3B WILLOW

K112271 · 3B Products, LLC · BZD · Feb 16, 2012 · Anesthesiology

Device Facts

Record IDK112271
Device Name3B WILLOW
Applicant3B Products, LLC
Product CodeBZD · Anesthesiology
Decision DateFeb 16, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5905
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Willow™ channels airflow noninvasively to a patient from a positive airway pressure (PAP) device or a bilevel system. The Willow™ is: (1) to be used by adult patients (> 66lb / 30 kg) for the treatment of sleep disordered breathing, such as obstructive sleep apnea (OSA), for whom positive airway pressure has been prescribed. (2) to be used for single-patient reuse in the home environment.

Device Story

3B Willow™ is a nasal pillow mask interface; directs airflow from positive airway pressure (PAP) or bilevel system to patient's nose. Held in place by adjustable headgear. Used in home environment by adult patients. Device channels prescribed therapeutic pressure (4-20 cmH2O) to treat sleep disordered breathing. Healthcare provider selects mask based on patient fit; patient self-administers therapy at home. Benefits include noninvasive delivery of prescribed airway pressure to maintain airway patency during sleep.

Clinical Evidence

Bench testing only. Performance characteristics (passive exhalation port flow, resistance to flow, dead space) compared against predicate ResMed Mirage Swift (K042403).

Technological Characteristics

Nasal pillow mask system; includes headgear. Connector: 22mm. Therapy pressure range: 4-20 cmH2O. Materials: Latex-free. Multi-size options. Non-sterile. Mechanical interface only; no software or electronics.

Indications for Use

Indicated for adult patients (> 66lb / 30 kg) with sleep disordered breathing, including obstructive sleep apnea (OSA), prescribed positive airway pressure therapy. For single-patient home use.

Regulatory Classification

Identification

A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K112271 Section 5: 510(k) Summary 3B Willow™ Nasal Mask FEB 1 6 2012 Section 5: 510(k) SUMMARY [As required by 21 CFR §807.92(c)] Date Prepared: July 27, 2011 Official Contact: Alex Lucio Managing Partner 3B Products, LLC 1142 N. Scenic Highway Lake Wales, FL 33853 Tel: (863) 676-5948 Email: alucio@3Bproducts.com | Device Trade Name: | 3B Willow™ Nasal Mask | | |---------------------------------------------|-------------------------------------|-----------------| | Device Common Name/<br>Classification Name: | Vented Nasal Mask | | | Classification: | 21 CFR §868.5905, 73 BZD (CLASS II) | | | Predicate Devices: | Manufacturer: | Resmed | | | Trade Name: | Mirage Swift II | | | 510(k) Number: | K042403 | Manufacturer: Respironics ComfortLite 2 Trade Name: 510(k) Number: K082558 Fisher & Paykel Manufacturer: Opus 360 Trade Name: K063036 510(k) Number: ## Device Description: The 3B Willow™ is a mask interface, of the nasal pillow variety, that directs airflow from a positive pressure device to the patient's nose. The mask is held in place with adjustable headgear that straps the mask to the face. The 3B Willow™ is safe when used under the conditions and purposes intended as indicated the labeling provided with the product. {2}------------------------------------------------ Section 5: 510(k) Summary 3B Willow™ Nasal Mask The 3B WillowTM is a prescription device supplied non-sterile. # Intended Use: The Willow™ channels airflow noninvasively to a patient from a positive airway pressure (PAP) device or a bilevel system. The Willow™ is: - (1) to be used by adult patients (> 66lb / 30 kg) for the treatment of sleep disordered breathing, such as obstructive sleep apnea (OSA), for whom positive airway pressure has been prescribed. - (2) to be used for single-patient reuse in the home environment. # Contraindications: None # Technological Characteristics Comparison: A comparative table of the Willow TM alongside the three identified predicate devices indicates that the Willow Nasal Pillows is substantially equivalent with the other three with respect to the form, performance, materials, structures and the ventilation characteristics. | Manufacturer | BMC | ResMed | Respironics | Fisher&Paykel | |---------------------|----------------------------------------|------------------------------------------------------------|-----------------------------------------|--------------------------------------------------| | Model | WillowTM<br>Nasal Pillows<br>System | Mirage Swift<br>II Nasal<br>Pillows<br>System<br>(K042403) | ComfortLiteTM 2<br>System<br>(K082558) | OpusTM 360 Nasal<br>Pillows Mask<br>(K063036) | | Pic | Image: BMC Willow Nasal Pillows System | Image: ResMed Mirage Swift II Nasal Pillows System | Image: Respironics ComfortLite 2 System | Image: Fisher&Paykel Opus 360 Nasal Pillows Mask | | Nasal pillows | Yes | Yes | Yes | Yes | | With headgear | Yes | Yes | Yes | Yes | | Latex free | Yes | Yes | Yes | Yes | | Multi Size | Yes | Yes | Yes | Yes | | Connector | 22mm | 22mm | 22mm | 22mm | | Therapy<br>Pressure | 4-20cmH20 | 4-20cmH20 | 4-30cmH20 | 3-25cmH20 | | Intentional | 4cmH20=20L/ | 4cmH20=20L | 4cmH20=22L/min | 5cmH20=22L/min | {3}------------------------------------------------ | leak | min<br>10cmH20=31.5<br>L/min<br>20cmH20=45L/min | /min<br>12cmH20=37<br>/min<br>20cmH20=49<br>L/min | 12cmH20=34L/min<br>20cmH20=42L/min | 11cmH20=33L/min<br>21cmH20=48L/min<br>25cmH20=52L/min | |------------|-------------------------------------------------|---------------------------------------------------|------------------------------------|-------------------------------------------------------| | Resistance | 1.0 cmH20 at 50L/min<br>3.1 cmH20 at 100L/min | 0.4 cmH20 at 50L/min<br>1.6 cmH20 at 100L/min | NA | 1.2 cmH20 at 50L/min<br>5.4 cmH20 at 100L/min | | Dead Space | 96ml | 91ml | 19-36ml(mask only) | 22ml(mask only) | To verify the substantial equivalence claim, the Willow™ was performance bench tested against the ResMed Mirage Swift (K042043). The results of the performance bench testing, along with a description of the testing protocol are found at Appendix A. In terms of performance characteristics (i.e. passive exhalation port flow, resistance to flow, and dead space), the two devices are substantially equivalent. {4}------------------------------------------------ Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 3B Products, LLC C/O Mr. Alex Lucio Managing Partner 1142 N. Scenic Highway Lake Wales, Florida 33853 FEB 1 6 2012 Re: K112271 Trade/Device Name: 3B Willow™ Nasal Mask Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator Regulatory Class: II Product Code: BZD Dated: February 4, 2012 Received: February 9, 2012 Dear Mr. Lucio: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device · Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Page 2 - Mr. Lucio {5}------------------------------------------------ Page 2 - Mr. Lucio Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to Sempilations ification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the I bailing) of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vours, h for Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health #### Enclosure {6}------------------------------------------------ 3B Willow™ Nasal Mask Section 4: Indications for Use K1112271 ## Indications for Use 510(k) Number: K112271 Device Name: 3B Willow™ Nasal Mask Indications For Use: The Willow™ channels airflow noninvasively to a patient from a positive airway pressure (PAP) device or a bilevel system The Willow ™ is: - (1) to be used by adult patients (> 66lb / 30 kg) for the treatment of sleep disordered breathing, such as obstructive sleep apnea (OSA), for whom positive airway pressure has been prescribed. - (2) to be used for single-patient reuse in the home environment. AND/OR Over-The-Counter Use _________________________________________________________________________________________________________________________________________________ × Prescription Use _ (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) y. Schulther (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: 以上1227/
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...