DIDECO COMPACT-A AND COMPACT-M AUTOTRANSFUSION SYSTEMS

K963759 · Sorin Biomedical, Inc. · CAC · Feb 20, 1998 · Anesthesiology

Device Facts

Record IDK963759
Device NameDIDECO COMPACT-A AND COMPACT-M AUTOTRANSFUSION SYSTEMS
ApplicantSorin Biomedical, Inc.
Product CodeCAC · Anesthesiology
Decision DateFeb 20, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5830
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Dideco Compact-A and Compact-M Autotransfusion Systems are intended to process, shed or collect blood for autologous transfusion. The Compact-A and Compact-M Autotransfusion Systems are to be used with disposables for the collection of shed blood and aspirated body fluids, and the separation of erythrocytes from other components of the aspirated blood prior to, during and/or after a surgical procedure. The Systems are also recommended to collect platelet-rich plasma (PRP) and/or platelet-poor plasma (PPP) from the patient’s whole blood immediately preoperative or intraoperative to a surgical procedure.

Device Story

The Dideco Compact-A and Compact-M are high-speed, lightweight automatic autotransfusion systems. The system includes a rolling cart, IV pole, and portable vacuum pump module. It functions by collecting shed blood and aspirated body fluids during or after surgery, or whole blood preoperatively/intraoperatively. The device processes these inputs to separate erythrocytes from other blood components or to isolate PRP/PPP. Used in clinical surgical settings, the system is operated by trained medical personnel. The output is processed autologous blood products returned to the patient, which reduces the need for donor blood transfusions and benefits the patient by minimizing transfusion-related risks.

Clinical Evidence

No clinical trials were performed. Substantial equivalence was established via bench and in vivo functional testing, including plasma free hemoglobin, ADP/collagen, pH testing, and cell/platelet counting to verify performance against the predicate device.

Technological Characteristics

High-speed automatic autotransfusion system; includes portable vacuum pump module, rolling cart, and IV pole. Operates via mechanical separation of blood components. Materials and flow rates are identical to the predicate Electromedics AT-1000. System is designed for use with disposables.

Indications for Use

Indicated for patients undergoing surgical procedures requiring autologous blood transfusion, including collection of shed blood, aspirated body fluids, and separation of erythrocytes. Also indicated for preoperative or intraoperative collection of platelet-rich plasma (PRP) and platelet-poor plasma (PPP) from whole blood.

Regulatory Classification

Identification

An autotransfusion apparatus is a device used to collect and reinfuse the blood lost by a patient due to surgery or trauma.

Predicate Devices

Submission Summary (Full Text)

{0} K963759 510(k) Notification FEB 20 1998 Sorin Biomedical Inc. Dideco Compact-A and Compact-M Autotransfusion Systems XIV. 510(k) SUMMARY A. Name and Address of Submitter Sorin Biomedical Inc. 17600 Gillette Avenue P.O. Box 19503 Irvine, California 92713-9503 B. Telephone and Fax Numbers of Submitter Telephone: (714) 250-8322 Fax: (714) 757-8644 C. Name of Contact Person Susan Reimers Manager, Clinical and Regulatory Affairs D. Date Summary was Prepared: September 16, 1996 E. Device Name Trade or Proprietary Name: Dideco Compact-A and Compact-M Autotransfusion Systems Common Name: Autotransfusion Apparatus Classification Name: Autotransfusion Devices F. Summary of Substantial Equivalence The Dideco Compact-A and Compact-M Autotransfusion Systems are substantially equivalent in intended use, materials, design, and performance characteristics to the Electromedics AT-1000. SORIN BIOMEDICAL INC. CONFIDENTIAL 22 {1} Sorin Biomedical Inc. 510(k) Notification Dideco Compact-A and Compact-M Autotransfusion Systems ## G. Device Description The Dideco Compact-A and Compact-M Autotransfusion Systems are composed of the following equipment: a high-speed lightweight automatic autotransfusion system including a rolling cart for the system with an IV pole and a portable vacuum pump module. ## H. Device Intended Use The Dideco Compact-A and Compact-M Autotransfusion Systems are intended to process, shed or collect blood for autologous transfusion. The Compact-A and Compact-M Autotransfusion Systems are to be used with disposables for the collection of shed blood and aspirated body fluids, and the separation of erythrocytes from other components of the aspirated blood prior to, during and/or after a surgical procedure. The Systems are also recommended to collect platelet-rich plasma (PRP) and/or platelet-poor plasma (PPP) from the patient’s whole blood immediately preoperative or intraoperative to a surgical procedure. ## I. Summary of Comparison of Technological Characteristics The technological characteristics of the Dideco Compact-A and Compact-M Autotransfusion Systems are similar to the Electromedics AT-1000 System. All three (3) systems are identical in respect to their process steps, method of operation, materials, and suggested flow rates. All three (3) are indicated for separation of Platelet-Poor Plasma (PPP) and Platelet-Rich Plasma (PRP) from whole blood collected from the patient. ## J. Summary of In-Vivo Tests Substantial equivalence was based on a comparison of test results from the following in vivo functional tests: - plasma free Hgb, ADP/Collagen and pH testing - cell and platelet counting. ## K. Conclusions Based upon the above information, Sorin Biomedical Inc. concludes that both the Dideco Compact-A and Compact-M Autotransfusion Systems and predicate Electromedics AT-1000 Autotransfusion System have comparable and acceptable performance for the indication of the collection of shed blood and aspirated body fluids, and the separation of erythrocytes from other components of the aspirated blood prior to, during and/or after a surgical procedure. The systems are also indicated for use to collect platelet-rich plasma (PRP) and/or platelet-poor plasma (PPP) from the patient’s whole blood immediately preoperative or intraoperative to a surgical procedure. SORIN BIOMEDICAL INC. CONFIDENTIAL {2} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 FEB 20 1998 Ms. Sharon Thompson Director, Regulatory Affairs/Quality Sorin Biomedical, Inc. 17600 Gillette Avenue P.O. Box 19503 Irvine, CA 92713-9503 Re: K963759 Dideco Compact and Compact A Autotransfusion Systems Regulatory Class: II (two) Product Code: 74 CAC Dated: January 30, 1998 Received: February 4, 1998 Dear Ms. Thompson: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {3} This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, ![img-0.jpeg](img-0.jpeg) Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4} K963759/A Page 1 of 1 510(k) Number (if known): K963759 Device Name: DIDECO COMPACT-A AND COMPACT-M AUTOTRANSFUSION SYSTEMS Indications For Use: The Dideco Compact-A and Compact-M Autotransfusion Systems are intended to process, shed or collect blood for autologous transfusion. The Compact-A and Compact-M autotransfusion systems are to be used with disposables for the collection of shed blood and aspirated body fluids, and the separation of erythrocytes from other components of the aspirated blood prior to, during and/or after a surgical procedure. The systems are also recommended to collect platelet-rich plasma (PRP) and/or platelet-poor plasma (PPP) from the patient's whole blood immediately preoperative to a surgical procedure. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use ☑ (Per 21 CFR 801.109) OR Over-The-Counter Use ☐ (Optional Format 1-2-96) (Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices 510(k) Number K963759
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...