K181053 · Respironics, Inc. · BZD · Jan 4, 2019 · Anesthesiology
Device Facts
Record ID
K181053
Device Name
Care Orchestrator
Applicant
Respironics, Inc.
Product Code
BZD · Anesthesiology
Decision Date
Jan 4, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5905
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Care Orchestrator is intended to support clinicians by tracking data on patients who are prescribed compatible therapy devices in accordance with the intended use of those therapy devices. Care Orchestrator provides remote patient data collection & viewing and is intended to be used by healthcare representatives (e.g., Physicians, Clinicians, Durable Medical Equipment providers) in conjunction with compatible non-life support therapy devices to adjust prescription and/ or performance settings. Care Orchestrator allows read-only access to patients. Care Orchestrator is intended to be used in hospital, institutional, provider, and home care settings.
Device Story
Care Orchestrator is a web-based data management system for sleep and respiratory therapy. It collects patient data from compatible therapy devices via wireless transfer or SD card. Healthcare representatives (physicians, clinicians, DME providers) use the system in hospital, institutional, or home settings to manage patient referrals, monitor compliance, generate reports, and adjust prescription settings for non-life support devices. The system provides a centralized database for patient management, task tracking, and device health monitoring. By enabling remote access to therapy data and settings, the device aims to improve clinical workflow efficiency and patient compliance management. It does not support remote setting changes for life-support devices.
Clinical Evidence
No clinical data. Safety and effectiveness were demonstrated through non-clinical software verification and validation testing, including system testing of data transfer (wireless and SD card), display of therapy/compliance reports, and prescription management functionality.
Technological Characteristics
Web-based software application; centralized database architecture. Supports wireless and SD card data transfer. Software level of concern: Moderate. Features include task management, configurable compliance/health rules, and reporting templates.
Indications for Use
Indicated for patients prescribed compatible non-life support therapy devices (sleep therapy and respiratory devices). Used by healthcare representatives (physicians, clinicians, DME providers) for remote data collection, viewing, and adjustment of prescription/performance settings.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 4, 2019
Respironics Inc. Ankitha Rao Regulatory Project Manager 1001 Murry Ridge Lane Murrysville, Pennsylvania 15668
Re: K181053
Trade/Device Name: Care Orchestrator Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: Class II Product Code: BZD, MNS, MNT, CBK, NOU, CAW Dated: December 5, 2018 Received: December 6, 2018
Dear Ankitha Rao:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to Mav 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
James J. Lee -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181053
Device Name Care Orchestrator
### Indications for Use (Describe)
Care Orchestrator is intended to support clinicians by tracking data on patients who are prescribed compatible therapy devices in accordance with the intended use of those therapy devices. Care Orchestrator provides remote patient data collection & viewing and is intended to be used by healthcare representatives (e.g., Physicians, Clinicians, Durable Medical Equipment providers) in conjunction with compatible non-life support therapy devices to adjust prescription and/ or performance settings. Care Orchestrator allows read-only access to patients. Care Orchestrator is intended to be used in hospital, institutional, provider, and home care settings.
Type of Use (Select one or both, as applicable)
| <span style="font-size: large;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------------------------------------------------------------------------------|
| <span style="font-size: large;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# TAB 5 510(K) SUMMARY
#### Submitter I.
| Official Contact | Ankitha Rao<br>Regulatory Project Manager<br>ankitha.rao@philips.com<br>Respironics Inc.<br>1740 Golden Mile Highway<br>Monroeville, PA 15146<br>Phone: 724-387-7442<br>Fax: 724-387-7490 |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date of Preparation | April 18, 2018 |
| II. Device | |
| Name of Device: | Care Orchestrator |
| Common/Usual Name: | Data Management System |
| Device Classification: | Class II |
| Classification Name/<br>Product Code: | Non-continuous ventilator (21 CFR 868.5905, Product Code BZD)<br>Continuous ventilator (21 CFR 868.5895, Product Codes MNS, MNT, CBK, NOU) |
Portable oxygen generator (21CFR868.5440, Product Code CAW)
#### III. Legally Marketed Predicate Device
K152356 Sapphire, Respironics Inc. Trade Name: Care Orchestrator
{4}------------------------------------------------
#### Device Description IV.
Care Orchestrator is a solution that will provide entities involved in a patient's therapy lifecycle with the ability to manage patients and referrals, control access to patient information and theract with billing information, enhance patient compliance management workflow, manage the resupply of medical equipment, and gain efficiencies in the overall Patient Therapy Workflow.
Care Orchestrator will support patient data management for sleep therapy devices (BZD, MNS, MNT) and respiratory devices (CBK, NOU, CAW). Care Orchestrator will support prescription updates for sleep therapy devices. There are no input flows to the respiratory device interfaces (Bluetooth and SD Card). Device prescription and settings are read-only for these respiratory device types.
#### V. Indications for Use
Care Orchestrator is intended to support clinicians by tracking data on patients who are prescribed compatible therapy devices in accordance with the intended use of those therapy devices. Care Orchestrator provides remote patient data collection & viewing and is intended to be used by healthcare representatives (e.g., Physicians, Durable Medical Equipment providers) in conjunction with compatible non-life support therapy devices to adjust prescription and/or performance settings. Care Orchestrator allows read-only access to patients. Care Orchestrator is intended to be used in hospital, institutional, provider, and home care settings.
#### Comparison of Technological Characteristics with the Predicate VI.
# Device
The modified Care Orchestrator has the following similarities to the legally marketed predicate device:
- Similar intended use ●
- . Same operating principle
- . Same technoloqy
- Same manufacturing (deployment) process .
| Feature/Function | Predicate Device<br>Device Name: Care<br>Orchestrator (Project: Sapphire)<br>510(k) Number: K152356<br>Manufacturer: Respironics Inc. | Subject Device<br>Device Name: Care<br>Orchestrator (modified)<br>510(k) Number: TBD<br>Manufacturer: Respironics<br>Inc. | Comments |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | Sapphire is intended to support<br>clinicians by tracking data on<br>patients who are prescribed<br>compatible therapy devices in | Care Orchestrator is<br>intended to support<br>clinicians by tracking data on<br>patients who are prescribed | Equivalent to<br>K152356. |
| | accordance with the intended<br>use of those therapy devices.<br>Sapphire provides remote<br>patient data collection & viewing<br>and is intended to be used by<br>healthcare representatives (e.g.,<br>Physicians, Clinicians, Durable<br>Medical Equipment providers) in<br>conjunction with compatible<br>non-life support therapy devices<br>to adjust prescription and/or<br>performance settings. Sapphire<br>allows read-only access to<br>patients. Sapphire is intended to<br>be used in hospital, institutional,<br>provider, and home care<br>settings. | compatible therapy devices<br>in accordance with the<br>intended use of those<br>therapy devices. Care<br>Orchestrator provides<br>remote patient data<br>collection & viewing and is<br>intended to be used by<br>healthcare representatives<br>(e.g., Physicians, Clinicians,<br>Durable Medical Equipment<br>providers) in conjunction<br>with compatible non-life<br>support therapy devices to<br>adjust prescription and/or<br>performance settings. Care<br>Orchestrator allows read-<br>only access to patients. Care<br>Orchestrator is intended to<br>be used in hospital,<br>institutional, provider, and<br>home care settings. | Remote setting<br>change<br>functionality is<br>not applicable<br>to Respironics<br>CBK, NOU and<br>CAW therapy<br>devices. |
| Application | Web based application | Web based application | Equivalent to<br>K152356. |
| Data storage | Data stored on centralized<br>database | Data stored on centralized<br>database | Equivalent to<br>K152356. |
| Functionality | Patient management Display therapy data Generate reports Settings management for<br>non-life supporting devices | Patient management Display therapy data Generate reports Settings management<br>for non-life supporting<br>devices | Equivalent to<br>K152356.<br>Remote setting<br>change<br>functionality is<br>not applicable<br>to Respironics<br>CBK, NOU and<br>CAW therapy<br>devices or any<br>Respironics<br>devices that<br>are intended<br>for life-support. |
| Settings<br>management | Ability to remotely change<br>device settings in non-life<br>support devices only. | Ability to remotely change<br>device settings in non-life<br>support devices only. | Equivalent to<br>K152356. |
| Reports | Detailed Report (includes<br>compliance information) | Detailed Report (includes<br>compliance information) | Equivalent to<br>K152356. |
{5}------------------------------------------------
# VII. Performance Data:
# Software Verification and Validation Testing
Software verification and validation testing was conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." This software was considered to have a
{6}------------------------------------------------
Moderate Level of Concern, since a failure or latent flaw in the software could result in minor harm to the patient.
### Non-Clinical Tests
Software verification and validation testing was performed on Care Orchestrator based on the product requirements. This testing included complete system testing to verify data transfer from therapy devices to Care Orchestrator, through both wireless data transfer and SD card data transfer. Once data was transferred, all tests confirmed that Care Orchestrator can display patient and device information, display therapy data including compliance and therapy reports and allow for a user to create and/or edit a patients prescription for an applicable therapy device.
The testing of Care Orchestrator verified that all product requirements have been met with passing test results. The verification and validation testing demonstrated the overall substantial equivalence of the Care Orchestrator system.
### Clinical Tests
Clinical tests were not required to demonstrate the safety and effectiveness of Care Orchestrator. Product functionality has been adequately assessed by non-clinical tests.
## Summary of additional feature(s) from the Care Orchestrator (K152356)
- . Portable oxygen generator (Product Code CAW) Support
- Support of additional therapy device models (Product codes BZD, MNT, MNS, CBK/NOU) .
- . Therapy data download via removable media (SD card) for Product codes CBK/NOU
- . Task Management
- Configurable compliance and health rules .
- . Enhancements to user and organization management capability including external authorizations
- Enhancements to reporting capability including custom report templates .
- Patient management enhancements including search capability, activity log, notes, and document attachments
- . Display of device health information
- . Product support console
- . Legacy application data migration and synchronization
The inclusion of these features has been assessed within the risk analysis and no additional safety risks have been found as a result of the inclusion of these features.
# VIII. Conclusion
The modified Care Orchestrator is as safe and as effective as the predicate device, Care Orchestrator (K152356) and is deemed substantially equivalent to the predicate device, Care Orchestrator (K152356).