The RespIn11 is intended to provide airway clearance therapy when external manipulation of the thorax is the physician's choice of treatment. Indications for this form of therapy are described by the American Association for Respiratory Care (AARC) in the Clinical Practices Guidelines for Postural Drainage Therapy' (1991). According to AARC guidelines, specific indications for external manipulation of the thorax include evidence or a suggestion of retained secretions, evidence that the patient is having difficulty with the secretion clearance, or presence of atelectasis caused by mucus plugging. In addition, the RespIn11 is also indicated for external manipulation of the thorax to promote airway clearance or improve bronchial drainage for purposes of collecting mucus for diagnostic evaluation.
Device Story
RespIn11 is a powered percussor for High Frequency Chest Wall Oscillation (HFCWO) therapy. System consists of a jacket connected to a vibration/pulsation generator using a multistage blower. Pulsations are transmitted to the patient's chest via an air pressure piston system in the vest. Compression-release cycles generate differential air speed in bronchial airways, creating a shearing effect to mobilize mucus for natural clearance or suction. Used in hospitals, clinics, and homes; operated by clinicians or patients. Output is mechanical oscillation of the chest wall. Therapy helps patients with airway obstruction by promoting bronchial drainage and secretion clearance.
Clinical Evidence
Bench testing only. Testing included frequency, pressure, pneumatic, and fault condition testing, alongside IEC 60601-1 safety and IEC 60601-1-2 electromagnetic compatibility testing. Results demonstrated performance equivalent to the predicate device.
Technological Characteristics
Pneumatic percussor system; electronic control unit with alphanumeric interface. Operates at 100-240 VAC, 50-60 Hz. Frequency range 5-20 Hz; pressure adjustable in 0-10 steps (max 50 mbar). Therapy duration 5-30 minutes. Complies with IEC 60601-1 and IEC 60601-1-2. No patient contact materials.
Indications for Use
Indicated for patients requiring airway clearance therapy due to retained secretions, difficulty with secretion clearance, or atelectasis caused by mucus plugging, or for bronchial drainage to collect mucus for diagnostic evaluation. Contraindicated in patients with ICP >20 mm Hg, recent spinal surgery/injury, bronchopleural fistula, pulmonary edema (CHF), large pleural effusions/empyema, pulmonary embolism, unstable head/neck injury, active hemorrhage/hemodynamic instability, distended abdomen, active/recent hemoptysis, uncontrolled airway/aspiration risk, subcutaneous emphysema, recent epidural/spinal anesthesia, suspected TB, lung contusion, bronchospasm, coagulopathy, chest wall pain, rib fractures, surgical wounds/skin grafts on thorax, recent esophageal surgery, or thoracic burns/wounds/infections.
Regulatory Classification
Identification
A powered percussor is a device that is intended to transmit vibration through a patient's chest wall to aid in freeing mucus deposits in the lung in order to improve bronchial drainage and that may be powered by electricity or compressed gas.
{1}------------------------------------------------
| 510(k) Summary | | |
|--------------------------|--|----------|
| Page 1 of 6<br>4/15/2012 | | K 121170 |
| RespInnovation SAS<br>611 Chemin de l'Adrech<br>SEILLANS, 83440 France | Tel - +33 6 17 81 08 01<br>Fax - +33 6 17 81 08 01 | JUL 13 2012 |
|------------------------------------------------------------------------|------------------------------------------------------|-------------|
| Official Contact: | Barrett R. Mitchell- President | |
| Proprietary or Trade Name: | RespIn11 Model | |
| Common/Usual Name: | Powered Percussor | |
| Classification Name/Code: | BYI - Powered Percussor<br>CFR 868.5 665<br>Class II | |
| Device: | RespIn11 | |
| Predicate Device: | Advanced Respiratory - The Vest K024309 | |
### Device Description:
RespIn11 is a medical device developed for an effective therapy of airway obstruction conditions. The device is made up of a jacket connected to a vibration/pulsation generator where the pulsations are obtained by the pressure differences of a multistage blower. These pulsations are transmitted to the subject's chest through an air pressure piston system specially designed and inserted into the front and rear cavities of the vest provided. The compression then release cycle on the chest wall generates a differential air speed in the bronchial airway in the lungs. This produces a shearing effect which pulls the mucus off the bronchial airway walls and to then be moved through mucociliary action into the larger upper airways to then be eliminated naturally through coughing or if necessary by external suction. This respiratory therapy is called HFCWO (High Frequency Chest Wall Oscillation).
### Indications for Use:
The RespIn11 is intended to provide airway clearance therapy when external manipulation of the thorax is the physician's choice of treatment. Indications for this form of therapy are described by the American Association for Respiratory Care (AARC) in the Clinical Practices Guidelines for Postural Drainage Therapy' (1991). According to AARC guidelines, specific indications for external manipulation of the thorax include evidence or a suggestion of retained secretions, evidence that the patient is having difficulty with the secretion clearance, or presence of atelectasis caused by mucus plugging. In addition, the RespIn11 is also indicated for external manipulation of the
{2}------------------------------------------------
### 510(k) Summary Page 2 of 6 4/15/2012
thorax to promote airway clearance or improve bronchial drainage for purposes of collecting mucus for diagnostic evaluation.
### Environment of Use:
Hospital, sub-intensive, clinical offices, home
### Contraindications
- Intracranial pressure (ICP) greater than 20 mm Hg .
- Recent spinal surgery or acute spinal injury .
- Bronchopleural fistula �
- Pulmonary edema associated with congestive heart failure .
- Large pleural effusions or empyema .
- Pulmonary embolism .
- Head and/or neck injury that has not yet been stabilized ●
- Active hemorrhage with hemodynamic instability .
- . Distended abdomen
- Active or recent gross hemoptysis ●
- Uncontrolled airway at risk for aspiration such as tube feeding or a recent meal .
- . Subcutaneous emphysema
- Recent epidural spinal infusion or spinal anesthesia .
- Suspected pulmonary tuberculosis .
- Lung contusion .
- Bronchospasm ●
- Coagulopathy .
- Complaint of chest wall pain ●
- Rib fractures, with or without flail chest (within 30 days after this may help . splint/stabilize)
- Surgical wound or healing tissue or recent skin grafts or flaps on the thorax .
- Recent esophageal surgery .
- Burns, open wounds, and skin infections on the thorax .
- Recent placement of transvenous or subcutaneous pacemaker (within 30 days) ●
### Summary of substantial equivalence
The RespIn11 was compared to the predicate The Vest (K024309) and is presented in the following comparison table.
{3}------------------------------------------------
remarket Notification 510(
ection 5 -- 510(k) Summary
·
espInnovation RespIn1
### 510(k) Summary
able of Compar
| Table of Comparison<br>Feature | Respinnovation RespIn11<br>510(k) | Advanced Respiratory (Hill-Rom) -<br>The Vest<br>510(k) K024309 | Comment |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Indications | The Respln11 is intended to provide<br>airway clearance therapy when external<br>manipulation of the thorax is the<br>physician's choice of treatment.<br>Indications for this form of therapy are<br>described by the American Association<br>for Respiratory Care (AARC) in the<br>Clinical Practices Guidelines for<br>Postural Drainage Therapy' (1991).<br>According to AARC guidelines, specific<br>indications for external manipulation of<br>the thorax include evidence or a<br>suggestion of retained secretions,<br>evidence that the patient is having<br>difficulty with the secretion clearance,<br>or presence of atelectasis caused by<br>mucus plugging. In addition, the<br>RespIn11 is also indicated for external<br>manipulation of the thorax to promote<br>airway clearance or improve bronchial<br>drainage for purposes of collecting<br>mucus for diagnostic evaluation. | The intended use of the Modified<br>Vest™ Airway Clearance System is the<br>same as the predicate device, which is to<br>provide airway clearance therapy when<br>external manipulation of the thorax is<br>the physician's choice of treatment.<br>Indications for this form of therapy are<br>described by the American Association<br>for Respiratory Care (AARC) in the<br>Clinical Practices Guidelines for<br>Postural Drainage Therapy' (1991).<br>According to AARC guidelines, specific<br>indications for external manipulation of<br>the thorax include evidence or a<br>suggestion of retained secretions,<br>evidence that the patient is having<br>difficulty with the secretion clearance,<br>or presence of atelectasis caused by<br>mucus plugging. In addition, the<br>Vest™ Airway Clearance System is<br>also indicated for external manipulation<br>of the thorax to promote airway<br>clearance or improve bronchial drainage<br>for purposes of collecting mucus for<br>diagnostic evaluation. | Equivalent |
| Prescriptive | Yes | Yes | Identical |
{4}------------------------------------------------
# Premarket Notification 510(k)
Section 5 – 510(k) Summary
## espInnovation RespIn1
| 510(k) Summary |
|----------------|
| Page 4 of 6 |
| 1/5/2012 |
| | | 0(k) Sunnial | |
|-------------|------------------------------------------|---------------------------------------------------------------------------------------------------|-----------|
| Feature | espinnovation RespIn1 | dvanced Respiratory (Hill-Rom | Commen |
| | 510(k) | | |
| | | The Vest | |
| echnolog | neumatic percussor, electronic contro | Pneumatic percussor, electronic control | ldentica |
| | | | |
| nvironment | ospital, sub-intensive, clinical offices | ospital, sub-intensive, clinical offices | Identica |
| പ്രഭം: | nome | nome | |
| Power | 0-240 VAC, 50-60 Hz, 4.7 A MAX | | Equivalen |
| | | | |
| tandard | EC 60601-1, IEC 60601-1-2 | 100 V AC to 230 V AC, 50 Hz to 60 Hz 3.4 A @ 100 VAC, 2.0 A @ 230 VAC_ IEC 60601-1, IEC 60601-1-2 | dentica |
| Alarms | None | None | Identica |
| | | | |
| ontrol Unit | Iphanumeri | Alphanumeric | ldentica |
| nterface | | | |
| erformance | equencies between 5 - 20 Hz | equencies between 5 – | Equivalen |
| requency | | | |
| erformance | ressure applied in 0-10 steps | essure applied in 0-10 steps | dentica |
| Control Unit | Alphanumeric | Alphanumeric | Identical |
|--------------------------------------------|---------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Interface | | | |
| Performance<br>Frequency | Frequencies between 5 - 20 Hz | Frequencies between 5 - 20 | Equivalent |
| Performance<br>Pressure | Pressure applied in 0-10 steps<br>(*max pressure = 50 mbar) | Pressure applied in 0-10 steps<br>(*max pressure = 50 mbar) | Identical |
| Performance<br>Therapy Times | Session therapy from 5 to 30 minutes<br>depending upon Dr.'s recommendations<br>per patient | From 0 to 60 minutes (* from the Vest<br>User Manual: common prescription<br>specify a treatment session time to be<br>between 10 and 30 minutes) | Equivalent |
| Patient Interface<br>Patient<br>Population | 2 jacket sizes: S/M, L/XXL<br>Not explicitly limited | 4 "vest" styles<br>Not explicitly limited | Equivalent |
| Operating<br>Temperature | 10°C to 34°C (50°F to 95 °F) | 50°F to 93°F (10°C to 34°C)<br>ambient temperature (* from Vest User<br>Manual) | Equivalent |
PDF Page 18 of 258
{5}------------------------------------------------
510(k) Summary Page 5 of 6 4/15/2012
### Differences Between Other Legally Marketed Predicate Devices
The RespIn! 1 is viewed as substantially equivalent to the predicate device because: The percussions provided by the RespIn11 is substantially equivalent to that commonly employed by other powered percussors.
### Indications -
The RespIn11 is intended to provide airway clearance therapy when external manipulation of the thorax is the physician's choice of treatment. Indications for this form of therapy are described by the American Association for Respiratory Care (AARC) in the Clinical Practices Guidelines for Postural Drainage Therapy' (1991). According to AARC guidelines, specific indications for external manipulation of the thorax include evidence or a suggestion of retained secretions, evidence that the patient is having difficulty with the secretion clearance, or presence of atelectasis caused by mucus plugging. In addition, the RespIn11 is also indicated for external manipulation of the thorax to promote airway clearance or improve bronchial drainage for purposes of collecting mucus for diagnostic evaluation.
Prescriptive - The RespIn11 is prescriptive as is the predicate.
Design and Technology - The RespIn11 has equivalent design and features as the predicate and has the identical technology to the predicate.
Performance and Specifications - The RespIn11 has equivalent specifications of performance as the predicate.
Compliance with standards - The RespIn11 and predicate device comply with IEC 60601-1 and IEC 60601-1-2.
Materials -No patient contact materials
### Environment of Use -
RespIn11 is for Hospital, sub-intensive, clinical offices, home, which is similar for predicate.
### Patient Population -
RespIn 1 is intended for patients that their clinician has determined that this treatment is appropriate. This is similar to the predicate.
{6}------------------------------------------------
### 510(k) Summary Page 6 of 6 4/15/2012
### Performance Testing
We have performed performance testing and the results demonstrated that the device performs as intended and is substantially equivalent to the performance of the predicate.
The tests included:
- Frequency Testing ●
- Pressure Testing .
- . Pneumatic Testing
- Fault condition testing .
- IEC 60601-1 Safety .
- IEC 60601-1-2 Electromagnetic Compatibility ●
The results of the testing demonstrate that the RespIn11 performs equivalent to the predicate.
### Conclusion
The RespIn11 has been demonstrated to be substantially equivalent to the predicate, The Vest (K024309), in indications for use, patient population, and environment for use, technology characteristics, specifications / performance and compliance with international standards.
{7}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three legs, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circle around the eagle. The eagle is black, and the text is also black. The logo is simple and recognizable.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
RespInnovation SAS C/O Mr. Paul Dryden ProMedic, Incorporated 24301 Woodsage Drive Bonita Springs, Florida 34134
JUL 13 2012
Re: K121170
Trade/Device Name: RespIn11 Regulation Number: 21 CFR 868.5665 Regulation Name: Powered Percussor Regulatory Class: II Product Code: BYI Dated: April 15, 2012 Received: April 17, 2012
Dear Mr. Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{8}------------------------------------------------
### Page 2 - Mr. Dryden
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Nh for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
### Indications for Use Statement
Page 1 of 1
### 510(k) Number:
(To be assigned)
Device Name:
RespIn11
### Indications for Use:
The RespIn11 is intended to provide airway clearance therapy when external manipulation of the thorax is the physician's choice of treatment. Indications for this form of therapy are described by the American Association for Respiratory Care (AARC) in the Clinical Practices Guidelines for Postural Drainage Therapy' (1991). According to AARC guidelines, specific indications for external manipulation of the thorax include evidence or a suggestion of retained secretions, evidence that the patient is having difficulty with the secretion clearance, or presence of atelectasis caused by mucus plugging. In addition, the RespInl 1 is also indicated for external manipulation of the thorax to promote airway clearance or improve bronchial drainage for purposes of collecting mucus for diagnostic evaluation.
Prescription Use X (Part 21 CFR 801 Subpart D) or
Over-the-counter use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) S
Concurrence of CDRH, Office of Device Evaluation (ODE)
L. Schullstrom
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K121170
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.