APEX MEDICAL CORP. ICH CPAP WITH PVA 9S-007XXX SERIES
K141522 · Apex Medical Corp. · BZD · Oct 6, 2014 · Anesthesiology
Device Facts
Record ID
K141522
Device Name
APEX MEDICAL CORP. ICH CPAP WITH PVA 9S-007XXX SERIES
Applicant
Apex Medical Corp.
Product Code
BZD · Anesthesiology
Decision Date
Oct 6, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5905
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
This device is intended to provide continuous positive airway pressure (CPAP) for the treatment of adult Obstructive Sleep Apnea (OSA) patients who are spontaneously breathing. It is intended for single patient reuse in the home environment.
Device Story
Device provides continuous positive airway pressure (CPAP) to treat adult OSA; home-use device for single-patient reuse. Incorporates Pressure Variation Algorithm (PVA) to provide exhalation pressure relief for patients struggling against high CPAP pressures. PVA reduces pressure during transition from inspiration to expiration; maintains reduced pressure during expiration; returns to therapeutic pressure at end of expiration. PVA automatically suspends if apneas detected; restarts after apneas disappear. Includes built-in heated humidifier to mitigate dry nose/throat symptoms. Device operates in CPAP or APAP modes. Healthcare providers use device settings to titrate therapy; patients use device at home to manage OSA symptoms.
Clinical Evidence
Bench testing only. Verification included safety and EMC testing per IEC 60601-1 and IEC 60601-1-2, side-by-side waveform testing against predicates, and algorithm analysis covering step functions, PVA waveforms across various breathing frequencies/volumes, and PVA performance during disordered breathing (apneas/hypopneas).
Technological Characteristics
CPAP device with 22mm air outlet; pressure range 4-20 cmH2O; 0.5 cmH2O increments. Features pressure ramp (0-45 min), pressure compensation, and altitude compensation. Includes integrated heated humidifier. Software-based Pressure Variation Algorithm (PVA) provides three levels of exhalation pressure relief. Operates in CPAP or APAP modes. Complies with IEC 60601-1 and IEC 60601-1-2.
Indications for Use
Indicated for adult patients with Obstructive Sleep Apnea (OSA) who are spontaneously breathing. Intended for home use.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
APEX MEDICAL XT Auto CPAP with PVA 9S-005720 (K112079)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the top half of the circle. Inside the circle is a stylized image of a human figure, with three overlapping profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 6,2014
Apex Medical Corporation Mr. Alan Chang Vice President No. 9, Minsheng Street Tucheng, New Taipei City 23679 TAIPEI
Re: K141522
Trade/Device Name: iCH CPAP with PVA 9S-007XXX Series Regulation Number: 21 CFR 868.5905 Regulation Name: Ventilator, Non-Continuous (Respirator) Regulatory Class: II Product Code: BZD Dated: September 2, 2014 Received: September 4, 2014
Dear Mr. Chang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Chang
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Susan Runno, DDS, mA
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K141522
#### Device Name
iCH CPAP with PVA 9S-007XXX Series
#### Indications for Use (Describe)
This device is intended to provide continuous positive airway pressure (CPAP) for the treatment of adult Obstructive Sleep Apnea (OSA) patients who are spontaneously breathing. It is intended for single patient reuse in the home environment.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
{3}------------------------------------------------
# 510(k) Summary- iCH CPAP with PVA 9S-007XXX series
| Date Prepared: | September 2, 2014 |
|-----------------------|-----------------------------------------------------------------------|
| Applicant name: | Apex Medical Corp. |
| Contact Person: | Alan Chang |
| Address: | No.9, Min Sheng St., Tu-Cheng, New Taipei City, 23679, Taiwan |
| Phone number: | 886-2-22683100 |
| Fax numbers: | 886-2-22686525 |
| Device name | Trade name: Apex Medical Corp. iCH CPAP with PVA 9S-007XXX Series |
| | Common name: CPAP |
| | Classification name: |
| | Non-continuous ventilator Class II in accordance with 21 CFR 868.5905 |
| Classification | VENTILATOR, NON-CONTINUOUS (RESPIRATOR) |
| | Regulation Number: 868.5905 |
| | Medical Specialty: Anesthesiology |
| | Product Code: 73 BZD |
| | Device Class: II |
| Predicate Device | APEX MEDICAL iCH CPAP 9S-007XXX series (K120035) |
| | APEX MEDICAL XT Auto CPAP with PVA 9S-005720 (K112079) |
| Reason for Submission | Modification of original APEX MEDICAL iCH CPAP 9S-007XXX series |
{4}------------------------------------------------
## Indications for Use:
This device is intended to provide continuous positive airway pressure (CPAP) for the treatment of adult Obstructive Sleep Apnea (OSA) patients who are spontaneously breathing. It is intended for single patient reuse in the home environment.
## Device Description:
The iCH CPAP with PVA 9S-007XXX Series are intended to provide continuous positive airway pressure for the treatment of adult obstructive sleep apnea (OSA). It is a modification of iCH CPAP 9S-007XXX (K120035) with the same algorithm of XT Auto with PVA 9S-005720 (K112079). It shares the same construction and auto adjustment algorithm with iCH CPAP 9S-007XXX but adds the identical Pressure Variation Algorithm with the predicate device XT Auto with PVA 9S-005720 (K112079). The Pressure Variation Algorithm provides a release pressure during expiration phase which is designed for patients who have difficulty exhaling against high CPAP pressure. It works by reducing the CPAP setting pressure in the normal breath status of expiratory phase from the treatment pressure in either CPAP or APAP mode.
A built-in heated humidifier of iCH CPAP with PVA 9S-007XXX Series are designed to increase the humidity of the air from the CPAP thereby relieving the symptoms of a dry nose and throat resulting from constant airflow that some patients may experience. The modified iCH CPAP series (with PVA) are from the iCH CPAP 9S-007XXX (K120035), but the algorithm for pressure variation algorithm(PVA) of the device comes from XT Auto with PVA (K112079).
The PVA Pressure Variation Alqorithm for the subject and predicated device (K112079) is an exhalation pressure relief function which is designed for patients who have difficulty exhaling against high CPAP pressure. Pressure can be reduced to three different levels (according to patient need) during the transition from Inspiration to Expiration phase. This reduced pressure level (depending on the PVA setting) will be maintained during Expiration phase and will return to the therapeutic pressure in the end of Expiration phase. In order to ensure the effectiveness of CPAP therapy, PVA will be automatically suspended if apneas are detected. After apneas disappear, PVA will re-start.
{5}------------------------------------------------
## Substantial Equivalence
The new device has the following similarities to the previously cleared predicate device(s)
- > Same intended use
- > Same operating principle
- A Same technology
- A Same manufacturing process
| Device<br>Characteristic | iCH CPAP with<br>PVA 9S-007XXX<br>series<br>(New Device) | Predicate1<br>iCH CPAP<br>9S-007XXX<br>(K120035) | Predicate2<br>XT Auto with<br>PVA 9S005720<br>(K112079) | Comment |
|-------------------------------------------------------|----------------------------------------------------------|--------------------------------------------------|---------------------------------------------------------|---------------------------------------------------------|
| Intended user | Adult | Adult | Adult | Equivalent |
| Outlook &<br>Construction | Image: iCH CPAP with PVA 9S-007XXX series | Image: iCH CPAP 9S-007XXX | Image: XT Auto with PVA 9S005720 | Subject is<br>identical to<br>Predicate 1<br>(K120035), |
| Air Outlet | 22mm | 22mm | 22mm | Equivalent |
| User | Single-user,<br>multi-use | Single-user,<br>multi-use | Single-user,<br>multi-use | Equivalent |
| Operating<br>Environment | +5 ~ 35 °C<br>15 ~ 95%<br>Non-condensing | +5 ~ 36 °C<br>10 ~ 95% | +5 ~ 35 °C<br>15 ~ 95%<br>Non-condensing | Equivalent |
| Pressure<br>Range | 4 ~ 20 cmH2O | 4 ~ 20 cmH2O | 4 ~ 20 cmH2O | Equivalent |
| Pressure<br>Increment | 0.5 cmH2O | 0.5 cmH2O | 0.5 cmH2O | Equivalent |
| Pressure Ramp<br>Time | 0~45 min, 5<br>minutes per step | 0~45 min, 5<br>minutes per step | 0~45 min, 5<br>minutes per step | Equivalent |
| Pressure<br>Compensate | Yes | Yes | Yes | Equivalent |
| Altitude<br>Compensate | Yes | Yes | Yes | Equivalent |
| Automatically<br>Titrates<br>Pressure in<br>APAP mode | Yes (iCH Auto) | YES (iCH Auto) | Yes | Identical |
{6}------------------------------------------------
| Device<br>Characteristic | iCH CPAP with<br>PVA 9S-007XXX<br>series<br>(New Device) | Predicate1<br>iCH CPAP<br>9S-007XXX<br>(K120035) | Predicate2<br>XT Auto with<br>PVA 9S005720<br>(K112079) | Comment |
|--------------------------------------------|----------------------------------------------------------|--------------------------------------------------|---------------------------------------------------------|-------------------------------------------------------|
| Expiration<br>Pressure<br>Release<br>(PVA) | Yes<br>Three Constant<br>Levels (1, 2, 3<br>cmH2O) | No | Yes<br>Three Constant<br>Levels(1,2,3<br>cmH2O) | Subject is<br>Identical to<br>Predicate2<br>(K112079) |
Table of comparison with predicate devices
Design verification tests were performed on the new device with the predicate device(s) as a result of the risk analysis and product requirements. The verified items are as follows:
- (1) Safety and EMC: according to iCH CPAP 9S-007XXX series (K120035) verification procedures, including FDA reviewer guidance 638.pdf, IEC 60601-1& IEC 60601-1-2.
- FDA Draft Reviewer Guidance for Ventilators (July 1995) (2)
- (3) FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 11, 2005)
- (4) Side by side waveform testing at various pressure and frequency levels with the predicate device(s).
- (5) Algorithm analysis; by comparing the test items as below:
- a. Step function
- PVA waveform with different breathing frequencies and flow volumes b.
- C. PVA feature handling conditions of disordered breathing (apneas and hypopneas)
- PVA waveform upper and lower pressure control d.
- PVA pressure average and tolerance error e.
- f. Performance characteristics of PVA
In conclusion, the above tests demonstrate that the iCH CPAP with PVA 9S-007XXX Series is as safe, as effective, and performs as well as the predicate device - XT Auto CPAP with PVA 9S-005720 (K112079). The relevant test reports are described in this submission. Therefore, we state that the iCH CPAP with PVA 9S-007XXX Series is substantially equivalent to the predicate device - XT Auto CPAP with PVA 9S-005720 (K112079).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.