Apex Medical Corp. iCH CPAP with PVA 9S-007XXX Series
Applicant
Apex Medical Corp.
Product Code
BZD · Anesthesiology
Decision Date
Sep 15, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5905
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
This device is intended to provide continuous positive airway pressure (CPAP) for the treatment of adult Obstructive Sleep Apnea (OSA) patients who are spontaneously breathing. It is intended for single patient reuse in the home environment.
Device Story
The iCH CPAP with PVA 9S-007XXX Series is a non-continuous ventilator providing continuous positive airway pressure (CPAP) or auto-titrating pressure (APAP) to treat adult Obstructive Sleep Apnea. The device takes ambient air, pressurizes it via an internal blower, and delivers it to the patient through a mask. It includes a built-in heated humidifier to reduce dryness. The device features Pressure Variation Algorithm (PVA) to provide pressure relief during expiration. It is intended for single-patient reuse in home environments. Healthcare providers prescribe the device and set parameters; patients operate the device daily. The device monitors breathing patterns to automatically titrate pressure and manage apnea/hypopnea events. Output is delivered as pressurized air; clinical benefit includes symptom relief and treatment of OSA.
Clinical Evidence
Bench testing only. No clinical data presented. Verification included safety/EMC (ANSI/AAMI ES60601-1, IEC 60601-1-2), firmware validation, noise testing (ISO 11201/7779), performance testing (ISO 17510-1, ISO 8185), PVA function comparison, and biocompatibility (ISO 10993).
Technological Characteristics
Non-continuous ventilator; 4-20 cmH2O pressure range; 22mm air outlet. Materials include plastic impeller. Power: 60W (24V DC). Features: heated humidifier, APAP auto-titration, PVA (3 levels), altitude/pressure compensation. Standards: ANSI/AAMI ES60601-1, IEC 60601-1-2, ISO 17510-1, ISO 8185, ISO 10993 series. Connectivity: Not specified.
Indications for Use
Indicated for adult patients with Obstructive Sleep Apnea (OSA) who are spontaneously breathing. Intended for home use.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized symbol. The symbol consists of three human profiles facing right, with flowing lines beneath them, representing the department's mission to protect the health of all Americans.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 15, 2015
Apex Medical Corp. Mr. Frank Lin Director of Quality Management Division No. 9, Min Sheng St. Tu-Cheng District New Taipei City, 23679 Taiwan
Re: K150111
Trade/Device Name: iCH CPAP with PVA 9S-007XXX Series Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (Ippb) Regulatory Class: Class II Product Code: BZD Dated: August 14, 2015 Received: August 17, 2015
Dear Mr. Frank Lin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K150111
Device Name iCH CPAP with PVA 9S-007XXX Series
Indications for Use (Describe)
This device is intended to provide continuous positive airway pressure (CPAP) for the treatment of adult Obstructive Sleep Apnea (OSA) patients who are spontaneously breathing. It is intended for single patient reuse in the home environment.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| <div> <span> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary- iCH CPAP with PVA 9S-007XXX series
Date Prepared: 8/14/2015
| Applicant name: | Apex Medical Corp. |
|-----------------------|--------------------------------------------------------------------------------|
| Contact Person: | Frank Lin |
| Address: | No.9, Min Sheng St., Tu-Cheng, New Taipei City, 23679, Taiwan |
| Phone number: | 886-2-22683100 |
| Fax numbers: | 886-2-22686525 |
| Device name | Trade name: Apex Medical Corp. iCH CPAP with PVA 9S-007XXX Series |
| | Common name: CPAP |
| | Classification name: |
| | Non-continuous ventilator Class II in accordance with 21 CFR 868.5905 |
| Classification | VENTILATOR, NON-CONTINUOUS (RESPIRATOR) |
| | Regulation Number: 868.5905 |
| | Medical Specialty: Anesthesiology |
| | Product Code: BZD |
| | Device Class: II |
| Predicate Device | APEX MEDICAL CORP. iCH CPAP with PVA 9S-007XXX series (K141522) |
| Reason for Submission | Modification of original APEX MEDICAL CORP. iCH CPAP with PVA 9S-007XXX series |
{4}------------------------------------------------
### Indications for Use
This device is intended to provide continuous positive airway pressure (CPAP) for the treatment of adult Obstructive Sleep Apnea (OSA) patients who are spontaneously breathing. It is intended for single patient reuse in the home environment.
## Device Description
The modified iCH CPAP with PVA 9S-007XXX Series is a modification of predicate iCH CPAP with PVA 9S-007XXX (K141522). It changed the material of impeller as well as PCB layout compare to the predicate device. A built-in heated humidifier of iCH CPAP with PVA 9S-007XXX Series is designed to increase the humidity of the air from the CPAP thereby relieving the symptoms of a dry nose and throat resulting from constant airflow that some patients may experience.
### Substantial Equivalence
The new device has the following similarities to the previously cleared predicate device(s)
- A Same intended use
- A Same operating principle
- A Same technology
- A Same manufacturing process
| Device<br>Characteristic | Modified<br>iCH CPAP with PVA<br>9S-007XXX series<br>(New Device) | Predicate<br>iCH CPAP with PVA<br>9S-007XXX<br>(K141522) | Comment |
|--------------------------|-------------------------------------------------------------------|----------------------------------------------------------|------------|
| Intended user | Adult | Adult | Equivalent |
| Air Outlet | 22mm | 22mm | Equivalent |
| User | single patient reuse | single patient reuse | Equivalent |
| Operating Environment | +5 ~ 35℃<br>15 ~ 95%<br>Non-condensing | +5 ~ 35℃<br>15 ~ 95%<br>Non-condensing | Equivalent |
| Pressure Range | 4 ~ 20 cmH2O | 4 ~ 20 cmH2O | Equivalent |
{5}------------------------------------------------
| Device<br>Characteristic | Modified<br>iCH CPAP with PVA<br>9S-007XXX series<br>(New Device) | Predicate<br>iCH CPAP with PVA<br>9S-007XXX<br>(K141522) | Comment |
|-------------------------------------------------|-------------------------------------------------------------------|----------------------------------------------------------|------------|
| Pressure Increment | 0.5 cmH2O | 0.5 cmH2O | Equivalent |
| Pressure Ramp Time | 0~45 min, 5 minutes<br>per step | 0~45 min, 5 minutes<br>per step | Equivalent |
| Pressure Compensate | Yes | Yes | Equivalent |
| Altitude Compensate | Yes | Yes | Equivalent |
| Automatically Titrates<br>Pressure in APAP Mode | Yes (iCH Auto) | Yes (iCH Auto) | Equivalent |
| Expiration Pressure<br>Release<br>(PVA) | Yes<br>Three Constant<br>Levels (C1, C2, C3) | Yes<br>Three Constant<br>Levels (C1, C2, C3) | Equivalent |
| Alarm System | Error 001~005<br>Warning 001~007,<br>Low P | Error 001~005<br>Warning 001~007,<br>Low P | Equivalent |
| Impeller Material of<br>Blower | Plastic | Aluminum | Equivalent |
| Power Supply | 60W (DC 24V, 2.5A) | 90W (DC 24V, 3.75A) | Equivalent |
| Heater Platform | 60W (DC 24V, 2.5A) | 90W (DC 24V, 3.75A) | Equivalent |
Table of comparison with predicate devices
Design verification tests were performed on the new device(s) based on the risk analysis and product requirements. In addition, the PVA function test was conducted with predicate device(s). The verified items are as follows:
- (1) Safety and EMC: according to ANSI/AAMI ES60601-1: 2005+C1:09+A2:10, IEC 60601-1-2: 2007 (Edition 3) and FDA reviewer guidance 638 (Excerpts Related to EMI from November 1993 Anesthesiology and Respiratory Devices Branch (includes EMI standard)).
- Firmware Validation: FDA Guidance for the Content of Premarket Submissions (2) for Software Contained in Medical Devices (May 11, 2005)
- Noise Test (Refer to ISO 11201:2010/ ISO 7779:2010) (3)
- Performance Tests (Refer to ISO 17510-1:2007): (4)
- a) Pressure Long-term Stability Test
- b) Dynamic Pressure Stability
- c) Static Pressure Stability
- d) Maximum Temperature
{6}------------------------------------------------
- e) Maximum Flow
- Limitation of Maximum Pressure f)
- g) Overflow Test
- h) Expected Useful Life Validation
- Humidification Performance (Refer to ISO 8185:2007) i)
- PVA Function Comparison Test (5)
- a) Endurance Operation and Dynamic Testing (Refer to ISO 17510-1:2007)
- b) APAP Auto titration & AHI Test
- c) The comparison for PVA Waveform with different breathing frequencies and flow volumes
- d) PVA feature handling conditions of disordered breathing (apneas and hypopneas)
- e) The comparison for PVA Waveform upper and lower pressure control
- Comparison for PVA Pressure Average and Tolerance Error f)
- g) Performance Characteristics of PVA Comparison
- Biocompatibility Assessment (6)
- a) Genotoxicity, Carcinogenicity and reproductive toxicity test (ISO 10993-3:2003)
- b) In vitro cytotoxicity test (ISO 10993-5:2009)
- c) Implantation test (ISO 10993-6:2007)
- d) Irritation and skin sensitization (ISO 10993-10:2010)
- e) Systemic toxicity (ISO 10993-11:2006)
- Particle Test: Refer US EPA PM2.5:1997 (7)
- VOC Test: Refer US EPA TO-15:1999 (8)
The modified iCH CPAP with PVA 9S-007XXX Series complies with the applicable voluntary and mandatory standards as following:
- (1) ANSI/AAMI ES60601-1:2005+C1:09+A2:10. Medical Electrical Equipment Part 1: General Requirements for Basic Safety and Essential Performance (FDA recognition number: 19-5)
- (2) IEC 60601-1-2:2007. Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests. (FDA recognition number: 19-1)
- (3) FDA reviewer guidance 638 (Excerpts Related to EMI from November 1993
{7}------------------------------------------------
Anesthesiology and Respiratory Devices Branch (includes EMI standard)
- (4) FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 11, 2005)
- (5) ISO 5367:2000. Breathing tubes intended for use with anaesthetic apparatus and ventilators. (FDA recognition number: 1-46)
- (6) ISO 11201:2010. Noise emitted by machinery and equipment -- Determination of emission sound pressure levels at a work station and at other specified positions in an essentially free field over a reflecting plane with negligible environmental corrections.
- (7) ISO 7779:2010. Measurement of airborne noise emitted by information technology and telecommunications equipment.
- (8) ISO 17510-1:2007. Sleep apnoea breathing therapy -- Part 1: Sleep apnoea breathing therapy equipment.
- (9) ISO 8185:2007. Respiratory tract humidifiers for medical use -- Particular requirements for respiratory humidification systems.
- (10) Determination Of Volatile Organic Compounds (VOCs) In Air Collected In Specially-Prepared Canisters And Analyzed By Gas Chromatography/Mass Spectrometry (GC/MS):1999.
- (11) US EPA National Ambient Air Quality Standards (NAAQS) PM2.5:1997.
- (12) ISO 10993-1:2009/(R) 2013. Biological Evaluation Of Medical Devices -- Part 1: Evaluation And Testing Within A Risk Management Process. (FDA recognition number: 2-156)
- (13) ISO 10993-3:2003. Biological evaluation of medical devices -- Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicity. (FDA recognition number: 2-175)
- (14) ISO 10993-5:2009. Biological evaluation of medical devices -- Part 5: Tests for in vitro cytotoxicity. (FDA recognition number: 2-153)
- (15) ISO 10993-6:2007. Biological evaluation of medical devices -- Part 6: Tests for local effects after implantation. (FDA recognition number: 2-120)
- (16) ISO 10993-10:2010. Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization. (FDA recognition number: 2-173)
- (17) ISO 10993-11:2006. Biological evaluation of medical devices -- Part 11: Tests for systemic toxicity. (FDA recognition number: 2-176)
- (18) IEC 60601-1-11:2010. Medical Electrical Equipment Part 1-11: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Requirements For Medical Electrical Equipment And Medical Electrical Systems
{8}------------------------------------------------
Apex Medical Corp.
Used In The Home Healthcare Environment. (FDA recognition number: 19-6)
In conclusion, the above tests demonstrate that the modified iCH CPAP with PVA 9S-007XXX Series perform as safe and effective as the predicate device - iCH CPAP with PVA 9S-007XXX (K141522). The relevant test reports are described in this submission. Therefore, we state that the modified iCH CPAP with PVA 9S-007XXX Series is substantially equivalent to the predicate device - iCH CPAP with PVA 9S-007XXX Series (K141522).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.