← Product Code [BWF](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BWF) · K150044

# O-PEP (K150044)

_Pari Respiratory Equipment, Inc. · BWF · May 27, 2016 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BWF/K150044

## Device Facts

- **Applicant:** Pari Respiratory Equipment, Inc.
- **Product Code:** [BWF](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BWF.md)
- **Decision Date:** May 27, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5690
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The PARI O-PEP is intended for use as a Positive Expiratory Pressure Device, and is designed to help patients improve secretion clearance. The PARI O-PEP is intended for adult and pediatric patients ages 5 and older, for use in home, hospital, and sub-acute institutions.

## Device Story

PARI O-PEP is a reusable, oscillating positive expiratory pressure (O-PEP) device for mobilizing lower respiratory tract secretions. Patient exhales into device against a movable ball; ball movement creates vibrations transmitted to lungs; vibrations open airways and mobilize mucus. Used in home, hospital, or sub-acute settings by patients under prescription. Device is mechanical; no software or electronics. Benefits include improved secretion clearance, strengthened respiratory tract, and reduced shortness of breath.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench performance testing comparing oscillation frequency and expiratory resistance against predicate devices, plus biocompatibility and cleaning efficacy validation.

## Technological Characteristics

Mechanical oscillating PEP device. Materials: rigid injection-molded polymer (POM). Principle: gravity-dependent ball-valve oscillation. Dimensions/form factor: handheld. Connectivity: none. Sterilization: non-sterile; reusable via soap and water cleaning. Biocompatibility: ISO 10993-1, 10993-5, 10993-10. Cleaning validation: AAMI TIR No. 30.

## Regulatory Identification

An incentive spirometer is a device that indicates a patient's breathing volume or flow and that provides an incentive to the patient to improve his or her ventilation.

## Predicate Devices

- Flutter D ([K972859](/device/K972859.md))
- Flutter ([K946083](/device/K946083.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a stylized emblem. The emblem consists of a caduceus-like symbol with three figures representing health and human services. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 27, 2016

PARI Respiratory Equipment, Inc. Michael Judge VP, Operations and Regulatory Affairs 2412 Pari Way Midlothian, Virginia 23112

Re: K150044

Trade/Device Name: PARI O-PEP Regulation Number: 21 CFR 868.5690 Regulation Name: Incentive Spirometer Regulatory Class: Class II Product Code: BWF Dated: Undated Received: April 27, 2016

Dear Michael Judge:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Tejashri Purohit-Sheth, M.D.

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Erin I. Keith M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known): K150044

Device Name: PARI O-PEP

## Indications for Use:

The PARI O-PEP is intended for use as a Positive Expiratory Pressure Device, and is designed to help patients improve secretion clearance. The PARI O-PEP is intended for adult and pediatric patients ages 5 and older, for use in home, hospital, and sub-acute institutions.

Prescription Use XXX (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page __ of (Posted November 13, 2003)

{3}------------------------------------------------

# PARI O-PEP Oscillating Positive Expiratory Pressure Device 510(k) Submission 510(k) Summary

#### Submitter Information

| Name:<br>Address: | PARI Respiratory Equipment, Inc.<br>2412 PARI Way<br>Midlothian, VA 23112 |
|-------------------|---------------------------------------------------------------------------|
| Phone Number:     | 804-253-7274 x269                                                         |
| Fax Number:       | 804-253-0260                                                              |
| Contact Name:     | Michael Judge                                                             |
| Date Prepared:    | May 26, 2016                                                              |
| Device Name       |                                                                           |
| Common Name:      | O-PEP (Oscillating Positive Expiratory Pressure) Device                   |

#### Trade Name: PARI O-PEP

Classification Name: Spirometer, Therapeutic (Incentive), 21 CFR 868.5690, Product Code BWF

#### Legally Marketed Predicate Device(s)

| Manufacturer                              | Device    | 510(k) Number     |
|-------------------------------------------|-----------|-------------------|
| Aptalis Pharma                            | Flutter D | K972859 (primary) |
| Bradstreet Clinical Research Assoc., Inc. | Flutter   | K946083           |

#### Device Description

The PARI O-PEP (Oscillating Positive Expiratory Pressure) is a small, single patient use, reusable oscillating PEP therapy device. The PARI O-PEP is a respiratory therapy device designed for temporary application to mobilize secretions or mucous in the lower respiratory tract, strengthen the respiratory tract and alleviate shortness of breath. The patient exhales repeatedly into the device against a movable ball, causing a vibration that is transmitted to the lungs. This causes the deeper respiratory tract to open. By doing this, secretions are mobilized and able to move up the airways. The device is non-sterile, prescription-use only, intended for use in hospital, clinic, or home environments.

#### Medical Condition Treated

The PARI O-PEP is intended to help patients with respiratory conditions clear mucus and secretions from their airways.

#### Indications For Use

The PARI O-PEP is intended for use as a Positive Expiratory Pressure Device, and is designed to help patients improve secretion clearance. The PARI O-PEP is intended for adult and pediatric patients ages 5 and older, for use in home, hospital, and sub-acute institutions.

#### Technological Characteristics Compared to Predicate Device

The PARI O-PEP is substantially equivalent to the predicate devices Flutter and Flutter D. Table 1 below is a summary of the comparison.

| Features                           | Proposed<br>PARI O-PEP                                                                                                                         | Primary Predicate<br>Flutter D<br>K972859                                                                                                                              | Secondary Predicate<br>Flutter<br>K946083                                                                                                |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use                | The PARI O-PEP is<br>intended for use as a<br>Positive Expiratory<br>Pressure Device, and is<br>designed to help patients<br>improve secretion | The efficacy of the Flutter<br>D as a mucus clearance<br>device for cystic fibrosis<br>patients is based on its<br>ability to 1) vibrate the<br>airways (which loosens | The Flutter provides<br>positive expiratory<br>pressure (PEP) therapy<br>for patients with<br>mucus-producing<br>respiratory conditions, |
| Product Code and<br>Regulation No. |                                                                                                                                                |                                                                                                                                                                        |                                                                                                                                          |
| Prescription Use                   | Rx Only                                                                                                                                        | Rx Only                                                                                                                                                                | Rx Only                                                                                                                                  |
| Environment of Use                 | Home, Hospital, Sub-<br>acute Institutions                                                                                                     | Not indicated                                                                                                                                                          | Not indicated                                                                                                                            |
| Patient Population                 | Pediatric<br>Adult                                                                                                                             | Patients with mucus-<br>producing respiratory<br>conditions – age<br>restrictions not specified                                                                        | Patients with mucus-<br>producing respiratory<br>conditions – age<br>restrictions not specified                                          |
| Contraindications                  | Pneumothorax, right<br>ventricular insufficiency,<br>coughing up blood and<br>tuberculosis.                                                    | Pneumothorax, overt<br>right-sided heart failure.                                                                                                                      | Pneumothorax, overt<br>right-sided heart<br>failure.                                                                                     |
| Software driven                    | No                                                                                                                                             | No                                                                                                                                                                     | No                                                                                                                                       |
| Materials in patient<br>contact    | Rigid injection-molded<br>polymer - POM.                                                                                                       | Rigid injection-molded<br>polymer                                                                                                                                      | Rigid injection-molded<br>polymer                                                                                                        |
| Oscillation<br>frequency           | Oscillation frequency of PARI O-PEP device is comparable to predicate Flutter<br>across a range of flow rates between 5 and 30 LPM.            |                                                                                                                                                                        |                                                                                                                                          |
| Expiratory<br>resistance           | Pressure level of PARI O-PEP device is comparable to predicate Flutter across a<br>range of flow rates between 5 and 30 LPM.                   |                                                                                                                                                                        |                                                                                                                                          |
| Cleaning and<br>disinfection       | Cleaning only, soap and<br>water.                                                                                                              | N/A                                                                                                                                                                    | Cleaning only, soap<br>and water.                                                                                                        |

Table 1

{4}------------------------------------------------

{5}------------------------------------------------

### Discussion of Differences

The design and intended use of the subject and predicate devices are similar, since both sets are oscillating resistance devices for prescription use, and aid patients with mucus secretion through the airways via Positive Expiratory Pressure. Product codes and regulation numbers for the subject device and predicate devices differ; however, devices under both regulations are used to transmit vibrations through the patient's chest wall to aid in freeing mucus and improving ventilation. Performance testing demonstrates that the PARI O-PEP and predicate Flutter D devices were tested to compare oscillation frequency and expiratory resistance. Results demonstrate that the PARI O-PEP device produces comparable oscillation frequencies across a range of flow rates and pressure ranges.

### Non-Clinical Test Summary

PARI O- PEP was tested to compare performance to the predicate Flutter device and satisfy applicable guidance, including:

- Expiratory resistance: PARI O-PEP expiratory pressure level was compared to the predicate Flutter device across a range of flow rates between 5 and 30 LPM and angular positions of 0 and 30 degrees, and was found to be substantially equivalent to the predicate device for all variables.
- . Oscillating frequency: PARI O-PEP oscillating frequency was compared to the predicate Flutter device across a range of flows between 5 and 30 LPM and angular positions of 0 and 30 degrees, and was found to be substantially equivalent to the predicate device for all variables.
- . Biocompatibility: Testing demonstrates that the subject device meets requirements in accordance with: ISO 10993-1: Biological Evaluation of medical devices – Part 1: Evaluation and testing within a ■
	- risk management process
	- . ISO 10993-5: Biological Evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity
	- . ISO 10993-10: Biological evaluation of medical devices – Part 10: Tests for irritation and skin sensitization
- . Cleaning efficacy validation: Testing demonstrates that the subject device meets requirements for cleaning of a reusable medical device in accordance with recognized guidelines, including:
	- 트 AAMI TIR No. 30, 2011: A Compendium of Processes, Materials, Test Methods, and Acceptance Criteria for Cleaning Reusable Medical Devices

#### Clinical Performance Summary

Clinical testing was not completed/is not required to show substantial equivalence.

#### Conclusion

The sponsor has demonstrated through performance testing, design and features, and non-clinical testing that the proposed device and predicates have been found to be substantially equivalent.

---

**Source:** [https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BWF/K150044](https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BWF/K150044)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
