RESUSCITATION UNIT 105

K103317 · Atom Medical Corporation · BZR · Apr 20, 2011 · Anesthesiology

Device Facts

Record IDK103317
Device NameRESUSCITATION UNIT 105
ApplicantAtom Medical Corporation
Product CodeBZR · Anesthesiology
Decision DateApr 20, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5330
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The Resuscitation Unit 105 PV and Resuscitation Unit 105 EV are intended to remove secretions from the neonatal / infant airways and to provide blended air and oxygen at a desired oxygen concentration in the range of 21-100% to a respiratory assist device connected to the patient such as a ventilator (including a manually operated one) and an oxygen tent.

Device Story

Resuscitation Unit 105 (models 105PV and 105EV) provides blended air/oxygen and suction for neonates/infants. Inputs: compressed air and oxygen; 105PV uses wall vacuum, 105EV uses internal venturi. Operation: manual blending of gases; pressure regulation via relief valve; monitoring via integral pressure gauge. Output: blended gas mixture to respiratory devices (tents, masks, ventilators) and suction for airway clearance. Used in NICU/delivery suites by trained clinicians. Output allows clinicians to manage oxygen concentration and airway pressure; benefits include improved respiratory support and secretion management.

Clinical Evidence

Bench testing only. Testing included flow accuracy, pressure gauge accuracy, O2 concentration (ISO 10651-5), and suction performance (ISO 10079-3: overfill, seal, aspiration, pressure, noise). Results demonstrated device performance met all pass/fail criteria and was equivalent to predicate devices.

Technological Characteristics

Gas-powered blender and suction unit. Features: pressure relief valve, pressure gauge, suction pressure gauge, dual outlets. 105PV: wall vacuum; 105EV: venturi vacuum. Gas sources: compressed air/oxygen (206-517 kPa). Blender range: 21-100% O2 (±3%). Flow rate: <15 lpm. Standards: ISO 10651-5 (resuscitators), ISO 10079-3 (suction). Mechanical/pneumatic operation; no software/electronics described.

Indications for Use

Indicated for neonates and infants requiring airway secretion removal or delivery of blended air/oxygen (21-100%) to respiratory assist devices (ventilators, oxygen tents, masks, cannulas). Prescription use only in hospitals, delivery suites, and NICUs.

Regulatory Classification

Identification

A breathing gas mixer is a device intended for use in conjunction with a respiratory support apparatus to control the mixing of gases that are to be breathed by a patient.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ .K103317 ## 510(k) Summary Page 1 of 8 5-Apr-11 APR 2 0 2011 | Atom Medical Corporation<br>3-18-15, Hongo, Bunkyo-ku<br>Tokyo, Japan 113-0033 | Tel - 011-81-3-3815-3632<br>Fax - 011-81-3-3812-3199 | |--------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Official Contact: | Tsuyoshi Sugino - Regulatory Affairs Manager | | Proprietary or Trade Name: | Resuscitation Unit 105 | | Common/Usual Name: | Air / oxygen blender | | Classification Name/Code: | BZR - Breathing gas mixer<br>CFR 868.5330 | | Device: | Resuscitation Unit 105EV<br>Resuscitation Unit 105PV | | Predicate Devices: | Bio-Med Devices_BMD_air/oxygen_blender_K925982<br>Precision Medical - Precision Blender - K053232<br>Atom Medical - V-505 Infant Warmer - K060461<br>GE Giraffe and Panda system - K070210 | ## Device Description: There are two (2) models of the Resuscitation Unit 105 air / oxygen blender with suction. - Resuscitation Unit 105PV . - . Resuscitation Unit 105EV They have the same basic design. Table 1 outlines the basic features and differences of each model. ## Indications for Use: The Resuscitation Unit 105 EV and Resuscitation Unit 105 PV are intended to remove secretions from the neonatal / infant airways and / or to provide blended air and oxygen at a desired oxygen concentration in the range of 21-100% to a respiratory assist device connected to the patient such as a ventilator (including a manually operated one) and an oxygen tent. Environment of Use: Hospitals, delivery suites, NICU {2}------------------------------------------------ ## 510(k) Summary Page 2 of 8 5-Apr-11 ## Table 1 Resuscitation Models 105PV and 105EV | Feature | Resuscitation Unit 105PV | Resuscitation Unit 105EV | |------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------| | Suction(Piping Type) | Yes | No | | Suction(Venturi Type) | No | Yes | | Oxygen / Air Blender | Yes | Yes | | Inlet Gas Sources | Oxygen compressed gas | Oxygen compressed gas | | | Air compressed gas | Air compressed gas | | | Suction compressed gas | | | Pressure Relief Valve | Yes | Yes | | Pressure Gauge | Yes | Yes | | Suction Pressure Gauge | Yes | Yes | | Outlet | Double outlets<br>(Outlet A/B) | Double outlets<br>(Outlet A/B) | | | Outlet A / B is for delivery of the air/oxygen gas mixture to a<br>variety of devices, i.e., oxygen tent, oxygen mask, nasal cannula,<br>manual resuscitator or manual ventilator, in-line humidifier jar,<br>etc. | | | | Airway Pressure tube and adapter can be connected at the patient<br>connected to measure and monitor the airway pressure which is<br>displayed on the integral pressure gauge as well as a pressure<br>regulating valve | | As outlined in Table 1, the Resuscitation Unit 105 EV and Resuscitation Unit 105 PV are identical except for the source for vacuum. Model 105PV is from a wall vacuum source and Model 105EV is from an internal venturi. Table 2 compares the Resuscitation Unit 105 EV and Resuscitation Unit 105 PV to the predicates: - Atom V-505 Infant warmer with optional oxygen / suction module (K060461) . - . Bio-Med Devices – BMD air/oxygen blender (K925982) a blender which is incorporated into the proposed devices - . Precision Medical – Precision Blender (K053232) which has the indications for use and product classification we are seeking. Table 3 compares the Resuscitation Unit 105 EV and Resuscitation Unit 105 PV to the predicates Atom V-505 Infant warmer with optional oxygen / suction module (K060461) and the GE Giraffe and Panda system (K070210) which incorporates suction function and features for the intended patient population and environment. {3}------------------------------------------------ 510((k) Summary S-Apr-11 # able of the Similarities and Differences to the Predicates for Air / Oxygen Blender feature | General Attributes | Resuscitation Unit<br>Unit 105PV<br>Unit 105EV | Atom V-505 Infant<br>Warmer with oxygen /<br>suction module K060461 | Bio-Med Devices<br>BMD air/oxygen<br>blender K925982 | Precision Medical<br>Precision Blender<br>K053232 | | |------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------| | Indications for Use | The Resuscitation Unit<br>105 EV and<br>Resuscitation Unit 105<br>PV are intended to remove<br>secretions from the<br>neonatal / infant airways or<br>to provide blended air and<br>oxygen at a desired oxygen<br>concentration in the range<br>of 21-100% to a respiratory<br>assist device connected to<br>the patient such as a<br>ventilator (including a<br>manually operated one)<br>and an oxygen tent. | The blender and suction<br>modules incorporated in the<br>V-505 Infant Warmer are<br>intended to remove<br>secretions from the neonatal<br>/ infant airways or to<br>provide blended air and<br>oxygen at a desired oxygen<br>concentration in the range of<br>21-100% to a respiratory<br>assist device connected to<br>the patient such as a<br>ventilator (including a<br>manually operated one) and<br>an oxygen tent. | There is no indications<br>for sue statement in the<br>FDA database but the<br>BMD air/oxygen<br>blender is intended to<br>provide blended air and<br>oxygen at a desired<br>oxygen concentration in<br>the range of 21-100% | Is intended to deliver<br>blended air and oxygen in<br>a hospital setting.<br>Oxygen concentrations<br>can be dialed in from 21<br>to 100%. The blender is<br>not intended as a life<br>supporting device. | | | Patient Population | Neonates and Infants | Neonates and Infants | Not specified | Not specified | | | Environment of Use | Hospital, delivery suites,<br>NICU | Hospital, delivery suites,<br>NICU | Hospital, delivery<br>suites, NICU | Hospital setting | | | Prescriptive | Yes | Yes | Yes | Yes | | | | | Page 4 of 8<br>5-Apr-11 | | | | | | Resuscitation Unit<br>Unit 105PV<br>Unit 105EV | Atom V-505 Infant<br>Warmer with oxygen /<br>suction module K0604618 | Bio-Med Devices<br>BMD air/oxygen<br>blender K925982 | Precision Medical<br>Precision Blender<br>K053232 | | | General Attributes (continued) | Outlet connections to | Oxygen tent<br>Oxygen mask<br>Cannula<br>Manual ventilator<br>Manual resuscitator | Oxygen tent<br>Oxygen mask<br>Cannula<br>Manual ventilator<br>Manual resuscitator | Oxygen tent<br>Oxygen mask<br>Cannula<br>Manual ventilator<br>Manual resuscitator<br>None supplied | Oxygen tent<br>Oxygen mask<br>Cannula<br>Manual ventilator<br>Manual resuscitator<br>None supplied | | | Accessories/Components | Airway pressure tube<br>Hoses and tubing<br>Humidifier jar<br>Suction jar | Hoses and tubing<br>Humidifier jar<br>Suction jar | | | | Technical specifications | Oxygen / air supply module | | | | | | | Gas sources | Air/O₂<br>300-500kPa | Air/O₂<br>206-517kPa | Air/O₂<br>206-517kPa | Air/O₂<br>206-517kPa | | | input pressure | | | | | | | Airway pressure<br>manometer<br>Range Accuracy | -20 to 80cmH₂O<br>±1% of full scale | N/A | N/A | N/A | | | Air/O₂ blender<br>Range<br>Accuracy | | 21-100%<br>±3%O₂ | 21-100%<br>±3%O₂ | 21-100%<br>±3%O₂ | | | Alarm | | Differential pressure<br>between oxygen and air<br>supply | Differential pressure<br>between oxygen and air<br>supply | Differential pressure<br>between oxygen and air<br>supply | | | Bleed flow rate | | 2.5-3.5 lpm | N/A | N/A | | | Gas flow rate range | < 15 lpm | < 15 lpm | ≥ 30 lpm | ≥ 30 lpm | | | | 510(k) Summary<br>5 of 8<br>5-Apr-11 | | | | | | Resuscitation Unit<br>Unit 105PV<br>Unit 105EV | Atom V-505 Infant<br>Warmer with oxygen /<br>suction module K060461 | Bio-Med Devices<br>BMD air/oxygen<br>blender K925982 | Precision Medical<br>Precision Blender<br>K053232 | | | Technical specifications (continued) | | | | | | | Relieving valve<br>(for preventing excessive<br>airway pressure) | (Default setting : 5.9kPa<br>(60cmH2O)) | N/A (no reliving valve) | N/A(no relieving valve) | N/A(no relieving valve) | | | Flow outlet | Dual flow outlet to connect to<br>a device requiring a gas<br>supply<br>Oxygen density of both is the<br>same. | Single flow outlet to connect<br>to device requiring a gas<br>supply | Flow outlet for<br>connecting to device<br>requiring a gas supply | Flow outlet for connecting<br>to device requiring a gas<br>supply | | | Standards | | | | | | | ISO 10651-5 Gas<br>powered resuscitators | Yes, applicable sections | N/A | N/A | N/A | | | ISO 10079-3 suction<br>equipment | Yes, applicable sections | N/A | N/A | N/A | | | Section 514 | None | None | None | None | | Page 52 --- {4}------------------------------------------------ 510(k) Summary . . Page 53 -. r . {5}------------------------------------------------ and the contraction of the comments of the control of . {6}------------------------------------------------ 510(k) Summary ## able of the Similarities and Differences to the Predicates for Suction features | General Attributes | | | | | |-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------| | Indications for Use | Resuscitation Unit<br>Unit 105PV<br>Unit 105EV | Atom V-505 Infant Warmer with<br>oxygen / suction module<br>K060461 | GE<br>Giraffe and Panda<br>K070210 | | | | The Resuscitation Unit 105 EV<br>and Resuscitation Unit 105 PV are<br>intended to remove secretions from<br>the neonatal / infant airways and to<br>provide blended air and oxygen at a<br>desired oxygen concentration in the<br>range of 21-100% to a respiratory<br>assist device connected to the patient<br>such as a ventilator (including a<br>manually operated one) and an<br>oxygen tent. | The blender and suction modules<br>incorporated in the V-505 Infant<br>Warmer are intended to remove<br>secretions from the neonatal / infant<br>airways and to provide blended air<br>and oxygen at a desired oxygen<br>concentration in the range of 21-<br>100% to a respiratory assist device<br>connected to the patient such as a<br>ventilator (including a manually<br>operated one) and an oxygen tent. | Provides the basic equipment<br>required for pulmonary<br>resuscitation of infants.<br>Pulmonary resuscitation includes<br>practices necessary to establish a<br>clear airway and provide oxygen<br>and air/oxygen mixtures and/or<br>manual ventilation to the infant.<br>(Includes a suction function and<br>feature) | | | Patient Population | Neonates and Infants | Neonates and Infants | Infant | | | Environment of Use | Hospital, delivery suites, NICU | Hospital, delivery suites, NICU | Hospital, delivery suites, NICU | | | Prescriptive | Persons trained in infant / neonate<br>resuscitation | Persons trained in infant / neonate<br>resuscitation | Persons trained in infant / neonate<br>resuscitation | | | Technical specifications | | | | | | Suction bottle capacity | Up to 1200 ml | Not specified | Not specified other than various<br>sizes | | | Dimensions | 200mm(W)x180 mm(D)x370mm (H) | Built-in to warmer, not standalone | 340mm(W)x250mm(D)x250mm(H) | | | Weight | 7kg | | 9 kg | | | Operating and Storage<br>limits | -20°C to 65°C<br>Up to 97% RH | -20°C to 65°C<br>Up to 97% RH | -20°C to 60°C<br>Up to 95% RH | | | 510(k) Summary<br>Page 7 of 8<br>5-Apr-11 | | | | | | | Resuscitation Unit<br>Unit 105PV<br>Unit 105EV | Atom V-505 Infant Warmer with<br>oxygen / suction module K060461 | | GE<br>Giraffe and Panda K070210 | | Technical specifications (continued) | | | | | | | Vacuum source | Model 105PV - wall suction<br>vacuum | Model 105PV - wall suction<br>vacuum | Wall suction<br>vacuum | | | Vacuum range | Model 105EV - venturi method<br>0-200 mmHg | 0-200 mmHg | 0-150 mmHg | | | Vacuum Flow range | ~ 20 lpm | N/A | 15 lpm | | | Standards | | | | | | ISO 10651-5 Gas<br>powered resuscitators | Yes, applicable sections | N/A | Yes, applicable sections | | | ISO 10079-3 - Suction<br>equipment | Yes, applicable sections | N/A | Yes, applicable sections | | | Section 514 | None | None | None | | Summary of substantial equivalence - Resuscitation Unit 105PV and Resuscitation Unit 105EV | | | | | | Indications for Use - The Resuscitation Unit 105 EV and Resuscitation Unit 105 PV have equivalent indications for use to the predicates. That<br>is as a breathing gas mixer (blender) with a suction feature. | | | |…
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...