The JLN-23XX Series Piston Compress Nebulizer (with models JLN-2300AS, JLN-2301AS, JLN-2317AS and JLN-2320AS)
Applicant
Shenzhen Homed Medical Device Co., Ltd.
Product Code
CAF · Anesthesiology
Decision Date
Sep 6, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5630
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The JLN-23XX Series Piston Compress Nebulizer (with models JLN-2300AS, JLN-2301AS, JLN-2317AS and JLN-2320AS) include an AC powered air compressor that provides a source of compressed air for home health care use. The compressor is used in conjunction with a jet (pneumatic) nebulizer to convert certain inhalable drugs into an aerosol form for inhalation by a patient. The device can be used with adult or pediatic patients (4 years and older).
Device Story
Piston-type air compressor; provides compressed air to power jet (pneumatic) nebulizers. Used in home healthcare; operated by patients or caregivers. Compressor connects to nebulizer set; converts liquid medication into aerosol for inhalation. Aerosol output features particles <5 microns. Healthcare provider prescribes medication; patient self-administers via inhalation to treat respiratory disorders. Benefits include effective delivery of inhaled drugs for respiratory conditions.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via cascade impactor testing (EN13544-1) comparing particle size distribution of the subject device and predicate using Ipratropium bromide, Ventolin, and Pulmicort. Results demonstrated equivalent performance with no significant difference in particle distributions.
Technological Characteristics
Piston-type air compressor; 120VAC, 60Hz power; 8 LPM free flow; max pressure ≥30 psig. Complies with IEC 60601-1 (safety), IEC 60601-1-2 (EMC), and IEC 60601-1-11 (home healthcare). Biocompatible materials per ISO 10993-1. Standalone device.
Indications for Use
Indicated for adult and pediatric patients (4 years and older) requiring aerosolized medication delivery via a jet (pneumatic) nebulizer in a home healthcare setting.
Regulatory Classification
Identification
A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.
Predicate Devices
Devilbiss Model 3655 Compressor/Nebulize (K020932)
Reference Devices
Besmed Nebulizer Set (Model No. PN-1128E) (K091272)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a series of human profiles.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 6, 2017
Shenzhen Homed Medical Device Co., Ltd Shengming Shi Manager Of Technical Regulation Department No.2 Building, Longgu Industrial Zone, Longhua Town Shenzhen, 518109 CN
Re: K161586
Trade/Device Name: The JLN-23XX Series Piston Compress Nebulizer (with models JLN-2300AS, JLN-2301AS, JLN-2317AS and JLN-2320AS) Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: Class II Product Code: CAF Dated: August 4, 2017 Received: August 7, 2017
Dear Shengming Shi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Michael J. Ryan -S
for Lori A. Wiggins, MPT, CLT Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K161586
#### Device Name
The JLN-23XX Series Piston Compress Nebulizer (with models JLN-2301AS, JLN-2317AS and JLN-2320AS)
#### Indications for Use (Describe)
The JLN-23XX Series Piston Compress Nebulizer (with models JLN-2300AS, JLN-2301AS, JLN-2317AS and JLN-2320AS) include an AC powered air compressor that provides a source of compressed air for home health care use. The compressor is used in conjunction with a jet (pneumatic) nebulizer to convert certain inhalable drugs into an aerosol form for inhalation by a patient. The device can be used with adult or pediatic patients (4 years and older).
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| <div> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## K161586
# 510(K) SUMMARY
#### Submitter:
Shenzhen Homed Medical Device Co., Ltd. No.2 Building, Longgu Industrial Zone, Longhua Town, Shenzhen, Guangdong, 518109, CHINA
#### Contact Person: ●
Shengming Shi Manager of Technical Regulation Department Tel: 86-755-29821675-8011 Fax: 86-755-29821953 Email: shifei@homedgroup.com
#### Date Prepared: ●
Aug 30, 2017
#### Name of the Devices:
- Device Common Name: Nebulizer (direct patient interface)
- Device Proprietary Name: JLN-23XX Series Piston Compress Nebulizer (with models JLN-2300AS, JLN-2301AS, JLN-2317AS and JLN-2320AS)
- Classification Name: Nebulizer (Direct Patient Interface)
- Regulation Number: 21CFR 868.5630 ●
- Product Code: CAF
- Review Panel: Anesthesiology
#### Legally Marketed Predicate Device(s):
K020932 Devilbiss Model 3655 Compressor/Nebulize
#### Indications for Use:
The JLN-23XX Series Piston Compress Nebulizer (with models JLN-2300AS,
{4}------------------------------------------------
JLN-2301AS, JLN-2317AS and JLN-2320AS) include an AC powered air compressor that provides a source of compressed air for home health care use. The compressor is used in conjunction with a jet (pneumatic) nebulizer to convert certain inhalable drugs into an aerosol form for inhalation by a patient. The device can be used with adult or pediatric patients (4 years and older).
#### Reference Guidance:
- Guidance for Industry and FDA Staff Format for Traditional and Abbreviated ● 510(k)s
- Guidance for Industry and FDA Staff - The 510(k) Program Evaluating Substantial Equivalence in Premarket Notifications [510(k)]
- Reviewer Guidance for Nebulizers, Metered Dose Inhalers, Spacers and Actuators
#### Recognized Consensus Standard used
ISO 10993-1 : 2009
Biological evaluation of medical devices
#### IEC 60601-1: 2005 + A1: 2012
Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.
#### IEC 60601-1-2 : 2014
Medical electrical equipment - - Part 1-2: General requirements for basic safety and essential performance -Standard: Electromagnetic disturbances - -Collateral Requirements and tests
#### IEC 60601-1-11: 2010 + CORR.1: 2011
Medical electrical equipment -- Part 1-11: General requirements for basic safety and essential performance --Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
#### Device Description:
The JLN-23XX Series Piston Compress Nebulizer (with models JLN-2300AS, JLN-2301AS, JLN-2317AS and JLN-2320AS) is designed and manufactured by Shenzhen Homed Medical Device Co., Ltd.
The JLN-23XX Series Piston Compress Nebulizer (with models JLN-2300AS. JLN-2301AS, JLN-2317AS and JLN-2320AS) is a small, piston-type air compressor, sized to provide the proper flow and pressure sufficient to power jet (pneumatic)
{5}------------------------------------------------
nebulizers. The unit is designed and manufactured to comply with electrical and mechanical safety standards applicable to this type of device.
When the compressor is used in conjunction with a therapeutic nebulizer set, the system converts liquid medication into an aerosol form that can be inhaled by the patient for the treatment of a variety of respiratory disorders. The JLN-23XX Series Piston Compress Nebulizer (with models JLN-2300AS, JLN-2301AS, JLN-2317AS and JLN-2320AS) produces an aerosol output with the majority of the aerosol by mass contained in particles less than 5 microns in diameter.
The Besmed Nebulizer Set (Model No. PN-1128E) - (K091272) which has been under FDA clearance will be used with JLN-23XX Series Piston Compress Nebulizer.
### Comparison to Predicate Devices:
The Substantial Equivalence Comparison Chat is provided as follows:
| Table1: Substantial Equivalence Comparison Chat | | |
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Descriptive Information | Proposed Device | Predicate Device |
| 510(K) Number | None | K020932 |
| Proprietary or Model Name | The JLN-23XX Series<br>Piston Compress<br>Nebulizer (with models<br>JLN-2300AS,<br>JLN-2301AS,<br>JLN-2317AS and<br>JLN-2320AS) | Devilbiss Model 3655<br>Compressor/Nebulize |
| Indication for Use | The JLN-23XX Series<br>Piston Compress<br>Nebulizer (with models<br>JLN-2300AS,<br>JLN-2301AS,<br>JLN-2317AS and<br>JLN-2320AS) include an<br>AC powered air<br>compressor that provides<br>a source of compressed<br>air for home health care<br>use. The compressor is<br>used in conjunction with a<br>jet (pneumatic) nebulizer<br>to convert certain | The DeVilbiss<br>compressor / nebulizer<br>Model 3655 includes an<br>AC powered air<br>compressor that provides<br>a source of compressed<br>air for home health care<br>use. The compressor is<br>used in conjunction with a<br>jet (pneumatic) nebulizer<br>to convert certain<br>inhalable drugs into an<br>aerosol form for<br>inhalation by a patient.<br>The device can be used |
Table1: Substantial Equivalence Comparison Chat
{6}------------------------------------------------
| | inhalable drugs into an | with adult or pediatric |
|----------------------------------------|-----------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|
| | aerosol form for | patients. |
| | inhalation by a patient. | |
| | The device can be used | |
| | with adult or pediatric | |
| | patients (4 years and | |
| | older). | |
| Compressor Type | Piston | Piston |
| Electrical Requirements | 120VAC, 60Hz | 115 VAC (±10%) 60 Hz |
| Power Consumption | 200VA | 140 watts maximum |
| Storage/Transport<br>Temperature Range | -25 to 70°C (-13 to 158°F) | -40° to +158°F (-40° to<br>+70°C) |
| Storage/Transport<br>Humidity | 10 - 95% | up to 95%<br>non-condensing |
| Operating Humidity | 10 - 95% non-condensing | up to 95%<br>non-condensing |
| Operating Temperature<br>Range | 10 to 40°C (50 to 104°F) | +40° to +104°F (+5 to<br>+40°C) |
| Compressor Flow (Free<br>flow) | 8LPM | 8 lpm |
| Max Pressure | 30 psig or greater / 206<br>Kpa | 35 psig (241 kPa) |
| Operating Pressure Range | 30 psig or greater | / |
| Noise Level | below 63dBA | 58-62 dBA |
| Biocompatibility | Biocompatible | Biocompatible |
| Guidance | Reviewer Guidance for<br>Nebulizers, Metered Dose<br>Inhalers, Spacers and<br>Actuators | Reviewer Guidance for<br>Nebulizers, Metered Dose<br>Inhalers, Spacers and<br>Actuators |
### Discussion of Non-Clinical Tests Performed:
The performance requirements for the JLN-23XX Series Piston Compress Nebulizer (with models JLN-2300AS, JLN-2301AS, JLN-2317AS and JLN-2320AS) were verified according to the following tests to make sure to be substantial Equivalent to the predicate device.
The JLN-23XX Series Piston Compress Nebulizer (with models JLN-2300AS, JLN-2301AS, JLN-2317AS and JLN-2320AS) met the requirements of IEC 60601-1 and IEC 60601-1-2. These testes demonstrate the electrical safety and EMC protection of proposed devices meet the same standard as predicate device.
{7}------------------------------------------------
Laboratory testing, include noise tests, pressure tests and flowrate tests, were conducted to validate and verify that the JLN-23XX Series Piston Compress Nebulizer (with models JLN-2300AS, JLN-2301AS, JLN-2317AS and JLN-2320AS) met all design specifications in various environments and was substantially equivalent to the predicate device.
The biocompatibility test were conducted on the JLN-23XX Series Piston Compress Nebulizer (with models JLN-2300AS, JLN-2301AS, JLN-2317AS and JLN-2320AS) along to make sure the proposed device substantially equivalent to the predicate device in biological aspect.
Regard to product performance, the particle size distribution test via Cascade Impactor (conducted according EN13544-1) of the JLN-23XX Series Piston Compress Nebulizer (with models JLN-2317AS) with the nebulizer set (K091272) was performed in comparison to the predicate device (K020932) with three drugs (Ipratropium bromide, Ventolin, and Pulmicort). The test has shown the proposed device consistent regard to repeatability tests for each three classes of drug, and also demonstrated equivalent performance ability as the predicate device (K020932) that no significant difference in the particle distributions. The Proposed devices have same performance characteristics as the predicate device (K020932) and also meet its product specification as well.
### Discussion of Clinical Tests Performed:
Not Applicable.
### Conclusions:
The JLN-23XX Series Piston Compress Nebulizer (with models JLN-2300AS, JLN-2301AS, JLN-2317AS and JLN-2320AS) has the same intended use and similar characteristics as the predicate device. Moreover, bench testing contained in this submission demonstrates that the JLN-23XX Series Piston Compress Nebulizer (with models JLN-2300AS, JLN-2301AS, JLN-2317AS and JLN-2320AS) is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.