← Product Code [BTT](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BTT) · K041963

# AMERICAN BANTEX HUMIDIFIER, DRY, BUBBLE, MODELS B9000, 3 PSI AND B9100, 6 PSI (K041963)

_American Bantex Corp. · BTT · Oct 20, 2004 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BTT/K041963

## Device Facts

- **Applicant:** American Bantex Corp.
- **Product Code:** [BTT](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BTT.md)
- **Decision Date:** Oct 20, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5450
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic

## Indications for Use

The American Bantex Humidifier is intended for use with oxygen concentrators or gas sources in a patients home, physicians office or hospital / institutional environment. The humidifier increases the moisture content of the airstream gases for administration directly to the patient.

## Device Story

The American Bantex Humidifier is a non-heated, bubble-type humidifier designed for use with oxygen concentrators or gas sources. It consists of a refillable plastic jar containing distilled or sterile water. Oxygen gas enters the device through a winged-nut inlet, passes through the water to increase moisture content, and exits via a standard tapered medical tubing connector for patient delivery. The device includes a 3 or 6-PSI safety valve that triggers an alarm if a kink or blockage occurs in the delivery system. It is intended for use in homes, clinics, or hospitals by patients or healthcare providers. The device is reusable for up to 30 days for a single patient.

## Clinical Evidence

Bench testing only. The device was tested to ensure it consistently performed within design parameters, demonstrating safety and effectiveness equivalent to the predicate device.

## Technological Characteristics

Non-heated bubble humidifier. Materials: polypropylene (jar), ABS (lid, check valve), brass (check valve), PVC (filter tube/bulb). Features: 3 or 6-PSI safety valve for blockage alarm, double-helix threaded jar, winged-nut inlet, tapered outlet. Single-patient use, non-sterile. No software or electronic components.

## Regulatory Identification

A respiratory gas humidifier is a device that is intended to add moisture to, and sometimes to warm, the breathing gases for administration to a patient. Cascade, gas, heated, and prefilled humidifiers are included in this generic type of device.

## Predicate Devices

- Airlife Bubble Humidifier ([K991484](/device/K991484.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# AMERICAN BANTEX CORPORATION

1815 Rollins Road Burlingame, California (650) 697 - 3545 (650) 697 - 3596 Tracy S. Best, Regulatory Affairs Consultant Preparation Date: July 10, 2004

### Summary of Safety and Effectiveness for the:

| Trade Name:          | American Bantex Humidifier                                                                 |
|----------------------|--------------------------------------------------------------------------------------------|
| Common Name:         | Bubble Humidifier, Dry                                                                     |
| Classification Name: | Respiratory gas humidifier, 21 CFR 868.5450, 73BTT, Direct Patient<br>Interface - Class II |

#### Legally Marketed Predicate Devices for Substantial Equivalence:

*K991484 - Airlife Bubble Humidifier, Allegiance Healthcare Corporation

Rationale for SE: The Airlife Bubble Humidifier is a Class II device that is intended to moisten the breathing gases that are delivered directly to the patient for breathing. This predicate device, along with the submitted device operate similarly by having the patient breathing gases pass through distilled water or sterile water in a refiliable plastic jar. Both devices are reusable for not more than 30 days. These are made from acrylic and polypropylene plastics and are intended for single patient use. The material types used in this and in the predicate devices are identical. The materials have previously been tested and accepted for biocompatibility and are widely accepted as industry material standards. Both the submitted device, and the predicate are latex free.

## Description of Submitted Device:

The American Bantex Humidifier is device that provides humidity in a non-heated pathway. The American Bantex Humidifier works with most all Oxygen Concentrators or to an oxygen source used in homes, doctors' office, hospital or institutional environments. The inlet uses an easy-to-grab winged styled nut that universally adapts to most all gas connections. The outlet is a push-on standard medical tubing connector with a tapered end. The American Bantex Humidifier is equipped with either a 3 or 6-PS1 safety valve that will alarm if there is a kink or blockage in the delivery system.

#### Intended Uses of the American Bantex Humidifier:

The American Bantex Humidifier is intended for use with Oxygen Concentrators in a patients home, physicians office or hospital / institutional environment. The humidifier increases the moisture content of the airstream gases for administration to the patient.

{1}------------------------------------------------

# Technological Characteristics and Substantial Equivalence:

| Component:           | American Bantex Humidifier | Airlife Bubble Humidifier |
|----------------------|----------------------------|---------------------------|
| Humidifier Bottle:   | Clear polypropylene        | Clear polypropylene       |
| Lid:                 | ABS                        | ABS                       |
| Check valve and lid: | Brass / ABS                | Brass / ABS               |
| Filter tube / bulb:  | PVC / PVC                  | PVC / ABS                 |
| 510(k) Approval      | This submission            | K991484                   |

Table I b: Comparison of American Bantex Humidifier to the predicate Airlife Bubble Humidifie

The characteristics, method of operation, accessory use and single patient use are equivalent to that of the predicate device. Both have a double helix threaded jar that is impact resistant and seals easily.

#### Conclusion:

The successful testing demonstrated the device consistently performed within its design parameters, is as safe and effective, and performs as well as, or better than, the predicated device.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the department's name around the perimeter. In the center of the seal is an emblem featuring a stylized caduceus, a symbol often associated with medicine and healthcare. The caduceus consists of a staff with two snakes coiled around it and wings at the top.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 2 0 2004

American Bantex Corporation C/O Mr. Tracy S. Best Regulatory Affairs Consultant 994 North Main Street Bountiful, Utah 84010

Re: K041963

Trade/Device Name: American Bantex Humidifier Regulation Number: 21 CFR 868.5450 Regulation Name: Respiratory Gas Humidifier Regulatory Class: II Product Code: BTT Dated: October 4, 2004 Received: October 7, 2004

Dear Mr. Best:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration,

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Best

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Chiu Lin, Ph.D.

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known):

Device Name: AMERICAN BANTEX HUMIDIFIER

Indications for Use:

The American Bantex Humidifier is intended for use with oxygen concentrators or gas sources in a patients home, physicians office or hospital / institutional environment. The humidifier increases the moisture content of the airstream gases for administration directly to the patient.

Single patient use only

Non-sterile

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

## Concurrence of CDRH, Office of Device Evaluation (ODE)

Ques

(Division Sian-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Device

510(k) Number. K041963

011

Page 1 of *_*_

---

**Source:** [https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BTT/K041963](https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BTT/K041963)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
