Predicate<br>Device | Subject<br>Device   | Predicate<br>Device |
| Total Mass<br>(μg)                                                        | 1248.8±57.8       | 1409.0±84.5         | 403.6±54.6        | 381.2±21.4          | 535.3±28.2          | 583.2±17.1          |
| Fine Particle<br>Fraction<br>(0.98-5.39µm<br>aerodynamic<br>diameter) (%) | 70.5±1.1          | 69.1±1.9            | 53.6±8.2          | 58.9±4.5            | 70.3±0.9            | 70.7±1.8            |
| Fine Particle<br>Mass (µg)                                                | 905.7±51.9        | 974.4±76.7          | 212.9±9.9         | 223.9±14.5          | 376.6±22.1          | 412.7±17.2          |
| MMAD (µm)                                                                 | 2.9               | 2.9                 | 5.0               | 4.6                 | 3.4                 | 3.1                 |
| GSD                                                                       | 2.2               | 2.2                 | 1.7               | 1.8                 | 2.2                 | 2.1                 |

### Table 3. Summary of Aerosol Characterization Test Results for Corrugated Tube with Mouthpiece (Subject Device) Used with MC300* Nebulizer and MC300* Nebulizer (K173367) (Predicate Device) at 13 psi (4 L/min)

| Metric                                                                    | Albuterol Sulfate |                     | Budesonide        |                     | Ipratropium Bromide |                     |
|---------------------------------------------------------------------------|-------------------|---------------------|-------------------|---------------------|---------------------|---------------------|
|                                                                           | Subject Device    | Predicate<br>Device | Subject<br>Device | Predicate<br>Device | Subject<br>Device   | Predicate<br>Device |
| Total Mass<br>(µg)                                                        | 1394.4±121.8      | 1348.8±158.9        | 358.1±16.5        | 383.3±21.8          | 513.3±40.3          | 555.4±35.8          |
| Fine Particle<br>Fraction<br>(0.98-5.39µm<br>aerodynamic<br>diameter) (%) | 71.6±3.5          | 72.0±2.5            | 59.5±3.9          | 59.4±7.0            | 74.4±1.1            | 72.1±2.4            |
| Fine Particle<br>Mass (µg)                                                | 995.9±67.0        | 971.3±124.8         | 213.0±16.5        | 227.8±22.8          | 381.8±28.2          | 400.8±32.2          |
| MMAD (µm)                                                                 | 2.8               | 2.4                 | 4.7               | 4.6                 | 2.4                 | 2.9                 |
| GSD                                                                       | 2.2               | 2.2                 | 1.7               | 1.9                 | 2.2                 | 1.9                 |

{7}------------------------------------------------

### Table 4. Summary of Aerosol Characterization Test Results for MC300* Nebulizer with Supine Accessory with Mask (Subject Device with Mask) and MC300* Nebulizer with Mask (Predicate Device with Mask) at 8 L/min

|                                                                            | Albuterol Sulfate              |                                  | Budesonide                     |                                  | Ipratropium Bromide            |                                  |
|----------------------------------------------------------------------------|--------------------------------|----------------------------------|--------------------------------|----------------------------------|--------------------------------|----------------------------------|
| Metric                                                                     | Subject<br>Device with<br>mask | Predicate<br>Device with<br>mask | Subject<br>Device with<br>mask | Predicate<br>Device with<br>mask | Subject<br>Device with<br>mask | Predicate<br>Device with<br>mask |
| Total Mass (µg)                                                            | 1396.7±124.7                   | 1435.4±92.2                      | 356.3±13.0                     | 333.5±22.8                       | 607.9±33.5                     | 570.5±29.8                       |
| Fine Particle<br>Fraction (0.98-<br>5.39µm<br>aerodynamic<br>diameter) (%) | 71.5±1.3                       | 69.5±2.1                         | 64.2±1.7                       | 61.3±7.1                         | 73.4±1.4                       | 70.3±1.9                         |
| Fine Particle<br>Mass (µg)                                                 | 999.7±103.9                    | 997.5±72.5                       | 228.8±9.6                      | 203.6±16.6                       | 446.1±26.0                     | 401.3±21.2                       |
| MMAD (µm)                                                                  | 2.6                            | 2.8                              | 4.3                            | 4.5                              | 2.6                            | 3.1                              |
| GSD                                                                        | 2.2                            | 2.2                              | 1.8                            | 1.8                              | 2.1                            | 2.0                              |

## 7.2. Biocompatibility Testing

Biological endpoints applicable to the corrugated tube with mouthpiece accessory are listed below. Materials were tested in accordance with ISO 10993-1 (2018) and the results satisfied the requirements. All in vitro and in vivo studies were performed and included the following battery of tests: Cytotoxicity, Sensitization, Intracutaneous Reactivity, Acute Systemic Toxicity, Genotoxicity, Extractables, and Chemical Characterization with a Biological Risk Assessment.

## Table 5. Summary of Biocompatibility Testing Conducted

| ISO Standard | Test/ Assessment                                |
|--------------|-------------------------------------------------|
| 10993-1      | Biological Risk Assessment                      |
| 10993-5      | Cytotoxicity Study Using the ISO Elution Method |
| 10993-3      | Genotoxicity: Bacterial Reverse Mutation Study  |
| 10993-3      | Genotoxicity: Mouse Lymphoma Assay              |
| 10993-11     | Acute Systemic Toxicity Study in Mice           |

{8}------------------------------------------------

| ISO Standard | Test/ Assessment                                                                                    |
|--------------|-----------------------------------------------------------------------------------------------------|
| 10993-10     | Intracutaneous Study in Rabbits                                                                     |
| 10993-10     | Guinea Pig Maximization Sensitization                                                               |
| 10993-17     | Establishment of allowable limits for leachable substances                                          |
| 10993-12     | Solvent and Extraction Condition Verification for ISO 10993-18 Chemical<br>Characterization Program |
| 10993-18     | Chemical characterization of materials                                                              |

## 7.3. Dry Gas Pathway Testing

To support the safe use of the corrugated tube with mouthpiece accessory in dry gas conditions, a worst-case assessment of volatile organic compounds (VOCs) and fine particles (particulate matter PM2.5) was conducted. Testing results and risk assessment demonstrated that exposure during use of the corrugated tube with mouthpiece accessory is unlikely to result in toxicological effects.

## 8. Clinical Performance Summary

Not applicable. The determination of substantial equivalence is not based on Clinical Performance data.

### 9. Conclusion

The non-clinical data demonstrate that the use of the corrugated tube with mouthpiece accessory used with MC300* Nebulizer (K173367) is safe and effective for use in patients for the delivery of aerosolized inhalation medications and is therefore substantially equivalent to the predicate device, MC300* Nebulizer, cleared under K173367.

---

**Source:** [https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/CAF/K203400](https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/CAF/K203400)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
