K233855 · Enchant Tek Co. , Ltd. · BWF · Feb 2, 2024 · Anesthesiology
Device Facts
Record ID
K233855
Device Name
AllPEP
Applicant
Enchant Tek Co. , Ltd.
Product Code
BWF · Anesthesiology
Decision Date
Feb 2, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5690
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Enchant AllPEP is intended for use as a Positive Expiratory Pressure (PEP) by patients aged 18 years and older. Patients must be capable of generating an exhalation flow of 10 lpm for 3-4 seconds to promote secretion clearance. This device may be used to prevent or reverse atelectasis and as an inspiratory, deep breathing positive exerciser. It may be used in hospital and clinical settings.
Device Story
AllPEP is a Positive Expiratory Pressure (PEP) device used in hospital and clinical settings to promote secretion clearance and prevent/reverse atelectasis. Device consists of housing, top cover, and internal swinger mechanism. Patient inhales room air through one-way valve; during exhalation, air is directed through upper portion of device where swinger mechanism oscillates air, creating vibration. Device provides PEEP and oscillations during exhalation. Healthcare providers use device to assist patients in respiratory therapy; output is physical vibration and pressure during exhalation. Benefits include improved airway clearance and lung expansion. Device is single-patient, multi-use, intended for replacement every 30 days.
Clinical Evidence
Bench testing only. Testing included aging/post-aging performance, cleaning validation, post-cleaning performance, frequency (Hz), amplitude (cmH2O), and pressure at min/max settings. Biocompatibility testing performed per ISO 10993-1 (cytotoxicity, sensitization, intracutaneous reactivity, material-mediated pyrogenicity, chemical characterization) and ISO 18562 (particulate matter, VOCs). Results met pre-defined acceptance criteria.
Technological Characteristics
PEP device featuring swinger mechanism for oscillation and one-way valve for inhalation. Materials meet ISO 10993-1 for external communicating, mucosal membrane contact, prolonged duration (>24h, <30 days). Performance metrics: frequency range 9.82-16.32 Hz and amplitude 3.24-9.17 cmH2O across 5-30 lpm flow rates. Mechanical operation; no energy source required. Single-patient, multi-use; cleaned with neutral dishwashing agent.
Indications for Use
Indicated for patients 18 years and older requiring secretion clearance or prevention/reversal of atelectasis; must be capable of generating 10 lpm exhalation flow for 3-4 seconds.
Regulatory Classification
Identification
An incentive spirometer is a device that indicates a patient's breathing volume or flow and that provides an incentive to the patient to improve his or her ventilation.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
February 2, 2024
Enchant Tek Co. Ltd. Leonard Ling QA/RA Supervisor No. 210 Xiangzhong Rd. Dongshan Township Yilan County, 26950 Taiwan
Re: K233855
Trade/Device Name: AllPEP Regulation Number: 21 CFR 868.5690 Regulation Name: Incentive Spirometer Regulatory Class: Class II Product Code: BWF Dated: December 4, 2023 Received: December 5, 2023
Dear Leonard Ling:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Rachana Visaria -S
Rachana Visaria, Ph.D. Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and
{2}------------------------------------------------
Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K233855
Device Name AllPEP
Indications for Use (Describe)
The Enchant AllPEP is intended for use as a Positive Expiratory Pressure (PEP) by patients aged 18 years and older. Patients must be capable of generating an exhalation flow of 10 lpm for 3-4 seconds to promote secretion clearance. This device may be used to prevent or reverse atelectasis and as an inspiratory, deep breathing positive exerciser. It may be used in hospital and clinical settings.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <span style="font-size:12px">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="font-size:12px">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| Date Prepared: | February 2, 2024 |
|----------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| Sponsor: | Enchant Tek Co. Ltd.<br>No. 210 Xiangzhong Rd.<br>Dongshan Township, Yilan County<br>Taipei, TW 26950<br>Tel +886- 03-959-3500 |
| Sponsor Contact: | Ling Leonard - QA/RA Supervisor |
| Submission Correspondent: | Paul Dryden<br>ProMedic, LLC |
| Proprietary or Trade Name: | AllPEP |
| Common/Usual Name: | PEP device |
| Classification Name: | Spirometer, therapeutic (incentive) |
| Product Code: | 21CFR 868.5690<br>BWF |
| Predicate Device: | DHD Acapella K991561 |
| Common/Usual Name: | Spirometer, therapeutic (incentive) |
| Classification Name: | 21CFR 868.5690 |
| Product Code: | BWF |
Device Description: AllPEP provides PEEP during user exhalation and features a mechanism which provides oscillations during exhalation as well.
Principle of Operation: The oscillating positive exhalation pressure device, AllPEP, includes a housing, a top cover, and an oscillating mechanism. The patient inhales room air through a one-way valve and when exhaling the exhaled gas is directed through the upper portion of the device which contains a swinger mechanism to oscillate the air, causing the air to vibrate.
### Indications for Use:
The Enchant AllPEP is intended for use as a Positive Expiratory Pressure (PEP) by patients aged 18 years and older. Patients must be capable of generating an exhalation flow of 10 lpm for 3-4 seconds to promote secretion clearance. This device may be used to prevent or reverse atelectasis and as an inspiratory, deep breathing positive exerciser. It may be used in hospital and clinical settings.
### Patient Population:
Patients 18 years and older who are capable of generating an exhalation flow of 10 1pm for 3-4 seconds.
{5}------------------------------------------------
### Environments of use:
For hospital and clinical settings.
### Substantial Equivalence Discussion
The table below compares the key features of the proposed with the identified predicate - DHD Acapella K991561. The comparison demonstrates that the subject device is substantially equivalent to the predicate device.
#### . Indications for Use -
The indications for use are similar for the subject device when compared to the predicate device.
Discussion - Since the predicate, newer cleared PEP devices have more details within the indications for use. We have added this information, it does not however raise different questions of safety or effectiveness.
#### . Technology and construction -
The technology and principle of operation is similar for the subject device when compared to the predicate device.
Discussion - While the predicate may have a different means of creating oscillation, there are a number of cleared PEP devices which also have slightly different ways of creating oscillations during the exhalation cycle. This difference does not raise different concerns of safety or effectiveness.
#### . Environment of Use -
The environments of use are similar to predicate which are hospital and clinical settings. Discussion - The environments of use are similar.
#### Patient Population -.
The patient population for the subject device is patients requiring secretion clearance and those who may have atelectasis. They should be capable of generating an exhalation flow of 10 lpm for 3-4 seconds.
Discussion - The subject and predicate device patient populations are similar.
### Non-Clinical Testing Summary -
### Bench testing -
The following tests were performed to demonstrate substantial equivalence:
- Aging and post-aging performance ●
- Cleaning validation ●
- Post-cleaning performance
- Frequency (Hz)
- Amplitude (cmH20) ●
- Pressure at min / max settings (cmH20) 0
Discussion - The test results are similar to the predicate and within pre-defined acceptance criteria.
{6}------------------------------------------------
## Biocompatibility -
According to ISO 10993-1 and the FDA Guidance related to biocompatibility from 2020, the subject device is considered:
- External Communicating
- Tissue Contacting
- Prolonged duration of use (>24 hours, <30 days) o
For those materials in direct contact, they would be considered:
- Surface Device
- Mucosal Membrane ●
- o Prolonged duration of use (>24 hours, <30 days)
Based upon ISO 10993-1, the sponsor has performed the following biocompatibility tests for the subject device:
- ISO 10993-5:2009 - Cytotoxicity
- ISO 10993-10:2010 Sensitization and Intracutaneous Reactivity ●
- ISO 10993-11:2017 Material-Mediated Pyrogenicity ●
- ISO 10993-18:2020 Chemical Characterization and Risk Assessment ●
- ISO 18562-2:2018 Particulate Matter testing ●
- ISO 18562-3:2018 Volatile Organic Compounds (VOC) with risk assessment ●
Discussion - The subject device was found to meet the applicable requirements for biocompatibility safety for the intended population.
### Discussion of Differences
There are no significant differences in indications for use, population, use environments, technological characteristics or performance which raise different concerns of safety compared to the predicate and thus should be found to be substantially equivalent.
### Substantial Equivalence Conclusion
Through performance testing, design, and non-clinical testing the proposed device and predicate have been found to be substantially equivalent.
{7}------------------------------------------------
| | Subject Device | Predicate Device | Subject and Predicate Comparison |
|-------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company | Enchant Tek Co., Ltd. | DHD Acapella | |
| 510(k) | K230622 | K991561 | |
| Classification No. &<br>Product code | 868.5690<br>BWF - spirometer, therapeutic (incentive) | 868.5690<br>BWF - spirometer, therapeutic (incentive) | Identical |
| Indications for Use | The Enchant AllPEP is intended for use as a<br>Positive Expiratory Pressure (PEP) by<br>patients 18 years and older. Patients must be<br>capable of generating an exhalation flow of<br>10 lpm for 3-4 seconds to promote secretion<br>clearance. This device may be used to<br>prevent or reverse atelectasis and as an<br>inspiratory, deep breathing positive<br>exerciser. It may be used in hospital and<br>clinical settings. | Indicated for use as a Positive Expiratory<br>Pressure (PEP) Device<br>improves clearance of secretions<br>may reduce the need for postural drainage<br>facilitates opening of airways in patients<br>with Cystic Fibrosis, COPD, asthma, and<br>lung diseases with secretory problems<br>may be used to prevent or reverse<br>atelectasis | Similar<br>More recent PEP clearances have<br>included the environment of use and<br>target population in the indications<br>statement. This does not change the<br>intended use. |
| Environments of Use | Hospital and clinical settings | Hospital, clinical and home care settings | Similar, but not for home use |
| Prescriptive | Yes | Yes | Similar |
| Patient Population | Patients requiring PEP therapy aged 18<br>years and older.<br>Patient capable of generating an exhalation<br>flow of 10 lpm for 3-4 seconds | Patients requiring PEP therapy | Similar |
| Single Patient, multi-use | Yes | Yes | Similar |
| Patient Interface | Mouthpiece | Mouthpiece | Similar |
| Basic Components | Swinging Flap valve which generates<br>oscillation during exhalation<br>One-way valve for inhalation<br>Mouthpiece | Rocker which generates oscillation during<br>exhalation<br>One-way valve for inhalation<br>Mouthpiece | Similar<br>The principle of operation is the<br>oscillation during airflow. This is similar<br>to the predicate device and does not raise<br>different questions of safety and<br>effectiveness |
| Mean frequency range<br>(Hz) | 5lpm - 9.82<br>10lpm - 9.96<br>15lpm - 11.85<br>20lpm - 13.01<br>25lpm - 14.89<br>30lpm - 16.32 | Not published | Similar based on bench testing of subject<br>device and predicate |
| | Subject Device | Predicate Device | Subject and Predicate Comparison |
| Mean amplitude<br>Pressure (cmH20) | 5lpm - 3.24<br>10lpm - 6.31<br>15lpm - 8.31<br>20lpm - 9.42<br>25lpm - 9.12<br>30lpm - 9.17 | Not published | Similar based on bench testing of subject<br>device and predicate |
| Lowest Pressure (cmH20) | 5lpm - 0.50<br>10lpm - 0.78<br>15lpm - 2.09<br>20lpm - 4.00<br>25lpm - 7.26<br>30lpm - 10.75 | Not published | Similar based on bench testing of subject<br>device and predicate |
| Highest Pressure<br>(cmH20) | 5lpm - 3.74<br>10lpm - 7.10<br>15lpm - 10.40<br>20lpm - 13.42<br>25lpm - 16.38<br>30lpm - 19.92 | Not published | Similar based on bench testing of subject<br>device and predicate |
| Cleaning method | Neutral dishwashing agent<br>Single patient, multi-use | Cleaning instructions not provided<br>Single patient, multi-use | Similar based on bench testing of subject<br>device and predicate |
| Performance testing | Aging and post-aging performance<br>Cleaning and post-cleaning<br>performance<br>Cleaning validation<br>Transportation - shock and vibration<br>Frequency (Hz)<br>Amplitude (cmH2O)<br>Pressure at min / max settings (cmH2O) | Frequency (Hz)<br>Amplitude (cmH2O)<br>Pressure at min I max settings (cmH2O) | Similar |
| Type of patient contact<br>with materials | Externally communicating/Tissue Surface<br>contact /mucosa<br>Prolonged duration (>24 hours, < 30 days) | Not published | Patients are instructed to replace device<br>every 30 days. |
{8}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.