Surgical cricothyrotomy set for securing the airway in case of upper airway obstruction or as ultima ratio if all other attempts to ventilate the patient have failed. ScalpelCric can be used with resuscitation bags or other standard ventilation systems.
Device Story
Surgical cricothyrotomy set for emergency airway access; used in hospital and pre-hospital settings (including military) by clinicians. Components include scalpel, bougie, cuffed tube, extension tube with swivel connector, syringe, and necktape. Procedure: incision of cricothyroid membrane; bougie insertion; tube placement over bougie into trachea; bougie removal; cuff inflation; tube fixation. Enables ventilation via resuscitation bags or standard systems. Benefits: provides life-saving airway access when other ventilation attempts fail.
Clinical Evidence
Bench testing only. No clinical data provided. Performance testing verified bonding, dimensions, cuff function, X-ray visibility, oxygen resistance, and functionality after simulated transport and accelerated aging (up to 5 years). Biocompatibility evaluated per ISO 10993-1 (cytotoxicity, irritation, sensitization, pyrogenicity, acute systemic toxicity).
Technological Characteristics
Single-use surgical cricothyrotomy kit. Components: scalpel, bougie (14 Fr), cuffed tube (6mm ID/9mm OD), extension tube with swivel connector, syringe (10ml), necktape. Materials evaluated for biocompatibility per ISO 10993. Sterilization: Ethylene Oxide (EO). Shelf life: 5 years. Mechanical assembly; no electronic or software components.
Indications for Use
Indicated for emergency airway access via the cricothyroid membrane in adults experiencing life-threatening dyspnoea that cannot be controlled in any other way.
Regulatory Classification
Identification
An emergency airway needle is a device intended to puncture a patient's cricothyroid membrane to provide an emergency airway during upper airway obstruction.
Predicate Devices
Melker Cuffed Emergency Cricothyrotomy Catheter Set - Surgical (K160200)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
VBM Medizintechnik GmbH Jannika Jaeger Regulatory Affairs Manager Einsteinstrasse 1 Sulz am Neckar, 72172 De
Re: K200190 Trade/Device Name: ScalpelCric Regulation Number: 21 CFR 868.5090 Regulation Name: Emergency Airway Needle
Regulatory Class: Class II Product Code: BWC, JOH, OGP Dated: September 25, 2020 Received: September 28, 2020
Dear Jannika Jaeger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Todd Courtney Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
### Device Name
ScalpelCric Crycothyrotomy set - Scalpel technique
### Indications for Use (Describe)
Emergency airway access via the cricothyroid membrane. Life-threatening dyspnoea that cannot be controlled in any other way.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the letters VBM in a sans-serif font. The letters are a bright blue color. The letters are large and take up most of the image.
# K200190
# VOLUME 003
# 510(k) Summary
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for VBM, which is written in a sans-serif font. The letters are large and bold, and they are all in the same blue color. The logo is simple and modern, and it is easy to read.
| DATE OF APPLICATION: | 10/26/2020 |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| APPLICANT: | VBM Medizintechnik GmbH<br>Einsteinstrasse 1<br>72172 Sulz a. N.<br>Germany<br>Tel: + 49 (0) 7454 9596 0<br>Fax: +49 (0) 7454 9596 33<br>E-Mail: info@vbm-medical.de |
| CONTACT PERSON: | Jannika Jaeger<br>Regulatory Affairs Manager<br>Tel.: + 49 (0) 7454 9596 660 |
E-Mail: jannika.jaeger@vbm-medical.de
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the letters VBM in a sans-serif font. The letters are a bright blue color. The background is white.
#### 1 Device Name
| Trade Name: | ScalpelCric |
|-----------------------------|-------------------------------------------------------|
| Common Name: | Cricothyrotomy Set<br>Needle, Emergency Airway |
| Device Classification Name: | Cricothyrotomy Kit<br>Tube tracheostomy and tube cuff |
# 2 Classification / Product Code
ScalpelCric can be classified according to following device name and product code:
Primary Product Code
| Device | Regulation<br>Description | Regulation<br>Medical<br>Specialty | Review Panel | Product<br>Code | Regulation<br>Number | Device<br>Classificatio<br>n |
|--------------------------------|--------------------------------|------------------------------------|----------------|-----------------|----------------------|------------------------------|
| Needle,<br>Emergency<br>Airway | Emergency<br>airway<br>needle. | Anesthesiology | Anesthesiology | BWC | 868.5090 | 2 |
### Additional Product ode
| Device | Regulation<br>Description | Regulation<br>Medical<br>Specialty | Review Panel | Product<br>Code | Regulation<br>Number | Device<br>Classificatio<br>n |
|---------------------------------------|-----------------------------------------|------------------------------------|----------------|-----------------|----------------------|------------------------------|
| Tube<br>tracheostomy<br>and tube cuff | Tracheostom<br>y tube and<br>tube cuff. | Anesthesiology | Anesthesiology | JOH | 868.5800 | 2 |
#### Predicate Device / Reference Device ന
| Device | Predicate Device | Reference<br>Device | 510(k)<br>Number | 510(k) Holder |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------|------------------|----------------------------------------------------------------------------------|
| Needle, Emergency<br>Airway | Melker Cuffed Emergency<br>Cricothyrotomy Catheter<br>Set -Percutaneous, Melker<br>Cuffed Emergency<br>Cricothyrotomy Catheter<br>Set - Surgical, Melker<br>Universal Cuffed<br>Emergency Cricothyrotomy<br>Catheter Set | --- | K160200 | COOK<br>INCORPORATED<br>750 DANIELS WAY<br>Bloomington,<br>Indiana 47404 |
| | --- | Airway<br>Connector<br>with Flex Tube | K942392 | Covidien Llc<br>formerly<br>MALLINCKRODT<br>DAR S.R.L.<br>15 HAMPSHIRE<br>STREET |
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the letters "VBM" in a bold, sans-serif font. The letters are a bright, light blue color. The letters are large and take up most of the frame.
| | | | | Mansfield, MA US 02048 |
|--|--|--|--|------------------------|
|--|--|--|--|------------------------|
#### 4 Device Description
Surgical Cricothyrotomy set for securing the airway in case of upper airway obstruction, respectively as "Ultima Ratio" if all other attempts to ventilate the patient failed.
The set consists of scalpel, bougie, tube, extension tube with swivel connector, syringe and necktape.
#### 5 Intended Use
Surgical cricothyrotomy set for securing the airway in case of upper airway obstruction or as ultima ratio if all other attempts to ventilate the patient have failed. ScalpelCric can be used with resuscitation bags or other standard ventilation systems.
Clinical benefit: ScalpelCric allows the ventilation of patients as ultima ratio.
Patient target group: Adults
Location of use: Hospital and pre-hospital conditions including military use
#### 6 Indication for Use
Emergency airway access via the cricothyroid membrane. Life-threatening dyspnoea that cannot be controlled in any other way.
#### 7 Mechanism of Action
The device is used for securing the airway in case of upper airway obstruction. After incision of the cricothyroid membrane a bouqie is inserted to allow an easy insertion of the tube over the bougie into the trachea. After bougie removal, inflation of the cuff and fixation of the tube the patient can be ventilated.
#### 8 Technological Characteristics
The technological characteristics of ScalpelCric are substantially equivalent to the technological characteristics of the predicate device or the reference device.
| Company | VBM<br>---<br>ScalpelCric,<br>Cricothyrotomy Set<br>(New Device) | Cook<br>Incorporated<br>---<br>Melker Cuffed<br>Emergency<br>Cricothyrotomy<br>Catheter Set -<br>Surgical<br>(Predicate<br>Device) | Covidien IIc<br>---<br>Airway<br>Connector with<br>Flex Tube<br>(Reference<br>Device) | Result | |
|-------------|-----------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|--------------------------------|
| Device Name | Needle, Emergency<br>Airway<br>Cricothyrotomy kit<br>Tube tracheostomy and<br>tube cuff | Needle, Emergency<br>Airway<br>Cricothyrotomy kit<br>Tube tracheostomy<br>and tube cuff | Airway Connector<br>with Flex Tube | - | |
| | Company | VBM<br>ScalpelCric,<br>Cricothyrotomy Set<br>(New Device) | Cook<br>Incorporated<br>Melker Cuffed<br>Emergency<br>Cricothyrotomy<br>Catheter Set –<br>Surgical<br>(Predicate<br>Device) | Covidien IIc<br>Airway<br>Connector with<br>Flex Tube<br>(Reference<br>Device) | Result |
| | Regulation Number | 868.5090<br>868.5800<br>868.5800 | 868.5090<br>868.5800<br>868.5800 | 868.5810 | Substantially<br>Equivalent(A) |
| | Class | 2 | 2 | 1 | Substantially<br>Equivalent(A) |
| | Code | BWC<br>OGP<br>JOH | BWC<br>OGP<br>JOH | BZA | Substantially<br>Equivalent(A) |
| | 510(k) number | --- | K160200 | K942392 | |
| | Tube /<br>Airway<br>Catheter | Inner<br>Diameter<br>6 mm | 5 mm | N.A. | Substantially<br>Equivalent(A) |
| | Tube /<br>Airway<br>Catheter | Outer<br>Diameter<br>9 mm | 22 French (7,2<br>mm) | N.A. | Substantially<br>Equivalent(A) |
| | Tube /<br>Airway<br>Catheter | Length<br>10cm | 9 cm | N.A. | Substantially<br>Equivalent(A) |
| | Bougie<br>/Loading<br>dilator | Outer<br>Diameter<br>14 French | 14 French | N.A. | Substantially<br>Equivalent |
| | Bougie<br>/Loading<br>dilator | Length<br>40 cm | 11 cm | N.A. | Substantially<br>Equivalent(A) |
| | Bougie<br>/Loading<br>dilator | Tip Shape<br>Blunt | Blunt | N.A. | Substantially<br>Equivalent |
| | Syringe | Volume<br>10 ml | 6 ml | N.A. | Substantially<br>Equivalent(A) |
| | Scalpel | Type<br>#10 | #11 | N.A. | Substantially<br>Equivalent(A) |
| | Necktape | Availability<br>Included | no Necktape<br>included | N.A. | Substantially<br>Equivalent(A) |
| | Extension<br>Tube with<br>Swivel<br>Connecto<br>r | Inner<br>Diameter -<br>patient<br>side<br>15 mm | 15 mm | | Substantially<br>Equivalent |
| | Extension<br>Tube with<br>Swivel<br>Connecto<br>r | Inner<br>Diameter -<br>machine<br>side<br>22 mm | no Extension Tube<br>included | 22 mm | Substantially<br>Equivalent |
| | Extension<br>Tube with<br>Swivel<br>Connecto<br>r | Length<br>extension<br>tube<br>13 cm | | 13 cm | Substantially<br>Equivalent |
### 8.1 Device Characteristics Table
{8}------------------------------------------------
### VBM Medizintechnik GmbH
ScalpelCric, Cricothyrotomy Set 510(k) Premarket Notification
Image /page/8/Picture/2 description: The image shows the letters VBM in a bold, sans-serif font. The letters are all capitalized and are a bright blue color. The letters are arranged horizontally, with the V on the left, the B in the middle, and the M on the right. The background is white.
A. Differences in relevant characteristics have been separately evaluated during Substantial equivalence discussion. They do not raise new questions regarding safety and efficacy.
### 8.2 Summary of Technological Characteristics
The proposed devices are substantially equivalent in terms of design, operating principles and intended use and have similar technological characteristics to the predicate devices.
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for VBM, which is written in a sans-serif font. The letters are large and bold, and they are all in the same blue color. The logo is simple and modern, and it is easy to read.
#### の Performance Data
## 9.1 Sterilization and Shelf Life
ScalpelCric Set is supplied in sterile condition. EO-sterilizations is used for the sterilization of the ScalpelCric Set. ScalpelCric Set has a shelf life of 5 years.
## 9.2 Biocompatibility
Devices of the ScalpelCric Set were evaluated for their biological safety according to ISO 10993-1. All relevant endpoints have been considered. Tests regarding cytotoxicity, irritation, sensitization, Pyrogen, material mediated pyrogenicity and acute systemic toxicity were performed during the biological safety evaluation. All conducted tests have been successful. The evaluation shows the biological safety of the ScalpelCric Set.
| Component | Type of physical Contact | | Duration of Contact |
|----------------|----------------------------------|----------------------------|---------------------|
| | Category | Contact | |
| Scalpel | External Communicating<br>device | Tissue/ Bone/ Dentine | < 24 hours |
| Bougie | External Communicating<br>device | Tissue/ Bone/ Dentine | < 24 hours |
| Cuffed Tube | External Communicating<br>device | Tissue/ Bone/ Dentine | < 24 hours |
| Extension Tube | Surface device | Mucosal membrane | < 24 hours |
| Neck tape | Surface device | Intact Skin | < 24 hours |
| Syringe | Without Patient Contact | Without Patient<br>Contact | N. A. |
### 9.2.1 Patient Contacting Components
# 9.2.2 Biocompatibility tests
| Tested component | Conducted Test | Result |
|------------------|--------------------------------|------------------------------------------------|
| Scalpel | Cytotoxicity | Non-cytotoxic |
| Scalpel | Material Mediated Pyrogenicity | No traces of material mediated<br>pyrogenicity |
| Scalpel | Acute Systemic Toxicity | No acute systemic toxic<br>characteristics |
| Scalpel | Pyrogen | Endotoxin free |
| Bougie | Irritation | No intracutaneous reactivity |
| Bougie | Sensitization | Non-sensitizer |
| Bougie | Cytotoxicity | Non-cytotoxic |
| Bougie | Material Mediated Pyrogenicity | No traces of material mediated<br>pyrogenicity |
| Bougie | Acute Systemic Toxicity | No acute systemic toxic<br>characteristics |
| Cuffed Tube | Cytotoxicity | Non-cytotoxic |
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the letters VBM in a large, sans-serif font. The letters are a bright, light blue color. The letters are arranged horizontally, with the V on the left, the B in the middle, and the M on the right. The background is white.
| | Material Mediated Pyrogenicity | No traces of material mediated pyrogenicity |
|----------------|--------------------------------|---------------------------------------------|
| | Acute Systemic Toxicity | No acute systemic toxic characteristics |
| | Sensitization | Non-sensitizer |
| | Irritation | No intracutaneous reactivity |
| | Pyrogen | Endotoxin free |
| Extension Tube | Cytotoxicity | Non-cytotoxic |
| | Irritation | No intracutaneous reactivity |
| | Sensitization | Non-sensitizer |
| | Pyrogen | Endotoxin free |
| | Material Mediated Pyrogenicity | No traces of material mediated pyrogenicity |
| | Acute Systemic Toxicity | No acute systemic toxic characteristics |
| Neck Tape | Cytotoxicity | Non-cytotoxic |
| Syringe | N.A. | N.A. |
# 9.3 Sterilization and Shelf Life
ScalpelCric is supplied in sterile condition and intended to be single-use. Validated EOsterilization procedure is used for sterilization. Residuals do not exceed the maximum amount defined by ISO 10993-7. The maximum shelf life of the Set is 5 years.
# 9.4 Bench Testing
The components of the ScalpelCric Set are extensively tested to prove the fulfillment of requirements defined during development. Tests could verify the suitability of all tested features.
| Object under Test | Features to be tested | Result |
|------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Bonding Connector | Bonding between connector and tube | Passed |
| Bonding Inflation Line | Bonding between inflation line and<br>tube | Passed |
| 15mm Connector | Dimensions of 15mm Connector | Passed |
| Luer Pilot Balloon | Dimensions of the female luer<br>connector in the pilot balloon | Passed |
| Cuff | Function of the Cuff | Passed |
| Scalpel | Function and design of the scalpel | Passed |
| Bougie | Bougie which shall fulfill all relevant<br>specification according [DI] and<br>[MS_Bougie]. | Passed |
| Tube | Function and dimensions of the tube | Passed |
| Tube | Tube is recognizable in X-Ray images | Passed |
| extension Tube | Function and dimension of the<br>extension tube | Passed |
| Neck Tape | Neck tape which shall fulfill all<br>relevant specification according [DI] | Passed |
| | and [MS_NT] after 4 times<br>sterilization | |
| Syringe | Function of the syringe | Passed |
| Tube with inflation line and cuff, | Function and features of the tube with<br>attached inflation line and pilot<br>balloon and with cuff | Passed |
| ScalpelCric | Biological compatibility | Passed |
| Material | Function and ingredient of the<br>ScalpelCric materials | Passed |
| Tube and extension Tube | Oxygen resistance of Tube and<br>Extension tube | Passed |
| Label | Device Label | Passed |
| Bougie | Device Label Bougie | Passed |
| IFU | IFU which shall contain all relevant<br>information regarding [DI] and [ER]. | Passed |
| ScalpelCric | ScalpelCric Set can be used in<br>Operating condition | Passed |
| ScalpelCric | Functionality after simulated transport | Passed |
| Transport Validation | Functionality after simulated transport | Passed |
| Safety | Verify implementation and<br>effectiveness of all Risk Control<br>Measures defined in the Risk Analysis | Passed |
| Suction Cap | Connecting tube with open suction cap<br>during ventilation | Passed |
| User Needs | User Needs for ScalpelCric, defined in<br>[SoP] | Passed |
| ScalpelCric | Usability | Passed |
| Headerbag | Requirements of DIN EN ISO 11607-1<br>and DIN EN ISO 11607-2. | Passed |
| Sterilization | Sterilization Requirements for<br>ScalpelCric | Passed |
| Business Needs | Business Needs for ScalpelCric | Passed |
| Label Sterile Packaging | Sterile packaging label which shall<br>contain all relevant information<br>regarding the [SFL], [DI] and [ER]. | Passed |
| Label Outer Packaging | Outer packaging label which shall<br>contain all relevant information<br>regarding the [SFL], [DL] and [ER] | Passed |
| Packaging | Packaging Requirements for<br>ScalpelCric | Passed |
| Accelerated Aging Syringe | Quality of the syringe after 5 years<br>accelerated aging | Passed |
| Accelerated aging of Pouch | Function of the sterile packaging after<br>5 years accelerated aging | Passed |
| Bougie | Bougie shall fulfill all relevant<br>specifications after 1 year accelerated<br>aging | Passed |
| Scalpel | Function of the scalpel after<br>accelerated aging 1 year | Passed |
| ScalpelCric | Scalpel Cric Set application oxygen<br>resistance after 1 year accelerated<br>aging | Passed |
| Surgicric | Function and dimension after 5 years<br>accelerated aging of the Surgicric and<br>the function of the sterile packaging. | Passed |
| | | |
| Surgicric | are identical to the devices in<br>ScalpelCric. | Passed |
| Surgicric | Function and dimension after 3 years<br>real time aging of the Surgicric<br>including Tubus and the function of<br>the sterile packaging. The tested tube<br>and sterile packaging are identical to<br>the devices in ScalpelCric. | Passed |
| Syringe | Quality of the syringe after 3 years<br>real time aging | Passed |
| Bougie | Bougie shall fulfill all relevant<br>specifications after 5 years<br>accelerated aging | Passed |
| Scalpel | Function of the scalpel after 5 years<br>accelerated aging | Passed |
| ScalpelCric Set | Application oxygen resistance after 5<br>years accelerated aging | Passed |
| Cuff Tube Collapse included in the<br>ScalpelCric after 5 years of<br>accelerated aging. | Test required by DIN EN ISO<br>5361:2013-03, Annex C: Resistance<br>of the VBM tracheal tube ventilation<br>lumen collapse with cuff inflated inside<br>a cylinder. | Passed |
{11}------------------------------------------------
### VBM Medizintechnik GmbH ScalpelCric, Cricothyrotomy Set 510(k) Premarket Notification
Image /page/11/Picture/1 description: The image shows the logo for VBM in a bright blue color. The letters are large and bold, taking up most of the frame. The background is white, which makes the blue letters stand out. The logo is simple and modern.
{12}------------------------------------------------
Image /page/12/Picture/1 description: The image shows the letters VBM in a bold, sans-serif font. The letters are all capitalized and are a bright blue color. The letters are arranged horizontally, with the V on the left, the B in the middle, and the M on the right. The background is white.
# 10 Substantial Equivalence Summary / Conclusion
Based on available 510(k) information herein provided, ScalpelCric, CricothyrotomySet is considered substantially equivalent to the predicate device "Melker Cuffed Emergency Cricothyrotomy Catheter Set - Surgical" in terms of intended use, technology and performance specifications.
The VBM component "Extension Tube with Swivel Connector" has been compared separately to the reference device "Airway Connector with Flex Tube". The substantially equivalence has been demonstrated in terms of clinical and biological characteristics as well as technical features.
The subject device and the predicate device have the same intended use and the differences in technological features do not raise different questions of safety and effectiveness.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.