K032809 · Respironics Healthscan, Inc. · CAF · Nov 13, 2003 · Anesthesiology
Device Facts
Record ID
K032809
Device Name
HOSPITAL SPACER
Applicant
Respironics Healthscan, Inc.
Product Code
CAF · Anesthesiology
Decision Date
Nov 13, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5630
Device Class
Class 2
Attributes
Therapeutic, Pediatric, 3rd-Party Reviewed
Indications for Use
ProChamber™ Valved Holding Chamber is intended for use in combination with Metered Dose Inhalers (MDI's) to assist in respiratory drug delivery. All patients using MDI's, i.e., children and adults, may use ProChamber ™. Settings for use include the home, clinics and hospitals. This device is intended for prescription use.
Device Story
ProChamber Valved Holding Chamber is a non-electrical accessory for Metered Dose Inhalers (MDIs). It consists of an MDI adapter, chamber, mouthpiece, duckbill valve, exhaust valve, and silicone cap. The device functions by attaching to an MDI to facilitate drug delivery; the one-way valve system and chamber design help minimize aerosol deposition in the oropharynx and ensure reproducible dose delivery. It is used by patients (children and adults) in homes, clinics, and hospitals under prescription. The device utilizes the MDI's own actuator to direct the aerosol plume toward the patient. It benefits patients by improving the efficiency of respiratory medication delivery compared to using an MDI alone.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, specifically particle size distribution and dose delivery reproducibility compared to the predicate device and MDI alone, following FDA guidance for spacers and actuators.
Technological Characteristics
Non-electrical valved holding chamber. Components include MDI adapter, chamber, mouthpiece, duckbill valve, exhaust valve, and silicone cap. Features one-way valve mechanism. Portable, manual operation. No software or electronic components.
Indications for Use
Indicated for children and adults using Metered Dose Inhalers (MDIs) to assist in respiratory drug delivery in home, clinic, or hospital settings.
Regulatory Classification
Identification
A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.
{0}------------------------------------------------
# NOV 1 3 2003
## Attachment #1
## 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is:_ K032809
#### Submitter's Identification: 1.
Respironics HealthScan, Inc. 41 Canfield Road Cedar Grove, NJ 07009
Contact: Ms. Lauren R. Ziegler
### Date Summary Prepared:
July 31, 2003
#### 2. Name of the Device:
ProChamber ™ Valved Holding Chamber
#### 3. Predicate Device Information:
K#992917, Aerochamber Plus™ Valved Holding Chamber, Trudell Medical International, London, Ontario, Canada
#### 4. Device Description:
The device consists of an MDI Adapter, a Chamber, a Mouthpiece, a Duckbill Valve, an Exhaust Valve and a Silicone Cap. There are no electrical or electronic components.
#### ട. Intended Use:
ProChamber™ Valved Holding Chamber is intended for use in combination with Metered Dose Inhalers (MDI's) to assist in respiratory drug delivery. All patients using MDI's, i.e., children and adults, may use ProChamber ™. Settings for use include the home, clinics and hospitals.
This device is intended for prescription use.
{1}------------------------------------------------
#### Comparison to Predicate Devices: 6.
The subject (ProChamber™) and predicate device (AeroChamber Plus™ – K#992917) are indicated for the same intended use. The devices are of similar size and shape and performance characteristics are basically the same. The main difference is the valve design. The subject device and predicate devices share the following characteristics:
- Similar dimensions .
- Portable .
- Reproducible dose delivery .
- Minimization of aerosol disposition in oropharynx .
- Spray direction of aerosol plume is toward patient .
- . One-way valve
- Utilizes MDI's own actuator .
### 7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:
The ProChamber ™ and the predicate device, the AeroChamber Plus™ , were tested for particle size distribution testing, as well as directly compared to and MDI alone without the spacer attached. A particle size comparison between the subject spacer and the attached MDI alone was tested, according to the methodology outlined in the FDA "Reviewer Guidance for Nebulizers, Metered Dose Inhalers, Spacers and Actuators, 10/93". Testing documentation showed that particle size range and reproducibility compared to the predicate device, was acceptable.
#### Discussion of Clinical Tests Performed: 8.
Not applicable
#### 9. Conclusions:
The subject device, the ProChamber™ Metered Dose Inhaler (MDI) Valved Holding Chamber, has a similar intended use and similar characteristics as the predicate device, the AeroChamber Plus™ Valved Holding Chamber. Moreover, non-clinical testing included in this submission demonstrated that any differences in their characteristics do not raise any new questions of safety or effectiveness. Thus, the subject device is substantially equivalent to the predicate device.
{2}------------------------------------------------
### Exhibit B
Page __ 1 __ of __ 1 __
510(k) Number (if known): __ K03 2809
Device Name ProChamber™ Valved Holding Chamber
### Indications For Use:
ProChamber™ Valved Holding Chamber is intended for use in combination with Metered Dose Inhalers (MDI's) to assist in respiratory drug delivery. All patients using MDI's, i.e., children and adults, may use ProChamber ™. Settings for use include the home, clinics and hospitals.
This device is intended for prescription use.
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
OR
Over-The Counter Use (Optional Format 1-2-96)
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the department's name around the perimeter. In the center of the seal is an abstract image of a bird-like figure, possibly representing an eagle or other national symbol. The image is in black and white.
### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 1 3 2003
Respironics Healthscan Inc. C/O Mr. Robert Mosenkis President CITECH 5200 Butler Pike Plymouth Meeting, PA 19462
Re: K032809
Trade/Device Name: ProChamber Valved Holding Chamber Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: Class II Product Code: CAF Dated: October 28, 2003 Received: October 29, 2003
Dear Mr. Mosenkis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice
{4}------------------------------------------------
Page 2 - Mr. Robert Mosenkis
requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours.
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### Exhibit B
Page 1 of
K032809 510(k) Number (if known): __
Device Name ProChamber™ Valved Holding Chamber
Indications For Use:
ProChamber™ Valved Holding Chamber is intended for use in combination with Metered Dose Inhalers (MDI's) to assist in respiratory drug delivery. All patients using MDI's, i.e., children and adults, may use ProChamber ™. Settings for use include the home, clinics and hospitals.
This device is intended for prescription use.
bthul
Division Sign-Off
esiology. General Hr
510(k) Number: K032809
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
OR
Prescription Use
(Per 21 CFR 801.109) √
Over-The Counter Use (Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.