K121642 · Apex Medical Corp. · BZD · Jun 6, 2013 · Anesthesiology
Device Facts
Record ID
K121642
Device Name
WIZARD 230 NASAL PILOW MASK
Applicant
Apex Medical Corp.
Product Code
BZD · Anesthesiology
Decision Date
Jun 6, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5905
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
WiZARD 230 Nasal Pillow Mask is intended to provide an interface for Continuous Positive Airway Pressure (CPAP) or bi-level therapy. This mask is intended for single-patient reuse in the home and multi-patient, multi-use in the hospital environment. This mask only can be used on patients greater than 30 kg for whom positive airway pressure (CPAP or bi-level system) has been prescribed.
Device Story
The WiZARD 230 Nasal Pillow Mask is a non-invasive patient interface for CPAP or bi-level therapy. It directs airflow from a positive pressure source to the patient's nostrils via a nasal pillow cushion. The device consists of a mask frame, cushion, elbow with integrated vent holes for CO2 flushing, and adjustable headgear (straps). It connects to standard air delivery tubing via a quick-release swivel connector. Used in home or hospital settings, the mask is secured to the patient's face by the user or a clinician. The vent holes ensure continuous air leak to minimize CO2 rebreathing. By providing a secure, comfortable seal, the device facilitates effective delivery of prescribed positive airway pressure, supporting treatment for patients with OSA.
Clinical Evidence
Bench testing only. The device underwent performance testing in compliance with ISO 17510-2, including CO2 rebreathing (normal and single fault conditions) and cleaning/disinfection protocols. Biocompatibility of all patient-contacting materials was verified per ISO 10993-1. No clinical data was required or provided.
Technological Characteristics
Materials: molded plastic, PU foam, fabric, nylon. Interface: nasal pillow cushion with frame. Energy: passive (pneumatic). Connectivity: standard conical connectors for air delivery tubing. Sterilization: supplied non-sterile. Standards: ISO 10993-1 (biocompatibility), ISO 17510-2 (performance/safety). Features: integrated vent holes on elbow for CO2 flushing, quick-release tube connector.
Indications for Use
Indicated for adult patients (>30 kg) with obstructive sleep apnea (OSA) requiring CPAP or bi-level positive airway pressure therapy. Contraindicated for use over open wounds prone to infection.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
{0}------------------------------------------------
K121642
JUN
6 2013
WiZARD 230 Nasa I Pillow Mask 510(k) Rev.003
## 2. 510 (K) SUMMARY
Date Prepared: May 31, 2012 (Revised on Jun 4, 2013)
Official Contact Frank Lin
## QA Manager
APEX Medical Corp.
No.9, Min Sheng St., Tu-Cheng, New Taipei City , 23679, Taiwan
Tel: +886-2-22685568
Fax:+886-2-22686525
Classification Reference 21 CFR 868.590 Product Code BZD non-continuous ventilator
Common/Usual Name CPAP Mask
Proprietary Name WiZARD 230 Nasal Pillow Mask
ResMed Swift™ FX Nasal Pillow Mask Predicate Device (K090244)
Reason for submission New device
#### Intended Use/Indications for use
| Indications for Use : | WiZARD 230 Nasal Pillow Mask is intended<br>to provide an interface for Continuous<br>Positive Airway Pressure (CPAP) or bi-level<br>therapy. This mask is intended for single-<br>patient reuse in the home and multi-patient,<br>multi-use in the hospital environment. This<br>mask only can be used on patients greater<br>than 30 kg for whom positive airway<br>pressure (CPAP or bi-level system) has<br>been prescribed. |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Patient Population : | Adults with OSA |
| Environment of Use : | Hospital, home |
| Contraindications : | The masks should not be placed over open<br>wounds that are prone to infection. |
#### Device Description
The WiZARD 230 Nasal Pillow Mask provides an interface such that airflow from a positive pressure source is directed to the patient's nostril. The mask is held in place with adjustable headgear that straps the mask to the face.
WiZARD 230 Nasal Pillow Mask is safe when used under the conditions and
Page 9 of 54
{1}------------------------------------------------
purposes intended as indicated in the labeling provided with the product.
WiZARD 230 Nasal Pillow Mask is a prescription device supplied nonsterile.
## Device Feature
Image /page/1/Picture/5 description: The image shows a diagram of a device with multiple numbered parts. The device includes a main body labeled as '9', with other components attached to it. A hose, labeled '5', connects to the main body via a series of smaller parts, including a connector labeled '4'. The diagram also highlights other parts such as '1', '2', '3', '6', '7', '8', '10', '11', and '12', indicating a detailed breakdown of the device's assembly.
| | Accessory | | Accessory |
|---|--------------------|----|-----------------------------|
| 1 | Mask Frame | 7 | Quick Release Button |
| 2 | Mask Cushion | 8 | Swivel Hose |
| 3 | Cushion Fixed Base | 9 | Side Strap |
| 4 | Elbow | 10 | Back Strap |
| 5 | Short Tube | 11 | Tube Retainer |
| 6 | Tubing Connector | 12 | Tube Retainer Fixed<br>base |
#### Technological Characteristics
WiZARD 230 Nasal Pillow Mask provides a secure interface via mask frame and cushion. This mask incorporate vent holes on elbow to provide continuous air leak to flush out and minimize the amount of CO2 rebreathed by the patient. The incorporation of these vent holes does not interfere with the intended performance of the mask.
WiZARD 230 Nasal Pillow Mask connects to an air delivery tube between the mask and the positive pressure source via standard conical connectors. The tube has a quick release structure makes it can easily connect/disconnect from positive pressure source.
Strap of WiZARD 230 Nasal Pillow Mask is constructed by using molded plastic and PU foam/fabric/nylon headgear. All the components are fabricated using materials deemed safe. Tube retain structure includes in strap to keep the air delivery tube at fix position.
Page 10 of 54
{2}------------------------------------------------
. .
## SE Comparative Table
,
.
| Features | Predicated Device | Proposed Device |
|---------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------|
| Trade Name | ResMed Swift™ FX Nasal Pillow<br>Mask | WIZARD 230 Nasal<br>Pillow Mask |
| Indications for Use | The Swift™ FX channels airflow<br>noninvasively to a patient from a<br>positive airway pressure (PAP)<br>device such as a continuous<br>positive airway pressure (CPAP)<br>or bilevel system.<br>The Swift™ FX is:<br>• To be used by adult patients (><br>66 lb/30 kg) for whom positive<br>airway pressure has been<br>prescribed<br>• Intended for single-patient re-<br>use in the home environment<br>and multipatient re-use in the<br>hospital/institutional<br>environment. | Identical |
| Environment of Use | Hospital, home | Same |
| Patient Population | Adult | Identical |
| Single patient,<br>Multi-use | Single patient multi-use | Same |
| Components | Cushion 、Elbow 、Breath Tube、<br>Strap(back/side) | Similar |
| Materials | Silicon 、PC、PP、<br>nylone/neoprene | Similar |
| Comparative<br>Testing for Safety<br>and Efficacy | Compliance to ISO 17510-2 | Same |
.
.
.
{3}------------------------------------------------
#### Summary of Test:
| Attribute | Requirement | Parameter | Result |
|------------------|----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|--------|
| Biocompatibility | All materials used in<br>the construction of<br>the mask shall be<br>compliant with ISO<br>10993-1 | All material which may<br>contact the patient or<br>the clinician must be<br>biocompatible | PASS |
| Performance | Overall performance<br>shall be compliant to<br>ISO17510-2 | Test items described in<br>ISO 17510-2 | PASS |
| Safety | Overall performance<br>shall be compliant to<br>ISO17510-2 | Test items including<br>cleaning/disinfection<br>and CO2 rebreathing<br>(normal and single fault<br>condition) | PASS |
| Shelf Life | Should be compliant<br>to product<br>specification | 5 years shelf life | PASS |
## Substantial equivalence
This premarket notification section 510(k) shows WiZARD 230 Nasal Pillow Mask substantially equivalent to ResMed Swift™ FX Nasal Pillow Mask.
The characteristics of the WiZARD 230 Nasal Pillow Mask (shown as below) are similar to the predicates device, ResMed Swift™ FX Nasal Pillow Mask. Both masks have the same intended use, environment of use and patient population. Based on verification testing, we conclude that although few minor differences in technological characteristics (e.g. sealing, side strape, mask assembly) of these two masks, but these differences do not affect the safety or effectiveness of the WiZARD 230 Nasal Pillow Mask. .
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure with outstretched wings.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 6, 2013
Apex Medical Corporation Mr. Frank Lin Quality Assurance Manager No. 9, Min Sheng Street Tu-Cheng City New Taipei City, Taiwan 23679
Re: K121642
Trade/Device Name: WiZARD 230 Nasal Pillow Mask Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: II Product Code: BZD Dated: May 24, 2013 Received: May 28, 2013
Dear Mr. Lin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical.Device. Amendments, or todevices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 – Mr. Lin
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Image /page/5/Picture/6 description: The image shows a signature above the text "Tejashri Purohit-Sheth, M.D., Clinical Deputy Director, DAGRID, FOR, Sincerely yours". The signature is illegible. The text is likely part of a letter or document.
Kwame Ulmer, M.S. Acting Division Director · Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{6}------------------------------------------------
## 1. INDICATIONS FOR USE STATEMENT
# Indications for Use
510(k) Number (if known): _ K | 2| 642
Device Name: WiZARD 230 Nasal Pillow Mask
Indications for Use:
WiZARD 230 Nasal Pillow Mask is intended to provide an interface for Continuous Positive Airway Pressure (CPAP) or bi-level therapy. This mask is intended for single-patient reuse in the home and multi-patient, multi-use in the hospital environment. This mask only can be used on patients greater than 30 kg for whom positive airway pressure (CPAP or bi-level system) has been prescribed.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
-- Concurrence of CDRH; Office of Device Evaluation (ODE)
Anya C. Harry

| Digitally signed by | Anya C. Harry |
|---------------------|------------------------------------------------------------------------------------------------------------|
| DN: | C=US, O=U.S. Government, OU=HHS, OU=FDA, OU=People, CN=Anya C. Harry, 0.9.2342.19200300.100.1.1=0011315590 |
| Date: | 2013.06.05 17:03:18 -04'00' |
(Division Sign-Off) (Division of Anestheslology, General Hospital Infection Control, Dental Devices
510(k) Number: k121642
Page 8 of 54
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.