K072019 · Activaero America, Inc. · CAF · Nov 5, 2007 · Anesthesiology
Device Facts
Record ID
K072019
Device Name
AKITA2 APIXNEB
Applicant
Activaero America, Inc.
Product Code
CAF · Anesthesiology
Decision Date
Nov 5, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5630
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The AKITA2 APIXNEB is a nebulizer system that will be used with patients for whom doctors have prescribed medication for nebulization in the home care, nursing home, subacute institution, or hospital environment. The AKITA2 APIXNEB is intended for patients 3 years and older who can coordinate breathing.
Device Story
AKITA2 APIXNEB is an electronic nebulizer system for aerosolizing liquid medications. System comprises an electrically powered compressor and a nebulizer handset (based on Pari eFlow technology) using a piezoelectric-vibrated perforated stainless steel membrane. Device operates by nebulizing medication only during the inhalation phase, guided by patient-specific breathing patterns stored on 'Smart Cards'. Used in home, nursing home, sub-acute, or hospital environments by patients 3 years and older. Healthcare providers program delivery parameters onto Smart Cards to control inhalation flow and timing. Output is a controlled aerosol mist; device optimizes drug delivery based on individual breathing patterns, potentially improving therapeutic efficacy and patient compliance.
Clinical Evidence
Bench testing only. Performance evaluations included flow performance, trigger pressure, on/off performance, nebulization time, time lag, timing parameters, life-time testing, cleaning performance, and smart card functionality. Additional testing included VOC/PM2.5 emissions, electrical safety, EMC, EMI, and mechanical/environmental testing. No clinical data provided.
Technological Characteristics
System includes an electrically powered compressor and a piezoelectric-vibrated perforated stainless steel membrane nebulizer handset. Features include inhalation-synchronized nebulization and programmable Smart Card interface for patient-specific breathing patterns. Operates on AC power. Designed for single-patient, multi-use. Software-controlled delivery parameters.
Indications for Use
Indicated for patients 3 years and older capable of coordinating breathing, requiring prescribed nebulized medication (excluding pentamidine) in home, nursing home, sub-acute, or hospital settings.
Regulatory Classification
Identification
A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.
{1}------------------------------------------------
K072019
Non-Confidential Summary of Safety and Effectiveness
Page 1 of 3 18-Sep-07
| Activaero America, Inc.<br>P.O. Box 351<br>Dublin, OH 43017-9684 | Tel - (614) 761-3555<br>Fax - (614) 761-3505 | |
|------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Official Contact: | William Zimlich - CEO | |
| Proprietary or Trade Name: | AKITA² APIXNEB | NOV - 5 200 |
| Common/Usual Name: | Nebulizer systems | |
| Classification Name: | Nebulizer (Direct Patient Interface)<br>CAF - 868.5630 | |
| Predicate Devices: | I-neb Insight -- K052941 -- Profile / Respironics<br>ProDose - K030747 - Profile / Respironics<br>eFlow™ - K033833 - Pari<br>AutoNeb -- K935693 - Vortran | |
### Device Description
The AKITA? APIXNEB nebulizer and the handset together constitute a single patient, multi-use, electronic nebulizer system designed to aerosolize liquid medications. The system includes and features:
- An electrically powered compressor which provides an air flow to the AKITA2 . APIXNEB nebulizer handset.
- . A nebulizer handset based upon the PARI e-Flow™, K033833, which uses piezoelectric vibration of a perforated stainless steel membrane (head) for the aerosol generation.
- . Single patient use, reusable
- Nebulization only during inhalation phase .
- . Smart Card series for defined patient breathing patterns
### Indications for Use
The AKITA2 APIXNEB is a nebulizer system that will be used with patients for whom doctors have prescribed medication (except pentamidine) for nebulization in the home care, nursing home, sub-acute institution, or hospital environment.
## Patient Population
The AKITA2 APIXNEB is intended for patients 3 years and older who can coordinate breathing.
### Environment of Use
Home care, nursing home, sub-acute institution, or hospital
### Contraindications
None
-
{2}------------------------------------------------
#### Non-Confidential Summary of Safety and Effectiveness
Page 2 of 3 18-Sep-07
To demonstrate substantial equivalence as well as safety and effectiveness as series of performance tests were done.
For substantial equivalence the predicate comparisons included:
- . General Attributes
- o Indications for Use
- ্ Patient Population
- o Environments of use
- Vortran AutoNeb K935693, I-neb K052941and Prodose AAD System -O K030747
- . Nebulizer performance
.
.
- Particle characterization ্ত
- Pari eFlow™ K033833 o
- Nebulization during inhalation
- Vortran AutoNeb K935693, I-neb K052941and Prodose AAD System o K030747
- Delivery based upon breathing patterns .
- 1-neb K052941and Prodose AAD System K030747 0
- . Algorithm to program breathing patterns
- I-neb K052941and Prodose AAD System K030747 ্ত
- in vitro data Activaero 0
- Controlled inhalation flow
- I-neb K052941 ୍
- . Use of programmable disc (Smart cards) for setting nebulization and delivery parameters
- O Prodose AAD System K030747 and I-neb K052941
For safety and effectiveness testing included:
- Performance of the AKITA2 APIXNEB system .
- Flow performance o
- o Trigger Pressure performance
- o On/Off performance
- Nebulization Time performance 0
- Time lag performance o
- Timing Parameters o
- Life Time test O
- Cleaning performance 0
- o Smart Card performance
- Inhalation / Exhalation resistance
- . VOC and PM25 testing
- Electrical safety, EMC, EMI, Mechanical and environmental testing ●
{3}------------------------------------------------
## Non-Confidential Summary of Safety and Effectiveness
Page 3 of 3 18-Sep-07
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
# Differences Between Other Legally Marketed Predicate Devices
The AKITA2 APIXNEB system is viewed as substantially equivalent to the predicates.
There are no significant differences that affect the safety or effectiveness of the intended device as compared to the predicate devices.
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the words "Public Health Service" in a simple, sans-serif font. The text is arranged on a single line, with each word clearly legible. The overall impression is clean and straightforward, suggesting an official or institutional context.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Activaero America, Incorporated C/O Mr. Paul Dryden President ProMedic, Incorporated 3460 Pointe Creck Court, # 102 Bonita Springs, Florida 34134
NOV - 5 2007
Rc: K072019
Trade/Device Name: AKITA2 APIXNEB Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: II Product Code: CAF Dated: September 18, 2007 Received: September 20, 2007
Dear Mr. Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Dryden
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation cntitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Neil R. Ogden
Chiu Lin. Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use Statement
Page 1 of 1
510(k) Number: K072019 (To be assigned)
AKITA2 APIXNEB Device Name:
Indications for Use:
The AKITA2 APIXNEB is a nebulizer system that will be used with patients for whom doctors have prescribed medication for nebulization in the home care, nursing home, subacute institution, or hospital environment.
The AKITA2 APIXNEB is intended for patients 3 years and older who can coordinate breathing.
Prescription Use XX (Part 21 CFR 801 Subpart D)
or
Over-the-counter use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
*(Division Sign-Off)*
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: KC)72016
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.