← Product Code [BZD](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BZD) · K091129

# QUATTRO LT (K091129)

_Resmed, Ltd. · BZD · Aug 3, 2009 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BZD/K091129

## Device Facts

- **Applicant:** Resmed, Ltd.
- **Product Code:** [BZD](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BZD.md)
- **Decision Date:** Aug 3, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5905
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic

## Indications for Use

The Quattro LT channels airflow noninvasively to a patient from a positive airway pressure (PAP) device such as a continuous positive airway pressure (CPAP) or bilevel system. The Quattro LT is: to be used by adult patients (>66lb / >30kg) for whom positive airway pressure has been prescribed intended for single-patient re-use in the home environment and multipatient re-use in the hospital/institutional environment.

## Device Story

Quattro LT is a vented full-face mask interface for noninvasive delivery of positive airway pressure (PAP) from CPAP or bilevel systems. Device directs airflow to patient's nose and mouth; held in place by adjustable headgear. Features include integrated vent holes for CO2 flushing and an antiasphyxia valve (AAV) for ambient air access if flow generator fails. Connects to standard air delivery hoses via ISO 5356-1:2004 conical connectors. Used in home (single-patient) or hospital (multipatient) environments. Healthcare providers prescribe the device; patients use it to receive PAP therapy for conditions requiring respiratory support. Benefits include noninvasive airway pressure delivery to treat sleep-disordered breathing or other respiratory conditions.

## Clinical Evidence

No clinical data. Bench testing was deemed sufficient to demonstrate safety and efficacy, consistent with the predicate devices. Performance data confirmed CO2 flushing and pressure-flow characteristics are substantially equivalent to predicates.

## Technological Characteristics

Molded plastic and silicone interface; fabric/nylon headgear. Standard conical connectors (ISO 5356-1:2004). Biocompatible materials (ISO 10993-1). Vented design for CO2 flushing. Integrated antiasphyxia valve. Non-sterile, prescription-only device.

## Regulatory Identification

A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.

## Predicate Devices

- Mirage Quattro ([K063122](/device/K063122.md))
- Ultra Mirage II Mask ([K050359](/device/K050359.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

1091129 AUG (13 2009

**RESMED**

.

Quattro LT Traditional 510K

:

:

|                                                          | 510(k) SUMMARY         [As required by 21 CFR 807.92(c)]                                                                                                                                                                                                                                                                  |
|----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared                                            | 15 April 2009                                                                                                                                                                                                                                                                                                             |
| Official Contact                                         | Mr. Steven Lubke         Director Regulatory Affairs                                                                                                                                                                                                                                                                      |
| Device Trade Name                                        | Quattro™ LT                                                                                                                                                                                                                                                                                                               |
| Device Common Name/         Classification Name          | Vented Full Face Mask;         Accessory to Noncontinuous Ventilator (IPPB)                                                                                                                                                                                                                                               |
| Classification                                           | 21 CFR 868.5905, 73 BZD (Class II)                                                                                                                                                                                                                                                                                        |
| Predicate Devices                                        | Mirage Quattro (K063122)         Ultra Mirage II Mask (K050359)                                                                                                                                                                                                                                                           |
| Description                                              | The Quattro LT provides an interface such that airflow from a         positive pressure source is directed to the patient's nose and         mouth. The mask is held in place with adjustable headgear that         straps the mask to the face.                                                                          |
|                                                          | The Quattro LT is safe when used under the conditions and         purposes intended as indicated in the labeling provided with the         product.                                                                                                                                                                       |
|                                                          | The Quattro LT is a prescription device supplied non-sterile.                                                                                                                                                                                                                                                             |
| Intended Use                                             | The Quattro LT channels airflow noninvasively to a patient from         a positive airway pressure (PAP) device such as a continuous         positive airway pressure (CPAP) or bilevel system.                                                                                                                           |
|                                                          | The Quattro LT is:         to be used by adult patients (>66lb / >30kg) for whom         positive airway pressure has been prescribed         intended for single-patient re-use in the home environment         and multipatient re-use in the hospital/institutional         environment.                               |
| Technological         Characteristics         comparison | Comparison with predicate Mirage Quattro         The new device and the Mirage Quattro predicate both provide         a seal via a silicone interface. Both masks are also offered in         various sizes to ensure adequate fit over the extended patient         population.                                          |
|                                                          | Both masks incorporate vent holes to provide continuous air         leak to flush out and minimize the amount of CO2 rebreathed by         the patient. The design of the mask components is such that         the incorporation of the vent-holes does not interfere with the         intended performance of the masks. |

15 April 2009

: : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : :

. .

.

{1}------------------------------------------------

RESMED

Both masks contain an antiasphyxia valve (AAV) to enable the patient to breathe fresh air in the event that airflow from the flow generator is impeded.

Both masks connect to a conventional air delivery hose between the mask and the positive airway-pressure source via standard conical connectors (ref: ISO 5356-1:2004)

Both masks are constructed using molded plastic and silicone components and fabric / nylon headgear. All components in both masks are fabricated using materials deemed safe (ref: ISO 10993-1).

Both the new device and the predicate are designed to operate on the same ResMed flow generator settings. The pressureflow characteristics and flow impedance of both devices are substantially equivalent.

Both the new device and the predicate device can be reused in the home and hospital / institution environment.

The main difference between the Quattro LT and the Mirage Quattro is the design/geometry and how individual components interface with each other. Both masks are designed and constructed under ResMed's 21 CFR Part 820 compliant Quality Management System.

Clinical Data Use of Full Face Masks with CPAP or Bilevel therapy is proven technology and is well accepted by the medical community. Bench testing is sufficient to demonstrate safety and efficacy of the Quattro LT, as was the case with the predicate devices.

Performance Data

## Substantial Equivalence Conclusion

Quattro LT is substantially equivalent to the predicate devices:

The CO2 performance of the new device and the predicate

- it has the same intended use; .
Comparison with predicate Ultra Mirage II

device are substantially equivalent.

- it has similar technological characteristics to both predicates;
- it does not raise new questions of safety and effectiveness:
- it is at least as safe and effective as the predicate devices Mirage Quattro and Ultra Mirage II.

{2}------------------------------------------------

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three tail feathers. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

#### Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

# AUG 0 3 2009

Resmed Limited C/O Mr. David D'Cruz Vice President, Clinical & Regulatory Affairs Resmed Corporation 14040 Danielson Street Poway, California 92064-6857

Re: K091129

Trade/Device Name: Quattro™ LT Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: II Product Code: BZD Dated: June 19, 2009 Received: June 24, 2009

### Dear Mr. D'Cruz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

#### Page 2- Mr. D'Cruz

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours,

for

Susan Runner, D.D.S., M.A. Acting Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

### Enclosure

{4}------------------------------------------------

#### Indication for Use

510(k) Number (if known):

Quattro™ LT

Indication for Use

Device Name:

The Quattro LT channels airflow noninvasively to a patient from a positive airway pressure (PAP) device such as a continuous positive airway pressure (CPAP) or bilevel system.

The Quattro LT is:

- to be used by adult patients (>66lb />30kg) for whom positive airway pressure has been . prescribed
- intended for single-patient re-use in the home environment and multipatient re-use in the . hospital/institutional environment.

L. Schuler

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: k 091129

Prescription Use X AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (Part 21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH; Office of Device Evaluation (ODE)

Page 1 of 1

15 April 2009

---

**Source:** [https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BZD/K091129](https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BZD/K091129)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
