The ResTraxx Data Center system is intended to augment the standard follow-up care of adult patients diagnosed with obstructive sleep apnea by providing wireless transmission and display of usage and therapeutic information. It is intended to be used in the home only and with compatible S7 Elite, AutoSet Spirit, AutoSet Respond & S8 Series CPAP Systems, positive airway pressure flow generators.
Device Story
ResTraxx Data Center system facilitates remote monitoring of obstructive sleep apnea (OSA) therapy. System consists of optional wireless modules (ResTraxx/S8 ResTraxx) that attach to ResMed S7/S8 series CPAP flow generators via rear expansion port. Modules automatically collect patient/machine usage data stored in flow generator memory. Data is transmitted via existing wireless messaging networks to a central Server System. Server stores data in a database and provides secure access for clinical review. Clinicians view transmitted usage and therapeutic information on a PC. System is intended for home use by patients; operated/monitored by clinicians. No patient access to the system. Benefits include improved follow-up care through remote access to therapeutic data, allowing clinicians to monitor patient compliance and therapy effectiveness without requiring in-person visits.
Clinical Evidence
Bench testing only. Design and verification activities were performed to confirm the system met all product and risk requirements. No clinical data was required or provided.
Technological Characteristics
System comprises wireless modules (ResTraxx/S8 ResTraxx) and a server-based software platform. Modules connect to CPAP flow generators via a docking mechanism at the rear expansion port. Connectivity utilizes existing wireless messaging networks. Software modules manage data retrieval, database storage, and secure clinician access via PC. No specific materials or ASTM standards cited.
Indications for Use
Indicated for adult patients diagnosed with obstructive sleep apnea (OSA) to augment standard follow-up care via wireless transmission and display of usage and therapeutic data. Intended for home use only with compatible ResMed CPAP flow generators.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "RESMED" in bold, black letters. The letters are capitalized and evenly spaced. The font appears to be sans-serif.
R
Section 4(6) Rule 506(b) Private Placement Notification
ResTraxx Data Center Traditional 510(k) Premarket Notification
| 510(k) Summary - ResTraxx Data Center | |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | JUL 1 5 2005 |
| Date Prepared | 5th April, 2005 |
| Official Contact | Dr Lionel King<br>V.P., Quality Assurance & Regulatory Affairs<br>ResMed Ltd<br>97 Waterloo Road<br>North Ryde, NSW 2113<br>Australia<br>Tel: +61 (2) 9886 5000<br>Fax: +61 (2) 9878 5517 |
| Classification Reference | 21 CFR 868.5905 |
| Product Code | 73 BZD |
| Common/Usual Name | CPAP System / Non continuous Ventilator (with accessory) |
| Proprietary Name | ResTraxx Data Center System |
| Predicate Device(s) | SomnoTraxx System (K030797) |
| Reason for submission | New Device |
| Indications for Use | The ResTraxx Data Center system is intended to augment the standard<br>follow-up care of adult patients diagnosed with obstructive sleep apnea by<br>providing wireless transmission and display of usage and therapeutic<br>information.<br><br>It is intended to be used in the home only and with compatible S7 Elite,<br>AutoSet Spirit, AutoSet Respond & S8 Series CPAP Systems, positive<br>airway pressure flow generators. |
{1}------------------------------------------------
# RESMED
### Substantial Equivalence
The new device has the following similarities to the previously cleared predicate device.
- Similar intended use A
- Same operating principle A
- Same technologies A
- Same manufacturing process A
Design and Verification activities were performed on the ResTraxx Data Center System as a result of the Design and product requirements. All tests confirmed the product met the acceptance criteria.
risk and product requirements. All tests confirmed the product to the prodicat risk and product requirements. 7 in tots sommitally Equivalent to the predicate device.
ResMed has determined that the new device is Substantially Equivalent used in the Reswed has delemined that the new oction to safety and effectiveness when used in the Res irax Data Center System nas not allery and outly antients. The new device complies with the
management of Obstructive Sleep Apnea (OSA) in adult patients. management of Obstructive Gloop Aprila (Coop) in the FDA guidance documents:
- FDA Draft Reviewer Guidance for Ventilators (July 1995) A
- FDA reviewer's and industry, Guidance for the content of premarket submissions for software A contained in medical devices, May 1998.
#### Intended Use
The ResTraxx Data Center system is intended to augment the standard follow-up care of adult patients The Res Fraxx Data Center System is intended to dagment the Standard of usage and therapeutic information.
lt is intended to be used in the home only and with compatible ResMed S7 Elite, AutoSet Spirit, AutoSet It is intended to be used in the nomo sing ative airway pressure flow generators.
#### Device Description
The performance and functional characteristics of the ResTraxx Data Center System includes all the The performance an the predicate device, SomnoTraxx System (K030797).
The ResTraxx Data Center System is designed to function with ResMed's S7™ Elite, AutoSet" Respond The Res Halx Dala Cellier Systems and S8 Series CPAP Systems for the transfer, storage, retireval and AutoSet Spint™ CPAF Systems and Oo Generators to the clinician, via wireless transmission and
and display of stored information from the flow generators to the care no and display of stored mornation from the liew generations of the system. There is no patient access to the system.
The ResTraxx Data Center System comprises two distinct components, ResTraxx or S8 ResTraxx and The ResTrax Data Center System comprises two also is transmitted via a wireless network, stored in the Server System. Data taken from the now gonorator is thanks and displayed on the Clinical reviewer's PC.
ResTraxx™ and S8 ResTraxx™ are optional wireless modules designed to attach to ResMed S7™ ResTraxx™ and 30 Nes Haxx__ are optional wirersos mow generators respectively using a Ellie, AutoSet Respond and AutoSel® Opine or So Senally connected watte existing
docking mechanism. This mechanism allows the device to be existing and SS Post docking filectialism. This nechanism anows and concerner works. When attached, the Res Traxx
expansion port located at the rear of the flow generator. When attached, the Re expansion port located at the rear of the now generator - within the flow generator's memory.
can automatically collect patient and machine information stored within the flow
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the word "RESMED" in bold, black font. The letters are capitalized and evenly spaced. The font appears to be sans-serif.
The ResTraxx and S8 ResTraxx sends information utilizing existing messaging networks providing wireless coverage to large portions of the US population.
Server System – The Server System consists of several functional software modules that are designed a Server System – The Server Official or of SesTraxx or SesTraxx or S8 ResTraxx and a
to retrieve information from ResMed flow generators through the ResTraxx and a to retrieve information from the information in a database and provide a secure into the wireless messaging network, store the mornation in a datababe and provide of the enternation.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/11 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 1 5 2005
ResMed Limited c/o Mr. David D'Cruz 14040 Danielson St. Poway, California 92064-6857
Re: K051314
Trade/Device Name: Res Traxx Data Center Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: II Product Code: BZD Dated: April 5, 2005 Received: May 20, 2005
Dear Mr. D'Cruz:
We have reviewed your Section 510(k) premarket notification of intent to market the device We have reviewed your Section 910(x) premaine is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encrosule) to regally manation in the Medical Device American on to commerce prior to May 20, 1970, the enaounce with the provisions of the Federal Food. Drug, devices that have been recials.ned in acceraanse was a premarket approval application (PMA).
and Cosmetic Act (Act) that do not require approval of a premarket approval appli and Cosmelle Act (Act) that to not require approvine capprovisions of the Act. The Y ou may, therefore, market the devices, books of the coursements for annual registration, listing of general controls provisions of the Fist labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is Classifica (SCC above) into existing major regulations affecting your device can
may be subject to such additional controls. Existing major research and li may be subject to such additional controller "Little 21, Parts 800 to 898. In addition, FDA may be found in the Ood of Pouch Snowning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. D'Cruz
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA s issualice of a subscribe with other requirements of the Act
that FDA has made a determination that your device of a cranice. You must that FDA has made a delerimiation that your are reas by other Federal agencies. You must or any Federal statures and regulations administed to: registration and listing (21
comply with all the Act's requirements, including, but not regultements as set comply with all the Act S requirements; mercements and in the practice requirements as set CFR Part 807); labeling (21 CFX 1 at 601); good manaxt 820); and if applicable, the electronic
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicab forth in the quality systems (Q5) regalatesin (Sections 531-542 of the Act); 21 CFF 1000-1050.
product radiation control provisions (Sections 531-542 of the Acchae Section S product radiation control provisions (occtions our device as described in your Section 510(k)
This letter will allow you to begin marketing your device of your device to a le This letter will anow you to begin manxemily of substantial equivalence of your device to a legally premarket notification. The PDA midning of succements of the more and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your de need on Also, please note the regulation entitled, and the move obtain contact the Office of Compliance at (210) 216 of Part 807.97). You may obtain
"Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain "Misbranding by reference to premarket itstitutions of the Act from the Division of Small
other general information on your responsibilities under the Act from pumber (800) 6 other general information on your responsible and at its toll-free number (800) 638-2041 or Manufacturers, International and Consulter Assistance of to corport/industry/support/index.html.
Sincerely yours,
Suytte of. Michael Onis
Chiú Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
... ..
Enclosure
{5}------------------------------------------------
ResTraxx Data Center Traditional 510(k) Premarket Notification
RESMED
## Indication for Use
510(k) Number (if known):
・ 、 ResTraxx Data Center
Scale
1:24,000
Device Name:
Indication for Use
The ResTraxx Data Center system is intended to augment the standard follow-up care of adult patients I he Hes Faxx Data Center system is inchient to dagment of the starter.
diagnosed with obstructive sleep apnea by providing wireless transmission and display of usage and therapeutic information.
It is intended to be used in the home only and with compatible S7 Elite, AutoSet Spirit, AutoSet Respond It is intended to be used in the norms shy any any pressure flow generators.
Prescription Use _____________________________________________________________________________________________________________________________________________________________
AND/OR
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________
(Part 21 CFR 801 Subpart D)
(Part 21 CFR 807 Subpart C)
(Please DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH; Office of Device Evaluation (ODE)
Qum Sylon
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, De
510(k) Number:
Page 1 of_ 1_
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.