K103410 · Childrens Medical Ventures, LLC · BZD · Mar 30, 2011 · Anesthesiology
Device Facts
Record ID
K103410
Device Name
NEOPAP SYSTEM
Applicant
Childrens Medical Ventures, LLC
Product Code
BZD · Anesthesiology
Decision Date
Mar 30, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5905
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The NeoPAP System is intended to provide continuous positive airway pressure (CPAP) for use in hospitals to treat newborns and infants, weighing less than 5 kg, with RDS or who are recovering from RDS. (Respiratory Distress Syndrome).
Device Story
NeoPAP System provides non-invasive nasal CPAP, constant flow, and resuscitation modes for infants < 5 kg. Inputs include air/oxygen gas supplies; system uses proportional solenoids to actively regulate pressure via Baby-Trak feature by monitoring pressure at patient's nose. Output is humidified air/oxygen mixture delivered via nasal cannula or mask. Operated by clinicians in hospital settings. Device features real-time pressure bar graph, alarm system for pressure/oxygen/occlusion/disconnect, and rotary knob for O2/pressure/flow adjustments. Software includes safety features like user confirmation for setting changes and auto-resetting occlusion alarms. Benefits include non-invasive respiratory support, reduced tissue damage via active pressure regulation, and improved clinical workflow through intuitive UI and standby mode. Compatible with external Fisher & Paykel MR850 humidifier.
Clinical Evidence
No clinical trials; evidence based on bench verification testing, software code reviews, unit/integration testing, biocompatibility, and clinical simulation (usability testing). Usability testing confirmed performance equivalent to predicate device. All test protocols met acceptance criteria.
Technological Characteristics
Non-continuous ventilator (BZD). Features proportional solenoids for pressure regulation, capacitive touch/rotary UI, and alarm system. Compatible with external Fisher & Paykel MR850 humidifier. Battery life: 2 hours. Patient interface: PI2 System (one-piece cannula/mask, corrugated tubing, exhalation body). Sterilization: Non-sterile. Connectivity: Standalone.
Indications for Use
Indicated for newborns and infants weighing < 5 kg with Respiratory Distress Syndrome (RDS) or recovering from RDS, for use in hospital settings.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
Predicate Devices
Guardian Neonate CPAP / Humidification System (K040862)
{0}------------------------------------------------
## TAB 5
K103410
NeoPAP System
MAR 3 0 2011
# 510(K) SUMMARY
| Official Contact | Scott Wright<br>Senior Regulatory Affairs Engineer<br>Children's Medical Ventures LLC<br>191 Wyngate Drive<br>Monroeville, PA 15146<br>Scott.Wright1@philips.com |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | 412-380-8812 t<br>724-681-2407 c<br>412-380-8850 f |
| Date of Submission | November 22, 2010 |
| Classification Reference | 21 CFR 868.5905 |
| Product Code | BZD - ventilator, non-continuous (respirator) |
| Common/Usual Name | CPAP System |
| Proprietary Name | NeoPAP System |
| Predicate Device(s) | Guardian Neonate CPAP / Humidification System (K040862) |
| Reason for submission | Modified Device |
# Substantial Equivalence
The modified device has the following similarities to the previously cleared predicate device:
- Similar intended use. 0
- Same operating principle. ם
- Same technology. 0
- ם Same manufacturing process.
Design verification tests were performed on the Children's Medical Ventures LLC NeoPAP System as a result of the risk analysis and product requirements. Children's Medical Ventures LLC has
{1}------------------------------------------------
determined that the modifications have no impact on the safety and effectiveness of the device. In summary, the device described in this submission is substantially equivalent to the predicate device.
The modified device complies with the applicable standards referenced in the Guidance for FDA Reviewers and Industry "Guidance for the Content of Pre-market Submissions for Software Contained in Medical Devices," May 2006.
Formal test protocols were written and executed to verify and validate the modified NeoPAP System. All tests that were created for the modified NeoPAP System had passing results with acceptance criteria successfully met, which demonstrates the safety & effectiveness of the system.
Testing included software code reviews, software unit testing, software integration testing, bench verification testing, biocompatibility testing, environmental testing, user manual/labeling inspection, drawing inspections, and a clinical simulation (usability testing confirmed that the NeoPAP System performs equivalently to the predicate device (Guardian Neonate CPAP / Humidification System, K040862). All tests had acceptable results or discrepancies that do not impact safety or effectiveness.
### Intended Use
The NeoPAP System is intended to provide continuous positive airway pressure (CPAP) for use in hospitals to treat newborns and infants, weighing less than 5 kg, with RDS or who are recovering from RDS. (Respiratory Distress Syndrome). The intended use of the NeoPAP System is unchanged from the predicate device (Guardian Neonate CPAP / Humidification System, commonly referred to as NeoPAP System, K040862).
### Device Description
The Children's Medical Ventures LLC NeoPAP System provides nasal CPAP (continuous positive airway pressure) as a non-invasive method of breathing support for infants and newborns with compromised respiratory systems who weigh less than 5 kg.
The NeoPAP System provides CPAP therapy delivered to patient through a nasal cannula or mask. The system has a feature called Baby-Trak which actively regulates pressure at all times by monitoring pressure at patient's nose and adjusting flow from oxygen and air gas supplies using proportional solenoids. The Baby-Trak feature eliminates the need for a closely-fitted patient interface that can damage fragile tissues.
In addition to CPAP mode, the NeoPAP System offers a constant flow mode where the device delivers a fixed flow of humidified air/oxygen mixture through the nasal prongs or mask. There is also
{2}------------------------------------------------
a resuscitation mode where the device delivers a fixed flow of unhumidified air/oxygen through the resuscitation port for use with a resuscitation or hyperinflation bag. The percentage of oxygen can be adjusted while in any mode by using a rotary knob on the user interface.
The NeoPAP System alarms when pressure levels or oxygen concentration are above or below alarm limits. The system also detects occlusion, disconnect, and other alarm conditions. The system displays a real-time bar graph that indicates delivered pressure.
# Predicate Comparison
The fundamental scientific technology of the NeoPAP System is unchanged from the predicate device (K040862). Children's Medical Ventures LLC has made the following changes to the previously cleared NeoPAP System to be considered for this submission:
- Flow Mode -- Flow Mode for the previously cleared system only delivered flow into an oxygen . hood, whereas the modified system enables the device to deliver a constant flow of oxygenated gas through a nasal cannula or mask. In the modified system, the process for adjusting flow settings was modified to match the process of adjusting pressure settings in CPAP Mode. Both processes now utilize a rotary knob.
- Resuscitation Mode The previously cleared system could only deliver 100% oxygen to the . patient while in Resuscitation Mode. Since research has shown that high levels of oxygen for an extended period of time can lead to eye damage, the modified system is capable of delivering a range of oxygen concentration between 21 – 100% while in Resuscitation Mode.
- Aesthetics The external design of the device has been modified to better align with the . Philips brand image.
- User Interface The user interface (UI) of the modified device was changed to emphasize . ease of use and to be more intuitive than the original design. Key changes are the addition of a flashing alarm bar, removal of the flow segment display, replacement of the majority of membrane keys with capacitive touch keys, and the addition of pressure, FiQ2, and battery status LEDs changing from white to red when a specific parameter causes an alarm. These changes allow for a less cluttered appearance and more visible feedback for the user to determine what parameter is in an alarm state.
- Occlusion Alarms Software was modified so that occlusion alarms only cause gas delivery . to stop in the case of high manifold or patient pressure. It is no longer halted in the case of flow readings being lower than expected. The software was also modified so that all occlusion alarms auto-reset. Once the trigger for the alarm is no longer present, the occlusion flag is clear and gas delivery will automatically restart if it had been halted.
{3}------------------------------------------------
- User Confirmation After Turning Rotary Knobs Software was modified so that the user must . confirm that settings should be changed after turning the O2 or pressure/flow setting control knobs. This will prevent inadvertent changes to settings.
- Humidification The original NeoPAP System utilized a Vapotherm humidification cartridge . which was internal to the device and a delivery circuit that utilized a heated water jacket to minimize rainout within the delivery line. However, due to Vapotherm's recall of the cartridge for potential of bacterial growth, the NeoPAP System was voluntarily recalled after less than a year on the market. For this reason, the modified NeoPAP System does not contain an internal humidifier, but is instead compatible with an external humidifier that is currently being demonstrated to be compatible with the NeoPAP System is the Fisher & Paykel MR850 Humidifier (K073706).
- Battery Life— The stand-alone battery life of the modified system is 2-hours instead of 1-hour t as in the original NeoPAP System. The longer battery life will be useful in the event that power is lost and a back-up generator is not available or that the device is used during intrahospital transport with no AC outlet available.
- Standby Mode The original NeoPAP System did not have a Standby Mode. It was added . to the modified NeoPAP System to accommodate system set-up prior to patient receiving therapy and care procedures where the patient is taken off of the device for a period of time.
- Configuration Menu For both the original and modified NeoPAP System it is necessary to . power off the device in order to enter the configuration menu. In order to allow the clinician to more easily adjust alarm parameters, several items were moved from the configuration menu to the delivery mode menu. The items that were moved are leak delay, high pressure delay, low pressure delay, pressure alarm band, and FiO2 alarm band.
- Device Packaging -- The original device was packaged using foam inserts. The modified . device utilizes suspension packaging which suspends the device in the airspace of the shipping container between two layers of highly resilient, low-slip film.
- Patient Interface Circuit The original NeoPAP System patient interface consisted of a . "generator" circuit into which the clinician would plug nasal prongs or nasal masks of an appropriate size for their patient. The tubing came over the top of the patient's head and between the eyes. The modified patient interface (commonly referred to as the PI2 System) consists of a one-piece cannula and one-piece mask. The tubing comes from both sides of the patient's head similar to an adult patient on oxygen. Additionally, a pressure tap was added to the most proximal part of the patient interface (at the patient) to allow the NeoPAP
{4}------------------------------------------------
System to better respond to changes in patient pressure to maintain the pressure level set by the clinician.
- Patient Interface Sizes/Shape Masks are still available in 2 sizes and are unchanged from . the original NeoPAP System in terms of overall footprint/outline. Cannulas are still available in 5 sizes and are unchanged from the original NeoPAP System in terms of prong spacing and prong diameter. The shape of the nasal prongs have been modified from a flare shape to a tapered shape for ease of manufacturing.
- Patient Interface Tubing and Connections –The pressure and delivery tubing in the original . NeoPAP patient interface was surrounded by a water jacket. It was connected via a slip fit to the patient tubing. The PI2 System utilizes corrugated delivery tubing packaged standard for Fisher & Paykel RT324 humidification circuits with a heated wire to minimize condensation so that a water jacket is not necessary. The pressure line is permanently bonded to the exhalation body to prevent misconnection. The patient tube is permanently bonded to a cork that plugs into the exhalation body.
- Patient Interface Exhalation The original NeoPAP patient interface had a series of holes in . the tubing located about 7 inches from the patient. Instead of holes in the tubing, the P12 has an exhalation body located about 15 inches from the patient with three large openings on different faces of the body so that the exhalation is not easily blocked regardless of orientation.
- Patient Interface Pop Off Valve The original NeoPAP patient interface had an umbrella .
- valve located directly at the patient in the generator which held the prongs or mask. The Pl2 . System utilizes the exact same valve made by the same supplier, but it is located in the exhalation body delivery air path.
- Patient Interface Bonnet The original NeoPAP patient interface included a cotton bonnet . available in 12 sizes to secure the "generator" circuit. The Pl2 System includes a bonnet available in 5 sizes with a hook and loop material to adjust the fit. It has a unique foam material on the inside lining to minimize shifting of the bonnet and adjustable bonnet clips to hold the PI2 tubing in place.
- Patient Interface Usage Type: The device manual of the origina! NeoPAP System specified . that the delivery tube, patient tube, and bonnet were single patient use while the nasal prongs and mask were single use. The device manual for the modified NeoPAP System indicates that entire patient interface is single patient use and lists acceptable agents/procedure for cleaning the exterior of the patient interface.
{5}------------------------------------------------
and the country of the country of the county of the county of the county of the county of
#### Traditional 510(k) Tab 5 – 510(k) Summary
Table 1 (follows) compares the NeoPAP System with the predicate device.
| Table 1 – Device Comparison | | |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Guardian Neonate CPAP /<br>Humidification System (K040862) | NeoPAP System |
| Intended Use | Intended to provide CPAP for use<br>in hospitals to treat newborns and<br>infants with RDS or are recovering<br>from RDS. (Respiratory Distress<br>Syndrome) May or may not include<br>humidification capabilities | Unchanged from K040862, other than<br>rewording.<br>The NeoPAP System is intended to<br>provide continuous positive airway<br>pressure (CPAP) for use in hospitals<br>to treat newborns and infants,<br>weighing less than 5 kg, with RDS or<br>who are recovering from RDS<br>(Respiratory Distress Syndrome). |
| Intended Environment of Use | Hospital | Unchanged from K040862. |
| Patient Population | Infants with or recovering from RDS<br>weighing less than 5 kg. | Unchanged from K040862. |
| Product Code | BZD | Unchanged from K040862. |
### Operating Mode Descriptions and Ranges
| Gas Delivered | Air / Oxygen Mixture | Unchanged from K040862. |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|
| CPAP Mode - Description | Device actively regulates pressure<br>at all times by monitoring pressure<br>and adjusting flow from air and O2<br>supplies using proportional<br>solenoids. | Unchanged from K040862. |
| CPAP Mode - Delivery<br>Location | Nasal prongs / mask. | Unchanged from K040862. |
| CPAP Mode - Range of<br>Pressure | 2 to 10 cmH2O | Unchanged from K040862. |
| CPAP Mode - %O2 Range | 21 – 100% | Unchanged from K040862. |
| Flow Mode - Description | Device delivers fixed flow of<br>humidified air/oxygen mixture. | Unchanged from K040862. |
스 """
. ---
:
{6}------------------------------------------------
の 2017年 10:00 PM 2017 10:
| Table 1 - Device Comparison | | |
|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Guardian Neonate CPAP /<br>Humidification System (K040862) | NeoPAP System |
| Flow Mode - Delivery<br>Location | Oxygen hood. | Nasal cannula / mask. |
| Flow Mode – Range of Flow | 5 to 15 L/min | 5 to 10 L/min |
| Flow Mode - %O2 Range | 21 - 100% | Unchanged from K040862. |
| Resuscitation Mode -<br>Description | Resuscitation Mode provides un-<br>humidified gas through the<br>resuscitation port for use with a<br>resuscitation or hyperinflation bag. | Unchanged from K040862. |
| Resuscitation Mode - Fixed<br>Flow | 10 L/min | Unchanged from K040862. |
| Resuscitation Mode - %O2<br>Range | 100% | 21 - 100% |
| Standby Mode | No | Yes |
| Humidification Method | Vapotherm Microporous Membrane | The Fisher & Paykel MR850 Humidifier<br>(K073706) has been demonstrated to<br>be compatible with the NeoPAP<br>System in this submission. |
| Range of Temperature of<br>Gas Delivered | 33 to 41 °C | The F&P MR850 humidifier does not<br>have communications with the NeoPAP<br>device so the temperature is set on the<br>humidifier instead of the NeoPAP<br>device. |
### User Interface / Display of Measured Data
| Circuit Pressure (bar graph<br>display) Range | $0-12 cmH_2O$ | Unchanged from K040862. |
|-----------------------------------------------|---------------|-------------------------------------------------------------------------------------------------------|
| %O₂ (window display) Range | 21 – 100% | Unchanged from K040862. |
| Gas Temperature Range | 10 to 50 °C | The F&P MR850 humidifier does not<br>have communication with the NeoPAP<br>device so the measured gas |
. . . . . . . . .
{7}------------------------------------------------
. . . . . .
| Tab 5 - 510(k) Summary | | |
|------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Table 1 – Device Comparison | | |
| | Guardian Neonate CPAP /<br>Humidification System (K040862) | NeoPAP System |
| | | temperature will only be displayed on<br>the humidifier and not on the NeoPAP<br>device. |
| Flow (bar graph) | Indicator only – indicates that<br>patient is inhaling / exhaling but<br>does not give actual flow reading. | Flow bar graph removed to prevent<br>confusion and provide a less cluttered<br>appearance. |
| Alarms | High / Low Air Supply Pressure<br>High / Low O₂ Supply Pressure<br>High / Low FiO2<br>High / Low Flow (Flow mode only)<br>High / Low Pressure (CPAP mode<br>only)<br>High / Low Temperature<br>High / Low Water Pressure<br>No Water Supply<br>No AC Power<br>No O₂ Sensor<br>Battery Not Present<br>Low Battery<br>Replace Battery<br>Pressure Unstable (CPAP mode<br>only)<br>Partial Occlusion<br>Occlusion<br>Leak<br>Unexpected Restart<br>Miscellaneous System Error Alarms | Unchanged from K040862 with the<br>exception that the NeoPAP device does<br>not have communication with the F&F<br>MR850 humidifier so the NeoPAP will<br>not alarm for high / low gas<br>temperature, high / low water pressure<br>or no water supply. However, the<br>humidifier has its own alarming system<br>for high / low gas temperature. |
| Alarm Visual Indicator | Backlit outline of a red bell. | Flashing alarm bar. Additionally,<br>pressure and FiO2 readings that are<br>typically displayed using white LEDs<br>change to red so that the user can<br>more easily determine which of the |
| Table 1 – Device Comparison | | |
| | Guardian Neonate CPAP /<br>Humidification System (K040862) | NeoPAP System |
| System Halt Conditions | Mains power disconnected<br>Air or O2 supply pressure over 93<br>psi<br>Internal battery voltage depleted<br>High pressure detected during<br>startup<br>Delivery tube disconnected during<br>operation<br>Ambient temperature out of range<br>Tube occluded during system start<br>up<br>Other miscellaneous halt conditions | Unchanged from K040862. |
| Input Types | Rotary knobs and membrane keys. | Rotary knobs, membrane keys, and<br>capacitive keys. |
| Configuration Menu Access | Turn off device and power on while<br>holding the Alarm Reset key. | Unchanged from K040862. |
| Configuration Menu Items | Alarm Sound Gap<br>Auto-Reset Setting<br>Patm<br>Language<br>Leak Delay<br>High Pressure Delay<br>Low Pressure Delay<br>Pressure Alarm Band<br>Temperature Alarm Band<br>FiO₂ Alarm Band | Alarm Sound Gap, Auto-Reset Setting,<br>Patm, and Language remain in<br>Configuration Menu, and Leak Delay,<br>High Pressure Delay, Low Pressure<br>Delay, Pressure Alarm Band, and FiO₂<br>Alarm Band where moved to the<br>Delivery Mode Menu so they could be<br>accessed without powering off the<br>device. Temperature Alarm Band item<br>does not exist in current software since<br>the F&P MR850 humidifier does not<br>have communication with the NeoPAP<br>device so it will not alarm for high / low<br>gas temperature. |
| Table 1 – Device Comparison | | |
| | Guardian Neonate CPAP /<br>Humidification System (K040862) | NeoPAP System |
| Power | AC and Battery | Unchanged from K040862. |
| Stand-alone Battery Life | 1 hour | 2 hours |
| Oxygen Sensor | Ventrex Oxygen Sensor (K963415) | Maxtec Oxygen Sensor (K972992). |
| Oxygen Sensor Location | Mixing block upstream of vapor<br>transfer cartridge | Moved to pressure manifold block to<br>allow FiO2 to also be measured in<br>Resuscitation Mode. |
| Patient Interface | | |
| Usage Type | Delivery tube, patient tube, bonnet,<br>are single patient use. Nasal<br>prongs / mask are single use. | All components are single patient use.<br>The entire patient interface can be<br>cleaned and reused on the same<br>patient. |
| Sterility Condition | Non-sterile | Unchanged from K040862. |
| Interface Design | Nasal prongs/mask is snapped into<br>"generator" which is connected to<br>dual lumen tubing that is routed<br>above the nose (secured by a clip<br>integral to the hat) to over the top of<br>the head. | Nasal cannula/mask is one-piece<br>molding that is connected to dual lume<br>pressure/delivery tube and single<br>lumen exhalation tube that are routed<br>to the sides of the head (secured by<br>bonnet clips) and then over the head. |
| Exhalation Management | Series of holes in tubing<br>approximately 7 inches from<br>patient. Tubing must be oriented<br>correctly so that holes are not<br>blocked. | Exhalation body approximately 15<br>inches from patient with 3 large<br>openings on different facets of the<br>exhalation body so that exhalation is<br>not easily blocked regardless of<br>orientation. |
| Patient Pressure<br>Measurement Location | At connection between<br>delivery/pressure tubing and patient<br>tube (approximately 10 inches from<br>patient). | At most proximal point of patient<br>interface (closest to patient) to provide<br>most accurate readings. |
| Pressure Relief Valve<br>Location | Located directly at the patient in the<br>"generator" which holds nasal | Located in the exhalation body delivery<br>air path approximately 15 inches |
| Table 1 - Device Comparison | | |
| Guardian Neonate CPAP /<br>Humidification System (K040862) | NeoPAP System | |
| prongs or mask | upstream from patient. | |
| Relief Valve Opening<br>Pressure (Measured at<br>Patient) | 18 cmH2O | 15 cmH2O |
| Nasal Cannula Sizes | Extra large, large, medium, small,<br>and extra small. | Unchanged from K040862. |
| Nasal Mask Sizes | Large and small. | Unchanged from K040862. |
| Bonnet Design | Cotton bonnet without mechanism<br>to adjust fit of bonnet. Laces<br>thread through bonnet eyelets to<br>secure patient interface. | Foam bonnet with hook and loop<br>mechanism to adjust fit of bonnet.<br>Adjustable bonnet clips to secure<br>patient interface. |
| Bonnet Sizes | Sizes 0 to 11. | Extra large, large, medium, small, and<br>extra small. |
and the comments of the production of the comments of the comments of the
.
:- -
{8}------------------------------------------------
. .
and the country of the country of the county
ﺴ
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
- --- · · · · ·
:
、
{9}------------------------------------------------
. . . . . . . .
- -- -- -
'
·
{10}------------------------------------------------
.
:
ﻣ
. .
:
.
. . . . . .
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
.
:
:
·
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a stylized eagle with three tail feathers and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circle around the eagle. The eagle is black, and the text is also black.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room-WO66-G609 Silver Spring, MD 20993-0002
Children's Medical Ventures, LLC C/O Mr. Scott Wright Senior Regulatory Affairs Engineer 191 Wyngate Drive Monroeville, Pennsylvania 15146
MAR 3 0 2011
Re: K103410
Trade/Device Name: NeoPAP System Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: II Product Code: BZD Dated: March 18, 2011 Received: March 23, 2011
Dear Mr. Wright:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{12}------------------------------------------------
Page 2 - Mr. Wright
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutlFDA/CentersOffices/CDRH/CDRHOffices/ucm] 15809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Nh for
Anthony Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{13}------------------------------------------------
### Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: _ _ _ NeoPAP System_
#### Indications for Use:
・・
The NeoPAP System is intended to provide continuous positive airway pressure (CPAP) for use in hospitals to treat newborns and infants, weighing less than 5 kg, with RDS or who are recovering from RDS. (Respiratory Distress Syndrome).
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
L. Schulte
on of Anesthesioner. General Infection Control and De
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.