K131982 · Respironics, Inc. · BZD · Oct 18, 2013 · Anesthesiology
Device Facts
Record ID
K131982
Device Name
REMSTAR AUTO A-FLEX HT
Applicant
Respironics, Inc.
Product Code
BZD · Anesthesiology
Decision Date
Oct 18, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 868.5905
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The REMstar Auto A-Flex HT delivers positive airway pressure therapy for the treatment of Obstructive Sleep Apnea in spontaneously breathing patients weighing over 30kg (66 lbs). It is for use in the home or hospital/institutional environment.
Device Story
Microprocessor-controlled blower-based positive pressure system; delivers CPAP and Auto CPAP therapy. Inputs: patient breathing signals. Operation: device detects breathing events (apneas/hypopneas) and automatically adjusts pressure; includes ramp, GPAP extended duration ramp, and pressure relief (FLEX technologies). Output: positive airway pressure delivered via patient circuit (tubing/mask). Includes heated humidifier and optional heated tubing for comfort. Used in home or hospital/institutional settings; operated by patients or clinicians. Healthcare providers use device data to monitor therapy; pressure adjustments benefit patients by maintaining airway patency during sleep.
Clinical Evidence
No clinical data. Safety and effectiveness demonstrated via non-clinical design verification testing and adherence to consensus standards (ISO 14971, ISO 10993-1, IEC 60601-1, IEC 60601-1-2, IEC 62304, ISO 5356-1).
Technological Characteristics
Microprocessor-controlled blower; CPAP/Auto CPAP modes. Materials: patient tubing (15mm/22mm/heated), heated humidifier plate. Connectivity: not specified. Sterilization: not specified. Software: IEC 62304 compliant. Standards: ISO 14971, ISO 10993-1, IEC 60601-1, IEC 60601-1-2, ISO 5356-1.
Indications for Use
Indicated for treatment of Obstructive Sleep Apnea in spontaneously breathing patients weighing >30kg (66 lbs).
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
{0}------------------------------------------------
### Tab 4
## 510(K) Summary of Safety & Effectiveness
| Official Contact | Michelle Brinker<br>Senior Manager, Regulatory Affairs<br>Respironics, Inc.<br>1740 Golden Mile Highway<br>Monroeville, PA 15146<br>Phone: 724-387-4146<br>OCT 1 8 2013 |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared | June 25, 2013 |
| Trade Name | REMstar Auto A-Flex HT |
| Common Name | CPAP System |
| Classification Name | Ventilator, non-continuous (respirator) (21 CFR 868.5905,<br>Product Code BZD) |
| Predicate Device(s) | Respironics REMstar Auto A-Flex HT (K113068) |
| Reason for Submission | The REMstar Auto A-Flex HT is the result of modifications made<br>to the REMstar Auto A-Flex HT (K113068) to include GPAP<br>extended duration ramp and patient-adjustable ramp time,<br>several auto CPAP therapy modalities/features, and humidifier<br>options. |
### Substantial Equivalence
The modified device has the following similarities to the previously cleared predicate device:
- . Same Intended Use
- . Same operating principle
- . Same technology
- . Same manufacturing process
Design verification tests were performed on the REMstar Auto A-Flex HT as a result of the risk analysis and product requirements. All tests were verified to meet the required acceptance criteria. Respironics has determined that the modifications have no impact on the safety and effectiveness of the device. In summary, the device described in this submission is substantially equivalent to the predicate device.
{1}------------------------------------------------
The REMstar Auto A-Flex HT delivers positive airway pressure therapy for the treatment of Obstructive Sleep Apnea in spontaneously breathing patients weighing over 30kg (66 lbs). It is for use in the home or hospital/institutional environment.
## Device Description
The REMstar Auto A-Flex HT is a microprocessor controlled blower based positive pressure system which is comprised of the therapy device, a heated humidifier and patient tubing (15mm, 22mm, or heated tubing).
The REMstar Auto A-Flex HT delivers CPAP and Auto CPAP therapy. With CPAP therapy, the device provides a continuous positive airway pressure throughout the entire therapy session. With Auto CPAP therapy, the device provides a positive airway pressure that automatically adjusts to the patient's needs as various breathing events, such as apneas and hypopneas, are detected.
In addition to the CPAP and Auto CPAP therapy, the REMstar Auto A-Flex HT incorporates several optional features to aid with patient comfort. These features include ramp, adjustable pressure relief (FLEX technologies), and humidification. Two ramp options are available, ramp and GPAP, in which therapy pressure gradually increases (ramps) to the prescription pressure. The therapy pressure increases to the prescription pressure over minutes with the standard ramp feature and over days with the GPAP extended duration ramp feature. Humidification options include both a heated humidifier and heated tubing. The heated humidifier adjusts the level of humidification by varying the temperature of a heated plate used to heat up a chamber of water. Optional heated tubing can then be used to maintain that air at a desired temperature until it reaches the patient's mask.
The REMstar Auto A-Flex HT is intended for use with a patient circuit that connects the device to a patient interface device (mask). A typical patient circuit consists of patient tubing (15mm, or heated tubing) and an exhalation device (if one is not present in the mask). When a heated humidifier is attached to the therapy device, the patient circuit connects to the air outlet port of the heated humidifier.
## Non-Clinical Tests
Design verification tests were performed for each modification to the REMstar Auto A-Flex HT based on the risk analysis. The testing confirmed that device met the pre-determined acceptance criteria. Results from the performance testing demonstrate that the REMstar Auto A-Flex HT meets its performance specifications, raises no new issues of safety or effectiveness, and is substantially equivalent to the identified predicate device.
The REMstar Auto A-Flex HT has been designed to meet the requirements of the following FDA Recognized Consensus Standards:
- . ISO 14971 Medical devices - Application of risk management to medical devices
{2}------------------------------------------------
- . ISO 10993-1 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
- . IEC 60601-1 Medical Electrical Equipment - Part 1: General Requirements for Safety
- IEC 60601-1-2 Medical Electrical Equipment Part 1-2: General Requirements for Safety; . Electromagnetic Compatibility - Requirements and Tests
- IEC 62304 Medical device software Software life cycle processes .
- . ISO 5356-1 Anaesthetic and respiratory equipment - Conical connectors: Part 1: Cones and sockets
# inical Tests
Clinical tests were not required to demonstrate the safety and effectiveness of the REMstar Auto A-Flex HT. Product functionality has been adequately assessed by non-clinical tests.
# Conclusion
The REMstar Auto A-Flex HT has passed all of the aforementioned non-clinical tests and required no clinical tests in order to demonstrate safety or effectiveness. It is therefore concluded that the REMstar Auto A-Flex HT is substantially equivalent to the predicate device in terms of safety and effectiveness.
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of an eagle or other bird-like figure with outstretched wings, rendered in a simple, bold line style.
Public Health Service
Food and Drug Administration 0903 New Hampshire Avenue Document Control Center - WO66-G60 Silver Spring, MD 20993-0002
#### October 18, 2013
Respironics, Incorporated Ms. Michelle Brinker Senior Manager, Regulatory Affairs 1740 Golden Mile Highway Monroeville PA 75146
Re: K131982
Trade/Device Name: REMstar Auto A-Flex HT Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (Intermittent Positive Pressure Breathing-IPPB) Regulatory Class: II Product Code: BZD Dated: September 17, 2013 Received: September 18, 2013
Dear Ms. Brinker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{4}------------------------------------------------
Page 2 - Ms. Brinker
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Mary Byner -S
Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K131982 510(k) Number (if known):
Device Name: REMstar Auto A-Flex HT
Indications for Use:
The REMstar Auto A-Flex HT delivers positive airway pressure therapy for the treatment of Obstructive Sleep Apnea in spontaneously breathing patients weighing over 30kg (66 lbs). It is for use in the home or hospital/institutional environment.
Prescription Use _ > (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Anya C. Harry-S
Digitally signed by Anya C. Harry-5
DN: cn=US, o=U.S. Government, ou=HHS,
ou=FDA, ou=People, cn=Anya C. Harry-5,
0.9.2342.19200300.100.1.1=0011315300
Date: 2011.10.18 12:37:23 -04'00'
(Division Sign-Off) Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices. 510(k) Number K131982
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.