K110471 · A Plus Medical · BZD · Oct 7, 2011 · Anesthesiology
Device Facts
Record ID
K110471
Device Name
BABI PLUS INFANT NASAL CANNULA SYSTEM
Applicant
A Plus Medical
Product Code
BZD · Anesthesiology
Decision Date
Oct 7, 2011
Decision
SESE
Submission Type
Special
Regulation
21 CFR 868.5905
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
Single patient use device intended for use with neonates, infants and children under 10 Kg requiring a nasal prong interface during intermittent or continuous gas flow therapy in the hospital critical care unit.
Device Story
Babi.Plus Infant Nasal Cannula System is a single-patient-use nasal cannula interface for neonates, infants, and children (≤ 10 Kg). Device features 8 sizes and 10 mm corrugated tubing for connection to gas flow therapy equipment. It provides a nasal prong interface for delivery of intermittent or continuous gas flow and includes a port to monitor delivered pressure at the nasal prong. Used in hospital critical care units by clinicians. Device facilitates respiratory support by delivering gas flow to the patient's airway; pressure monitoring allows clinicians to verify delivered therapy, aiding in respiratory management and patient stabilization.
Clinical Evidence
Bench testing only. Dimensional analysis and performance testing (resistance to gas flow in inspiratory and expiratory circuits) were conducted to compare the subject device against the predicate. Results confirmed similar performance characteristics.
Technological Characteristics
Single-patient-use nasal cannula; 8 sizes; 10 mm corrugated tubing; integrated pressure monitoring port. Materials are identical to the predicate. Non-powered, passive interface device.
Indications for Use
Indicated for neonates, infants, and children ≤ 10 Kg requiring nasal prong interface for intermittent or continuous gas flow therapy in hospital critical care units. Contraindicated for patients > 10 Kg or those not requiring nasal prong interface.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for A Plus Medical. The logo features a stylized letter "a" enclosed in a circle on the left. To the right of the "a" is the text "A Plus Medical" in a bold, sans-serif font. Below "A Plus Medical" is the tagline "Solutions for Respiratory Care" in a smaller font.
K110471
#### 510(k) Summary 5.0.
February 18, 2010 Date:
# Owner:
A Plus Medical 5431 Avenida Encinas, STE G Carlsbad, CA 92008-4411 Tel: + 760-930-4025
+ 760-930-0040 Fax:
Owner/Operator Number: 10023166
# Official Contact:
Thomas C. Loescher + 760-930-4025 Tel: Fax: + 760-930-0040
Trade Names: Babs.Plus" Infant Nasal Cannula System
# Common/Usual Name:
Nasal Cannula, Infant Nasal Cannula
# Classification Name:
Ventilator, non-continuous (respirator) Device Name: Product Code: BZD Regulation: 868.5905 II Device Class:
Device: The Babi*Plus Nasal Cannula
#### Predicate Devices:
K093716 Number: Product Name: Babs.Plus" Infant Nasal Cannula System A Plus Medical Manufacturer: Product Codes: C100 & C102
#### Device Description:
Single patient use nasal cannula offered in 8 different sizes which have been specifically designed for patents ≤ 10 Kg. The device includes a short length of 10 mm corrugated tubing allowing connection to a variety of devices. The device also provides a means to monitor delivered pressure at the nasal prong.
#### Indications for Use:
Single patient use device intended for use with neonates, infants and children under 10 Kg requiring a nasal prong interface during intermittent or continuous gas flow therapy in the hospital critical care unit.
> 5411 Avenida Encinas, STE G Carlsbad, CA 92008-4411 2. +740-930-4025 Fax: +760-930-0040 ~ 13~
OCT - 7 2011
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for A Plus Medical. The logo features the letter "a" in a circle, followed by the words "A Plus Medical" in bold, sans-serif font. Below the company name is the tagline "Solutions for Respiratory Care" in a smaller font. The logo is simple and professional, and it conveys the company's focus on medical solutions for respiratory care.
### Contraindications:
Patients not requiring a nasal prong interface during intermittent or continuous gas flow therapy. Patients > 10 Kg.
# Patient Population:
Patient population of neonate (premature infant), infant and child
#### Environment of Use:
Hospital Critical Care Unit
# Comparative of Technological Characteristics:
The Babl*Plus Nasal Cannula is substantially equivalent in indications for use, environment of use, patient population, material and function to the identified predicate. Bench testing and dimensional analysis confirmed that the Babi*Plus Nasal Cannula and predicate device have similar performance characteristics. Testing was performed and the below identified differences were demonstrated:
. Resistance to gas flow in the inspiratory circuit.
| | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | 11 | 12 |
|---------------------------------------------------|------|------|------|-----|-----|-----|-----|-----|-----|-----|-----|-----|
| Gas Flow | | | | | | | | | | | | |
| Predicate | 0 | 0.1 | 0.2 | 0.3 | 0.4 | 0.5 | 0.7 | 0.9 | 1.1 | 1.3 | 1.6 | 1.9 |
| Proposed Device | <0.1 | <0.1 | 0.1 | 0.2 | 0.2 | 0.3 | 0.5 | 0.6 | 0.8 | 0.9 | 1.1 | 1.3 |
| Resistance to gas flow in the expiratory circuit. | | | | | | | | | | | | |
| Gas Flow | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | 11 | 12 |
| Predicate | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 0.1 | 0.1 | 0.1 | 0.1 | 0.1 |
| Proposed Device | <0.1 | <0.1 | <0.1 | 0.1 | 0.1 | 0.2 | 0.2 | 0.3 | 0.3 | 0.4 | 0.5 | 0.6 |
#### Conclusion:
.
The revised Babi*Plus Nasal Cannula is substantially equivalent to the predicate of Application K093716 because:
- Both are made from identical material.
- . Both have substantially equivalent performance.
- . Both have identical indications for use.
- . Both are used in identical patient populations.
- . Both are used in identical environments of use.
5411 Avenida Encinas, STE G Carlsbad, CA 92008-4411 28. +740-930-4025 Fax: +760-930-0040 ~ 14~
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an emblem featuring a stylized abstract design, resembling an eagle or bird in flight. The emblem is composed of three curved lines that converge at the bottom.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Thomas C. Loescher, R.R.T President A Plus Medical 5431 Avenida Encinas. Suite G Carlsbad, California 92008
OCT - 7 2011
Re: K110471
Trade/Device Name: Babi.Plus™ Infant Nasal Cannula System Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatorv Class: II Product Code: BZD Dated: September 7, 2011 Received: September 8, 2011
Dear Mr. Loescher:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
# Page 2- Mr. Loescher
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm 115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kesia Alexander
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology. General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
fr
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for A Plus Medical. The logo consists of the letter "a" in a circle, followed by the words "A Plus Medical" in bold font. Below the company name is the text "Solutions for Respiratory Care" in a smaller font.
# Indications for Use Statement
510(k) Number:
(To be assigned)
Device Name:
Babi.Plus"Infant Nasal Cannula System
Indications for Use:
Single patient use device intended for use with neonates, infants and children under 10 Kg requiring a nasal prong interface during intermittent or continuous gas flow therapy in the hospital critical care unit.
Prescription Use X (Part 21 CFR 801 Subpart D)
or Over-the-counter use
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
L. Schultze
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: u110471