HAEMONETICS CARDIOVASCULAR PERIOPERATIVE AUTOTRANSFUSION SYSTEM
Applicant
Haemonetics Corp.
Product Code
CAC · Anesthesiology
Decision Date
Jan 4, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5830
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Haemonetics® Cardiovascular Perioperative Autotransfusion System (cardioPAT™) is indicated for use to salvage red blood cells from blood lost intraoperatively and postoperatively during cardiovascular surgical procedures, where the expected rate of processing of salvaged blood and fluid aspirated from the surgical site is less than or equal to two liters per hour. Autotransfusion is indicated for patients who meet at least one of the following criteria: - The patient is expected to lose sufficient blood in the perioperative period, so as to require red blood cell transfusion, and autotransfusion will likely reduce or eliminate the need for allogeneic blood transfusion. - Religious beliefs cause the patient to refuse allogeneic transfusion, but accept autologous transfusion. - Compatible allogeneic blood is not available. - The patient is unable to donate sufficient quantities of autologous blood prior to surgery to adequately cover the anticipated transfusion requirement. - The patient or physician prefers perioperative autotransfusion rather than preoperative autologous donation or transfusion of allogeneic blood.
Device Story
Cardiovascular Perioperative Autotransfusion System (cardioPAT) collects/processes RBCs lost during/after cardiovascular surgery. System comprises electromechanical device and sterile single-use disposable set; mounts on IV pole. Used in OR/recovery room by clinical staff. Processes salvaged blood/fluid at rates ≤ 2 L/hr. Recycles patient's own blood to reduce/eliminate need for allogeneic transfusion. Benefits include conservation of blood resources and avoidance of allogeneic transfusion risks. Healthcare providers use salvaged RBCs for patient reinfusion based on clinical need.
Clinical Evidence
Bench testing only. Study TP- & TR-DIS-02028 evaluated cardioPAT performance using simulated use conditions with blood pools of varying hematocrit (5%, 15%, 40%). Primary endpoints included mean percent washout (heparin, albumin, hemoglobin) and mean percent RBC recovery. Results showed mean washout of 98.9% (heparin), 99.1% (albumin), 97.3% (hemoglobin), and 85.9% RBC recovery. Comparison to predicate OrthoPAT data showed <2% difference in washout and <3% difference in RBC recovery, confirming performance equivalence.
Technological Characteristics
Electromechanical autotransfusion apparatus; portable IV-pole mounted form factor. Consists of reusable electromechanical unit and sterile single-use disposable set. Operates by collecting and processing salvaged blood via centrifugation/washout. No specific material standards or software architecture details provided.
Indications for Use
Indicated for patients undergoing cardiovascular surgery with anticipated perioperative blood loss requiring RBC transfusion. Applicable when allogeneic blood is unavailable, refused for religious reasons, or when preoperative autologous donation is insufficient/not preferred. Contraindications: None stated.
Regulatory Classification
Identification
An autotransfusion apparatus is a device used to collect and reinfuse the blood lost by a patient due to surgery or trauma.
Predicate Devices
Haemonetics® Orthopedic Predicate Perioperative Autotransfusion System (OrthoPAT®) (K962475)
Related Devices
K053000 — HAEMONETICS CARDIOVASCULAR PERIOPERATIVE AUTOTRANSFUSION SYSTEM (CARDIOPAT) · Haemonetics Corp. · Nov 21, 2005
K992723 — ORTHOPEDIC PERIOPERATIVE AUTOTRANSFUSION (ORTHOPAT) SYSTEM · Transfusion Technologies Corp. · Oct 18, 1999
K122262 — ORTHOPAT ADVANCE PERIOPERATIVE AUTOTRANSFUSION SYSTEM · Haemonetics Corp. · Oct 25, 2012
K962475 — ORTHOPEDIC PERIOPERATIVE AUTOTRANSFUSION (ORTHOPAT) SYSTEM · Transfusion Technologies Corp. · Feb 20, 1997
K982660 — ORTHOPEDIC PERIOPERATIVE AUTOTRANSFUSION (ORTHOPAT) SYSTEM · Transfusion Technologies Corp. · Mar 30, 1999
Submission Summary (Full Text)
{0}------------------------------------------------
K043/27
# Section 10: 510(k) Summary
| Statement | This summary of 510(k) safety and effectiveness information is being submitted<br>in accordance with the requirements of SMDA 1990 and 21 CFR 807.92 |
|--------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter | Haemonetics Corporation<br>400 Wood Road<br>Braintree, MA. 02184-9114 |
| Company<br>Contact | Gabriel J. Muraca, Jr.<br>RA Project Manager<br>Haemonetics Corporation<br>355 Wood Rd.<br>Braintree, MA. 02184-9114 |
| Device Name | Proprietary Name:<br>Cardiovascular Perioperative Autotransfusion System<br>(cardioPATTM)<br>Common Name: Autotransfusion apparatus<br>Classification Name: Autotransfusion apparatus (74 CAC) |
#### The currently marketed predicate device is the Haemonetics® Orthopedic Predicate Perioperative Autotransfusion System (OrthoPAT®). Device
{1}------------------------------------------------
### The cardioPAT system is designed to provide perioperative autotransfusion for patients undergoing cardiovascular surgery. The system consists of an Device electromechanical device and a sterile single-use disposable set, which together Description collect and process red blood cells (RBCs) lost during and after surgery. It is a small portable system which mounts on an IV pole. It is designed to be used in the operating room to recycle blood lost during cardiovascular surgical procedures and in the recovery room to recycle blood lost after surgery, where the expected rate of processing of salvaged blood and fluid aspirated from the surgical site is less than or equal to two liters per hour.
#### The Haemonetics® Cardiovascular Perioperative Autotransfusion System Indications for (cardioPAT™) is indicated for use to salvage red blood cells from blood lost Use intraoperatively and postoperatively during cardiovascular surgical procedures, where the expected rate of processing of salvaged blood and fluid aspirated from the surgical site is less than or equal to two liters per hour.
Autotransfusion is indicated for patients who meet at least one of the following criteria:
- The patient is expected to lose sufficient blood in the . perioperative period, so as to require red blood cell transfusion, and autotransfusion will likely reduce or eliminate the need for allogeneic blood transfusion.
- Religious beliefs cause the patient to refuse allogeneic . transfusion, but accept autologous transfusion.
- Compatible allogeneic blood is not available. .
- The patient is unable to donate sufficient quantities of . autologous blood prior to surgery to adequately cover the anticipated transfusion requirement.
- The patient or physician prefers perioperative autotransfusion . rather than preoperative autologous donation or transfusion of allogeneic blood.
{2}------------------------------------------------
#### The quality of salvaged red blood cells returned to the donor during autotransfusion procedures with both the current OrthoPAT and modified Performance cardioPAT systems is acceptable with respect to measured markers.
The following table is a summary of Tables 1, 2 and 3 from the test report TP-& TR-DIS-02028, Exhibit 5. It shows the mean percent washout and the mean percent red blood cell recovery for pools of different hematocrit blood that were processed under simulated use conditions by the cardioPAT system.
| Mean<br>Washout | | | Mean<br>Red Blood Cell<br>Recovery | |
|-----------------|-----------------------------|-----------------------------|------------------------------------|-------|
| Hematocrit | Supernate<br>Heparin<br>(%) | Supernate<br>Albumin<br>(%) | Supernate<br>Hemoglobin<br>(%) | (%) |
| 5% | 99.88 | 99.90 | 99.38 | 77.83 |
| 15% | 99.68 | 99.80 | 98.18 | 88.77 |
| 40% | 97.19 | 97.67 | 94.37 | 91.00 |
| Mean | 98.9 | 99.1 | 97.3 | 85.9 |
## Table 1: Data Summary* for cardioPAT
*From TP- & TR-DIS-02028
Table 2 compares the combined means of the cardioPAT results from Table 1 and compares them to the combined mean test results previously obtained, using similar pools of blood, which were processed on the predicate OrthoPAT system (Test Report V0007, K962475, provided in Exhibit 5).
### Table 2: Combined Pool Results Comparison
| | Mean<br>Washout | | | Mean<br>Red Blood Cell<br>Recovery |
|-----------------------------|-----------------------------|-----------------------------|----------------------------------|------------------------------------|
| Data Source | Supernate<br>Heparin<br>(%) | Supernate<br>Albumin<br>(%) | Supernatant<br>Hemoglobin<br>(%) | (%) |
| OrthoPAT<br>V0007 (SW 3.0c) | 98.9 | 98.9 | 98.4 | 83.8 |
| cardioPAT<br>TP-DIS-02028 | 98.9 | 99.1 | 97.3 | 85.9 |
| % Difference | + 0.0 | + 0.2 | - 1.1 | + 2.1 |
{3}------------------------------------------------
This comparison shows that there is less than 2% difference in combined mean percent (%) washout values and less than 3% difference between the mean percent red blood cell recovery values. Because these percent differences are very small, they are not considered clinically significant. The cardioPAT system performance test results are within the expected operating efficiencies of the predicate OrthoPAT device, which demonstrates performance equivalence.
### Substantial Equivalence
The substantial equivalence of the cardioPAT System is supported by its similarities in intended use, technological characteristics, and performance as compared to the currently marketed OrthoPAT system.
Gabriel J. Munoz Jr.
Gabriel J. Muraca, Jr. RA Project Manager Haemonetics Corporation
Date: November 10, 2004
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles three abstract human figures or lines, possibly representing people or services provided by the department.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN - 4 2005
Haemonetics Corporation c/o Mr. Gabriel J. Muraca, Jr. RA Project Manager 355 Wood Road Braintree, MA 02184-9114
K043127 Re:
K043127
Haemonetics® Cardiovascular Perioperative Autotransfusion System (cardioPATTM) Regulation Number: 21 CFR 868.5830 Regulation Name: Autotransfusion Apparatus Regulatory Class: Class II (two) Product Code: CAC Dated: November 10, 2004 Received: November 12, 2004
Dear Mr. Muraca:
We have reviewed your Section 510(k) premarket notification of intent to market the device indication we have reviewed your Section 9 ro(if) presidentially equivalent (for the indications felerenced and nave determined the are are are are are and one sparketed in interstate for use stated in the cherosure) to regars actment date of the Medical Device Amendments, or to commerce prior to May 28, 1976, the enational with the provisions of the Federal Food. Drug.
devices that have been reclassified in accordance with the provisions of any inst devices that have been recalismod in asses approval of a premarket approval application (PMA). and Cosment Act (Act) that do not require appt to the general controls provisions of the Act. The You may, merelore, market the device, becarements for annual registration, listing of general controls provisions of the rise labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (500 abor of the Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may be found in the Oous or ents concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Gabrie! J. Muraca, Jr.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r lease be advised that i Drivisation that your device complies with other requirements of the Act that I Dri has made a actives and regulations administered by other Federal agencies. You must of any reactal statutes and regaranents ancluding, but not limited to: registration and listing (21 Comply with an the 11et 31equirements, and manufacturing practice requirements as set CFN in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic forth in the quality systems (QD) roggistions 531-542 of the Act); 21 CFR 1000-1050. product lastation oona of pro recom (2) .
This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will anow yourse ough finding of substantial equivalence of your device to a legally premarket notification: "The PDF interessification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please rryou desire specific at (301) 594-4646. Additionally, for questions on the comact the Office of Compilance and evice, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to (301) 594-4057. Also, prease nets roganison way obtain. Other general information on your premarket hourseation - (2) of the abe obtained from the Division of Small Manufacturers, Itsponsional and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Donna R. Vochner
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
#### K043127 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name:
# Haemonetics® Cardiovascular Perioperative Autotransfusion System (cardioPATTM)
### Indications for Use:
The Haemonetics® Cardiovascular Perioperative Autotransfusion System (cardioPAT™) is I IIe Haemonches® Cardio resource > errells from blood lost intraoperatively and postoperatively muring cardiovascular surgical procedures, where the expected rate of processing of salvaged during cardio rasoural bargical site is less than or equal to two liters per hour.
Autotransfusion is indicated for patients who meet at least one of the following criteria:
- The patient is expected to lose sufficient blood in the perioperative period, so as . to require red blood cell transfusion, and autotransfusion will likely reduce or eliminate the need for allogeneic blood transfusion.
- Religious beliefs cause the patient to refuse allogeneic transfusion, but accept ● autologous transfusion.
- Compatible allogeneic blood is not available. .
- The patient is unable to donate sufficient quantities of autologous blood prior to . surgery to adequately cover the anticipated transfusion requirement.
- The patient or physician prefers perioperative autotransfusion rather than . preoperative autologous donation or transfusion of allogeneic blood.
and/or X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) ______________________________________________________________________________________________________________________________________________________________________
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
R. bechner
(Division Sign-Off)
Division of Cardiovascular Devices
510(K) Number_KO43127
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.