The ConchaTherm® Neptune® is a respiratory humidifier designed to heat and humidify respiratory gases delivered via endotracheal tubes, nasal cannula or face masks to adult, pediatric, infant and neonatal patients. This system may be used with either conventional (non-heated wire) breathing circuits or compatible (21-volt) Hudson Respiratory Care Incorporated (RCI) heated-wire circuits. The ConchaTherm® NeptuneTM can be used with ventilators, continuous flow systems, oxygen diluters and blenders, adjustable nebulizer adapters for aerosol therapy, or nonflammable anesthesia gases to help maintain patient body temperature. When used with the Hudson RCI ConchaTherm Neptune Heated Humidifier and Hudson RCI ventilator circuits, the ConchaSmart Column provides heated humidification for patients with and without an artificial airway in place. To provide a continuous flow of heated and humidified gas to spontaneously breathing patients.
Device Story
ConchaTherm Neptune is an active heated humidifier providing heat/moisture to medical gases for invasive/non-invasive ventilation. System components include ConchaSmart Column (disposable humidifier cartridge) and Comfort Flo Humidification System (heated wire breathing circuit). Device accepts input from ventilators, flow systems, or blenders; transforms gas via heated cartridge; outputs heated/humidified gas to patient. Used in clinical settings by healthcare providers. Output helps maintain patient body temperature and airway hydration. System features include low water notification, auto-settings for clinician defaults, and compatibility with 21-volt heated-wire circuits. Benefits include improved patient comfort and respiratory gas conditioning.
Clinical Evidence
Bench testing only. Compliance verified against ISO 8185 (respiratory gas humidifiers), IEC 60601-1 (general safety), IEC 60601-1-8 (alarm systems), and IEC 60601-1-4 (programmable electrical medical systems). Biocompatibility testing performed per ISO 10993-3, -5, -6, -10. Sterility validated per ISO 11137-1/2. Shelf life and useful life testing conducted. Software verification and validation performed.
Technological Characteristics
Active heated humidifier; 21-volt heated-wire circuit compatibility. Materials biocompatibility per ISO 10993. Features: low water notification, auto-settings mode, side-mounting capability. Compliant with ISO 8185:2007. Sterilization per ISO 11137. Software-controlled alarm and temperature regulation.
Indications for Use
Indicated for adult, pediatric, infant, and neonatal patients requiring heated and humidified respiratory gases via endotracheal tubes, nasal cannula, or face masks. Used with ventilators, continuous flow systems, oxygen diluters/blenders, or anesthesia gases. Comfort Flo system indicated for spontaneously breathing patients.
Regulatory Classification
Identification
A respiratory gas humidifier is a device that is intended to add moisture to, and sometimes to warm, the breathing gases for administration to a patient. Cascade, gas, heated, and prefilled humidifiers are included in this generic type of device.
{0}------------------------------------------------
## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
## A. Name, Address, Phone and Fax Number of Applicant
Teleflex Medical, Incorporated 2917 Weck Drive Research Triangle Park, NC 27709 USA Phone: 919-433-8050 919-433-4996 Fax:
## B. Contact Person
Amanda Webb Regulatory Affairs Specialist
## C. Date Prepared
February 5, 2013
## D. Device Name
| Trade Name: | ConchaTherm Neptune Heated Humidifier |
|-----------------------|---------------------------------------|
| Common Name: | Respiratory Gas Humidifier |
| Product Code: | BTT |
| Regulation Number: | 868.5450 |
| Classification: | II |
| Classification Panel: | Anesthesiology |
| Trade Name: | ConchaSmart Column |
| Common Name: | Respiratory Gas Humidifier |
| Product Code: | BTT |
| Regulation Number: | 868.5450 |
| Classification: | II |
| Classification Panel: | Anesthesiology |
| Trade Name: | Comfort Flo Humidification System |
| Classification Name: | Respiratory Gas Humidifier |
| Product Code: | BTT |
| Regulation Number: | 868.5450 |
Anesthesiology
II
Classification:
Classification Panel:
{1}------------------------------------------------
## E. Predicate Device
This submission demonstrates substantial equivalence to the following predicate devices:
- . ConchaTherm Neptune Heated Humidifier - K063758
- . Minimal Compliance Column - K993355
- Comfort Flo Humidification System K061736 ●
### F. Device Description
The ConchaTherm Neptune Heated Humidifier is an active heated humidifier designed to provide heat and moisture to the medical gases delivered to a patient through continuous flow, invasive ventilation or non-invasive ventilation.
The ConchaSmart Column is an accessory to the ConchaTherm Neptune Heated Humidifier. It is a single use, disposable humidifier cartridge which is intended for neonatal, infant, pediatric, and adult patients requiring heated humidification. The ConchaSmart Column is used by inserting the column into the center of the ConchaTherm Neptune.
The Comfort Flo Humidification System is intended to deliver heated and humidified respiratory gases to spontaneously breathing patients. The Comfort Flo Humidification System is designed to be used in conjunction with the ConchaTherm Neptune Heated Humidifier for neonate/infant, pediatric, and adult populations. The Comfort Flo Humidification System includes a Heated Wire Breathing Circuit, a ConchaSmart Column, and various Class I exempt accessories.
## G. Indications for Use
#### ConchaTherm Neptune Heated Humidifier:
The ConchaTherm® Neptune TM is a respiratory humidifier designed to heat and humidify respiratory gases delivered via endotracheal tubes, nasal cannula or face masks to adult, pediatric, infant and neonatal patients. This system may be used with either conventional (non-heated wire) breathing circuits or compatible (21-volt) Hudson Respiratory Care Incorporated (RCI) heated-wire circuits.
The ConchaTherm® NeptuneTM can be used with ventilators, continuous flow systems, oxygen diluters and blenders, adjustable nebulizer adapters for aerosol therapy, or nonflammable anesthesia gases to help maintain patient body temperature.
#### ConchaSmart Column:
When used with the Hudson RCI ConchaTherm Neptune Heated Humidifier and Hudson RCI ventilator circuits, the ConchaSmart Column provides heated humidification for patients with and without an artificial airway in place.
#### Comfort Flo Humidification System:
To provide a continuous flow of heated and humidified gas to spontaneously breathing patients.
{2}------------------------------------------------
## H. Technological Characteristics Comparison to the predicate
The proposed ConchaTherm Neptune Heated Humidifier, ConchaSmart Column and Comfort Flo Humidification System are substantially equivalent to the predicate devices listed above in that the indications for use, the intended use, and fundamental scientific technology remain unchanged. The following tables summarize the technological differences between the proposed and predicate devices.
### ConchaTherm Neptune
| Different Technological Characteristic | Predicate Device |
|------------------------------------------------------------------|--------------------------------------------------------|
| Ability to side mount the unit | Only allowed for rear mounting the unit |
| Heated Wire Subglottic: 32° - 40°C<br>(Intubated Mode) | Heated Wire Subglottic: 32° - 39°C<br>(Intubated Mode) |
| Supraglottic (Non-Invasive): 28° - 37°C | Supraglottic (Non-Invasive): 30° - 37°C |
| Continuous Flow Applications 1-60LPM | Continuous Flow Applications 1-40LPM |
| Compliant with ISO 8185:2007 | Compliant with ISO 8185:1997 |
| Low Water Notification | Does not have a low water notification |
| Auto Settings Mode - allows clinician to<br>set default settings | Does not allow clinician to set default<br>settings |
#### ConchaSmart Column
| Different Technological Characteristic | Predicate Device |
|-----------------------------------------|--------------------------------------|
| Includes ability to trigger low water | Does not interact with humidifier to |
| notification in the ConchaTherm Neptune | provide a low water notification |
| Labeled for 30 day Useful Life | Not labeled for useful life |
## Comfort Flo Humidification System
| Different Technological Characteristic | Predicate Device |
|--------------------------------------------------------------------------------|----------------------------------------------------------------------------|
| Includes above ConchaSmart Column | Included the Comfort Flo Column |
| Additional product offering with a<br>corrugated heated wire breathing circuit | Only included a smooth bore PVC heated<br>wire breathing circuit |
| Shelf Life on corrugated Comfort Flo<br>Humidification System | Does not include a shelf life on full<br>Comfort Flo Humidification System |
| Rated flow for adults is 1-60 LPM | Rated flow for adults is 1-40LPM |
| Labeled for 30 day Useful Life | Tested but not labeled for useful life |
## I. Performance Data
The following testing was performed on the proposed devices.
- ISO 8185 .
- IEC 60601-1 .
- IEC 60601-1-8 .
- IEC 60601-1-4
- Biocompatibility .
- o ISO 10993-3
- o ISO 10993-5
{3}------------------------------------------------
- o ISO 10993-6
- o ISO 10993-10
- . Sterility
- o ISO 11137-1
- o ISO 11137-2
- . Shelf life
- Useful life .
- Software Verification and Validation ●
## J. Conclusion
The device data and test results demonstrate that the device is as safe and as effective as the predicate device and therefore substantially equivalent.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three wing-like shapes. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the bird symbol. The logo is presented in black and white.
Public Health Service
Food and Drug Administration 0903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 6, 2014
Teleflex Medical Incorporated Ms. Amanda Webb Regulatory Affairs Specialist 2917 Weck Drive Research Triangle Park, NC 27709
Re: K131912
Trade/Device Name: ConchaTherm Neptune Heated Humidifier ConchaSmart Column Comfort Flo Humidification System Regulation Number: 21 CFR 868.5450 Regulation Name: Respiratory Gas Humidifier Regulatory Class: II Product Code: BTT Dated: February 10, 2014 Received: February 11, 2014
Dear Ms. Webb:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Ms. Webb
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Image /page/5/Picture/7 description: The image contains a combination of text and graphic elements. The text includes "Tejashri Purohit-Sheth, M.D.", "Clinical Deputy Director", and "DAGRID". There is also a signature present. The graphic elements include a geometric shape resembling a stylized triangle or pyramid, and the word "FOR" is visible in the lower right corner.
Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
Sincerely yours,
{6}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known) K 131912
#### Device Name
Comfort Flo Humidification System
Indications for Use (Describe)
To provide a continuous flow of heated and humidified gas to spontaneously breathing patients.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY -్లు
గ్రామం నుండి 10 కి.మీ. దూర Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Image /page/6/Picture/12 description: The image shows a logo with the letters "FM" in a stylized, geometric font. The letters are outlined with multiple layers, creating a three-dimensional effect. The overall design is modern and abstract.
## Anya C. Harry -S 2014.03.05 14:12:41 -05'00'
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (1/14)
15C Publishing Sen ices (101) 43-6740
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{7}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known) K131912
Device Name ConchaSmart Column
#### Indications for Use (Describe)
When used with the Hudson RCI ConchaTherm Neptune Heated Humidifier and Hudson RCI ventilator circuits, the ConchaSmart Column provides heated humidification for patients without an artificial airway in place.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY . " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " "
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Image /page/7/Picture/12 description: The image shows the letters FDA in a stylized, outlined font. The letters are formed by multiple parallel lines, giving them a layered or stacked appearance. The overall design is simple and graphic, with a focus on the bold, geometric shapes of the letters.
## Anya C. Harry -S 2014.03.05 14:10:51 -05'00'
, - Fran
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete. and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (1/14)
SC Datalishing Sansaies (101) 441-6740
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{8}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known) K131912
Device Name
ConchaTherm Neptune Heated Humidifier
#### Indications for Use (Describe)
The ConchaTherm® Neptune® is a respiratory humidifier designed to heat and humidify respiratory gases delivered via endotracheal tubes, nasal cannula or face masks to adult, pediatric, infant and neonatal patients. This system may be used with either conventional (non-heated wire) breathing circuits or compatible (21-volt) Hudson Respiratory Care Incorporated (RCI) heated-wire circuits.
The ConchaTherm® Neptune® can be used with ventilators, continuous flow systems, oxygen diluters and blenders, adjustable nebulizer adapters for aerosol therapy, or non-flammable anes to help maintain patient body temperature.
| | Comments of the contribution of the first and the program and the program and the comments of the comments of ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | |
|--|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| | Type of Use (Select one or both, as applicable) | |
| | | |
| | | |
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY . . . Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
# Anya C. Harry -S 2014.03.05 14:07:24 -05'00'
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
* An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (1/14)
PSC Publishing Services (NDI) 443-46740 EF
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.