K173205 · O-Two Medical Technologies, Inc. · BZR · Aug 29, 2018 · Anesthesiology
Device Facts
Record ID
K173205
Device Name
Equinox Advantage
Applicant
O-Two Medical Technologies, Inc.
Product Code
BZR · Anesthesiology
Decision Date
Aug 29, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5330
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Equinox® Advantage Nitrous Oxide/Oxygen Analgesic Gas Mixing and Delivery System is indicated for administering an adjustable mixture of Nitrous Oxide analgesic gas and Oxygen, on demand, to a conscious, spontaneously breathing patient.
Device Story
Portable, pneumatically powered device for on-demand delivery of adjustable Nitrous Oxide/Oxygen analgesic gas mixtures. Inputs: N2O and O2 cylinders via pressure regulators. Operation: User selects O2 concentration (25-100%) via knob; gas delivery triggered by patient inspiratory effort. Includes Oxygen Enrichment function (30 L/min 100% O2) for circuit flushing. Used in pre-hospital (ambulance) and in-hospital (ER, Labor and Delivery) settings by clinicians. Features integrated pneumatic alarm system (audible/visual) for low input pressure (<40 PSI) and automatic shut-off if O2 input falls below 35 PSI. Output: Analgesic gas mixture to patient via face mask or mouthpiece. Benefits: Provides adjustable pain relief while preventing hypoxic mixtures; enhances patient safety through fail-safe mechanisms.
Clinical Evidence
Bench testing only. Comparative performance testing against predicate (K113687) evaluated inspiratory/expiratory resistance, volume-time, pressure-time, and flow-time waveforms, alarm characteristics, and gas concentration accuracy. Additional bench testing conducted per ISO 11195 and standards for electrical safety and essential performance.
Technological Characteristics
Pneumatically powered gas mixer. Materials: Brass (nickel-plated), anodized aluminum, stainless steel, ABS, PC/ABS, polycarbonate, silicone, polypropylene, PVC, polyurethane. Connectivity: None. Dimensions/Form: Portable. Standards: ISO 11195:1995, ISO 80601-2-13:2011. Sensing: Pressure sensors for gas inputs. Actuation: Demand valve triggered by patient inspiratory effort.
Indications for Use
Indicated for conscious, spontaneously breathing patients requiring pain relief. Contraindicated in patients with hypersensitivity to medication, head injuries with impaired consciousness, maxillofacial injuries, pneumothorax (artificial, traumatic, or spontaneous), air embolism, middle ear occlusion/infection, decompression sickness, or abdominal distension/intestinal obstruction.
Regulatory Classification
Identification
A breathing gas mixer is a device intended for use in conjunction with a respiratory support apparatus to control the mixing of gases that are to be breathed by a patient.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white. Underneath the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.
O-Two Medical Technologies Inc. David Zhang QA Manager 45A Armthorpe Road Brampton, Ontario L6T 5M4
#### Re: K173205
Trade/Device Name: Equinox Advantage Nitrous Oxide/Oxygen Analgesic Gas Mixing and Delivery System Regulation Number: 21 CFR 868.5330 Regulation Name: Breathing Gas Mixer Regulatory Class: Class II Product Code: BZR Dated: July 30, 2018 Received: August 1, 2018
### Dear David Zhang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Todd D. Courtney -S
Tina Kiang, Ph.D. for Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K173205
#### Device Name
Equinox® Advantage Nitrous Oxide/Oxygen Analgesic Gas Mixing and Delivery System
Indications for Use (Describe)
The Equinox® Advantage Nitrous Oxide/Oxygen Analgesic Gas Mixing and Delivery System is indicated for administering an adjustable mixture of Nitrous Oxide analgesic gas and Oxygen, on demand, to a conscious, spontaneously breathing patient.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: overline;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510 (k) Summary
| Submitter's Name & Address: O-Two Medical Technologies | |
|---------------------------------------------------------|----------------------------|
| | 45A Armthorpe Road |
| | Brampton, Ontario, L6T 5M4 |
| | Canada |
| | Tel: 905-792-6896 |
| Official Contact: | David Zhang |
|----------------------------|------------------------------------------------------------------------------------------------------------------------------|
| Date of Revision: | 2018-08-17 |
| Proprietary or Trade Name: | Equinox <sup>®</sup> Advantage Nitrous Oxide/Oxygen Analgesic Gas Mixing and Delivery System |
| Common/Usual Name: | Mixer, Breathing Gases, Anesthesia Inhalation |
| Classification Name: | Breathing Gas Mixer (21 CFR 868.5330, product code: BZR) |
| Device Class: | Class II |
| Classification Panel: | Anesthesiology |
| Predicate Devices: | Equinox <sup>®</sup> Relieve <ul><li>Manufactured by O-Two Medical Technologies Inc.</li><li>510(k) number K113687</li></ul> |
#### Device Description:
Equinox® Advantage Nitrous Oxide/Oxygen Analgesic Gas Mixing and Delivery System is a portable pneumatically powered device intended to administer an adjustable mixture of Nitrous Oxide analgesic gas and Oxygen mixture to provide pain relief, on demand, to a conscious, spontaneously breathing, patient.
The device provides two input connection with nitrous oxide and oxygen cylinders through pressure regulators. The device has a control knob for turning ON or OFF the device. When it is turned ON, the output of N2O/O2 gas mixture will only be administered by an inspiratory effort by the patient.
{5}------------------------------------------------
A selection knob is used to select desired N20/02 gas mixture. The adjustable range is from 25% O2 (NJO 75%) to 100% O2 (N3O 0%), with the minimum of 25% oxygen output to eliminate the possibility of delivering a hypoxic mixture.
The Oxygen Enrichment Function is designed to provide 30 L/min of 100% oxygen in order to flush any residual gas from the patient circuit following the completion of the patient treatment.
The built-in pneumatically powered alarm system will generate both visual and audible alarms should either nitrous oxide or oxygen input fall below 40 PSI, and the device will be automatically shut off should oxygen input falls below 35 PSI.
The main accessories for the proposed device are a disposable Patient Circuit and a Mouth Piece or a universal Face Mask.
#### Indications for Use:
The Equinox® Advantage Nitrous Oxide/Oxygen Analgesic Gas Mixing and Delivery System is indicated for administering an adjustable mixture of Nitrous Oxide analgesic gas and Oxygen, on demand, to a conscious, spontaneously breathing patient.
#### Patient Population:
Conscious, spontaneous breathing patients requiring pain relief.
### Contraindications:
- Hypersensitivity to the medication
- Head injuries with impaired consciousness ●
- Maxillofacial injuries ●
- Artificial, traumatic or spontaneous pneumothorax ●
- Air embolism ●
- Middle ear occlusion, ear infection ●
- Decompression sickness ●
- Abdominal distension / intestinal obstruction ●
NOTE: Nitrous Oxide/Oxygen (N20/02) mixtures must never be used in any condition where air is trapped in the body and expansion (up to 3x original size) would be dangerous. For example, it will exacerbate pneumothorax and increase pressure from any intracranial air. Air in any other cavities such as the sinuses, middle ear and gut may also expand.
#### Environment of Use:
Pre-hospital (ambulance) use and in-hospital use (ER. Labor and Delivery etc.)
{6}------------------------------------------------
| Characteristic | Proposed<br>Equinox® Advantage | Predicate K113687<br>Equinox® Relieve | Substantial<br>Equivalence |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|
| Intended Use | Administer an adjustable mixture of<br>N2O/O2, on demand, to a conscious,<br>spontaneously breathing patient | Administer 50/50% N2O/O2 mixture, on<br>demand, to a conscious, spontaneously<br>breathing patient | Equivalent |
| Environments<br>of use | Pre-hospital use and in-hospital<br>environment | Pre-hospital use and in-hospital<br>environment | Yes |
| Patient<br>population | Conscious, spontaneous breathing<br>patients requiring pain relief | Conscious, spontaneous breathing<br>patients requiring pain relief | Yes |
| Contra-<br>Indications | Hypersensitivity to the medication Head injuries with impaired<br>consciousness Maxillofacial injuries Artificial, traumatic or spontaneous<br>pneumothorax Air embolism Middle ear occlusion, ear infection Decompression sickness Abdominal distension / intestinal<br>obstruction | Hypersensitivity to the medication Head injuries with impaired<br>consciousness Maxillofacial injuries Artificial, traumatic or spontaneous<br>pneumothorax Air embolism Middle ear occlusion, ear infection Decompression sickness Abdominal distension / intestinal<br>obstruction | Yes |
## Comparative table - Intended use:
# Comparative table - Operating principle:
| Characteristic | Proposed<br>Equinox® Advantage | Predicate K113687<br>Equinox® Relieve | Substantial<br>Equivalence |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|
| Gas input and<br>regulating | 50- 70 PSI Oxygen and Nitrous Oxide<br>regulated by internal regulators | 50- 70 PSI Oxygen and Nitrous Oxide<br>stabilized by internal regulators | Yes |
| Gas mixing | Oxygen and Nitrous Oxide are<br>equalized and mixed at variable ratios | Oxygen and Nitrous Oxide are<br>equalized and mixed at 1:1 ratio | Similar-<br>(mixing ratio is<br>adjustable) |
| Gas delivery | Gas mixture delivery is proportional to<br>patient's demand. | Gas mixture delivery is proportional to<br>patient's demand. | Yes |
| Monitoring<br>and alarms | The input pressure from each gas input<br>is measured by pressure sensors and<br>Audible/Visual alarms activated should:<br>- N2O input is below 40 PSI;<br>- O2 input is below 40 PSI | The input pressure from each gas input<br>is measured by pressure sensors and<br>Audible/Visual alarms activated should:<br>- N2O input is below 40 PSI;<br>- O2 input is below 40 PSI | Yes |
| Safety features | In the event of O2 supply failure,<br>1.<br>the protection circuit will shut off<br>the device: | 1.<br>In the event of O2 supply failure,<br>the protection circuit will shut off<br>the device: | Yes |
| | 2.<br>In the event of N2O supply failure,<br>Oxygen will be supplied as per ISO<br>11195:1995 Sec.12.4 | 2.<br>In the event of N2O supply failure,<br>the protection circuit will shut off<br>the device | Similar-<br>(Modification<br>made as per ISO<br>11195 Sec.12.4) |
{7}------------------------------------------------
# Comparative table- Materials
| Materials | Proposed<br>Equinox® Advantage | Predicate K113687<br>Equinox® Relieve | Substantial<br>Equivalence |
|---------------|----------------------------------------------------------------------|----------------------------------------------------------------------|----------------------------|
| Metal parts | Brass-Nickel plated, Anodized<br>Aluminum, Stainless Steel | Brass-Nickel plated, Anodized<br>Aluminum, Stainless Steel | Yes |
| Plastic parts | ABS, PC / ABS blends, Polycarbonate,<br>Silicone, Polypropylene, PVC | ABS, PC / ABS blends, Polycarbonate,<br>Silicone, Polypropylene, PVC | Yes |
| Tubing | Polyurethane | Polyurethane | Yes |
# Comparative table- Safety features
| Characteristic | Proposed<br>Equinox® Advantage | Predicate K113687<br>Equinox® Relieve | Substantial<br>Equivalence |
|-----------------|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|
| Safety features | - CGA gas specific connection to prevent<br>misconnection of gas supply; | - CGA gas specific connection to prevent<br>misconnection of gas supply; | Yes |
| | - Activated only by patient inspiratory<br>effort; | - Activated only by patient inspiratory<br>effort; | Yes |
| | - Audible & visual alarms; | - Audible & visual alarms; | Yes |
| | - Input Gas Fail Safe: Shut off the device<br>if supply of O2 dropped below 35 PSI | - Input Gas Fail Safe: Shut off the device<br>if supply of either N2O or O2 dropped<br>below 35 PSI | Similar-<br>(Modification<br>made as per ISO<br>11195 Sec.12.4) |
## Comparative table- Technological characteristics/specifications of performance
| Characteristic | Proposed<br>Equinox® Advantage | Predicate K113687<br>Equinox® Relieve | Substantial<br>Equivalence |
|---------------------------------|--------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|------------------------------------------------|
| Input gas | O2 and N2O | O2 and N2O | Yes |
| Input pressure | 50 to 70 PSI | 50 to 70 PSI | Yes |
| Output mixture<br>concentration | Adjustable Oxygen: 25%-100% | Preset 50/50% (V/V) O2 / N2O | Equivalent-<br>(mixing ratio is<br>adjustable) |
| Oxygen<br>Enrichment | 30 L/min oxygen | Not available | New feature |
| Alarms | Low input audible and visual alarms<br>for both gases if either or both inputs<br>dropped below 40 PSI | Low input audible and visual alarms<br>for both gases if either or both inputs<br>dropped below 40 PSI | Yes |
{8}------------------------------------------------
| Safety features | - Activated only by patient inspiratory effort;<br>- CGA connection to prevent misconnection of gas supply;<br>- Audible & visual alarms;<br>- Power Fail Safe: Shut off the device if supply of O2 drops below 35 PSI | - Activated only by patient inspiratory effort;<br>- CGA connection to prevent misconnection of gas supply;<br>- Audible & visual alarms;<br>-Power Fail Safe: Shut off the device if supply of O2 drops below 35 PSI<br>-Power Fail Safe: Shut off the device if supply of N2O drops below 35 PSI | Similar-<br>(Modification made as per ISO 11195 Sec.12.4) | |
|-----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|-----|
| | Patient support mode | Demand valve | Demand valve | Yes |
| | Peak Flow on Demand (L/min) | 120 minimum | 120 minimum | Yes |
| | Accessories | - Disposable patient circuit with 30 mm Scavenging port and Universal PVC Face Mask;<br>- Oxygen supply hose;<br>- Nitrous Oxide supply hose | - Disposable patient circuit with 30 mm Scavenging port and Universal PVC Face Mask;<br>- Oxygen supply hose;<br>- Nitrous Oxide supply hose | Yes |
### Substantially equivalence to the predicate devices:
The proposed device (Equinox® Advantage) is equivalent to the predicate device in Intended use, Patient populations, Environment of use and Contra-indications.
It not only incorporates parts which are identical to the Equinox® Relieve/K113687, but also use the same materials as the Equinox® Relieve/K113687.
Furthermore, Equinox® Advantage uses the identical operating principle as the Equinox® Relieve.
Safety features and range of performance specifications of the Equinox® Advantage are also found equivalent to those on the Equinox® Relieve/K113687.
The proposed Equinox® Advantage is substantially equivalent to the predicate Equinox® Relieve/ K113687 with the following minor modifications:
| Modification | Equinox Relieve | The proposed | Standard |
|--------------|-----------------|-------------------|------------|
| made | (K113687 | Equinox Advantage | compliance |
{9}------------------------------------------------
| Mixing<br>system | Gas mixing<br>ratio | Fixed (Oxygen 50%) | Modified to adjustable mixing ratio<br>(Oxygen 25% - 100%) | ISO 11195:1995<br>Sec.3.1 |
|------------------------------|--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------|
| Delivery<br>system | Oxygen<br>enrichment | Not available | Oxygen enrichment (30 L/min)<br>added | ISO 80601-2-<br>13:2011 Sec.<br>201. 101.8 |
| Safety/<br>Warning<br>system | Output failure<br>shutoff | When O2 output pressure is<br>much lower than N2O, Shut<br>Off Sensor will shut off the<br>device to prevent delivering a<br>hypoxic mixture. | "Output failure shutoff" function<br>removed to enable adjustable mixing<br>ratio.<br>Minimum 25% oxygen setting<br>prevents delivering a hypoxic<br>mixture. | ISO 11195:1995<br>Sec.11.1 |
| | Power failure<br>shutoff-<br>Nitrous oxide | If the Nitrous Oxide input<br>pressure drops to 35 PSI, the<br>Demand Valve and the Mixer<br>will be shut off. | "Nitrous oxide failure shutoff"<br>function removed.<br>When Nitrous Oxide input drops to<br>35 PSI, with low pressure N2O<br>alarm on, the proposed device<br>continues to operate with higher<br>oxygen output. If there is no N2O<br>input, the device will deliver 100%<br>oxygen. | ISO 11195:1995<br>Sec. 12.4 |
As all the modifications made, meet the requirements of the current technical standards with safety measures also in place to prevent from introducing new potential risks, there is not any significant differences that raise different questions of safety or effectiveness of the intended device as compared to the predicate devices.
#### Summary of Performance Testing:
A comparative performance testing was performed on Equinox® Advantage and Equinox® Relieve. The result of the test demonstrates substantial equivalence of the proposed to the predicate.
The summary of the test results follows:
- Both mixers delivered equivalent Inspiratory Resistance, Expiratory Resistance, volume-time, pressure-time and flow- time wave forms under the same ventilation mode;
- Both mixers showed equivalent "Low Oxygen supply" and "Low Nitrous Oxide supply" alarm characteristics;
- Both mixers showed equivalent "Oxygen Failure Shutoff" characteristic; -
- When Equinox® Advantage is set at 50% Oxygen mixing ratio, both units delivered equivalent Oxygen concentration and Nitrous Oxide concentration.
We have also performed bench testing as per ISO 11195 and other applicable standards with respect to Electrical Safety & Essential performance of the proposed device and its accessory.
#### Conclusion:
{10}------------------------------------------------
The results of the above comparative performance testing as well as bench testing demonstrate that the proposed Equinox® Advantage is as safe, as effective and performs as well as the legally marketed predicate devices - Equinox® Relieve (K113687).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.