The Humidifier Heater Base PMH7000PLUS/PMH7000 is intended to add moisture and to warm the breathing gasses for administration to a patient. The Humidifier Heater Base, PMH7000PLUS/PMH7000 is intended for use on adult at hospital and home.
Device Story
Humidifier heater base; provides respiratory humidification/warming for adult patients; used in hospital and home. Device inputs: temperature data from dual temperature probe. Operation: dual temperature servo control system regulates gas temperature at humidifying chamber outlet and patient delivery point; heater wire in breathing circuit provides additional control. Output: warmed, humidified gas. Operator: clinician or patient. Safety features: audible/visual alarms for high/low temperature, probe disconnect, heater wire fault; automatic shutdown after 10 minutes of alarm condition. Modification from predicate (K080149) includes updated front/side panel design for maneuverability and improved heater wire interface connector.
Clinical Evidence
Bench testing only. No clinical data required. Testing included mechanical, electrical, and thermal safety, electromagnetic compatibility (IEC 60601-1, IEC 60601-1-2), and performance/accuracy testing per ISO 8185.
Technological Characteristics
Heated Passover humidifier; dual-sensor temperature probe; dual temperature servo control; electronic PID algorithm microprocessor; inspiratory and expiratory heater wire support. Complies with IEC 60601-1, IEC 60601-1-2, and ISO 8185. Connectivity: heater wire interface connector.
Indications for Use
Indicated for adult patients in hospital or home settings requiring warmed, humidified respiratory gases.
Regulatory Classification
Identification
A respiratory gas humidifier is a device that is intended to add moisture to, and sometimes to warm, the breathing gases for administration to a patient. Cascade, gas, heated, and prefilled humidifiers are included in this generic type of device.
{1}------------------------------------------------
**MAY 01 2014**
K132857
Image /page/0/Picture/1 description: The image shows the logo and contact information for Pacific Medico Co., LTD. The address is 2-15-10 Hongo, Bunkyo-ku, Tokyo 113-0033 Japan. The phone number is +81-3-3818-0573, and the fax number is +81-3-3818-7392. The website address is www.pacific-medico.com.
# 510(k) Summary
## Sumitter/510(k) Holder
PACIFIC MEDICO
Pacific Medico Co., Ltd. 2-6-4 Hongo Bunkyo-ku, Tokyo Tel: +81 3-3818-0573 Fax: +81 3-3818-7392 Registration Number: 10028261
Contact Person: Yoshio Toyama
Date of Preparation: March 20, 2014
# Manufacturing Facility
Pacific Denshi Co., Ltd. 650-6, Okaisshiki, Sizuoka, 410-0012, Japan Registration #:3007584704
## Device Name
Classification Reference: 21 CFR 868.5450 Product Code: BTT Device Trade Name: Humidifier Heater Base PMH7000PLUS/PMH7000 Classification Name: Respiratory Gas Humidifier Classification: Class II
#### Predicate Device
The predicate device is Humidifier Heater Base, PMH7000 with 510(k) number, K080149 and the concurrence date December 18, 2008.
#### Device Description
The Humidifier Heater Base, PMH7000PLUS/PMH7000 is a humidifier incorporating with ventilator and provides respiratory humidification through patient breathing circuit for adult at hospital and home.
Humidifier Heater Base, PMH7000PLUS/PMH7000 was designed to comply with applicable portions of the following standards.
PACIFIC MEDICO CO., LTD. 2-15-10 Hongo, Bunkyo-ku, Tokyo 113-0033 Japan Phone: +81-3-3818-0573 Fax: +81-3-3818-7392
Yoshio Toyama, General Manager Regulatory Affairs Product Development Department Yoshio.Toyama@pacific-medico.com
{2}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Pacific Medico. The logo features a stylized letter "P" above the words "PACIFIC MEDICO". There are horizontal lines on either side of the words "PACIFIC MEDICO".
Medical Electrical Equipment - General Requirements IEC 60601-1 Electromagnetic Compatibility - Requirements and Tests IEC 60601-1-2 Humidifiers for Medical Use 150 8185 ISO 14971 Medical Devices
This heater base features a dual temperature servo control system which regulates the temperature of the gas delivered to the patient and the gas at outlet of humidifying chamber. It continuously monitors temperatures by a dual temperature probe. The heater base warms and humidifies the gas through the humidifier chamber and in addition a heater wire in the patient breathing circuit further controls same gas temperature. The operator is allowed to set the temperature of two locations: temperature of the chamber outlet gas and the gas immediately before delivered to the patient.
The Humidifier Heater Base PMH7000PLUS/PMH7000 is equipped with alarms which activate audible and visual indicators to alert the operator of adverse conditions. Alarms are provided for high and low airway temperature, high and low chamber outlet temperature probe disconnect or fault condition, heater wire fault connection, and the light identifies the area of alarm.
The Humidifier Heater Base PMH7000PLUS/PMH7000 includes several safety features such as prevention of excessive heating which can be harmful to the patient or can cause damage to the heater base itself. To ensure patient safety, the heater base will shut down under any alarm conditions if it continues for 10 minutes.
The Humidifier Heater Base are available in two models, PMH7000PLUS and PMH7000. PMH7000PLUS/PMH7000 has both inspiratory and expiratory heater wires for use in the inspiratory and expiratory limbs of breathing circuit. The use of the expiratory heater wire with PMH7000PLUS/PMH7000 is to control humidity in the expiratory limb and to reduce moisture returned to the ventilator.
The materials used have not been changed.
#### Description of Device Modification
The subject device has the same technological characteristics and indications for use, material composition, intended use as predicate; the hardware, and mechanical aspects of Humidifier Heater Base. PMH7000PLUS/PMH7000 have been updated to current technology equivalent to the cleared device (Humidifier Heater Base, PMH7000, K080149 dated December 18, 2008) as described below. The subject device is substantially equivalent to the predicate device.
PACIFIC MEDICO CO., LTD. 2-15-10 Hongo, Bunkyo-ku, Tokyo 113-0033 Japan Phone: +81-3-3818-0573 Fax: +81-3-3818-7392
Yoshio Toyama, General Manager Regulatory Affairs Product Development Department Yoshio.Toyama@pacific-medico.com
·2/5
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Pacific Medico. The logo features a stylized letter 'P' above the words 'PACIFIC MEDICO'. There are two horizontal lines on either side of the words 'PACIFIC MEDICO'.
PACIFIC MEDICO CO., LTD.
2-15-10 Hongo, Bunkyo-ku,
Tokyo 113-0033 Japan
Phone: +81-3-3818-0573
Fax: +81-3-3818-7392
www.pacific-medico.com
Humidifier Heater Base, PMH7000PLUS/PMH7000 are modified version of Humidifier Heater Base, PMH7000.
The modification is to change the front and side panel design in order to realize the easy maneuverability and, and to change the interface for heater wire connector in order to facilitate easy and simple connectivity.
Modifications which have been made are described below.
#### PMH7000PLUS:
- 1) changing the front and side panel design
- 2) changing the interface connector for heater wire
- 3) improving the maneuverability.
## PMH7000:
- 1) changing the front and side panel design
- 2) changing the interface connector for heater wire
Because of the improvements of the maneuverability from predicate device of PMH7000, the upgraded model of PMH7000 is added under the name of PMH7000PLUS. The purpose of adding this name of PMH7000PLUS is to identify the differentiation from the predicate device.
#### Intended Use
The Humidifier Heater Base PMH7000PLUS/PMH7000 is intended to add moisture and to warm the breathing gasses for administration to a patient.
The Humidifier Heater Base, PMH7000PLUS/PMH7000 is intended for use on adult at hospital and home.
The intended use has not been changed.
## Technological Characteristics
The Technological Characteristics of Humidifier Heater Base PMH7000PLUS/ PMH7000 are equivalent to the predicate devices Humidifier Heater Base PMH7000 (available in two models, the PMH7000D and the PMH7000S).
PMH7000PLUS/PMH7000 are equivalent to the PMH7000 in terms of: type (heated Passover humidification), configuration (chamber, heated wire breathing circuits, dual -sensor temperature probe), power usage (same heater system ratings),
PACIFIC MEDICO CO., LTD. 2-15-10 Hongo, Bunkyo-ku, Tokyo 113-0033 Japan Phone: +81-3-3818-0573 Fax: +81-3-3818-7392
Yoshio Toyama, General Manager Requlatory Affairs Product Development Department Yoshio. Toyama@pacific-medico.com
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a company logo and contact information. The logo is a circular emblem with the letters "TOV" inside. The text next to the logo includes the company name "PACIFIC M", followed by the address "2-15-10 He", and the city and postal code "Tokyo 113". The contact information includes a phone number starting with "+8", a fax number starting with "+81-3", and a website address starting with "www.pacific".
performance (same temperature and humidity output), control method (electronic and PID algorithm microprocessor), and uses equivalent materials and some common components.
# Non Clinical Tests
PACIFIC MEDICO
Non-clinical testing of Humidifier Heater Base PMH7000PLUS/ PMH7000 has been carried out covering mechanical, electrical and thermal safety, environmental conditions and electromagnetic compatibility, functional verification, and performance capacity and accuracy.
PMH7000PLUS/ PMH7000 meets the requirements of the IEC 60601-1 and IEC 60601-1-2 electro medical and EMC standards, and the relevant USA deviations to these in ANSI/AAMI ES60601-1. It complies with performance and safety requirements of the ISO 8185 particular standard for Humidification Systems.
#### Substantial Equivalence
The modified PMH7000PLUS/PMH7000 have the following similarities to those which previously received 510(k) concurrence (PMH7000).
- · have the same intended use,
- · use the same operating principle,
- · incorporate the same basic humidifier design,
- · have the same shelf life, and
- · packaged using the same materials and process.
The comparison table is provided as a summary of the technological characteristics relative to the predicate devices. This is to demonstrate that Pacific Medico PMH7000PLUS/PMH7000 Humidifier Heater Base for respiratory therapy (heater base) is substantially equivalent to the predicate device.
#### Conclusion
The testing carried out for Humidifier Heater Base PMH7000PLUS/PMH7000 indicates that it meets design and performance functional requirements. The proposed device meets the requirements of international and USA medical electrical equipment standards for safety, and key performance and safety requirements from the particular standard for humidification systems.
This information indicates that Humidifier Heater Base PMH7000PLUS/PMH7000 is substantially equivalent to the predicate devices.
- 1) The modification does not affect the intended use or alter the fundamental scientific technology.
PACIFIC MEDICO CO., LTD. 2-15-10 Hongo, Bunkyo-ku, Tokvo 113-0033 Japan Phone: +81-3-3818-0573 Fax: +81-3-3818-7392
Yoshio Toyama, General Manager Regulatory Affairs Product Development Department Yoshio. Toyama@pacific-medico.com
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the contact information for Pacific Medico Co., LTD. The address is 2-15-10 Hongo, Bunkyo-ku, Tokyo 113-0033 Japan. The phone number is +81-3-3818-0573, and the fax number is +81-3-3818-7392. The website is www.pacific-medico.com.
2) Based on the results of the comparison and reproducibility studies described in this 510(k) submission, it is concluded that the Humidifier Heater Base, PMH7000PLUS/PMH7000 is equivalent to the predicate device as an aid for mitigation of dryness of airway.
- 3) Under the above technological comparison, discussion on substantial equivalence, it is demonstrated that the PMH7000PLUS/PMH7000 is equivalent to the predicate device.
- 4) It is determined that the Performance Testing Clinical is not required because of the above comparison.
- 5) Verification and validation testing have been performed which demonstrate that PMH7000PLUS/PMH7000 function as intended and is substantially equivalent to the predicate devices for the intended use.
PACIFIC MEDICO CO., LTD. 2-15-10 Hongo, Bunkyo-ku, Tokyo 113-0033 Japan Phone: +81-3-3818-0573 Fax: +81-3-3818-7392
== PACIFIC MEDICO =
Yoshio Toyama, General Manager Regulatory Affairs Product Development Department Yoshio.Toyama@pacific-medico.com
રાદ
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized design featuring three abstract shapes resembling human figures or birds in flight.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administr 0003 New Hammehire A Silver Spring, MD 20993.
May 1. 2014
Pacific Medico Co., LTD. Mr. Yoshio Toyama. General Manager Regulatory Affairs, Product Development Department 2-6-4 Hongo. Bunkyo-ku Tokyo 113-0033 Japan
Re: K132857
Trade/Device Name: Humidifier Heater Base PMH7000PLUS/PMH7000 Regulation Number: 21 CFR 868.5450 Regulation Name: Respiratory Gas Humidifier Regulatory Class: 11 Product Code: BTT Dated: March 31. 2014 Received: April 1. 2014
Dear Mr. Toyama:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
# Page 2 - Mr. Toyama
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
**Tejashri Purohit-Sheth, M.D.**
Clinical Deputy-Director
DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) KI32857
Device Name
Humidifier Heater Base PMH7000PLUS/PMH7000
Indications for Use (Describe)
The Humidifier Heater Base PMH7000PLUS/PMH7000 is intended to add moisture and to warm the breathing gases for administration to the patient.
The Humidifier Heater Base, PMH7000PLUS/PMH7000 is intended for use on adult at home.
Type of Use (Select one or both, as applicable) [X] Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
: : } : . : 보(월) 후 2 FOR FDA USE ONLY : 上一篇: : 第 : ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ ・ . . . . : Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
A
2
1
Anya C. Harry -S 2014.05.05 15:58:13 -04'00'
PSC Publishing Scrvices (10)) 443-6740
EF
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
{9}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.