← Product Code [BZD](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BZD) · K960518

# THE PURITAN-BENNETT 318 PLUS NASAL CPAP SYSTEM (K960518)

_Puritan Bennett Corp. · BZD · May 9, 1996 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BZD/K960518

## Device Facts

- **Applicant:** Puritan Bennett Corp.
- **Product Code:** [BZD](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BZD.md)
- **Decision Date:** May 9, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5905
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic

## Indications for Use

The intended use of the 318 Plus Nasal CPAP System is to provide Continuous Positive Airway Pressure (CPAP) between 3 and 18 cm H2O to spontaneously breathing adult patients for the treatment of Obstructive Sleep Apnea in a homecare environment.

## Device Story

The 318 Plus Nasal CPAP System provides continuous positive airway pressure (3-18 cm H2O) to treat obstructive sleep apnea. It operates via an AC-powered motor with a feedback loop to maintain constant speed, compensating for circuit pressure losses without a dedicated pressure transducer. Homecare providers configure settings; patients operate the device via a guide, using a ramp function to gradually increase pressure from a lower starting point to the prescribed level. A pressure display shows current settings. An optional compliance meter tracks usage by monitoring motor current fluctuations during patient breathing. The device is intended for home use by adult patients; it is not life-supporting. Accessories include patient tubing, masks, and headgear.

## Clinical Evidence

No clinical data submitted. Bench testing only, including electrical performance, mechanical/environmental testing, and static/dynamic pressure testing, demonstrated performance equivalent to the predicate C314 device.

## Technological Characteristics

CPAP device; 3-18 cm H2O pressure range; AC mains powered (100-240 VAC, 50/60 Hz) with automatic switching power supply; motor-driven with feedback loop for speed control; non-software driven; non-sterile; multiple-use device/filter/humidifier; single-patient-use accessories.

## Regulatory Identification

A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.

## Predicate Devices

- Puritan-Bennett C314 Nasal CPAP System ([K952292](/device/K952292.md))
- Puritan-Bennett C318 Nasal CPAP System ([K903764](/device/K903764.md))
- Puritan-Bennett 335 Respiratory Support System ([K942210](/device/K942210.md))
- Healthdyne Tranquility Quest Compact Nasal CPAP System Model 7300 ([K943509](/device/K943509.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

K960518

# SECTION 16. 510(k) SUMMARY

MAY - 9 1996

## (1) Company Information

Owner/Operator: Puritan-Bennett Corporation
Registration Number 9919005

Manufacturing Site: Puritan-Bennet Corporation
9728 Pflumm Road
Lenexa, KS 66215
Registration Number 1933149

Official Correspondent: C. Marshall Smith
Phone: 913-495-4845
Direct Dial: 913-495-4845
Fax: 913-394-8764

Date of Summary Preparation: 2 February 1996

## (2) Device Name:

Proprietary Name: Puritan-Bennett 318 Plus Nasal CPAP System

Common Name: CPAP Machine

Classification Name: Noncontinuous Ventilator, per 21 CFR 868.5905 and 73 BZD

## (3) Equivalent Devices:

Primary Predicate Device: Puritan-Bennett C314 Nasal CPAP System, K952292

Other Predicate Devices: Puritan-Bennett C318 Nasal CPAP System, K903764
Puritan-Bennett 335 Respiratory Support System, K942210 (pending)
Healthdyne Tranquility Quest Compact Nasal CPAP System Model 7300, K943509

35

{1}

# (4) Device Description:

The 318 Plus Nasal CPAP System is a device used to provide Continuous Positive Airway Pressure between 3 and 18 cm H₂O.

The device is a modification to the Puritan-Bennett C314 Nasal CPAP System, which adds some additional features.

The device is powered by AC Mains from 100 to 240 VAC, 50/60 Hz by utilizing a standard automatic switching power supply similar to the one in the PB335 Respiratory Support System.

The device is set up for use by the homecare provider using the Set-up Instructions provided with the device. It is operated using the Patient Guide.

The pressure is set to the prescription with the use of a patient circuit, with the pressure being measured at the patient end of the circuit. By using this method, losses associated with the patient circuit are compensated for and thus the device does not require a pressure transducer. The motor maintains a constant speed through the feedback loop from the motor to the motor drive circuit.

The device has a ramp function. This allows the patient to reduce the pressure when first going to bed to make it more comfortable to fall asleep. The available ramp times are 0, 5, 10 or 20 minutes. The ramp duration is set up by the homecare provider by use of the Ramp Duration Switches. The ramp starting pressure can be adjusted by the patient using the Ramp Starting Pressure Adjustment switches. The ramp feature can be activated and deactivated by pressing the Ramp Button.

Both the prescription pressure and the ramp pressure can be seen in the Pressure Display on the top housing.

The optional Compliance Meter measures patient compliance by measuring changes in the motor current as the patient breathes on the device.

The accessories, i.e., the patient tubing, patient masks and headgear are the same ones used with the C314 and PB335.

The device is not for life-supporting or life-sustaining situations.

The device is for prescription use.

The device itself, the air filter and the humidifier are for multiple use. The other accessories, i.e., the patient tube, nasal masks and headgear are for single patient use.

The device is not software-driven.

Neither the device nor its accessories are sterile.

36

{2}

(5) Intended Use:

The intended use of the 318 Plus Nasal CPAP System is to provide Continuous Positive Airway Pressure (CPAP) between 3 and 18 cm H2O to spontaneously breathing adult patients for the treatment of Obstructive Sleep Apnea in a homecare environment. This is the same intended use as for the C314 Nasal CPAP System.

(6) Technological Characteristics:

The device has the same technological characteristics as the primary predicate device, inasmuch as the two devices use either the same or similar components and the two devices function in the same manner.

(7) Nonclinical Tests Submitted:

The device was tested in accordance with the "Electrical Performance Testing" and "Mechanical and Environmental Testing" sections of Appendix A of the "Draft - Reviewer Guidance for Premarket Notifications, Anesthesiology and Respiratory Branch" document dated November 1993. The device passed all of the tests.

Static and dynamic pressure testing was also performed. The results of the testing demonstrated that the 318 Plus performed in the same manner as the primary predicate device, the C314.

The device was also tested when powered by a 12 volt DC power supply using the optional DC to AC inverter. The results of the testing demonstrated that the device's performance was the same as when powered by AC Mains.

(8) Clinical Tests Submitted:

None.

(9) Conclusions from Tests:

As stated in Section (7) above, all of the testing demonstrated that the 318 Plus is as safe and effective and performs as well or better than the primary predicate device, the C314.

---

**Source:** [https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BZD/K960518](https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BZD/K960518)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
