GMX Series Medical Air/Oxygen Blender is designed to dispense a continuous blend of Medical Air and Oxygen via outlet ports to infant, pediatric, and adult patients.
Device Story
GMX Series Medical Air/Oxygen Blender is a pneumatically powered device used in institutional settings by trained personnel under physician direction. It accepts medical air and oxygen via inlet connectors with filters and check valves. A two-stage pressure balancing module equalizes gas source pressures; a proportioning module mixes gases based on the selected oxygen percentage. Output is a continuous blend of air and oxygen delivered to patients. A bypass poppet mechanism triggers an audible alarm if inlet pressure differentials exceed specified limits or if a gas source is disconnected, while simultaneously routing the higher-pressure gas to the outlet. A bleed valve is used for low-flow accuracy. The device provides clinicians with precise oxygen concentration control, aiding in respiratory support and patient oxygenation management.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including performance verification, risk analysis, and compliance with ISO 11195, USP <788>, ASTM D 5466-01, and IEC 60601-1 mechanical testing requirements.
Technological Characteristics
Pneumatically powered gas mixer; no electrical components. Models: GMX30U-AIR/O2 and GMX120U-AIR/O2. Features two-stage pressure balancing module, proportioning module, bypass poppet alarm, and bleed valve. Connectors: NIST or DISS. Complies with ISO 11195:1995. Operating temperature: 10°C-40°C. Accuracy: ±3%.
Indications for Use
Indicated for infant, pediatric, and adult patients requiring a continuous blend of medical air and oxygen.
Regulatory Classification
Identification
A breathing gas mixer is a device intended for use in conjunction with a respiratory support apparatus to control the mixing of gases that are to be breathed by a patient.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around the perimeter. In the center of the seal is a stylized image of three human profiles facing to the right, with flowing lines beneath them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 19, 2015
Gentec (Shanghai) Corporation % Mike Gu Regulatory Manager Guangzhou Osmunda Medical Device Consulting Co., Ltd 7th Floor, 982 Congyun Road Baiyun District, 510420 China
Re: K142291
Trade/Device Name: GMX Series Medical Ait/oxygen Blender Regulation Number: 21 CFR 868.5330 Regulation Name: Breathing Gas Mixer Regulatory Class: Class II Product Code: BZR Dated: February 13, 2015 Received: February 18, 2015
Dear Mr. Gu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Page 2 - Mr. Mike Gu
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
*Tejashri Purohit-Sheth, M.D.*
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K142291
Device Name GMX Series Medical Air/Oxygen Blender
Indications for Use (Describe)
The GMX Series Medical Air/Oxygen Blender is designed to dispense a continuous blend of Medical Air and Oxygen via outlet ports to infant, pediatric, and adult patients.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Date: | May 18,2015 | | | | |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----|-------|-----|----------|
| Submitter: | GENTEC (SHANGHAI) CORPORATION<br>No. 1988 Yushu Road, Songjiang District, Shanghai,<br>201699 ,China | | | | |
| Primary Contact Person: | Mike Gu<br>Regulatory Manager<br>Guangzhou Osmunda Medical Device Consulting Co., Ltd<br>Tel: +86-20-62321333<br>Fax: +86-20-86330253 | | | | |
| Secondary Contact<br>Person: | Lifeng Miao<br>Quality Engineer<br>GENTEC (SHANGHAI) CORPORATION Technology Co., Ltd<br>Tel: +86-21- 67727123-149<br>Fax: +86-21-67727176-1759 | | | | |
| Device Trade Name: | GMX Series Medical Air/Oxygen Blender | | | | |
| Common/Usual Name: | Oxygen Blender | | | | |
| Classification Name:<br>Product Code: | breathing<br>BZR | gas | mixer | CFR | 868.5330 |
| Predicate Device(s): | Precision Blender (K053232) | | | | |
| Device Description: | The GMX Series Medical Air/Oxygen blender is a restricted<br>medical device intended for use by qualified and trained<br>personnel under the direction of a physician in institutional<br>environments where delivery of air/oxygen mixtures is<br>required. | | | | |
| | The GMX Series Medical Air/Oxygen blender includes two<br>models: GMX30U-AIR/O2, GMX120U-AIR/O2. | | | | |
| | The GMX Series Medical Air/Oxygen blender is designed,<br>tested, and will be manufactured in accordance with the<br>following standard: | | | | |
| | ISO 11195 Gas mixers for medical use - Stand-alone gas<br>mixers 1995. | | | | |
| Intended Use: | GMX Series Medical Air/Oxygen Blender is designed to<br>dispense a continuous blend of Medical Air and Oxygen via<br>outlet ports to infant, pediatric, and adult patients. | | | | |
1 / 4
{5}------------------------------------------------
- The medical oxygen/air enters the two-stage pressure Technology: balancing module through the inlet connectors, each containing a filter and check valves. In the balancing module, the pressures of both gas sources are adjusted and equalized. Then the gases flow into the proportioning module and are mixed according to the oxygen percentage selected.
When the difference in pressure between the two inlet gasses exceeds the prescribed value ( or one gas source is disconnected), the bypass poppet which blocks the flow of both gases will move to one side and create a path for the high pressure gas to flow into the alarm channel and generate the audible alarm. Meanwhile, the gas with the higher pressure will flow directly to the outlet port.
When the desired output flow is below the specified range, the bleed valve needs to be activated to ensure the accuracy of oxygen concentration.
Determination of The proposed device has the same technological Substantial Equivalence: characteristics (i.e. design, energy source) as the predicate device. Both devices accept room air and 100% oxygen gas and, via mechanical means, adjust the molar fraction of oxygen gas in the room air, and blended gas is delivered to the patient, they have identical specifications for the delivery of blended gas and similar specifications for the audible alarm (the pressure differential specification is identical). Both devices are pneumatically powered (supply source, not included) and they produce an audible alarm when the difference in pressure between the two inlet gasses exceeds the prescribed value, or one gas source is disconnected. Neither device is intended to be used sterile, nor do either device contain electrical components.
Please see Table below for more details.
| Item | Precision Blender | GMX Series<br>Air/Oxygen<br>Blender |
|------------|----------------------------------|-------------------------------------|
| Dimensions | 4.9"*2.3"*4.1"<br>(125*57*104mm) | 137.5*140*165mm |
| Weight | 1.34 kg | 1.5Kg |
- 2 / 4
{6}------------------------------------------------
| Oxygen %<br>Range | 21-100% | 21-100% |
|---------------------------------------|-------------------------------------------------------|----------------------------------------------------|
| Accuracy | $\pm$ 3% | $\pm$ 3% |
| Energy Source | Air Supply | Air Supply |
| Supply<br>Pressure | 30-75 Psi | 30-72.5 psi |
| Max Flow<br>(GMX120U-<br>AIR/O2) | ≥120 LPM | ≥120 LPM |
| Pressure Drop<br>(GMX120U-<br>AIR/O2) | ≤6psi at 50Psi<br>inlet<br>Pressure and<br>40lpm flow | ≤6psi at 50Psi<br>inlet pressure and<br>40lpm flow |
| Alarm/Bypass<br>Reset | ≤6psi | ≤6psi |
| Alarm<br>Intensity | 0.3meter≥80db | 1foot≥80db |
| Operating<br>Temperature | 15°C-40°C | 10°C-40°C |
| Connector | NIST or DISS | NIST or DISS |
| Max Flow<br>(GMX30U-<br>AIR/O2) | ≥30 LPM | ≥30LPM |
| Pressure Drop<br>(GMX30U-<br>AIR/O2) | ≤6psi at 50Psi<br>inlet<br>Pressure and<br>10lpm flow | ≤6psi at 50Psi<br>inlet pressure and<br>10lpm flow |
The following quality assurance measures were applied to the development of the system:
- Risk Analysis ●
- Requirements Reviews
- Design Reviews
- . Raw materials verification
- Final acceptance testing (Validation) .
- Performance testing (Verification) ●
3 / 4
{7}------------------------------------------------
## Summary of Non-Clinical Tests:
The Medical Air/Oxygen Blender complies with voluntary standards :
- -ISO 11195 Gas mixers for medical use – Stand-alone gas mixers 1995;
- -USP <788>:2012—Particulate Matter in Injections;
- ASTM D 5466-01(2007): Standard Test Method for - Determination of Volatile Organic Chemicals in Atmospheres (Canister Sampling Methodology);
- Mechanical Testing as per IEC 60601-1 Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance.
### Summary of Clinical Tests:
The subject of this premarket submission, Medical Air/Oxygen Blender, did not require clinical studies to support substantial equivalence.
The proposed device is Substantially Equivalent (SE) to the predicate device which is US legally market device.
- GENTEC (SHANGHAI) CORPORATION considers the GMX Conclusion: Series Medical Air/Oxygen Blender to be as safe, as effective, and performance is substantially equivalent to the predicate device(s).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.