← Product Code [BWF](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BWF) · K133873

# TRIBALL INCENTIVE SPIROMETER (K133873)

_Besmed Health Business Corp · BWF · Apr 4, 2014 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BWF/K133873

## Device Facts

- **Applicant:** Besmed Health Business Corp
- **Product Code:** [BWF](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BWF.md)
- **Decision Date:** Apr 4, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5690
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic

## Indications for Use

The Besmed Incentive Spirometer is intended as an inspiratory deep breathing positive exerciser. Intended for single-patient, multi-use in a hospital or home care setting.

## Device Story

TriBall Incentive Spirometer is a mechanical, non-sterile, single-patient, multi-use device used for respiratory therapy. It consists of a housing, three balls, flexible tubing, and a mouthpiece. The patient inhales through the mouthpiece, creating a vacuum that causes the balls to rise within the chamber. The height of the balls provides a visual indicator of the patient's inspiratory flow rate (600, 900, and 1200 cc/sec). Used in hospitals or home care settings, the device serves as a 'lung exerciser' to help patients improve respiratory capacity through repeated use. The visual feedback acts as an incentive for the patient to increase their inspiratory effort, aiding in recovery or strengthening of respiratory function.

## Clinical Evidence

Bench testing only. Comparative performance testing evaluated inspiratory rate/volume accuracy (600, 900, 1200 cc/sec) against the predicate, demonstrating accuracy within +/- 5%. Biocompatibility testing (ISO 10993) for cytotoxicity, sensitization, and intracutaneous irritation was performed. Environmental testing included high/low temperature, humidity, and drop testing to confirm performance specifications.

## Technological Characteristics

Mechanical 3-ball system; flow-based measurement; includes housing, balls, tubing, mouthpiece, and gross particulate filter. Materials tested per ISO 10993 for external communicating (indirect gas pathway) and surface contact. No energy source required. Single-patient, multi-use, disposable.

## Regulatory Identification

An incentive spirometer is a device that indicates a patient's breathing volume or flow and that provides an incentive to the patient to improve his or her ventilation.

## Predicate Devices

- Hudson RCI TriFlo ([K781831](/device/K781831.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# APR 0 4 2014

## 510(k) Summary Page 1 of 4

| Date Prepared:                                                                                              | 03-Apr-2014                                                             |  |
|-------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|--|
| Besmed Health Business Corp.<br>No. 5, Lane 116, Wu-Kong 2nd Rd,<br>Wu-Ku District, New Taipei City, Taiwan | Tel - 011-886-2-2290-3959<br>Fax - 011-886-2-2299-9076                  |  |
| Official Contact:                                                                                           | Winnie Chung<br>Regulatory Affairs Associate                            |  |
| Proprietary or Trade Name:                                                                                  | TriBall Incentive Spirometer                                            |  |
| Common/Usual Name:                                                                                          | Incentive spirometer                                                    |  |
| Classification Name:                                                                                        | 21CFR 868.5690<br>BWF - spirometer, therapeutic (incentive)<br>Class II |  |
| Predicate Devices:                                                                                          | K781831 - Hudson RCI TriFlo                                             |  |

## Device Description:

The Besmed TriBall Incentive Spirometer has a flexible tube and mouthpiece which the patient inhales through. This tube connects to the inspiration port of the unit. The chamber has a series of balls which upon the patient inhaling, creates a vacuum, which causes the balls to rise, they reflect the inspiratory flow rate in cc/sec (600, 900 and 1200).

It is a single patient, multi-use, disposable, non-sterile device.

As a patient improves their respiratory capacity or gets stronger their inspiratory flow rate raises the balls. The principle is to provide the user a visual indicator of their inspiratory flow rate and help them to improve the respiratory function with repeated uses. Incentive spirometers are commonly referred to as "lung exercisers".

The TriBall achieves the therapeutic intent by helping the patient to improve their inspiratory flow rate. It is a relative improvement device that as described offers the "incentive" to improve. This is the identical therapeutic intent of all predicate flow / volume based incentive spirometers.

#### Indications for Use:

The Besmed Incentive Spirometer is intended as an inspiratory deep breathing positive exerciser.

Intended for single-patient, multi-use in a hospital or home care setting.

### Patient Population:

Patients requiring inspiratory exercise.

## Environments of use:

Hospital and home care settings

{1}------------------------------------------------

510(k) Summary

Comparison to Predicate

| Attribute                 | Hudson RCI<br>K781831                                                                                                  | Proposed<br>Besmed TriBall                                                                                                                                                                                                          |
|---------------------------|------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use       | Intended as an inspiratory deep breathing positive exerciser.                                                          | The Besmed Incentive Spirometer is intended as an inspiratory deep breathing positive exerciser.                                                                                                                                    |
| Environments of use       | Intended for single-patient, multi-use in a hospital or home care setting.<br>Home care settings and hospitals         | Intended for single-patient, multi-use in a hospital or home care setting.<br>Home care settings and hospitals                                                                                                                      |
| Prescriptive              | Yes                                                                                                                    | Yes                                                                                                                                                                                                                                 |
| Patient population        | Patients requiring inspiratory exercise                                                                                | Patients requiring inspiratory exercise                                                                                                                                                                                             |
| Single patient, multi-use | Yes                                                                                                                    | Yes                                                                                                                                                                                                                                 |
| Patient interface         | Mouthpiece                                                                                                             | Mouthpiece                                                                                                                                                                                                                          |
| Basic components          | Housing<br>3 balls<br>Tubing<br>Mouthpiece<br>Gross particulate filter                                                 | Housing<br>3 balls<br>Tubing<br>Mouthpiece<br>Gross particulate filter                                                                                                                                                              |
| Flow / Volume range       | 600 / 900 / 1200 cc/sec                                                                                                | 600 / 900 / 1200 cc/sec                                                                                                                                                                                                             |
| Performance testing       | Inspiratory Rate / Volume accuracy<br>600 - 16.7%<br>900 - 1.6%<br>1200 - 11.3%<br>Accuracy specification - not stated | Inspiratory Rate / Volume accuracy<br>600 - 0.5%<br>900 - 0.3%<br>1200 - 0.1%<br>Accuracy specification - +/- 5%<br>Age Testing<br>Pre and post- exposure<br>Environmental Testing<br>High/Low and Humidity conditions<br>Drop test |

{2}------------------------------------------------

## 510(k) Summary Page 3 of 4 03-Apr-2014

#### Substantial Equivalence Discussion

Table 1 compares the key features of the proposed Besmed TriBall incentive spirometer with the identified predicate and demonstrates that the device can be found to be substantially equivalent.

In summary one can conclude that substantial equivalence is met based upon the following:

#### Indications for Use -

The indications for use are identical for the proposed device when compared to the predicate -K781831 - Hudson RCI incentive spirometer.

Discussion - Each device is indicated for use volumetric inspiratory deep breathing positive exerciser.

#### Technology and construction -

The design, components, shape, size, etc. are equivalent to the predicate - K 781831 - Hudson RCI incentive spirometer.

Discussion - The design is a 3 ball system that upon the patient inhaling raises the balls to indicate the volume of inspired air.

#### Environment of Use -

The environments of use are identical to predicate - K781831 - Hudson RCI incentive spirometer.

Discussion - The environments of use are identical to the predicate K781831 - Hudson RCI incentive spirometer.

#### Patient Population -

The patient population of patients requiring inspiratory exercise is equivalent to the predicate -K781831 - Hudson RCI incentive spirometer.

Discussion - The patient populations are equivalent to the predicate - K781831 - Hudson RCI incentive spirometer.

#### Non-Clinical Testing Summary -

#### Materials:

We have performed ISO 10993 testing on the component materials of the TriBall Incentive spirometer which is considered as External Communicating (Indirect gas pathway) and Surface Contact (direct skin) with the patient which means the following tests were performed.

- Cytotoxicity .
- Sensitization .
- Intracutaneous / Irritation .

{3}------------------------------------------------

## 510(k) Summary Page 4 of 4 03-Apr-2014

## Environmental:

The proposed device was exposed to various environmental conditions of high and low temperatures over time and the performance evaluated and compared before and after these tests to confirm that the proposed device met it performance specifications.

## Performance Testing including Comparative:

We performed comparative volume accuracy to the predicate and the results demonstrated equivalent (or better) performance demonstrating the proposed device is equivalent to the -K781831 - Hudson RCI incentive spirometer.

Multiple samples of each device were tested multiple times and then evaluated for consistency of performance.

## Substantial Equivalence Conclusion -

The proposed device has been found to be substantially equivalent to the predicate. Differences between the proposed device and the predicate do not raise new questions of safety or efficacy.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes representing the department's mission to promote health, well-being, and human services. The eagle is positioned to the right of the department's name, which is written in a circular pattern around the logo.

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 4, 2014

Besmed Health Business Corp C/O Mr. Paul Dryden, President ProMedic, Inc. 24301 Woodsage Drive Bonita Springs, FL 34134

Re: K133873

Trade/Device Name: TriBall Incentive Spirometer Regulation Number: 21 CFR 868.5690 Regulation Name: Incentive spirometer Class: II · Product Code: BWF Dated: March 5, 2014 Received: March 6, 2014

Dear Ms. Chung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract tiability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

## Page 2 - Ms. Chung

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note

the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Tejashri Purohit-Sheth, M.D.
Clinical Deputy Director
DAGRID

FOR

FOR

Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.

510(k) Number (if known)

K33873

Device Name Besmed TriBall Incentive Spirometer

Indications for Use (Describe)

The Besmed Incentive Spirometer is intended as an inspiratory deep breathing positive exerciser.

Intended for single-patient, multi-use in a hospital or home care setting.

Type of Use (Select one or both, as applicable)

🇿 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

| FOR FDA USE ONLY                                                             |                                                                                                                                                                                                        |
|------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) |                                                                                                                                                                                                        |
| K133873                                                                      | Anya C. Harry -S                                                                                                                                                                                       |
|                                                                              | Digitally signed by Anya C. Harry -S<br>DN: c=US, o=U.S. Government,<br>ou=HHS, ou=FDA, ou=People,<br>cn=Anya C. Harry -S,<br>0.9.2342.19200300.100.1.1=001131559<br>Date: 2014.04.03 11:45:44 -04'00' |

FORM FDA 3881 (9/13) PDF Page 41 of 266 Page 4.2 EF

---

**Source:** [https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BWF/K133873](https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BWF/K133873)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
