← Product Code [BZD](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BZD) · K150685

# Skynector CPAP Mask (K150685)

_Sky Wise Medical Instrument (Shen Zhen) Co., Ltd. · BZD · Aug 24, 2015 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BZD/K150685

## Device Facts

- **Applicant:** Sky Wise Medical Instrument (Shen Zhen) Co., Ltd.
- **Product Code:** [BZD](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BZD.md)
- **Decision Date:** Aug 24, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5905
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic

## Indications for Use

The Skynector CPAP Mask channels airflow non-invasively to a patient from a positive airway pressure device such as a continuous positive airway pressure (CPAP) or bi-level system. The Skynector CPAP Mask is to be used by adult patients (>66lb />30kg) for whom positive airway pressure has been prescribed. The Skynector CPAP Mask is intended for single patient re-use in the home environment and/or multi-patient reuse in the hospital /institutional environment.

## Device Story

Skynector CPAP masks (full-face and nasal models) serve as patient interfaces for CPAP or bi-level ventilation. Devices consist of hard polycarbonate frames and soft silicone cushions/pads, secured by nylon/polyester headgear. Masks channel airflow from a flow generator to the patient's nose and/or mouth. Full-face models include an anti-asphyxia valve and intentional vent holes to minimize CO2 re-breathing. Used in home or hospital settings by patients under prescription. Healthcare providers use the mask to facilitate prescribed pressure therapy; the interface ensures a seal to maintain required pressure levels (4-20 cmH2O). Benefits include non-invasive delivery of respiratory support for sleep apnea or related conditions.

## Clinical Evidence

No clinical data. Substantial equivalence is supported by bench testing, including ISO 17510-2 performance testing (leak rate, pressure drop, anti-asphyxia valve function, CO2 re-breathing, inspiratory/expiratory resistance) and biocompatibility testing (ISO 10993 series).

## Technological Characteristics

Materials: Polycarbonate (bracket, swivel, connector, elbow), silicone rubber (cushion, pad, valve), PVC (pick off port), nylon/polyester (headgear). Design: Face interface with headgear. Connectivity: None (mechanical interface). Sterilization: Provided non-sterile. Standards: ISO 5356-1, ISO 10993-3/5/6/10, ISO 17510-2, ASTM D3045-92.

## Regulatory Identification

A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.

## Predicate Devices

- Mirage Quattro Full Face Mask ([K113127](/device/K113127.md))
- ComfortGel Nasal Mask ([K092835](/device/K092835.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized caduceus-like symbol with three human profiles facing right, representing health and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 24, 2015

Sky Wise Medical Instrument (Shenzhen) Co., Ltd. C/O Field Fu Official Correspondent Shenzen Joyantech Consulting Co., Ltd. 4th Floor, Jinhui Building Nanhai BLVD. Nanshan District Shenzhen, Guang Dong, 518000 CHINA

Re: K150685

Trade/Device Name: Skynector CPAP Mask Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator Regulatory Class: Class II Product Code: BZD Dated: July 17, 2015 Received: July 23, 2015

#### Dear Mr. Fu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply 

{1}------------------------------------------------

with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

*Tejashri Purohit-Sheth, M.D.*

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Erin I. Keith M.S. Director Division of Anesthesiology, General Hospital. Respiratory. Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K150685

Device Name Skynector CPAP Mask

#### Indications for Use (Describe)

The Skynector CPAP Mask channels airflow non-invasively to a patient from a positive arway pressure device such as a continuous positive airway pressure (CPAP) or bi-level system.

The Skynector CPAP Mask is to be used by adult patients (>661b />30kg) for whom positive airway pressure has been prescribed.

The Skynector CPAP Mask is intended for single patient re-use in the home environment and/or multi-patient reuse in the hospital / institutional environment.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# SECTION 05 510(k) Summary

This summary of 510(K) safety and effectiveness information is submitted as Required by requirements of SMDA and 21 CFR §807.92.

## 1 Administrative Information

| Date of Summary<br>prepared          | Feb, 28, 2014                                                                                                                                                                                                                                                                                                              |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer<br>information          | Company title:<br>Sky Wise Medical Instrument (Shenzhen) Co.<br>Ltd.<br>Company address:<br>No 12 South Ping Xi Road Xinsheng<br>Community, Longgang Street, Longgang<br>District, Shenzhen, Guangdnog, China<br>Phone: +86-755-28491103<br>Fax: +86-755-28494339<br>Contact Person: Hu hanhan<br>E-mail: 942526346@qq.com |
| Submission<br>Correspondent          | Shenzhen Joyantech Consulting Co., Ltd.<br>4th Floor, Jinhui Building, Nanhai BLVD,<br>Nanshan District, Shenzhen, Guangdong,<br>China.<br>Contact person: Mr. Field.Fu<br>E-Mail: cefda13485@163.com<br>QQ: 670312758<br>Website: www.cefda.com                                                                           |
| Image: Joyantech logo                |                                                                                                                                                                                                                                                                                                                            |
| Establishment registration<br>number | No.                                                                                                                                                                                                                                                                                                                        |

### 2 Device Information

| Type of 510(k)<br>submission: | Traditional         |
|-------------------------------|---------------------|
| Trade Name:                   | Skynector CPAP Mask |
| Model:                        | FM-02, NM-03        |

{5}------------------------------------------------

Version:A/0

| Product: Skynector CPAP Mask |  |
|------------------------------|--|

Ventilator, Non-Continuous (Respirator). Classification name: Anesthesiology Review Panel: BZD Product Code: Device Class: ll 868.5905 Regulation Number: |

### 3 Predicate Device Information

| Sponsor:       | ResMed Limited;<br>Respironics, Incorporated.            |
|----------------|----------------------------------------------------------|
| Device:        | Mirage Quattro Full Face Mask;<br>ComfortGel Nasal Mask. |
| 510(K) Number: | K113127, K092835                                         |

### 4 Device Description

The Skynector CPAP masks are designed based on human facial shape and structure, and the operating characteristics during application, they are injection mold of hard PC plastic and soft silicone rubber materials, with high level of air tightness, convenient operation and comfortable to wear, etc. They are used for the interfacing devices that are used to provide users with Continuous Positive Airway Pressure (CPAP) ventilation or biphasic positive airway pressure (BiPAP) ventilation treatment.

### 5 Intended Use

The Skynector CPAP Mask channels airflow non-invasively to a patient from a positive airway pressure device such as a continuous positive airway pressure (CPAP) or bi-level system.

The Skynector CPAP Mask is to be used by adult patients (>661b />30kg) for whom positive airway pressure has been prescribed.

The Skynector CPAP Mask is intended for single patient re-use in the home environment and/or multi-patient reuse in the hospital /institutional environment.

### 6 Indications for Use

{6}------------------------------------------------

The Skynector CPAP Mask channels airflow non-invasively to a patient from a positive airway pressure device such as a continuous positive airway pressure (CPAP) or bi-level system.

The Skynector CPAP Mask is to be used by adult patients (>661b />30kg) for whom positive airway pressure has been prescribed.

The Skynector CPAP Mask is intended for single patient re-use in the home environment and/or multi-patient reuse in the hospital /institutional environment.

# 7 Technological characteristics of the proposed device compared to the predicate device

The proposed device and the Predicate device have the same intended use, design principle, and similar material composition. The differences don't impact safety and effectiveness of the device. The proposed device is substantially equivalent to the predicate devices. The compared details are listed in tables below:

|                                       | Subject Device:                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Predicate Device:                                                                                                                                                                                                                                                                                                                   | Remarks:                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                   |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                       | Device:<br>Skynector CPAP Full-face Mask<br>Manufacturer:<br>Sky Wise Medical<br>Instrument(Shenzhen) Co.,Ltd.                                                                                                                                                                                                                                                                                                                                                         | Device:<br>Mirage Quattro Full<br>FaceMask<br>Manufacturer:<br>ResMed LTD.                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                   |
|                                       | 510(k) Number:<br>To be determined                                                                                                                                                                                                                                                                                                                                                                                                                                     | 510(k) Number:<br>K113127                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                   |
| Intended<br>Use                       | The Skynector CPAP Mask<br>channels airflow non-<br>invasively to a patient from a<br>positive airway pressure<br>device such as a continuous<br>positive airway pressure<br>(CPAP) or bi-level system.<br>The Skynector CPAP Mask is<br>to be used by adult patients<br>(>66lb / >30kg) for whom<br>positive airway pressure has<br>been prescribed.                                                                                                                  | The Mirage Quattro channels<br>airflow non-invasively to a<br>patient from a positive<br>airway pressure device such<br>as a continuous positive<br>airway pressure (CPAP) or<br>bi-level system.<br>The Mirage Quattro is to be<br>used by adult patients (>30<br>kg) for whom positive<br>airway pressure has been<br>prescribed. | Both masks are<br>noninvasive and<br>the meanings of<br>intended use are<br>same.                                                                                                                                                            |                                                                                                                                                                                                                                                                                   |
|                                       | Subject Device:                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Predicate Device:                                                                                                                                                                                                                                                                                                                   | Remarks:                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                   |
|                                       | intended for single patient re-<br>use in the home environment<br>and/or multi-patient reuse in<br>the hospital /institutional<br>environment.                                                                                                                                                                                                                                                                                                                         | single-patient re-use in the<br>home environment and<br>multi-patient re-use in the<br>hospital/institutional<br>environment.                                                                                                                                                                                                       |                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                   |
| Patient<br>Population                 | Patients(>30 kg)                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Patients > 66 lbs/30 kg                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                   |
| Environment<br>of use                 | Home or Hospital/institutional<br>Environment                                                                                                                                                                                                                                                                                                                                                                                                                          | Home or<br>Hospital/institutional<br>Environment                                                                                                                                                                                                                                                                                    | Same                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                   |
| Product<br>Code                       | BZD                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | BZD                                                                                                                                                                                                                                                                                                                                 | Same                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                   |
| Provided<br>Sterile or<br>Non-Sterile | Provided Non-Sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Provided Non-Sterile                                                                                                                                                                                                                                                                                                                | Same                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                   |
| Patient<br>Usage<br>Type              | Single Patient Use in the<br>home or multi-patient use in<br>the hospital/institutional<br>environment.                                                                                                                                                                                                                                                                                                                                                                | Single Patient Use in the<br>home or multi-patient use in<br>the hospital/institutional<br>environment.                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                   |
| Design                                | Face interface and headgear                                                                                                                                                                                                                                                                                                                                                                                                                                            | Face interface and headgear                                                                                                                                                                                                                                                                                                         | Same                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                   |
| Material                              | FM-02:<br>bracket, swivel, connector,<br>elbow: Polycarbonate;<br>forehead pad, cushion, valve:<br>Silicone Rubber;<br>pick off port:<br>Polyvinylchloride;<br>faceplate: Polypropylene;<br>headgear: nylon, polyester.<br>FM-03:<br>bracket, faceplate, swivel,<br>connector, support arm<br>assembly, elbow:<br>Polycarbonate:<br>forehead pad, cushion, valve:<br>Silicone Rubber;<br>pick off port:<br>Polyvinylchloride;<br>headgear: nylon, polyester.<br>FM-05: | Moulded plastic and silicone<br>components and fabric /<br>nylon headgear.                                                                                                                                                                                                                                                          | All the<br>materials<br>composed the<br>Skynector<br>CPAP masks<br>were evaluated<br>per ISO 10993-<br>1.<br>Based on the<br>testing<br>performed, the<br>differences will<br>not affect the<br>safety or<br>effectiveness of<br>the device. |                                                                                                                                                                                                                                                                                   |
|                                       | Subject Device:                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Predicate Device:                                                                                                                                                                                                                                                                                                                   | Remarks:                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                   |
|                                       | bracket, faceplate, swivel,<br>connector, support arm<br>assembly, cushion clip,<br>elbow: Polycarbonate;<br>forehead pad, cushion, valve:<br>Silicone Rubber;<br>pick off port:<br>Polyvinylchloride;<br>headgear: nylon, polyester.                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                   |
| Number of<br>Pick off ports:          | One                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Two                                                                                                                                                                                                                                                                                                                                 | The design<br>difference will<br>not affect the<br>safety or<br>effectiveness of<br>the device.                                                                                                                                              |                                                                                                                                                                                                                                                                                   |
| Shape and<br>size of<br>faceplate     | The shape of the faceplate<br>follows the contours of the<br>face allowing clearance of<br>facial features, covering the<br>mouth and nose for therapy.<br>The faceplate attaches to a<br>silicone cushion via a<br>retaining ring.                                                                                                                                                                                                                                    | The shape of the faceplate<br>and cushion clip offered in<br>various sizes to ensure<br>adequate fit over the<br>extended patient population.                                                                                                                                                                                       | The design<br>difference will<br>not affect the<br>safety or<br>effectiveness of<br>the device.                                                                                                                                              |                                                                                                                                                                                                                                                                                   |
| Shape and<br>size of<br>cushion       | The cushions of the FM-02<br>and FM-05 are made of single<br>layer silicone with damping<br>groove to increase the<br>comfort.<br>The cushion of the FM-03 is<br>made of double layer silicone<br>to increase the comfortable fit.                                                                                                                                                                                                                                     | The previously cleared mask<br>provides characteristics seal<br>via silicone interface. The<br>previously cleared mask<br>offered in various sizes to<br>ensure adequate fit over the<br>extended patient population.                                                                                                               | Base on the<br>testing of<br>leakage, the<br>design<br>modification<br>will not affects<br>the safety or<br>effectiveness of<br>the device.                                                                                                  |                                                                                                                                                                                                                                                                                   |
| Safety Valve                          | Contain an anti-asphyxia<br>valve (AAV) to enable the<br>patient to breathe fresh air in<br>the event that airflow from the<br>flow generator is impeded.                                                                                                                                                                                                                                                                                                              | Contain an anti-asphyxia<br>valve (AAV) to enable the<br>patient to breathe fresh air in<br>the event that airflow from the<br>flow generator is<br>impeded.                                                                                                                                                                        | Same                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                   |
| Exhalation<br>device                  | No accessory exhalation<br>device is required. The full-<br>face masks incorporate<br>intentional vent holes to                                                                                                                                                                                                                                                                                                                                                        | No accessory exhalation<br>device is required. The<br>masks incorporate vents<br>to provide continuous air<br>flow to flush out and<br>minimize the amount of CO2                                                                                                                                                                   | Same                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                   |
|                                       | Subject Device:                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Predicate Device:                                                                                                                                                                                                                                                                                                                   | Remarks:                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                   |
|                                       | reduce carbon dioxide re-<br>breathing.                                                                                                                                                                                                                                                                                                                                                                                                                                | re-breathed by the patient.                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                   |
| Anatomical<br>Sites                   | Mouth and nose                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Mouth and nose                                                                                                                                                                                                                                                                                                                      | Same                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                   |
| Patient<br>Circuit<br>Connection      | 22 mm entrainment valve<br>elbow                                                                                                                                                                                                                                                                                                                                                                                                                                       | 22 mm entrainment valve<br>elbow                                                                                                                                                                                                                                                                                                    | Same                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                   |
| Pressure<br>Range                     | 4-20 cmH2O                                                                                                                                                                                                                                                                                                                                                                                                                                                             | 4-40 cmH2O                                                                                                                                                                                                                                                                                                                          | Verification<br>testing of the<br>modified device<br>did not raise<br>any new<br>questions of<br>safety and<br>efficacy.                                                                                                                     |                                                                                                                                                                                                                                                                                   |
| Number of<br>Mask Sizes               | FM-02: Three- small,<br>medium, and large<br>FM-03: Three- small,<br>medium, and large<br>FM-05: Three- small,<br>medium, and large                                                                                                                                                                                                                                                                                                                                    | Four- extra small, small,<br>medium, and large                                                                                                                                                                                                                                                                                      | The difference<br>will not affect<br>its function and<br>effectiveness.                                                                                                                                                                      |                                                                                                                                                                                                                                                                                   |
| Mask<br>Deadspace                     | size                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | FM-02<br>Full-<br>face<br>mask                                                                                                                                                                                                                                                                                                      | Mirage<br>Quattro Full<br>Face Mask                                                                                                                                                                                                          | The dead space<br>is calculated by<br>the identical<br>methods. The<br>dead space has<br>no pass/fail<br>criteria,<br>reportable value<br>only. It appears<br>a little larger<br>than the<br>predicate<br>device, but it<br>will not affect<br>its function and<br>effectiveness. |
|                                       | L                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 358                                                                                                                                                                                                                                                                                                                                 | 242                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                   |
|                                       | M                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 311                                                                                                                                                                                                                                                                                                                                 | No<br>information<br>available                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                   |
|                                       | S                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 280                                                                                                                                                                                                                                                                                                                                 | No<br>information<br>available                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                   |

### Table 1: Device Comparison Table to K113127

{7}------------------------------------------------

Product: Skynector CPAP Mask

{8}------------------------------------------------

SEC_005:001_510K Summary

Product: Skynector CPAP Mask

{9}------------------------------------------------

Product: Skynector CPAP Mask

{10}------------------------------------------------

| Version:A/0 |
|-------------|
|             |

|                                | Subject Device:                  |        | Predicate Device: | Remarks:                                                                                                                                                                                                                                                                                                                                   |                                           |                                                                                                                                                                                         |
|--------------------------------|----------------------------------|--------|-------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Leak rate                      |                                  |        |                   | This leak rate is<br>provided for the<br>healthcare<br>professional to<br>determine if it is<br>compatible with<br>CPAP or bi-<br>level therapy<br>device. It<br>appears the leak<br>rates of CPAP<br>mask is similar<br>with the Mirage<br>Quattro Full<br>Face Mask, leak<br>rate increases<br>stably when air<br>pressure<br>increases. |                                           |                                                                                                                                                                                         |
| Pressure<br>Drop               | Specifications<br>and models     |        | Flow<br>(L/min)   | Drop in<br>Pressure<br>(cmH2O)                                                                                                                                                                                                                                                                                                             | Mirage<br>Quattro<br>Full Face<br>Mask    | drop is<br>measured by the<br>same test<br>configuration. It<br>appears the<br>pressure drop of<br>the CPAP Full-<br>face mask is a<br>little larger than                               |
|                                | Full-Face<br>mask                | FM-02L | 100L/min          | 1.5                                                                                                                                                                                                                                                                                                                                        | 50L/min                                   | face mask is a<br>little larger than                                                                                                                                                    |
|                                |                                  | FM-02M | 50L/min           | 0.5                                                                                                                                                                                                                                                                                                                                        | 0.1cmH2O                                  | the Mirage<br>Quattro Full                                                                                                                                                              |
|                                |                                  | FM-02S | 50L/min           | 0.5                                                                                                                                                                                                                                                                                                                                        | 0.4cmH2O                                  | Face Mask.<br>The pressure                                                                                                                                                              |
| Valve<br>Open to<br>Atmosphere | FM-02 Full-Face<br>mask:0.5cmH2O |        |                   |                                                                                                                                                                                                                                                                                                                                            | Mirage Quattro<br>Full Face Mask:1.1cmH2O | drop is nominal<br>and it will not<br>affect its<br>function and<br>effectiveness.<br>The valve will<br>fully open to<br>atmosphere<br>when the mask<br>pressure reaches<br>the opening |

{11}------------------------------------------------

## SEC_005:001_510K Summary

Product: Skynector CPAP Mask

Version:A/0

|                                 | Subject Device:                  | Predicate Device:                         | Remarks:                                                                                                    |
|---------------------------------|----------------------------------|-------------------------------------------|-------------------------------------------------------------------------------------------------------------|
|                                 |                                  |                                           | pressure and<br>remains fully<br>open as the<br>patient inhales<br>and exhales.                             |
| Valve<br>Close to<br>Atmosphere | FM-02 Full-Face<br>mask:0.8cmH2O | Mirage Quattro<br>Full Face Mask:1.6cmH2O | The valve will<br>fully close to<br>atmosphere<br>when the<br>activation<br>pressure is met<br>or exceeded. |

### Table 2: Device Comparison Table to K092835

| Table 2: Device Comparison Table to K092835 |                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                   |
|---------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
|                                             | Subject Device:                                                                                                                                                                                                                                                                                                                                                                                  | Predicate Device:                                                                                                                                                                                                                                                                                                                                                                                                    | Remarks:                                                                          |
|                                             | Device:<br>Skynector CPAP Nasal Mask<br>Manufacturer:<br>Sky Wise Medical<br>Instrument(Shenzhen)<br>Co.,Ltd.<br>510(k) Number:<br>To be determined                                                                                                                                                                                                                                              | Device:<br>ComfortGel Nasal Mask<br>Manufacturer:<br>Respironics, Inc.<br>510(k) Number:<br>K092835                                                                                                                                                                                                                                                                                                                  |                                                                                   |
| Intended<br>Use                             | The reusable CPAP masks<br>with headgear are intended<br>for use as patient interface<br>for a CPAP or bi-level<br>system by adult<br>patients(>30 kg) for whom<br>CPAP or bi-level therapy<br>has been prescribed. The<br>nasal mask covers the nose,<br>and it is to be for single-<br>patient re-use in the home<br>and multi-patient re-use in<br>the hospital/institutional<br>environment. | The ComfortGel Nasal Mask is<br>intended to provide an<br>interface for application of<br>CPAP or bi-level therapy to<br>patients. The mask is for single<br>patient use in the home or<br>multi-patient use in the<br>hospital/institutional<br>environment. The mask is to be<br>used by patients (>66lbs/30kg)<br>for whom CPAP or bi-level<br>therapy has been prescribed<br>using a CPAP or bi-level<br>system. | Both masks are<br>noninvasive and the<br>meanings of<br>intended use are<br>same. |
| Patient<br>Population                       | Patients(>30 kg)                                                                                                                                                                                                                                                                                                                                                                                 | Patients > 66 lbs/30 kg                                                                                                                                                                                                                                                                                                                                                                                              | Same                                                                              |
| Environment<br>of use                       | Home or<br>Hospital/institutional<br>Environment                                                                                                                                                                                                                                                                                                                                                 | Home or Hospital/institutional<br>Environment                                                                                                                                                                                                                                                                                                                                                                        | Same                                                                              |
| Product<br>Code                             | BZD                                                                                                                                                                                                                                                                                                                                                                                              | BZD                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                                                                              |

{12}------------------------------------------------

SEC_005:001_510K Summary

Product: Skynector CPAP Mask

|                                       | Subject Device:                                                                                                                                                                                                                                                                                                                                                                                                                                               | Predicate Device:                                                                                                                                                                                                                               | Remarks:                                                                                                                                                                                                                         |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Provided<br>Sterile or<br>Non-Sterile | Provided Non-Sterile                                                                                                                                                                                                                                                                                                                                                                                                                                          | Provided Non-Sterile                                                                                                                                                                                                                            | Same                                                                                                                                                                                                                             |
| Patient<br>Usage<br>Type              | Single Patient Use in the<br>home or multi-patient use in<br>the hospital/institutional<br>environment.                                                                                                                                                                                                                                                                                                                                                       | Single Patient Use in the home<br>or multi-patient use in the<br>hospital/institutional<br>environment.                                                                                                                                         | Same                                                                                                                                                                                                                             |
| Design                                | Nasal interface and<br>headgear                                                                                                                                                                                                                                                                                                                                                                                                                               | Nasal interface and headgear                                                                                                                                                                                                                    | Same                                                                                                                                                                                                                             |
| Material                              | NM-01 :<br>bracket, faceplate, swivel,<br>support arm assembly,<br>quick clip, elbow:<br>Polycarbonate;<br>forehead pad, cushion:<br>Silicone Rubber;<br>headgear: nylon, polyester;<br>pick off port:<br>Polyvinylchloride.<br>NM-03:<br>bracket, faceplate, swivel,<br>support arm assembly,<br>cushion clip, elbow:<br>Polycarbonate;<br>forehead pad, cushion:<br>Silicone Rubber;<br>headgear: nylon, polyester;<br>pick off port:<br>Polyvinylchloride. | Faceplate: Polycarbonate<br>Face seal: Gel cushion with<br>polyester overlay<br>Cushion Flap: Silicone<br>Exhalation Elbow:<br>Polycarbonate<br>Exhalation Port: Delrin<br>Forehead Pad: Silicone<br>Headgear: UBL, Urethane<br>Foam, and Lycra | All the materials<br>composed the<br>Skynector CPAP<br>masks were<br>evaluated per ISO<br>10993-1.<br>Based on the testing<br>performed, the<br>modification will<br>not affect the safety<br>or effectiveness of<br>the device. |
| Number of<br>Pick off ports:          | One                                                                                                                                                                                                                                                                                                                                                                                                                                                           | None                                                                                                                                                                                                                                            | The design<br>difference will not<br>affect the safety or<br>effectiveness of the<br>device.                                                                                                                                     |
| Shape and<br>size of<br>faceplate     | The shape of the faceplate<br>follows the contours of the<br>face allowing clearance of<br>facial features, covering the<br>nose only for therapy.<br>The faceplate attaches to a<br>silicone cushion via a<br>retaining ring.                                                                                                                                                                                                                                | The shape of the faceplate<br>follows the contours of the face<br>allowing clearance of facial<br>features, covering the nose<br>only for therapy.<br>The faceplate attaches to a<br>silicone sealing flap via a<br>retaining ring.             | Same                                                                                                                                                                                                                             |

{13}------------------------------------------------

SEC_005:001_510K Summary

Product: Skynector CPAP Mask

|                                  | Subject Device:                                                                                                 | Predicate Device:                                                                                     | Remarks:                                                                                                                                        |  |                                                                                                             |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|--|-------------------------------------------------------------------------------------------------------------|
| Shape and<br>size of<br>cushion  | The cushions of the NM-01<br>and NM-03 are made of<br>double layer silicone to<br>increase the comfortable fit. | Gel cushion to allow for a<br>more comfortable fit and<br>improved mask performance<br>(fewer leaks). | The materials of<br>cushion are<br>different. The<br>design modification<br>will not affect the<br>safety or<br>effectiveness of the<br>device. |  |                                                                                                             |
| Safety Valve                     | N/A- nasal mask                                                                                                 | N/A- nasal mask                                                                                       | Same                                                                                                                                            |  |                                                                                                             |
| Exhalation<br>device             | No accessory exhalation<br>device is required. An<br>exhalation vent is<br>integrated.                          | No accessory exhalation device<br>is required. 54 pin holes<br>exhalation ports are integrated.       | Based on the testing<br>performed, the<br>design difference<br>will not affect the<br>safety or<br>effectiveness of the<br>device.              |  |                                                                                                             |
| Anatomical<br>Sites              | Nose                                                                                                            | Nose                                                                                                  | Same                                                                                                                                            |  |                                                                                                             |
| Patient<br>Circuit<br>Connection | 22 mm exhalation elbow                                                                                          | 22 mm exhalation elbow                                                                                | Same                                                                                                                                            |  |                                                                                                             |
| Pressure<br>Range                | 4-20 cmH2O                                                                                                      | 4-30 cmH2O                                                                                            | Verification testing<br>of the modified<br>device did not raise<br>any new questions<br>of safety and<br>efficacy.                              |  |                                                                                                             |
| Number of<br>Mask Sizes          | NM-01: Four- petite, small,<br>medium, and large;<br>NM-03: Three- small,<br>medium, and large.                 | Four- petite, small, medium,<br>and large                                                             | The difference will<br>not affect its<br>function and<br>effectiveness.                                                                         |  |                                                                                                             |
| Mask<br>Dea…

---

**Source:** [https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BZD/K150685](https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BZD/K150685)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
