K112824 · Quality Life Technologies Co;Ltd · CAF · Jun 21, 2012 · Anesthesiology
Device Facts
Record ID
K112824
Device Name
QUALITY LIFE TECHNOLOGIES COMPRESSOR NEBULIZER
Applicant
Quality Life Technologies Co;Ltd
Product Code
CAF · Anesthesiology
Decision Date
Jun 21, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Quality Life Technologies Compressor Nebulizer is an AC powered air compressor that provides a source of compressed air for home health care use. The product is used in conjunction with a jet (pneumatic) nebulizer to convert liquid medication into an aerosol form with particles smaller than 5 microns diameter for inhalation into the patient's respiratory tract for the treatment of respiratory disorders. The target population for this device consists of adult patients suffering from, but not limited to asthma, cystic fibrosis, and chronic obstructive pulmonary disease. Further, additional applications for aerosolized medication are constantly under, investigation and this device may be deemed suitable for such applications as prescribed. The intended environment for use of the product is in the patient's home on the order of a physician.
Device Story
Device is AC-powered air compressor; provides compressed air to drive jet (pneumatic) nebulizer. Nebulizer converts liquid medication into aerosol mist; particles <5 microns diameter. Used in home environment; operated by patient on physician order. Healthcare provider prescribes medication; patient inhales aerosolized medication into respiratory tract. Benefits include targeted delivery of medication for respiratory disorders. Device includes tubing, nebulizer cup, insert, cap, and mouthpiece.
Clinical Evidence
No clinical data. Bench testing only. Performance testing included flow rate, pressure, and aerosol particle size distribution (MMAD) for various medications (Albuterol Sulfate, Ipratropium Bromide, Cromolyn Sodium). Biocompatibility testing performed per ISO 10993 Part-1. Electrical safety and electromagnetic compatibility testing performed per IEC 60601-1 and IEC 60601-1-2.
Technological Characteristics
AC-powered air compressor; pneumatic nebulization principle. Max flow rate 9.9 l/min; max pressure 40.0 psig. Medication cup volume max 5.0 ml. Components include tubing, nebulizer cup, insert, cap, mouthpiece. Biocompatible materials. Complies with IEC 60601-1 and IEC 60601-1-2.
Indications for Use
Indicated for adult patients with respiratory disorders including asthma, cystic fibrosis, and chronic obstructive pulmonary disease (COPD) requiring aerosolized medication delivery via a jet nebulizer in a home setting.
Regulatory Classification
Identification
A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.
Predicate Devices
DEVILBISS MODEL 4650 COMPRESSOR NEUBULIZER (K931015)
HAND HELD PHEUMATIC POWERED (DISPOSABLE) NEBULIZER (K870027)
{0}------------------------------------------------
# 510(k) Summary k / 12824
JUN 2 1 2012
## Summary Prepared Date: 04/12/2012
### Submission Sponsor:
#### Quality Life Technologies Co., Ltd
No.5 Lao Fu Wu Road, Huang Wu Industrial District, Dong Keng Town, Dongguan, Guangdong, P.R.China 523447
### Submission Correspondent:
Mr. Leon Lu Director of Quality and Requlatory Affairs
## MEDevice Services, LLC
3500 South Dupont Highway Dover, Delaware 19901 USA Phone: 1-877-202-1588 Fax: 1-888-202-8884 Email: info@medeviceservices.com Website: www.medeviceservices.com
#### Trade/Device Name:
Quality Life Technologies Compressor Nebulizer, Models NBA02-XXX Common or Usual Name: Nebulizer (direct patient interface) Device Class: II Classification Name: Nebulizer (direct patient interface) Regulation Number: 21 CFR 868.5630 Product Code: CAF Review Panel: Anesthesiology
### Predicate Device:
- K931015, DEVILBISS HEALTH CARE, INC. THUNGER TIGER CORP. . � DEVILBISS MODEL 4650 COMPRESSOR NEUBULIZER
- K870027, SALTER LABS . HAND HELD PHEUMATIC POWERED (DISPOSABLE) NEBULIZER
{1}------------------------------------------------
- K052438, EMG Technology Co., Ltd . EMG Handle Nebulizer NB02 Series
## Device Description:
Our Quality Life Technologies Compressor Nebulizer works to convert the medication to a high-quality mist of fine particles and spray liquid medications in aerosol form into gases that are directly delivered to the patient.
## Intended Use:
The Quality Life Technologies Compressor Nebulizer is an AC powered air compressor that provides a source of compressed air for home health care use. The product is used in conjunction with a jet (pneumatic) nebulizer to convert liquid medication into an aerosol form with particles smaller than 5 microns diameter for inhalation into the patient's respiratory tract for the treatment of respiratory disorders.
The target population for this device consists of adult patients suffering from, but not limited to asthma, cystic fibrosis, and chronic obstructive pulmonary disease. Further, additional applications for aerosolized medication are constantly under, investigation and this device may be deemed suitable for such applications as prescribed.
The intended environment for use of the product is in the patient's home on the order of a physician.
| Descriptive<br>Information | Proposed Device | Predicate Device |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | NONE | K931015 |
| Manufacturer | Quality Life Technologies Co., Ltd | DEVILBISS HEALTH CARE, INC |
| Proprietary or<br>Model Name | Quality Life Technologies<br>Compressor Nebulizer, Models<br>NBA02-XXX | DEVILBISS MODEL 4650<br>COMPRESSOR NEUBULIZER |
| Indication for Use | The Quality Life Technologies<br>Compressor Nebulizer is an AC<br>powered air compressor that<br>provides a source of compressed<br>air for home health care use. The<br>product is used in conjunction with<br>a jet (pneumatic) nebulizer to<br>convert liquid medication into an<br>aerosol form with particles smaller<br>than 5 microns diameter for | The DeVilbiss compressor/nebulizer is an<br>AC powered air compressor that provides<br>a source of compressed air for home<br>health care use. The product is used in<br>conjunction with a jet (pneumatic)<br>nebulizer to convert liquid medication into<br>an aerosol form with particles smaller<br>than 5 microns diameter for inhalation into<br>the patient's respiratory tract for the<br>treatment of respiratory disorders. |
## Comparison to Predicate Devices:
{2}------------------------------------------------
| inhalation into the patient's<br>respiratory tract for the treatment of<br>respiratory disorders. | The target population for this device<br>consists of adult patients suffering from,<br>but not limited to asthma, cystic fibrosis, |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| The target population for this device<br>consists of adult patients suffering<br>from, but not limited to asthma,<br>cystic fibrosis, and chronic<br>obstructive pulmonary disease.<br>Further, additional applications for<br>aerosolized medication are<br>constantly under investigation and<br>this device may be deemed suitable<br>for such applications as prescribed. | and chronic obstructive pulmonary<br>disease. Further, additional applications<br>for aerosolized medication are constantly<br>under investigation and this device may<br>be deemed suitable for such applications<br>as prescribed.<br>The intended environment for use of the<br>product is in the patient's home on the<br>order of a physician. |
| The intended environment for use<br>of the product is in the patient's<br>home on the order of a physician. | |
.
・
:
t
.
:
.
・・
・
·
| Descriptive<br>Information | Proposed Device | Predicate Device |
|------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | NONE | K870027 |
| Manufacturer | Quality Life Technologies Co., Ltd | SALTER LABS |
| Proprietary or<br>Model Name | Quality Life Technologies Compressor<br>Nebulizer, Models NBA02-XXX | HAND HELD PHEUMATIC<br>POWERED (DISPOSABLE)<br>NEBULIZER |
| Indication for Use | The Quality Life Technologies<br>Compressor Nebulizer is an AC<br>powered air compressor that provides<br>a source of compressed air for home<br>health care use. The product is used in<br>conjunction with a jet (pneumatic)<br>nebulizer to convert liquid medication<br>into an aerosol form with particles<br>smaller than 5 microns diameter for<br>inhalation into the patient's respiratory<br>tract for the treatment of respiratory<br>disorders. | Hand Held Pheumatic Powered<br>(Disposable) Nebulizer is to convert<br>liquid medication into an aerosol form<br>with particles smaller than 5 microns<br>diameter for inhalation into the<br>patient's respiratory tract for the<br>treatment of respiratory disorders |
| | The target population for this device<br>consists of adult patients suffering<br>from, but not limited to asthma, cystic<br>fibrosis, and chronic obstructive<br>pulmonary disease. Further, additional<br>applications for aerosolized medication<br>are constantly under investigation and<br>this device may be deemed suitable for<br>such applications as prescribed. | |
| | The intended environment for use of<br>the product is in the patient's home on<br>the order of a physician. | |
| Principle | Aerosolizes liquid into aerosol form by<br>compressing air. | Aerosolizes liquid into aerosol form by<br>compressing air |
| MMAD | <5 micron<br>Albuterol Sulfate (2.5 mg/3.0ml): 2.21 | <5 micron<br>Albuterol Sulfate (2.5 mg/3.0ml): 3.13 |
·
{3}------------------------------------------------
| | Ipratropium Bromide (0.5 mg/2.5ml):<br>3.04 | Ipratropium Bromide (0.5 mg/2.5ml):<br>3.10 |
|-------------------|-------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| | Cromolyn Sodium (20mg/2.0ml): 2.87 | Cromolyn Sodium (20mg/2.0ml): 2.40 |
| Accessories | Tubing, Nebulizer Cup, Nebulizer<br>Insert, Nebulizer Cap, Mouthpiece | Mouthpiece, T-piece, Cap, Medication<br>Cup, Nebulizer Air-Inlet Connector,<br>Baffle, Tubing |
| Target population | Adult patients | Both adult and pediatric patients |
| Environment use | In the patient's home on the order of a<br>physician. | In the patient's home on the order of a<br>physician. |
| Design | Medication Cup Volumes: Max.<br>5.0ml<br>Device features: handy | Medication Cup Volumes: Max.<br>5.0ml<br>Device features: handy |
| Performance | Flow rate of compressor:<br>Max. flow rate of compressor is 9.9<br>(l/min)<br>Max. pressure is 40.0 (psig) | Flow rate of compressor:<br>Max. flow rate of compressor is 10.0<br>(l/min)<br>Max. pressure is 40.0 (psig) |
| | Percent of drug does delivered<br>which falls in the respirable<br>fraction: | Percent of drug does delivered<br>which falls in the respirable<br>fraction: |
| | Albuterol Sulfate (2.5mg/3ml): 24.5%<br>Ipratropium Bromide (0.5mg/2.5ml):<br>47.7%<br>Cromolyn Sodium (20mg/2.0ml):<br>18.1% | Albuterol Sulfate (2.5mg/3ml): 53.5%<br>Ipratropium Bromide (0.5mg/2.5ml):<br>48.0%<br>Cromolyn Sodium (20mg/2.0ml):<br>29.5% |
| Biocompatibility | Biocompatible | Biocompatible |
| Descriptive<br>Information | Proposed Device | Predicate Device |
|------------------------------|------------------------------------------------------------------------|----------------------------------|
| 510(k) Number | NONE | K052438 |
| Manufacturer | Quality Life Technologies Co., Ltd | EMG Technology Co., Ltd |
| Proprietary or<br>Model Name | Quality Life Technologies<br>Compressor Nebulizer, Models<br>NBA02-XXX | EMG Handle Nebulizer NB02 Series |
| Materials | Same | Same |
## Discussion of Non-Clinical Tests Performed:
The performance requirements for Quality Life Technologies Compressor Nebulizer were met:
The patient contact components of the Quality Life Technologies Compressor Nebulizer met the requirements of ISO 10993 Part-1, "Biological Evaluation of Medical Devices".
Quality Life Technologies Compressor Nebulizer met the requirements of IEC 60601-1 and IEC 60601-1-2.
{4}------------------------------------------------
The performance of the Quality Life Technologies Compressor Nebulizer was tested by third party lab and demonstrated substantially equivalent to that of the predicate device.
Therefore, we conclude that the Quality Life Technologies Compressor Nebulizer is both safe and effective for its intended use.
## Discussion of Clinical Tests Performed:
None
### Conclusion:
The proposed device is as safe and effective as the predicate devices. The proposed device has the same intended uses and indications, similar technological characteristics, and principles of operation as its predicate device. The minor differences between the proposed device and its predicate devices raise no new issues of safety or effectiveness. Thus, Quality Life Technologies Compressor Nebulizer, Models NBA02-XXX are substantially equivalent to its predicate devices.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle with three legs, representing the department's mission to protect the health of all Americans and provide essential human services. The seal is surrounded by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Quality Life Technologies Company, Limited C/O Mr. Leon Lu Director, of Quality and Regulatory Affairs MEDevice Services, LLC 3500 South Dupont Highway Dover, Delaware 19901
·JUN 2 1 2012
Re: K112824
Trade/Device Name: Quality Life Technologies Nebulizer Compressor, Models NBA02-XXX Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: II Product Code: CAF Dated: June 12, 2012 Received: June 15, 2012
Dear Mr. Lu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
#### Page 2- Mr. Lu
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address · http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Nh for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices
Office of Device Evaluation
Center for Devices and Radiological Health
#### Enclosure
{7}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known):
Device Name: Quality Life Technologies Nebulizer Compressor, Models NBA02-XXX
#### Indications for Use:
The Quality Life Technologies Compressor Nebulizer is an AC powered air compressor that rnovides a source of compressed air for home health care use. The product is used in provided a ooutro of compatic) nebulizer to convert liquid medication into an aerosol form with particles smaller than 5 microns diameter for inhalation into the patient's respiratory fract for the treatment of respiratory disorders.
The target population for this device consists of adult patients suffering from, but not limited to asthma, cystic fibrosis, and chronic obstructive pulmonary disease. Further, additional assilind, oyotto librools) in medication are constantly under investigation and this device may be deemed suitable for such applications as prescribed.
The intended environment for use of the product is in the patient's home on the order of a physician.
Over-The-Counter Use AND/OR Prescription Use × (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
L. Schultheis
Division of Anesthestology, General Hospital Infection Control, Dental Devices
510(k) Number: KC.172824
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.