The FOX MOBILE inhalation system is a nebulizer system that will be used with patients for whom doctors have prescribed medication for nebulization in the home care, nursing home, sub-acute institution, or hospital environments. It is intended for patients 3 years and older who can coordinate breathing.
Device Story
FOX MOBILE is a handheld, single-patient, multi-use vibrating mesh nebulizer system. It delivers aerosolized medication to patients; system includes base unit, disposable reservoir, cap, and mouthpiece. Device features flow limitation valve (LIMIX) and air control. Operation is breathe-activated; device detects patient inhalation via pressure signal to trigger synchronized drug delivery on demand. User receives feedback via LEDs during inhalation. Inhalation times are adjustable via pre-sets. Device is battery-powered. Used in home, nursing home, sub-acute, or hospital environments. Healthcare providers prescribe the medication; patients operate the device. Benefits include synchronized, on-demand delivery of aerosolized medication, potentially improving drug deposition efficiency.
Clinical Evidence
Bench testing only. Comparative particle characterization testing was performed using a Cascade Impactor at adult (15 lpm) and pediatric (12 lpm) flow rates. Metrics included MMAD, GSD, Total Delivered Dose, Total Respirable Dose, and Total Output Rate for three drugs (Sultanol, CromoHEXAL, Atrovent). Results demonstrated performance equivalent to the predicate AKITA2 APIXNEB. Material biocompatibility was verified per ISO 10993-1.
Technological Characteristics
Vibrating mesh nebulizer technology. Materials tested per ISO 10993-1 (cytotoxicity, sensitization, genotoxicity, implantation, systemic/subchronic toxicity). Battery-powered. Features flow limitation valve (LIMIX) and LED user feedback. Synchronized, breathe-activated delivery triggered by pressure signal. Dimensions/form factor: handheld. Connectivity: none specified. Sterilization: not specified.
Indications for Use
Indicated for patients 3 years and older who can coordinate breathing, for whom medication has been prescribed for nebulization in home, nursing home, sub-acute, or hospital settings.
Regulatory Classification
Identification
A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of three human profiles facing right, arranged in a way that they appear to be interconnected. The profiles are rendered in black and have a flowing, abstract design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the profile graphic.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 9, 2015
Vectura GmbH c/o Paul Dryden Consultant Robert-Koch-Allee 29 Gauting, Germany 82131
Re: K142059
Trade/Device Name: FOX MOBILE Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer (Direct Patient Interface) Regulatory Class: Class II Product Code: CAF Dated: December 11, 2014 Received: December 12, 2014
#### Dear Mr. Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Dryden
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
*Tejashri Purohit-Sheth, M.D.*
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
K142059
Device Name
FOX MOBILE
Indications for Use (Describe)
The FOX MOBILE inhalation system is a nebulizer system that will be used with patients for whom doctors have prescribed medication for nebulization in the home care, nursing home, sub-acute institution, or hospital environments. It is intended for patients 3 years and older who can coordinate breathing.
Type of Use (Select one or both, as applicable)
XX Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
FORM FDA 3881 (9/13)
EF PSC Publishing Services (301) 443-6740
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
{3}------------------------------------------------
Attachment # 26
# 510(k) Summary
{4}------------------------------------------------
## 510(k) Summary
Page 1 of 6
| Date Prepared: | 11-Dec-14 |
|----------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|
| Vectura GmbH<br>Robert-Koch-Allee 29<br>82131 Gauting, GERMANY | Tel - +49 89 89 79 69 Ext 19<br>Fax - +49 89 89 79 69 22 |
| Official Contact: | Raimund Gleixner<br>Director of Regulatory Affairs, Devices |
| Proprietary or Trade Name: | FOX MOBILE |
| Common/Usual Name: | Nebulizer (Direct Patient Interface) |
| Classification Name: | Nebulizer (Direct Patient Interface)<br>Product Classification – CAF<br>21 CFR 868.5630<br>Class II |
| Predicate Devices: | K072019 - Activaero (Vectura) - AKITA2 APIXNEB<br>K935693 - Vortran - AutoNeb |
#### Device Description:
The FOX MOBILE is a single patient, multi-use, handheld inhalation system to deliver medications which are to be aerosolized. The system includes:
- A vibrating mesh nebulizer
- LEDs for user feedback
- . Flow limitation valve (LIMIX)
- Air control
#### Indications for Use:
The FOX MOBILE inhalation system is a nebulizer system that will be used with patients for whom doctors have prescribed medication for nebulization in the home care, nursing home, subacute institution, or hospital environments. It is intended for patients 3 years and older who can coordinate breathing.
#### Comparison to Predicates
We have chosen two (2) predicates for our substantial equivalence claim. The following is a rationale for this selection.
- K072019 - Activaero - AKITA2 APIXNEB
- K935693 Vortran AutoNeb ●
Table 1 is a table which highlights the reason for selecting each predicate.
{5}------------------------------------------------
## 510(k) Summary Page 2 of 6
11-Dec-14
| Substantial Equivalent Elements | K072019 | K935693 |
|-----------------------------------------------------------|---------------------|-------------------|
| | Activaero - AKITA2 | Vortran - AutoNeb |
| | APIXNEB | |
| Indications for Use | X | X |
| Environment of Use | X | X |
| Patient Population | X | X |
| Technology of vibrating mesh for<br>nebulizing drugs | X | Jet nebulizer |
| Synchronized delivery of nebulizer drug | X | X |
| Drug delivery on demand | X | X |
| Adjustable inhalation times that are<br>breathe activated | X | X |
| | Pre-set Smart Cards | Manual |
| Patient can adjust inhalation times | | X |
## Table 2 Compares the FOX MOBILE to Activaero - AKITA2 APIXNEB (K072019)
| Attribute | Activaero<br>AKITA2 APIXNEB (K072019) | Proposed<br>FOX MOBILE |
|--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The AKITA2 APIXNEB is a nebulizer<br>system that will be used with patients for<br>whom doctors have prescribed<br>medication (except pentamidine) for<br>nebulization in the home care, nursing<br>home, sub-acute institution, or hospital<br>environment.<br><br>The AKITA2 APIXNEB is intended for<br>patients 3 years and older who can<br>coordinate breathing. | The FOX MOBILE inhalation system<br>is a nebulizer system that will be used<br>with patients for whom doctors have<br>prescribed medication for nebulization<br>in the home care, nursing home, sub-<br>acute institution, or hospital<br>environments.<br><br>It is intended for patients 3 years and<br>older who can coordinate breathing. |
| Single patient, multi-use | Yes | Yes |
| Basic components | Control unit<br>Disposable reservoir with cap<br>Disposable aerosol generator<br>Disposable Mouthpiece | Base unit<br>Disposable reservoir with nebulizer<br>Disposable Cap<br>Disposable Mouthpiece |
| Technology, Features and Specifications | | |
| Nebulizer technology | Vibrating mesh | Vibrating mesh |
| Synchronized delivery of<br>nebulized drug | Yes | Yes |
| Drug delivery on demand | Yes | Yes |
| Inhalation Volume (times) | 2 – 8 seconds | 1.5 to 8 seconds |
| Gas source | Compressor | Self-generated by user |
| Flow rate | 15 lpm | 15 lpm |
| Power source | Mains | Battery AC |
{6}------------------------------------------------
#### 510(k) Summary Page 3 of 6
| Attribute | Activaero<br>AKITA2 APIXNEB (K072019) | Proposed<br>FOX MOBILE |
|---------------------|---------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance testing | Particle characterization<br>Comparison results found to be<br>equivalent | Particle characterization<br>Mechanical<br>Environmental<br>Simulated lifetime cycle (cleaning)<br>Differential Pressure<br>IEC 60601-1 plus deviations<br>IEC 60601-1-2<br>IEC 60601-1-6<br>IEC 60601-1-11 |
### Table 3 - Compares FOX MOBILE to Predicate - Vortran - AutoNeb (K935693)
| Attribute | Predicate<br>Vortran AutoNeb<br>K935693 | Proposed<br>FOX MOBILE |
|--------------------------------------------|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | Not listed but a general purpose<br>nebulizer | The FOX MOBILE inhalation system is a<br>nebulizer system that will be used with<br>patients for whom doctors have prescribed<br>medication for nebulization in the home care,<br>nursing home, sub-acute institution, or<br>hospital environments.<br><br>It is intended for patients 3 years and older<br>who can coordinate breathing. |
| Environments of use | Not specified but includes home care<br>setting | home care, nursing home, sub-acute<br>institution, or hospital environment |
| Patient population | All - not specified | patients 3 years and older who can coordinate<br>breathing |
| Technology | Jet nebulizer | Vibrating mesh |
| Operational Flow Rate | 1.5-16 LPM | 15 LPM |
| Synchronized delivery of<br>nebulized drug | Yes | Yes |
| Mode of Operation | Breathe activated | Breathe activated |
| Drug delivery triggered by | Patient Inhalation<br>Pressure signal | Patient Inhalation<br>Pressure signal |
| Adjustable Inhalation<br>times | 0.5 - 9.9 sec - patient adjustable | 1.5 - 8 sec - patient adjustable |
### Substantial Equivalence Discussion
Tables 2 to 3 above compare the key features of the proposed FOX MOBILE with the identified predicates and demonstrates that the device can be found to be substantially equivalent. In summary one can conclude that substantial equivalence is met based upon the following:
{7}------------------------------------------------
#### 510(k) Summary Page 4 of 6 11-Dec-14
#### Indications for Use -
The indications for use are identical for the proposed device when compared to the predicate -K072019 - AKITA2 APIXNEB. The predicate K935693 - AutoNeb does not have published indications but is known to be similar as a general purpose nebulizer. Discussion - Each device is indicated for use as a general purpose nebulizer.
## Technology and construction -
The nebulizer technology is identical to the predicate – K072019 - AKITA2 APIXNEB. The basic design, fabrication, etc. are equivalent to the predicates as a handheld nebulizer. Discussion - The design, vibrating mesh nebulizer technology, controlled inhalation flow and adjustable inhalation times are all similar to the predicate K072019 - AKITA A APIXNEB. There is a difference in the power source between the 2 devices. The FOX MOBILE is operated by a rechargeable battery while the predicate K072019 - AKITA2 APIXNEB is powered by standard AC power. The use of battery power to operate a device is common and this difference does not raise any new safety concerns.
#### Patient Control Features -
The ability of the user to change or adjust inhalation times is substantially equivalent to the predicate K935693 - AutoNeb.
In addition the design of drug delivery on demand, synchronized with inhalation, controlled inhalation flow, user feedback as to how they are performing during inhalation are very similar to the predicate K072019 - AKITA2 APIXNEB while some features are equivalent to the predicate K935693 – AutoNeb.
Discussion - This difference of adjusting the inhalation times between the FOX MOBILE and the predicate K935693 - AutoNeb is that the FOX MOBILE has pre-sets that the user selects, while the predicate AutoNeb, the user makes manual adjustments between the high and low inhalation time range. This difference in how the patient adjusts the inhalation time does not raise new safety issues.
#### Environment of Use -
The environments of use are identical to both predicates - K072019 - AKITA2 APIXNEB and K935693 - AutoNeb.
Discussion - Both devices are used in the home care, nursing home, sub-acute institution, or hospital environments settings.
#### Patient Population -
The patient population is identical to the predicate - K072019 - AKITA2 APIXNEB. Discussion - The patient populations are equivalent to K072019 - AKITA2 APIXNEB. The patient populations are believed to be equivalent to the predicate K935693 – AutoNeb but there is little available information.
{8}------------------------------------------------
#### 510(k) Summary Page 5 of 6 11-Dec-14
## Non-clinical Testing Summary –
#### Particle Characterization -
We performed comparative particle characterization testing via Cascade Impactor and the results demonstrated equivalent performance to the predicate K072019 - AKITA2 APIXNEB.
#### Materials -
We have tested the materials per ISO 10993-1 and the results supported the material satisfying the requirements. Testing included, Cytotoxicity, Sensitization, Genotoxicity, Implantation, Systemic Toxicity, Subchronic toxicity, Leachable and Extractable at 50°C for 72 hours, VOC and PM2.5 with a complete Risk Based Assessment.
Discussion - We have tested the materials which are commonly used in nebulizers.
| Table 4 – Summary of Comparative Particle Characterization for FOX MOBILE vs. |
|----------------------------------------------------------------------------------|
| AKITA² APIXNEB (K072019) – Adult flow rate – 15 lpm (Confidence interval of 95%) |
| Particle characterization | Drug | FOX MOBILE | AKITA2 APIXNEB<br>(K072019) |
|-------------------------------------------------|------------|--------------|-----------------------------|
| MMAD (um) | Sultanol | 4.0 ± 0.1 | 4.6 ± 0.1 |
| | CromoHEXAL | 4.0 ± 0.1 | 4.6 ± 0.1 |
| | Atrovent | 4.0 ± 0.1 | 4.3 ± 0.1 |
| GSD | Sultanol | 1.65±0.01 | 1.70+0.06 |
| | CromoHEXAL | 1.69±0.03 | 1.72±0.15 |
| | Atrovent | 1.67+0.02 | 1.63+0.03 |
| Total Delivered Dose by<br>Device (mg) | Sultanol | 2.22 ± 0.08 | 2.23 ± 0.10 |
| | CromoHEXAL | 17.8 ± 0.3 | 18.4 ± 0.5 |
| | Atrovent | 0.48 ± 0.01 | 0.48 ± 0.01 |
| Total Respirable Dose<br>(< 5 um) (%) | Sultanol | 67.7% ± 1.1 | 56.3% ± 3.2 |
| | CromoHEXAL | 66.2% ± 2.0 | 56.5% ± 9.0 |
| | Atrovent | 67.8% ± 2.4 | 64.9% ± 1.7 |
| Total Respirable Dose<br>(<5 um) (mg) | Sultanol | 1.49 ± 0.06 | 1.23 ± 0.07 |
| | CromoHEXAL | 11.77 ± 0.44 | 10.43 ± 1.87 |
| | Atrovent | 0.31 ± 0.02 | 0.30 ± 0.00 |
| Total Output Rate (TOR)<br>(mg/min) | Sultanol | 0.66 ± 0.03 | 0.80 ± 0.14 |
| | CromoHEXAL | 6.01 ± 0.19 | 7.50 ± 1.38 |
| | Atrovent | 0.15 ± 0.01 | 0.16 ± 0.02 |
| Coarse Particle Dose<br>>4.7 microns - % | Sultanol | 37.0 % ±1.2 | 49.1 % ±3.1 |
| | CromoHEXAL | 38.5 % ±2.0 | 48.7 % ±8.6 |
| | Atrovent | 37.2 % ±2.5 | 45.4 % ±2.0 |
| Fine Particle Dose<br><4.7 microns (mg) | Sultanol | 1.40 ± 0.06 | 1.13 ± 0.04 |
| | CromoHEXAL | 10.92 ± 0.43 | 9.45 ± 1.78 |
| | Atrovent | 0.30 ± 0.02 | 0.26 ± 0.01 |
| Ultra-Fine Particle Dose<br><1.0 microns – (mg) | Sultanol | <LoD | <LoD |
| | CromoHEXAL | | Limit of Detection |
| | Atrovent | | |
{9}------------------------------------------------
### 510(k) Summary Page 6 of 6 11-Dec-14
#### Table 5 - Summary of Comparative Particle Characterization for FOX MOBILE vs. AKITA2 APIXNEB (K072019) - Pediatric flow rate - 12 lpm (Confidence interval of 95%)
| Particle characterization | Drug | FOX MOBILE | AKITA2 APIXNEB |
|---------------------------|------------|-------------------------------------------------|---------------------|
| | | | (K072019) |
| MMAD (um) | Sultanol | 3.77 + 0.09 | 4.35 + 0.28 |
| | CromoHEXAL | 3.77 + 0.1 | 4.09 + 0.07 |
| | Atrovent | 3.68 + 0.09 | 4.31 + 0.24 |
| GSD | Sultanol | 1.73+0.04 | 1.80+0.08 |
| | CromoHEXAL | 1.58+0.39 | 1.74+0.06 |
| | Atrovent | 1.75+0.02 | 1.73+0.10 |
| Emitted Dose (Total | Sultanol | 2.19 + 0.03 | 2.19 + 0.06 |
| Delivered Dose) (mg) | CromoHEXAL | 18.15 ± 0.32 | 18.51 + 1.12 |
| | Atrovent | 0.47 + 0.01 | 0.48 + 0.01 |
| Total Respirable Dose | Sultanol | 70.5% ± 1.5 | 59.6% +6.0 |
| (< 5 um) (%) | CromoHEXAL | 69.1% ± 1.5 | 65.8% + 0.00 |
| | Atrovent | 71.1% + 1.7 | 56.6% ± 0.00 |
| Total Respirable Dose | Sultanol | 1.53 + 0.06 | 1.30 + 0.11 |
| (0.5 - 5 um) (mg) | CromoHEXAL | 12.53 + 0.40 | 12.13 + 1.20 |
| | Atrovent | 0.32 ± 0.03 | 0.30 ± 0.00 |
| Total Output Rate (TOR) | Sultanol | 0.62 + 0.02 | 0.78 + 0.18 |
| (mg/min) | CromoHEXAL | 5.75 + 0.32 | 6.77 + 0.23 |
| | Atrovent | 0.14 + 0.01 | 0.17 + 0.03 |
| Coarse Particle Dose | Sultanol | 33.8 % ±1.6 | 45.0 % +5.8 |
| >4.7 microns - % | CromoHEXAL | 35.0 % +1.7 | 39.5 % +2.5 |
| | Atrovent | 33.0 % +1.8 | 44.8 % +4.5 |
| Fine Particle Dose | Sultanol | 1.45 + 0.05 | 1.20 + 0.10 |
| <4.7 microns (mg) | CromoHEXAL | 11.80 + 0.42 | 11.21 + 1.13 |
| | Atrovent | 0.32 ± 0.01 | 0.26 + 0.02 |
| Ultra-Fine Particle Dose | Sultanol | <lod< td=""><td><lod< td=""></lod<></td></lod<> | <lod< td=""></lod<> |
| <1.0 microns - (mg) | CromoHEXAL | | Limit of Detection |
| | Atrovent | | |
### Substantial Equivalence Conclusion -
The sponsor has demonstrated through performance testing, design and features, and non-clinical testing that the proposed device and predicate have been found to be substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.