K090829 · Pari Respiratory Equipment, Inc. · BWF · Jul 21, 2009 · Anesthesiology
Device Facts
Record ID
K090829
Device Name
PARI PEP S
Applicant
Pari Respiratory Equipment, Inc.
Product Code
BWF · Anesthesiology
Decision Date
Jul 21, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5690
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The PARI PEP S is intended for use as a Positive Expiratory Pressure Device, and is designed to help patients exercise their lungs properly and improve secretion clearance. The PARI PEP S may be used by itself or in conjunction with aerosol drug therapy. The PARI PEP S is intended for adult and pediatric patients, for use in home, hospital, and sub-acute institutions,
Device Story
PARI PEP S is a reusable, single-patient-use positive expiratory pressure (PEP) device; operates via fixed orifice to create resistance during exhalation; incorporates one-way valve for low-resistance inhalation; includes pressure monitoring port for clinician/user feedback; used in hospital, clinic, or home environments; operated by patients or clinicians; facilitates lung exercise and secretion clearance; can be used concurrently with aerosol drug delivery; benefits patients by improving airway clearance.
Clinical Evidence
No clinical data; bench testing only. Performance compared to predicate PARI PEP regarding expiratory and inspiratory resistance.
Technological Characteristics
Fixed-orifice PEP device; one-way valve for low-resistance inhalation; pressure monitoring port; reusable; non-sterile; materials identical to PARI PEP predicate.
Indications for Use
Indicated for adult and pediatric patients requiring lung exercise and secretion clearance; usable as standalone device or with aerosol drug therapy in home, hospital, or sub-acute settings.
Regulatory Classification
Identification
An incentive spirometer is a device that indicates a patient's breathing volume or flow and that provides an incentive to the patient to improve his or her ventilation.
{0}------------------------------------------------
K090829
JUL 21 2009
# PARI PEP S Positive Expiratory Pressure Device 510(k) Submission 510(k) Summary
#### Submitter Information
| Name: | PARI Respiratory Eqiupment, Inc. |
|----------------|---------------------------------------------|
| Address: | 2943 Oak Lake Blvd.<br>Midlothian, VA 23112 |
| Phone Number: | 804-253-7274 x269 |
| Fax Number: | 804-639-7244 |
| Contact Name: | Michael Judge |
| Date Prepared: | March 6, 2009 |
#### Device Name
| A 1 2 7 2 2 1 6 2 6 2 6 2 6 2 2 3 6 2 2 3 6 2 2 3 2 3 2 3 2 3 2 3 2 3 2 3 2 3 2 3 2 3 2 3 2 3 2 3 2 3 2 3 2 3 2 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------|
| Common Name: | PEP (Positive Expiratory Pressure) Device |
| Trade Name: | PARI PEP S |
| Classification Name: | Incentive Spirometer, 21 CFR 868.5690, Product Code BWF |
| | |
#### Legally Marketed Predicate Device(s)
| Manufacturer | Device | 510(k) Number |
|----------------------------------|----------|---------------|
| PARI Respiratory Equipment, Inc. | PARI PEP | K972042 |
| Smiths Medical, Inc. | TheraPEP | K983467 |
#### Device Description
The PARI PEP S is a small, single patient use, reusable PEP therapy device for use as a stand alone PEP device or for use with aerosol drug delivery. The device is non-sterile, prescription-use only, intended for use in hospital, clinic, or home environments.
#### Indications For Use
The PARI PEP S is intended for use as a Positive Expiratory Pressure Device, and is designed to help patients exercise their lungs properly and improve secretion clearance. The PARI PEP S may be used by itself or in conjunction with aerosol drug therapy. The PARI PEP S is intended for adult and pediatric patients, for use in home, hospital, and sub-acute institutions,
#### Technological Characteristics Compared to Predicate Device
The PARI PEP S. PARI PEP and TheraPEP are all fixed orifice, positive expiratory pressure devices where a choice of orifices are provided allowing a clinician to select the appropriate settings to accommodate a patient breathing pattern. All devices incorporate a one way valve to allow low resistance inhalation and exhalation through the selected restrictive orifice. All three devices include a pressure monitoring port to allow user or clinician monitoring of the expiratory pressure. PARI PEP and TheraPEP may all be used as stand alone devices or may be used in conjunction with aerosol therapy.
The PARI PEP S uses materials identical to the PARI PEP system.
#### Non-Clinical Test Summary
PARI PEP S was tested to compare performance to the predicate PARI PEP device, including:
- Expiratory resistance: PARI PEP S is comparable to the predicate devices. 트
- 트 Inspiratory Resistance: PARI PEP S is comparable to the predicate devices
#### Clinical Performance Summary
Clinical testing was not completed/is not required to show substantial equivalence.
#### Conclusions from Testing
PARI PEP S meets performance requirements and raises no new issues of safety or effectiveness.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a snake entwined around it, and three horizontal bars above it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the caduceus symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Michael Judge Director of Quality Assurance and Regulatory Affairs PARI Respiratory Equipment, Incorporated 2943 Oak Lake Boulevard Midlothian, Virginia 23112
JUL 2 1 2009
Re: K090829
Trade/Device Name: PARI PEP S Regulation Number: 21 CFR 868.5690 Regulation Name: Incentive Spirometer Regulatory Class: II Product Code: BWF Dated: July 1, 2009 Received: July 2, 2009
Dear Mr. Judge:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
## Page 2 - Mr. Judge
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/ CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Susan Runner, D.D.S., M.A
Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{3}------------------------------------------------
# Indications for Use
# 510(k) Number (if known): N/A
Device Name: PARI PEP S
### Indications for Use:
The PARI PEP S is intended for use as a Positive Expiratory Pressure Device, and is designed to help patients exercise their lungs properly and improve secretion clearance. The PARI PEP S may be used by itself or in conjunction with aerosol drug therapy. The PARI PEP S is intended for adult and pediatric patients, for use in home, hospital, and sub-acute institutions.
Prescription Use XXX (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
L. Schuler
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: KD908629
Page of (Posted November 13, 2003)
Page 4.2