← Product Code [BZD](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BZD) · K153563

# AirFit F20 (K153563)

_Resmed, Ltd. · BZD · May 13, 2016 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BZD/K153563

## Device Facts

- **Applicant:** Resmed, Ltd.
- **Product Code:** [BZD](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BZD.md)
- **Decision Date:** May 13, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5905
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The AirFit F20 is a non-invasive accessory used for channeling airflow (with or without supplemental oxygen) to a patient from a positive airway pressure (PAP) device such as a continuous positive airway pressure (CPAP) or bilevel system. The AirFit F20 is: to be used by patients weighing more than 66lb (30kg) for whom positive airway pressure therapy has been prescribed intended for single-patient reuse in the home environment and multi-patient reuse in the hospital/institutional environment.

## Device Story

AirFit F20 is a non-invasive, vented full-face mask for delivering positive airway pressure therapy. It covers the patient's mouth and nose, creating a seal to direct airflow from a PAP device (CPAP or bilevel) to the patient. The system consists of a silicone cushion, frame, elbow, and adjustable headgear. The elbow includes a vent array for CO2 flushing and an anti-asphyxia valve (AAV) for safety if airflow is interrupted. The device is used in home or hospital settings by patients prescribed PAP therapy. It connects to standard air delivery hoses via a conical connector. The mask provides a stable interface to ensure effective pressure delivery, benefiting patients by facilitating respiratory therapy. It is supplied non-sterile and is intended for single-patient reuse at home or multi-patient reuse in clinical environments.

## Clinical Evidence

No clinical data. Substantial equivalence was demonstrated through non-clinical bench testing, including CO2 rebreathing, AAV performance, flow resistance, and mechanical integrity testing per ISO 17510-2. Biocompatibility was verified per ISO 10993 standards.

## Technological Characteristics

Vented full-face mask; materials include silicone, molded plastic, and polyamide/polyurethane/polyester headgear. Biocompatibility per ISO 10993-1. Features include anti-asphyxia valve, vent array, and standard conical connector. Designed per ISO 17510-2. Non-sterile, reusable.

## Regulatory Identification

A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.

## Predicate Devices

- Quattro Air ([K123979](/device/K123979.md))

## Reference Devices

- AirFit N10 ([K132887](/device/K132887.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. The logo is simple and professional, reflecting the department's role in public health and welfare.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 13, 2016

ResMed Ltd. % Larissa D'Andrea Director, Government & Regulatory Affairs ResMed Corp. 9001 Spectrum Center Boulevard San Diego, California 92123

Re: K153563

Trade/Device Name: AirFit F20 Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (Ippb) Regulatory Class: Class II Product Code: BZD Dated: April 8, 2016 Received: April 11, 2016

Dear Larissa D'Andrea:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

*Tejashri Purohit-Sheth, M.D.*

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Form Approved: OMB No. 0910-0120

Expiration Date: January 31, 2017

See PRA Statement below.

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known)

K153563

Device Name AirFit F20

Indications for Use (Describe)

The AirFit F20 is a non-invasive accessory used for channeling airflow (with or without supplemental oxygen) to a patient from a positive airway pressure (PAP) device such as a continuous pressure (CPAP) or bilevel system.

The AirFit F20 is:

• to be used by patients weighing more than 66 lb (30 kg) for whom positive airway pressure therapy has been prescribed • intended for single-patient reuse in the home environment and multi-patient reuse in the hospital/institutional environment.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) SUMMARY [As required by 21 CFR 807.92(c)] Date Prepared May 4, 2016 Company Name / Owner ResMed Ltd 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia Submitter Name Ms. Johanna Wright Senior Regulatory Affairs Manager ResMed Ltd Tel: +612 8884 1000 Fax: +612 8884 2004 johanna.wright@resmed.com.au Official Contact Ms Larissa D'Andrea Director, Government & Regulatory Affairs ResMed Corp 9001 Spectrum Center Blvd San Diego CA 92123 USA Tel: +1 858 836 6837 Fax: +1 858 836 5519 Larissa.D'Andrea@resmed.com Device Trade Name AirFit F20 Device Common Name Vented Full Face Mask Noncontinuous Ventilator (IPPB) (21 CFR 868.5905, Classification Name Product Code BZD) Predicate Device Quattro Air (K123979) Reference Device AirFit N10 (K132887) Device Description The AirFit F20 is an externally placed vented mask covering the mouth and the nose of the patient. It provides a seal such that pressure from a positive pressure source is directed to the patient's nose and/or mouth. The mask connects via a standard conical connector to a conventional air delivery hose, which in turn connects to the positive pressure source. The mask is held in place with an adjustable headgear that straps the mask to the face. The AirFit F20 mask system comprises four subassemblies: cushion, frame, elbow and headgear. The elbow incorporates the vent array and anti-asphyxia valve safety features. The cushion and headgear are available in various sizes to fit a wide patient population.

AirFit F20 is a prescription device supplied non-sterile.

{4}------------------------------------------------

| Intended Use                                                 | The AirFit F20 is a non-invasive accessory used for channeling<br>airflow (with or without supplemental oxygen) to a patient from a<br>positive airway pressure (PAP) device such as a continuous<br>positive airway pressure (CPAP) or bilevel system.<br><br>The AirFit F20 is:<br>to be used by patients weighing more than 66lb (30kg) for<br>whom positive airway pressure therapy has been prescribed intended for single-patient reuse in the home environment and<br>multi-patient reuse in the hospital/institutional environment.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|--------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Similarities and<br>Differences with the<br>Predicate Device | The AirFit F20 device has the following similarities to the<br>previously cleared predicated device (Quattro Air (K123979):  same intended use same operating principle similar design and materials which incorporates:  a silicone interface providing a seal around the nose and the mouth; vent holes providing continuous air leak to flush out and minimize the amount of CO2 re-breathed by the patient; anti-asphyxia valves (AAV) to enable the patient to breathe fresh air in the event that airflow from the flow generator is impeded; elbow which connects to a conventional air delivery hose via standard conical connector; molded plastic and silicone components and polyamide/polyurethane/polyester headgear - all deemed biologically safe (ref: ISO 10993-1); Both masks are offered in various sizes to ensure adequate fit over the extended patient population.   similar performance i.e both masks operate on the same flow generator settings and have similar pressure-flow characteristics and flow impedance. same operating environments i.e reuse in the home and hospital / institution environments similar manufacturing processes |
|                                                              | The main differences between the AirFit F20 and the previously<br>cleared predicated device (Quattro Air (K123979)) include mask<br>component design and geometry:<br>The AirFit F20 frame incorporates an upper arm positioned under the eyes whilst the Quattro Air frame includes a forehead support. The AirFit F20 vent array is located on the elbow assembly, whereas the Quattro Air vent array is located on the cushion assembly. For sealing performance, the AirFit F20 silicone cushion employs a novel feature to anchor and stabilise the interface compared with the traditional Quattro Air silicone cushion design.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                                                              | Verification and validation testing has demonstrated that these                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |

{5}------------------------------------------------

differences do not raise new questions of safety or effectiveness.

Non-Clinical Data The AirFit F20 was designed and tested in accordance with ISO 17510-2: Sleep apnoea breathing therapy - Part 2: Masks and application accessories.

> The scope of non-clinical testing conducted to support the substantial equivalence claim of AirFit F20 with the predicate device included:

- CO2 rebreathing ●
- . AAV performance (activation/deactivation, inspiratory /expiratory resistance, inadvertent activation/deactivation, response to extreme humidification)
- . Total mask flow
- Flow resistance ●
- Through impedance ●

Mechanical integrity and performance of the new device was also verified to simulated normal and reasonable abuse scenarios including: home cleaning; multi-patient reuse; transportation and storage.

Biocompatibility testing was conducted in accordance with ISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 10993-12 and ISO 10993-17 on new materials used in the manufacture of the AirFit F20 with patient exposure classifications permanent external communicating device (tissue) and /or permanent skin contact.

Testing confirmed that the new device met the predetermined acceptance criteria and the performance of the AirFit F20 is substantially equivalent to Quattro Air (K123979).

Clinical Data Clinical data was not relied upon to demonstrate substantial equivalence to the predicate device. Bench testing demonstrates that the new AirFit F20 device performs in an equivalent manner and is as safe and as effective as the predicate device.

#### Substantial Equivalence The new AirFit F20 is substantially equivalent to the predicate Conclusion device:

- it has the same intended use -
- it has similar technological characteristics
- it has similar performance characteristics
- the difference do not raise any new questions of safety or effectiveness
- it is at least as safe and as effective as the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BZD/K153563](https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BZD/K153563)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
