← Product Code [BTT](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BTT) · K141353

# HUDSON RCI AQUAPAK HUMIDIFIER ADAPTOR (K141353)

_Teleflexmedical, Inc. · BTT · Aug 1, 2014 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BTT/K141353

## Device Facts

- **Applicant:** Teleflexmedical, Inc.
- **Product Code:** [BTT](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BTT.md)
- **Decision Date:** Aug 1, 2014
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 868.5450
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic

## Indications for Use

The Hudson RCI® AquaPak® Prefilled Humidifier adds humidity to a patient's breathing gases.

## Device Story

Device adds humidity to dry medical gas for patient inhalation; prevents respiratory mucosa drying/irritation. System consists of prefilled sterile water/saline reservoir (190ml-650ml) and adaptor component. Adaptor connects to flow-metered gas source (50 psi oxygen); gas bubbles through internal micro-diffuser to produce water vapor. Features audible pressure relief valve to indicate gas flow restriction. Used in hospitals, sub-acute/long-term care facilities, and home environments. Operated by clinicians or patients. Single-use device; provides therapeutic humidity output (at least 10mg H2O/L).

## Clinical Evidence

Bench testing only. Tests included packaging integrity (ISTA 2A), ink adherence (ASTM F2252), cannula pop-off, bottle burst, occlusion, spitting, and humidification output (mg/L). Results verify performance is substantially equivalent to predicates.

## Technological Characteristics

Materials: Polypropylene reservoir, HDPE/Surlyn packaging. Principle: Gas bubbling through micro-diffuser. Energy: Pneumatic (50 psi oxygen source). Flow rate: 1.5–10 LPM. Features: Audible pressure relief valve (5–30 psi). Sterilization: Reservoirs via reverse osmosis/distillation/aseptic fill; adaptors non-sterile. Single-use.

## Regulatory Identification

A respiratory gas humidifier is a device that is intended to add moisture to, and sometimes to warm, the breathing gases for administration to a patient. Cascade, gas, heated, and prefilled humidifiers are included in this generic type of device.

## Predicate Devices

- Hudson RCI® AquaPak® Prefilled Humidifier ([K780562](/device/K780562.md))
- Hudson RCI® AquaPak® Prefilled Humidifier ([K833974](/device/K833974.md))
- CareFusion Airlife® Prefilled Humidifier ([K853146](/device/K853146.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 1, 2014

Teleflex Medical, Inc. Mr. James A. Cochie Sr. Regulatory Affairs Specialist 2917 Weck Drive Research Triangle Park, NC 27709

Re: K141353

Trade/Device Name: Hudson RCI® AquaPak® Prefilled Humidifier Regulation Number: 21 CFR 868.5450 Regulation Name: Respiratory gas humidifier Regulatory Class: Class II Product Code: BTT Dated: June 30, 2014 Received: July 2, 2014

Dear Mr. Cochie:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2 - Mr. Cochie

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

*Tejashri Purohit-Sheth, M.D.*

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K141353

Device Name Hudson RCI® AquaPak® Prefilled Humidifier

#### Indications for Use (Describe) The Hudson RCI® AquaPak® Prefilled Humidifier adds humidity to a patient's breathing gases.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Image /page/2/Picture/12 description: The image shows the FDA logo. The logo is a stylized version of the letters "FDA". The "F" is on the left, the "D" is in the middle, and the "A" is on the right. The logo is in a light gray color.

# Anya C. Harry -S 2014.07.31 12:36:04 -04'00'

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## Section 24 - 510(k) Summary

## 510(k) SUMMARY

# Hudson RCI® AquaPak® Prefilled Humidifiers

## A. Name, Address, Phone and Fax Number of Applicant

Teleflex Medical, Incorporated 2917 Weck Drive Research Triangle Park, NC 27709 USA Phone: 919-433-8083 919-433-4996 Fax:

#### B. Contact Person

James Cochie Sr. Regulatory Affairs Specialist

#### C. Date Prepared

July 31, 2014

#### D. Device Name

Trade Name: Hudson RCI® AquaPak® Prefilled Humidifier

Common Name: Humidifier

Classification Name: Humidifier, Respiratory Gas (Direct Patient Interface), CFR -868.5450, Class II

#### E. Device Description

Hudson RCI® AquaPak® Prefilled Humidifiers add water vapor (molecular H₂O) to a dry medical gas that is suspended in the gas to be inhaled by the patient.

The Hudson RCI® AquaPak® Prefilled Humidifiers are designed to add humidity to inspired gas, helping to control the drying and irritation of the respiratory mucosa. The humidifiers feature a micro-diffuser that produces smaller bubbles and greater surface agitation, allowing for a quiet operation and therapeutic humidity output.

Prefilled sterile reservoirs for AquaPak® Humidifiers come in four sizes; 190ml, 340ml, 540ml and 650ml. Each reservoir must be used with a suitable adaptor component, which connects the system to a flow-metered gas source and provides humidifier functionality.

{4}------------------------------------------------

#### Section 24 – 510(k) Summary

Adaptor types provided with Hudson RC1® AquaPak® Prefilled Humidifiers feature an audible pressure relief valve, which indicates a restriction of gas flow to the user.

## F. Indications for Use

The Hudson RCI® AquaPak® Prefilled Humidifier adds humidity to a patient's breathing gases.

#### G. Target Population

This device is intended for use on any patient connected to ventilation devices

#### H. Environments of Use

This device is intended for hospital, sub-acute facilities, long-term care facilities and in a home care environment.

This product is single use only.

#### I. Contraindications

There are no known contraindications.

#### J. Comparative Characteristics

The proposed Hudson RCI® AquaPak® Prefilled Humidifier is substantially equivalent to the predicate device:

| Comparative<br>Characteristics | Proposed Device:<br>Hudson RCI®<br>AquaPak® Prefilled<br>Humidifier                                              | Predicate Device:<br>Hudson RCI®<br>AquaPak® Prefilled<br>Humidifier                                             | Predicate Device:<br>CareFusion Airlife®<br>Prefilled Humidifier                                                                |
|--------------------------------|------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer                   | Teleflex Medical,<br>Inc.                                                                                        | Teleflex Medical,<br>Inc.                                                                                        | CareFusion                                                                                                                      |
| 510(k) Number                  | TBD                                                                                                              | K780562 – Prefilled<br>Humidifier<br>K833974 –<br>Humidifier Adaptor                                             | K853146                                                                                                                         |
| Indications for Use            | The Hudson RCI®<br>AquaPak® Prefilled<br>Humidifier adds<br>humidity to a<br>patient's breathing<br>gases.       | The Hudson RCI®<br>AquaPak® Prefilled<br>Humidifier adds<br>humidity to a<br>patient's breathing<br>gases.       | The CareFusion<br>Airlife® Prefilled<br>Humidifiers provide<br>a cost-effective<br>solution for sterile<br>water inhalation.    |
|                                |                                                                                                                  |                                                                                                                  | (www.caefusion.com)                                                                                                             |
| Principle of Operation         | Gas bubbles through built-in diffuser in reservoir                                                               | Gas bubbles through built-in diffuser in reservoir                                                               | Gas bubbles through reservoir                                                                                                   |
| Gas source                     | 50 Psi oxygen regulated via a flow meter                                                                         | 50 Psi oxygen regulated via a flow meter                                                                         | 50 Psi oxygen regulated via a flow meter                                                                                        |
| Technology and Features        | Flow rate: 1.5 – 10 LPM<br>Audible notification of occlusion: 5 – 30 psi<br>Humidity output: At least 10mg H2O/L | Flow rate: 1.5 – 10 LPM<br>Audible notification of occlusion: 5 – 30 psi<br>Humidity output: At least 10mg H2O/L | Flow rate: 4 LPM for audible notification<br>Audible notification of occlusion: Not specified<br>Humidity output: Not specified |
| Sterilization                  | Adaptor – Non-sterile<br>Sterile Water/Saline Reservoirs - Reverse osmosis, distillation, and aseptic fill       | Adaptor – Gamma Irradiation<br>Sterile Water/Saline Reservoirs - Reverse osmosis, distillation, and aseptic fill | Adaptor – Non-sterile<br>Sterile Water/Saline Reservoirs – Steam or dry heat                                                    |
| Single Use                     | Yes                                                                                                              | Yes                                                                                                              | Adaptor - Single Patient Use<br>Sterile Water/Saline Reservoirs – Single Use                                                    |
| Shelf Life                     | Adaptors – N/A Sterile Water/Saline – 2 years from date of manufacture                                           | Adaptors – 5 years from date of manufacture<br>Sterile Water/Saline – 2 years from date of manufacture           | Adaptors - N/A Sterile Water/Saline – 2 years from date of manufacture                                                          |
| Packaging                      | Adaptor Top web: Coated High Density Polyethylene (HDPE)<br>Adaptor Bottom web: Surlyn Amcor C-060               | Adaptor Top web: Tyvek<br>Adaptor Bottom film: Surlyn Amcor C-060<br>Bottle – Sealed                             | Adaptor - Heat sealed polybag<br>Bottle - Sealed polypropylene                                                                  |
|                                | Bottle - Sealed<br>polypropylene                                                                                 | polypropylene                                                                                                    |                                                                                                                                 |

Teleflex Medical, Inc.

{5}------------------------------------------------

# Section 24 – 510(k) Summary

{6}------------------------------------------------

## Section 24 - 510(k) Summary

## K. Non-clinical Comparative Performance Testing

Bench testing has been performed to verify that the performance of the proposed Hudson RCI® AquaPak® Prefilled Humidifier is substantially equivalent to the predicate device, and that the Hudson RC1® AquaPak® Prefilled Humidifier will perform as intended.

| Test<br>Performed                                       | Reference to Standard (if<br>applicable) | Principle of Test                                                                                                                                                            |
|---------------------------------------------------------|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Packaging<br>Integrity Tests                            | ISTA 2A                                  | Determines the integrity of the packaged<br>device against known shipping tests                                                                                              |
| Ink Adherence<br>Tests                                  | ASTM F2252                               | Determines ink adherence properties of<br>the printed film used in packaging the<br>device                                                                                   |
| Cannula Pop-<br>Off, Bottle<br>Burst,<br>Occlusion Test | N/A                                      | Determines strength of the tubing<br>connector, the reservoir does not burst<br>when nares are occluded and the relieve<br>valve performs as expected to relieve<br>pressure |
| Spitting Test                                           | N/A                                      | Determines the expected outcome that<br>the humidifier will not spit or flood the<br>tubing at the maximum flow rate of 10<br>LPM                                            |
| Humidification<br>Output Test                           | N/A                                      | Determines the rate of humidification at<br>which the liquid is humidified and<br>emitted in milligrams per liter                                                            |

#### L. Substantial Equivalence

The proposed Hudson RCI® AquaPak® Prefilled Humidifier is substantially equivalent in intended use, design, performance and principles of operation to the identified predicate devices cleared under 510(k) K780562 and K833974. The differences between the proposed Hudson RCI® AquaPak® Prefilled Humidifier and the predicate devices are minor (sterile to non-sterile packaging) and raise no new issues of safety and efficacy. The proposed Hudson RCI® AquaPak® Prefilled Humidifier is substantially equivalent to the currently marketed predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BTT/K141353](https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BTT/K141353)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
