VENTSTAR HEATED, INFINITY ID BREATHING CIRCUIT HEATED

K102618 · Draeger Medical AG & Co. KG · BTT · Dec 10, 2010 · Anesthesiology

Device Facts

Record IDK102618
Device NameVENTSTAR HEATED, INFINITY ID BREATHING CIRCUIT HEATED
ApplicantDraeger Medical AG & Co. KG
Product CodeBTT · Anesthesiology
Decision DateDec 10, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5450
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

VentStar Dual Heated / VentStar Heated Disposable inspiratory heated breathing circuit with humidifier chamber for connection to a Fisher & Paykel MR850 humidifier, for conveying moistened breathing gas between the humidifier and adult patients with a body weight of at least 40 kg (88 lbs). VentStar Heated (N) / VentStar Heated (N) basic Disposable inspiratory heated breathing circuit for connection to a Fisher & Paykel MR850 humidifier, for conveying moistened breathing gas between the humidifier and neonatal patients with a body weight of up to 5 kg (11 lbs). Infinity ID Breathing Circuit Heated (N) Inspiratory heated breathing circuit for connection to a Fisher & Paykel MR850 humidifier, for moping moistened breathing gas between the humidifier and neonates with a body weight of up lo 5 kg (11 Ibs) Intended for single-use only. Breathing circuit with integrated transponder. The transponder serves as a carrier of product-specific data for processing by Dräger Infinity ID equipment. Infinity ID Breathing Circuit Dual Heated / Infinity ID Breathing Circuit Heated Heated breathing circuit for connection to a Fisher & Paykel MR850 humidifier, for conveying moistened breathing gas between the humidfier and adults with a body welght of at least 40 kg (88 moration broaining goo Doching circuit with integrated transponder. The transponder. The transponder serves as a carrier of product-specific data for processing by Dräger Infinity ID equipment.

Device Story

Heated breathing gas hose systems for neonatal or adult ventilation in ICUs; consists of inspiratory/expiratory limbs and y-piece. Inspiratory limbs contain heating wires to minimize condensation when connected to Fisher & Paykel MR850 humidifier; expiratory limbs include water traps. Infinity ID variants include RFID transponders storing hose characteristics (patient type, diameter, resistance, compliance, shelf life) for automatic readout by compatible Dräger ventilators. Used by clinicians in hospital settings to deliver humidified gas; helps maintain airway moisture and reduce condensation during ventilation.

Clinical Evidence

Bench testing only. Testing included electrical and thermal safety, inspiratory and expiratory limb performance, system compatibility with Dräger ventilators and Fisher & Paykel MR850 humidifier, and material biocompatibility per ISO 10993-1.

Technological Characteristics

Heated breathing circuits with integrated heating wires; Infinity ID models include RFID transponders for data storage. Materials evaluated per ISO 10993-1. Compatible with Fisher & Paykel MR850 humidifier. Single-use design.

Indications for Use

Indicated for adult patients (≥40 kg) or neonatal patients (≤5 kg) requiring mechanical ventilation; used to convey moistened breathing gas from a Fisher & Paykel MR850 humidifier to the patient.

Regulatory Classification

Identification

A respiratory gas humidifier is a device that is intended to add moisture to, and sometimes to warm, the breathing gases for administration to a patient. Cascade, gas, heated, and prefilled humidifiers are included in this generic type of device.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the word "Dräger" in a bold, sans-serif font. Above the word "Dräger" is the text "K102618". The text "K102618" is written in a handwritten style. #### ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------DEC: 1 0 2010 ## Applicants Name and Address: Dräger Medical AG & Co. KG Moislinger Allee 53-55 23542 Lübeck Germany ## Manufacturer Name and Address: Dräger Medical AG & Co. KG Moislinger Allee 53-55 23542 Lübeck Germany #### Establishment Registration Number : 9611500 ## Contact Person: Ulrich Schröder Director Regulatory & Clinical Affairs Tel. No.: 011 49 (451) 882-3648 Fax No .: 011 49 (451) 882-3018 # Applicants US Contact Person Joyce Kilrov Vice President, Processes, Quality & Requlatory Tel. No .: (215) 660-2626 Fax No .: (215) 721-5424 # Date submission was prepared: 2010/09/03 #### Device Name: Common Usual Name: Heated Breathing Circuits Proprietary Name: VentStar Heated / Infinity ID Breathing Circuit Heated Product Code: BTT Classification Name: humidifier, respiratory gas, (direct patient interface) 21 CFR 868.5450 Regulation Number: Class: 11 {2}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the word "Dräger" in a bold, sans-serif font. The word is black and the background is white. The two dots above the "a" are also black. | 510(k) number | Trade name | Company | |---------------|------------------------------------------------------------------------|----------------------------------| | K073706 | Respiratory Humidifier, Model MR850, heated system RT200, RT225, RT212 | Fisher & Paykel Healthcare, Ltd. | | K033710 | | | | K020332 | | | | K983112 | | | | K934140 | MR290 single use humidification chamber | Fisher & Paykel Healthcare, Ltd. | # Legally Marketed Devices to which Substantial Equivalence is claimed: # Device Description: The 510(k) comprises heated breathing gas hose systems including a humidifier chamber for the ventilation of neonatal or adult patients in Intensive Care Units. The hose systems consist of two breathing hoses (inspiratory and expiratory limb) with an additional y- piece. Inside the inspiratory and in case of the dual heated systems also in the expiratory limb of the hose system there is a heating wire which can be connected to a Fisher & Paykel humidifier MR850 to heat up the breathing gas to minimise condensation. An expiratory limb without heating wire includes a water trap. There are additional sensor ports where the temperature sensors of the Fisher & Paykel humidifier MR850 (e.g. Fisher & Paykel 900MR868) can be placed in. In case of the neonatal system, the hose is equipped with an incubator extension for use inside an incubator. The Infinity ID heated breathing circuits are based on the heated breathing circuits additionally equipped with an RFID identification tag. The tag stores characteristics of the breathing hose (e.g. patient type, hose diameter, resistance and compliance values, manufacturing date and shelf life) and is automatically be readout when connected to the outlets of a ventilator device which supports this function. # Intended Use: | Device Name | Intended Use | |-------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | VentStar Dual Heated<br>/ VentStar Heated | Disposable inspiratory heated breathing circuit with humidifier chamber for<br>connection to a Fisher & Paykel MR850 humidifier, for conveying moistened<br>breathing gas between the humidifier and adult patients with a body weight of<br>at least 40 kg (88 lbs). | | VentStar Heated (N) /<br>VentStar Heated (N)<br>basic | Disposable inspiratory heated breathing circuit for connection to a Fisher &<br>Paykel MR850 humidifier, for conveying moistened breathing gas between the<br>humidifier and neonatal patients with a body weight of up to 5 kg (11 lbs). | | Infinity ID Breathing<br>Circuit Heated (N) | Inspiratory heated breathing circuit for connection to a Fisher & Paykel MR850<br>humidifier, for conveying moistened breathing gas between the humidifier and<br>neonates with a body weight of up to 5 kg (11 lbs). Intended for single-use only.<br>Breathing circuit with integrated transponder. The transponder serves as a<br>carrier of product-specific data for processing by Dräger Infinity ID equipment. | | Infinity ID Breathing<br>Circuit Dual Heated /<br>Infinity ID Breathing<br>Circuit Heated | Heated breathing circuit for connection to a Fisher & Paykel MR850 humidifier,<br>for conveying moistened breathing gas between the humidifier and adults with<br>a body weight of at least 40 kg (88 lbs). Intended for single-use only. Breathing<br>circuit with integrated transponder. The transponder serves as a carrier of<br>product-specific data for processing by Dräger Infinity ID equipment. | {3}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the word "Dräger" in a bold, sans-serif font. The word is black and the background is white. The two dots above the "a" are also black. # Summary of Testing The following testing has been performed on the breathing systems: - Electrical and thermal safety - - Inspiratory and expiratory limb performance ー - System Compatibility with Dräger ventilators and humidifier MR850 - - ﺖ Materials - o All materials used have been evaluated acc. to tests outlined in ISO 10993-1 # Conclusion: The intended use and general construction as the predicate devices remain the same. The design of the heated breathing circuits is identical in fit, form and function to marketed products named in the table above. It has been shown that product performance is given within the range the device can be used by clinicians. The technological characteristics and the results of the performance data demonstrated that the heated breathing hoses issued no new risks during design verification and validation which could question device use. In accordance with the Federal Food and Cosmetic Act and 21 CFR Part 807, based on the information provided in this premarket notification Dräger Medical AG & Co. KG concludes that the heated breathing circuits are safe, effective and substantially equivalent to the predicate devices as described in this application. {4}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized emblem. The emblem consists of a symbol that resembles an abstract caduceus or a stylized representation of human figures. # DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Draeger Medical AG & Company KG C/O Ms. Joyce Kilroy Draeger Medical System, Incorporated 3135 Quarry Road Telford, Pennsylvania 18969 # DEC 1 0 2010 Re: K102618 Trade/Device Name: VentStar Heated / Infinity ID Breathing Circuit Heated Regulation Number: 21 CFR 868.5450 Regulation Name: Respiratory Gas Humidifier Regulatory Class: II Product Code: BTT Dated: September 3, 2010 Received: September 13, 2010 Dear Ms. Kilrov: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ Page 2- Ms. Kilroy Please be advised that FDA's issuance of a substantial equivalence determination does not, mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, 1 please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ # Indications for Use DEC 1 0 2010 # 510(k) Number: K102618 Device Name: VentStar Heated / Infinity ID Breathing Circuit Heated Indications for Use: #### VentStar Dual Heated / VentStar Heated Disposable inspiratory heated breathing circuit with humidifier chamber for connection to a Fisher & Paykel MR850 humidifier, for conveying moistened breathing gas between the humidifier and adult patients with a body weight of at least 40 kg (88 lbs). #### VentStar Heated (N) / VentStar Heated (N) basic Disposable inspiratory heated breathing circuit for connection to a Fisher & Paykel MR850 humidifier, for conveying moistened breathing gas between the humidifier and neonatal patients with a body weight of up to 5 kg (11 lbs). #### Infinity ID Breathing Circuit Heated (N) Inspiratory heated breathing circuit for connection to a Fisher & Paykel MR850 humidifier, for moping moistened breathing gas between the humidifier and neonates with a body weight of up lo 5 kg (11 Ibs) Intended for single-use only. Breathing circuit with integrated transponder. The transponder serves as a carrier of product-specific data for processing by Dräger Infinity ID equipment. # Infinity ID Breathing Circuit Dual Heated / Infinity ID Breathing Circuit Heated Heated breathing circuit for connection to a Fisher & Paykel MR850 humidifier, for conveying moistened breathing gas between the humidfier and adults with a body welght of at least 40 kg (88 moration broaining goo Doching circuit with integrated transponder. The transponder. The transponder serves as a carrier of product-specific data for processing by Dräger Infinity ID equipment. AND/OR Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Eyaluation (ODE) L. Schutte (Division Sign-Off) Division of Anesthesiology, General പ്രശ്യമപ്പും 1 Infection Control, Dental Devices 4102618 510(k) Number: _
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...