Sleep Apnea Breathing Therapy Mask: YF-01 Full Face Mask, YF-02 Full Face Mask, YF-03 Full Face Mask, YN-02 Nasal Mask, YN-03 Nasal Mask, YP-01 Nasal Pillow Mask
Sleep Apnea Breathing Therapy Mask: YF-01 Full Face Mask, YF-02 Full Face Mask, YF-03 Full Face Mask, YN-02 Nasal Mask, YN-03 Nasal Mask, YP-01 Nasal Pillow Mask
Applicant
Suzhou Yuyue Medical Technology Co., Ltd.
Product Code
BZD · Anesthesiology
Decision Date
Jul 2, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5905
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The sleep apnea breathing therapy mask is a non-invasive accessory used for channeling airflow (with or without supplemental oxygen) to a patient from a positive airway pressure (PAP) device such as a continuous positive airway pressure (CPAP) or bi-level system. -The mask is to be used by patients (weighing>30kg), intended for single patient reuse in the home environment. -The mask is intended for single-patient use in the hospital or institutional environment.
Device Story
Non-invasive interface for CPAP/BPAP devices; delivers pressurized air to prevent apnea. Components: frame, cushion, elbow, swivel, headgear. Elbow includes vent for exhaust and anti-asphyxiation valve. Swivel allows 360-degree rotation to prevent tube winding. Connects to standard 22mm tubing. Used in home or hospital settings by patients. Disassembled without tools for cleaning. Benefits: provides secure seal for effective PAP therapy delivery; improves patient compliance through ergonomic design.
Clinical Evidence
Bench testing only. Compliance verified against ISO 17510:2015 (mask performance), ISO 5356-1:2004 (connectors), ISO 10993 series (biocompatibility), and ISO 18562 series (breathing gas pathway safety).
Technological Characteristics
Non-invasive mask interface; 22mm conical connector per ISO 5356-1. Materials evaluated per ISO 10993-1, -5, -10 and ISO 18562-1, -2, -3, -4. Pressure range 4-30 cmH2O (full face/nasal) or 4-20 cmH2O (nasal pillows). Passive device; no energy source or software.
Indications for Use
Indicated for patients weighing >30kg requiring non-invasive positive airway pressure (PAP) therapy (CPAP or bi-level) for sleep apnea. Intended for single-patient reuse in home or hospital/institutional settings.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right is the FDA logo, with the acronym "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in a larger, bolder font, and "ADMINISTRATION" in a smaller font below.
July 2, 2021
Suzhou Yuyue Medical Technology Co., Ltd. Sun Wei R&D engineer No. 9 Jinfeng Road, Suzhou Science & Technology Town Suzhou, Jiangsu 215163 China
Re: K203347
Trade/Device Name: Sleep Apnea Breathing Therapy Mask: YF-01 Full Face Mask, YF-02 Full Face Mask, YF-03 Full Face Mask, YN-02 Nasal Mask, YN-03 Nasal Mask, YP-01 Nasal Pillows Mask. Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: Class II Product Code: BZD Dated: May 28, 2021 Received: June 3, 2021
Dear Sun Wei:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Rachana Visaria, Ph.D. Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K203347
#### Device Name
Sleep Apnea Breathing Therapy Mask: YF-02 Full Face Mask, YF-03 Full Face Mask, YN-02 Nasal Mask, YN-03 Nasal Mask, YP-01 Nasal Pillows Mask
#### Indications for Use (Describe)
The sleep apnea breathing therapy mask is a non-invasive accessory used for channeling airflow (with or without supplemental oxygen) to a patient from a positive airway pressure (PAP) device such as a continuous positive airway pressure (CPAP) or bi-level system.
-The mask is to be used by patients (weighing>30kg), intended for single patient reuse in the home environment. -The mask is intended for single-patient use in the hospital or institutional environment.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| <span style="font-size: 16px;"> <input checked="true" type="checkbox"/> </span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 16px;"> <input type="checkbox"/> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Company Name/Owner | Suzhou Yuyue Medical Technology Co., Ltd. |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact person/Author | Sun Wei (Mr.) |
| Date prepared | 06/30/2021 |
| Contact details | Address: No.9 Jinfeng Road, Suzhou Science &<br>Technology Town, 215163 Suzhou, Jiangsu,<br>PEOPLE'S REPUBLIC OF CHINA<br>Tel: +86-512-67373001<br>Fax: +86-512-67373008 |
| Trade name | Sleep Apnea Breathing Therapy Mask:<br>YF-01 Full Face Mask, YF-02 Full Face Mask,<br>YF-03 Full Face Mask, YN-02 Nasal Mask,<br>YN-03 Nasal Mask, YP-01 Nasal Pillows<br>Mask |
| Common name | Sleep Apnea Breathing Therapy Mask |
| Classification name | Non Continuous Ventilator (IPPB)<br>Class II (21 CFR §868.5905)<br>Product code BZD (Anaesthesiology) |
| Predicate device | Full Face Mask: AirFit F20(K170924)<br>Nasal Mask: AirFit N20(K161978)<br>Nasal Pillows Mask: Swift™ FX Nasal<br>Pillows (K090244) |
{4}------------------------------------------------
## Description
These masks act as an interface for CPAP/BPAP device and deliver pressurized air to prevent apnea from occurring. The masks consist of several parts which includes frame, cushion, elbow, swivel, and headgear. The frame provides support for the mask, the cushion provides the seal, the elbow is designed to provide a vent for exhaust gas and a valve to prevent asphyxiation. The swivel can be rotated 360 degrees, which can help prevent pipe winding. The size of the connector port of the swivel is consistent with ISO standards and can be connected to any tube which conforms to the same standard. The headgear is used to fasten the mask to the face. The masks can be disassembled without any tools.
#### Indications for use
The sleep apnea breathing therapy mask is a non-invasive accessory used for channeling airflow (with or without supplemental oxygen) to a patient from a positive airway pressure (PAP) device such as a continuous positive airway pressure (CPAP) or bi-level system.
-The mask is to be used by patients (weighing>30kg), intended for single patient reuse in the home environment.
-The mask is intended for single-patient use in the hospital or institutional environment.
## Substantial equivalence comparison with predicate device
Detailed substantial comparison was made between subject device and predicate device. Please refer to below tables for details.
{5}------------------------------------------------
# Table 1: Comparison table between full face mask and AirFit F20
| | Subject Device | Predicate Device | | |
|--------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Comparison Elements | YF-01 Full Face Mask<br>YF-02 Full Face Mask<br>YF-03 Full Face Mask | AirFit F20 | Comment | |
| Classification<br>Regulation | 868.5905 | 868.5905 | Identical | |
| Product Code | BZD | BZD | Identical | |
| 510(k) Number | K203347 | K170924 | | |
| Indications for use | The sleep apnea breathing therapy mask is a<br>non-invasive accessory used for channeling<br>airflow (with or without supplemental<br>oxygen) to a patient from a positive airway<br>pressure (PAP) device such as a continuous<br>positive airway pressure (CPAP) or bi-level<br>system.<br><br>-The mask is to be used by patients<br>(weighing>30kg), intended for single patient<br>reuse in the home environment.<br><br>-The mask is intended for single-patient use<br>in the hospital or institutional environment. | The AirFit F20 is a non-invasive<br>accessory used for channeling airflow<br>(with or without supplemental oxygen)<br>to a patient from a positive airway<br>pressure (PAP) device such as a<br>continuous positive airway pressure<br>(CPAP) or bilevel system.<br><br>The AirFit F20 is:<br>• to be used by patients weighing more<br>than 66 lb (30 kg) for whom positive<br>airway pressure therapy has been<br>prescribed<br>• intended for single-patient reuse in<br>the home environment and<br>multi-patient reuse in the | Different<br><br>The inclusion of "single-<br>patient reuse" is a<br>narrower indication to the<br>"multi-patient, multi-use"<br>in the hospital /institutional<br>environment and does not<br>alter the intended use of<br>the device. Safety and<br>effectiveness issue will not<br>be raised. | |
| Principle | Positive airway pressure(PAP) | Positive airway pressure(PAP) | Identical | |
| Patient usage type | single-patient reuse in the home<br>environment. single-patient use in the<br>hospital or institutional environment. | single-patient reuse in the home<br>environment and multi-patient reuse in<br>the hospital/institutional | Different<br>The inclusion of “single-<br>patient reuse” is a narrower<br>indication to the “multi-<br>patient, multi-use” in the<br>hospital/institutional<br>environment and does not<br>alter the intended use of<br>the device. Safety and<br>effectiveness issue will not<br>be raised. | |
| Pressure source action<br>body site | Mouth and Nose | Mouth and Nose | Identical | |
| Technical Specifications | | | | |
| Breathing Tube<br>connection | 22mm conical connector | 22mm conical connector | Identical | |
| Pressure Range | 4-30 cmH2O | 3-40 cmH2O | Different<br>Safety and effectiveness<br>issue will not be raised. | |
| Pressure-Flow Cure | | | Similar, Safety and<br>effectiveness issue will not<br>be raised. | |
| Dead space (largest) | 208mL | 240mL | Different<br>The dead space of our masks<br>is less than AirFit F20. Both<br>meet relevant standard<br>requirement. Safety and<br>effectiveness issue will not<br>be raised. | |
| Resistance to Flow | at 50 L/min: 0.5 hPa<br>at 100 L/min:1.5 hPa | at 50 L/min: 0.2hPa<br>at 100 L/min:0.6 hPa | Different<br>The Resistance of our mask<br>is higher than AirFit F20.<br>Both meet relevant<br>standard requirement.<br>Safety and effectiveness<br>issue will not be raised. | |
| Inspiratory and<br>expiratory resistance<br>with the<br>Non-Rebreathing<br>Valve<br>open-to-atmosphere | Inspiratory at 50 L/min: ≤2 cmH₂O<br>Expiratory at 50 L/min: ≤2 cmH₂O | Inspiratory at 50 L/min: 0.6 cmH₂O<br>Expiratory at 50 L/min: 0.7 cmH₂O | Different<br>Both meet Clause 2.1 of ISO<br>17510:2015 which need the<br>pressure is less than 10<br>cmH₂O. Safety and<br>effectiveness issue will not | |
| pressure | | | be raised. | |
| Sound | DECLARED DUAL-NUMBER NOISE EMISSION<br>VALUES in accordance with ISO 4871. The A-<br>weighted sound power level of the mask is<br>less than 38 dBA, with uncertainty 3 dBA. The<br>A-weighted sound pressure level of the mask<br>at a distance of 1 m is less than 30 dBA, with<br>uncertainty 3 dBA. | DECLARED DUAL-NUMBER NOISE<br>EMISSION VALUES in accordance with<br>ISO 4871. The A-weighted sound power<br>level of the mask is 30 dBA, with<br>uncertainty of 3 dBA. The A-weighted<br>sound pressure level of the mask at a<br>distance of 1 m is 23 dBA, with<br>uncertainty of 3 dBA. | Different<br>The sound of our masks is<br>more than AirFit F20, but<br>the sound pressure level of<br>our masks is less than 30<br>dBA, they are quiet and<br>don't interfere with sleep.<br>Safety and effectiveness<br>issue will not be raised. | |
| Operating<br>environment | +5°C to +40°C (41°F to 104°F)<br><br>10%~90% relative humidity<br>non-condensing | 41°F to 104°F(5°C to +40°C)<br><br>15% to 95% non-condensing | Temperature is identical.<br>Humidity is different, but<br>very similar. Safety and<br>effectiveness issue will not<br>be raised. | |
| Storage and transport<br>environment | -20°C to +60°C (-4°F to 140°F)<br>10%~90% relative humidity<br>non-condensing | -4°F to 140°F (-20°C to +60°C)<br>Up to 95% non-condensing | Temperature is identical.<br>Humidity is different, but<br>very similar. Safety and<br>effectiveness issue will not<br>be raised. | |
| Comparison Elements | Subject Device | Predicate Device | Reference device | Comment |
| | YN-02 Nasal Mask<br>YN-03 Nasal Mask | AirFit N20 | BMC-NM | |
| Classification<br>Regulation | 868.5905 | 868.5905 | 868.5905 | Identical |
| Product Code | BZD | BZD | BZD | Identical |
| 510(k) Number | K203347 | K161978 | K133009 | |
| Indications for use | The sleep apnea breathing therapy<br>mask is a non-invasive accessory used<br>for channeling airflow (with or<br>without supplemental oxygen) to a<br>patient from a positive airway<br>pressure (PAP) device such as a<br>continuous positive airway pressure<br>(CPAP) or bi-level system.<br>-The mask is to be used by patients<br>(weighing>30kg), intended for single<br>patient reuse in the home<br>environment.<br>-The mask is intended for single-<br>patient use in the hospital or<br>institutional environment. | The AirFit N20 channels airflow<br>noninvasively to a patient from<br>a positive airway pressure (PAP)<br>device such as a continuous<br>positive airway pressure (CPAP)<br>or bilevel device.<br>The AirFit N20 is:<br>• to be used by patients<br>weighing more than 66lb<br>(>30kg) for whom positive<br>airway pressure has been<br>prescribed.<br>• intended for single-patient re-<br>use in the home environment<br>and<br>multi-patient re-use in the<br>hospital/institutional environment | The BMC-NM Nasal Mask channel<br>airflow noninvasively to a patient<br>from a positive airway pressure<br>device such as a continuous<br>positive airway pressure (CPAP)<br>or Bi-level system.<br>The BMC-NM Nasal Mask are:<br>• To be used by adult patients<br>(66lbs/>30kg) for whom positive<br>airway pressure has been<br>prescribed.<br>• Intended for single-patient<br>reuse in home environment and<br>multi-patient re-use in the<br>hospital/institutional<br>environment. | Different<br>The inclusion of "single-<br>patient reuse" is<br>a<br>narrower indication to<br>the<br>"multi-patient,<br>multi-use" in the<br>hospital /institutional<br>environment and does<br>not alter the intended<br>use of the device.<br>Safety and effectiveness<br>issue will not be raised. |
| Principle | Positive airway pressure(PAP) | Positive airway pressure(PAP) | Positive airway pressure(PAP) | Identical |
| Patient usage type | single-patient reuse in the home<br>environment. single-patient use in the<br>hospital or institutional environment. | single patient re-use in the<br>home environment and multi-<br>patient re-use in the<br>hospital/institutional | Single-patient reuse in home<br>environment and multi-patient<br>re-use in the hospital/institutional<br>environment. | Different<br>The inclusion of<br>a narrower indication |
| | | Environment. | | to the "multi-patient,<br>multi-use" in the<br>hospital/institutional<br>environment and does<br>not alter the intended<br>use of the device.<br>Safety and<br>effectiveness issue will<br>not be raised. |
| Pressure source action<br>body site | Nose | Nose | Nose | Identical |
| Technical Specifications | | | | |
| Breathing Tube<br>connection | 22mm conical connector | 22mm conical connector | 22mm entrainment valve<br>elbow | Identical |
| Pressure Range | 4-30 cmH2O | 4-30 cmH2O | 4 to 30 hPa | Identical |
| Pressure-Flow Cure | Image: Pressure-Flow Cure | Image: Pressure-Flow Cure | 4hPa=19L/min<br>12hPa=34L/min<br>2hPa=50L/min<br>30hPa=68L/min | Similar<br>Safety and effective-<br>ness issue will not be<br>raised. |
| Dead space (largest) | 142mL | 103.3mL | 145ml | Different<br>The dead space of our<br>masks is bigger than<br>AirFit N20. Both meet<br>relevant standard<br>requirement.<br>In addition, the dead |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
{9}---------------…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.