Reinforced Tube: Rusch Reinforced Endotracheal Tubes are designed for nasal intubation and are indicated for airway management. The correct designation (oral, nasal or oral/nasal) is printed on the tube and unit package. Reinforced Endotracheal Tubes may be used to reduce the potential for kinking whenever an unusual positioning of the head or neck is required following intubation. The Reinforced Endotrached to be used on all patients requiring ventilation, pediatric and adult. Safety Clear and Safety Clear Pediatric Tubes: Rusch Endotracheal Tubes, Cuffed and Uncuffed, with or without Murphy Eye are for oral or nasal intubation. They are indicated for airway management. The Rusch Endotracheal Tubes are intended to be used on all patients requiring ventilation, pediatric and adult.
Device Story
Sterile, single-use endotracheal tubes inserted via nose or mouth to establish patent airway for ventilation. Devices consist of PVC tube, cuff, inflation line, pilot balloon, one-way valve, and 15mm connector. Reinforced version includes embedded stainless-steel spiral for kink resistance and X-ray visualization; Safety Clear versions include radiopaque line. Used in clinical settings where intubation is performed; operated by healthcare professionals. Output is a maintained airway for mechanical ventilation. Benefits include secure airway access, reduced kinking in difficult positions, and DEHP-free material construction.
Clinical Evidence
Bench testing only. Performance verified via visual inspection, dimensional testing, bonding strength, tube curvature, cuff burst/herniation/diameter, kink resistance, tube collapse, and tracheal seal testing. Biocompatibility testing conducted per ISO 10993-1, including cytotoxicity, sensitization, irritation, acute systemic toxicity, material-mediated pyrogenicity, implantation, subacute systemic toxicity, genotoxicity, chemical characterization (extractables/leachables), and particulate/VOC analysis.
Indicated for airway management in pediatric and adult patients requiring ventilation via oral or nasal intubation. Reinforced tubes are indicated for cases requiring unusual head/neck positioning to reduce kinking risk. No contraindications.
Regulatory Classification
Identification
A tracheal tube is a device inserted into a patient's trachea via the nose or mouth and used to maintain an open airway.
{0}------------------------------------------------
February 11, 2025
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Teleflex Incorporated Bouse Angela Senior Manager Regulatory Affairs-Product Management 3015 Carrington Mills Blvd Suite 600 Morrisville, North Carolina 27560
Re: K241451
Trade/Device Name: Rusch Endotracheal Tubes (Reinforced): Rusch Endotracheal Tubes (Safety Clear); Rusch Endotracheal Tubes (Safety Clear Pediatric) Regulation Number: 21 CFR 868.5730 Regulation Name: Tracheal Tube Regulatory Class: Class II Product Code: BTR Dated: May 22, 2024 Received: January 3, 2025
Dear Bouse Angela:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}------------------------------------------------
the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Bradley Q. Quinn -S
Bradley Quinn Assistant Director DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K241451
#### Device Name
Rusch Endotracheal Tubes (Reinforced); Rusch Endotracheal Tubes (Safety Clear); Rusch Endotracheal Tubes (Safety Clear Pediatric)
#### Indications for Use (Describe)
Reinforced Tube: Rusch Reinforced Endotracheal Tubes are designed for nasal intubation and are indicated for airway management. The correct designation (oral, nasal or oral/nasal) is printed on the tube and unit package. Reinforced Endotracheal Tubes may be used to reduce the potential for kinking whenever an unusual positioning of the head or neck is required following intubation. The Reinforced Endotrached to be used on all patients requiring ventilation, pediatric and adult.
Safety Clear and Safety Clear Pediatric Tubes: Rusch Endotracheal Tubes, Cuffed and Uncuffed, with or without Murphy Eye are for oral or nasal intubation. They are indicated for airway management. The Rusch Endotracheal Tubes are intended to be used on all patients requiring ventilation, pediatric and adult.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(k) Summary
#### A. Name, Address, Phone and Fax Number of Applicant
Teleflex Medical, Inc. 3015 Carrington Mill Blvd Morrisville, NC 27560 USA Phone: 919-332-8189
#### B. Contact Person
Angela Bouse Principal Regulatory Affairs Specialist
#### C. Date Prepared
February 6, 2025
# D. Device Name
#### Subject Device 1
| Trade Name: | Rusch Endotracheal Tubes (Reinforced |
|-----------------------|--------------------------------------|
| Common Name: | Tracheal Tube |
| Product Code: | BTR |
| Regulation Number: | 21 CFR 868.5730 |
| Classification Name: | Tube, Tracheal (W/Wo Connector) |
| Classification: | II |
| Classification Panel: | Anesthesia |
#### Subject Device 2
| Trade Name: | Rusch Endotracheal Tube (Safety Clear) |
|-----------------------|----------------------------------------|
| Common Name: | Tracheal Tube |
| Product Code: | BTR |
| Regulation Number: | 21 CFR 868.5730 |
| Classification Name: | Tube, Tracheal (W/Wo Connector) |
| Classification: | II |
| Classification Panel: | Anesthesia |
#### Subject Device 3
| Trade Name: | Rusch Endotracheal Tubes (Safety Clear Pediatric |
|-----------------------|--------------------------------------------------|
| Common Name: | Tracheal Tube |
| Product Code: | BTR |
| Regulation Number: | 21 CFR 868.5730 |
| Classification Name: | Tube, Tracheal (W/Wo Connector) |
| Classification: | II |
| Classification Panel: | Anesthesia |
{5}------------------------------------------------
# E. Predicate Device
K990619 Rusch Reinforced Endotracheal (or Tracheal) Tube K961837 Rusch Oral/Nasal Tracheal Tube/Safety Clear Plus Tracheal Tube K993786 Rusch Oral/Nasal (Safety Clear Plus) Tracheal Tube
#### F. Device Description
An endotracheal tube is a device that is inserted into the trachea via the nose or the mouth to establish a patent airway to allow ventilation. The proposed Teleflex Medical Rusch Endotracheal Tubes are sterile, single use devices that are made non made with DEHP. The tracheal tubes contain a compatible cuff, inflation line, pilot balloon and one-way valve. A radiopaque line is incorporated into the full length of the Safety Clear tracheal tube, while the Reinforced tube contains an embedded stainless-steel spiral that allows for X-ray visualization. Each tracheal tube is supplied with an appropriately sized 15mm connector.
#### G. Indications for Use
See Substantial Equivalence Table Below
## H. Contraindications
None
## I. Substantial Equivalence
The subject devices are substantially equivalent to the predicate device with respect to intended use, technology, and construction. The differences between the predicate and the subject devices are minor and any risks have been mitigated through testing. The table below summarizes the differences between the subject and predicate device.
| Features | Teleflex Medical<br>Rusch Reinforced<br>Endotracheal Tube<br>(Subject Device 1) | Rusch/Teleflex Medical<br>Reinforced Endotracheal<br>Tube<br>(Predicate K990619) | Comments |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|----------|
| Classification Name | Tube, Tracheal (w/wo connector) | Same | Same |
| Product Code | BTR | Same | Same |
| Regulation Number | 868.5730 | Same | Same |
| Indications for Use | Rusch Reinforced<br>Endotracheal Tubes are<br>designed for oral or nasal<br>intubation and are indicated<br>for airway management.<br>The correct designation<br>(oral, nasal or oral/nasal) is<br>printed on the tube and unit | Same | Same |
| Substantial Equivalence Comparative Table: Reinforced Endotracheal Tube (Subject | | | | |
|----------------------------------------------------------------------------------|--|--|--|--|
| Device 1) | | | | |
{6}------------------------------------------------
| | package. Reinforced<br>Endotracheal Tubes may be<br>used to reduce the potential<br>for kinking whenever an<br>unusual positioning of the<br>head or neck is required<br>following intubation. The<br>Reinforced Endotracheal<br>Tubes are intended to be<br>used on all patients<br>requiring ventilation,<br>pediatric and adult. | | |
|-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
| Environment of<br>Use | Locations where ET<br>intubation may be<br>performed | Not stated | Same |
| Patient Population | Pediatric and Adult | Not stated | Same |
| Contraindications | None | Same | Same |
| Single Use | Yes | Same | Yes |
| Size Range | 3.5 mm – 10.0 mm | 2.5 mm – 11.0 mm | Within cleared<br>range |
| Cuffed | Yes | Same | Same |
| Radiopaque | Yes | Same | Same |
| Connection to<br>ventilation source | 15 mm connector | Same | Same |
| Method of<br>Sterilization | Ethylene Oxide 10-6 SAL | Same | Same |
| Biocompatibility | Materials have been tested<br>per ISO 10993-1 | Same | Same |
| Made without<br>DEHP | Yes | No | While slightly<br>different, the<br>prosed device is<br>made without<br>the known<br>hazardous<br>substance |
| Sterile | Yes | Same | Same |
| Eye | Murphy | Same | Same |
| Graduations | Multiple cm markings | Same | Same |
| Shaft (main tube) | PVC | Same | Same |
| Features | Teleflex Medical<br>Rusch Safety Clear<br>Endotracheal Tube<br>(Subject Device 2) | Rusch/Teleflex Medical<br>Oral/Nasal Tracheal<br>Tube/Safety Clear Plus<br>Tracheal Tube<br>(Predicate K961837) | Comments |
| Classification<br>Name | Tube, Tracheal (w/wo<br>connector) | Same | Same |
| Product Code | BTR | Same | Same |
| Regulation<br>Number | 868.5730 | Same | Same |
| Indications for Use | Rusch Endotracheal Tubes,<br>Cuffed and Uncuffed, with<br>or without Murphy Eye are<br>for oral or nasal intubation.<br>They are indicated for<br>airway management. The<br>Rusch Endotracheal Tubes<br>are intended to be used on<br>all patients requiring<br>ventilation, pediatric and<br>adult. | Same | Same |
| Environment of<br>Use | Locations where ET<br>intubation may be<br>performed | Not stated | Same |
| Patient Population | Pediatric and Adult | Not stated | Same |
| Contraindications | None | Same | Same |
| Single Use | Yes | Same | Yes |
| Size Range | 4.5 mm - 10.00 mm | 4.5 mm - 11.0 mm | Within cleared<br>range |
| Cuffed | Yes | Same | Same |
| Radiopaque | Yes | Same | Same |
| Connection to<br>ventilation source | 15 mm connector | Same | Same |
| Method of<br>Sterilization | Ethylene Oxide 10-6 SAL | Same | Same |
| Biocompatibility | Materials have been tested<br>per ISO 10993-1 | Same | Same |
| Made without<br>DEHP | Yes | No | While slightly<br>different, the |
| | | | prosed device is |
| | | | made without |
| | | | the known |
| | | | hazardous |
| | | | substance |
| Sterile | Yes | Same | Same |
| Eye | Murphy | Same | Same |
| Graduations | Multiple cm markings | Same | Same |
| Shaft (main tube) | PVC | Same | Same |
| Cuff | PVC | Same | Same |
{7}------------------------------------------------
#### Substantial Equivalence Comparative Table: Safety Clear Endotracheal Tube (Subject Device 2)
{8}------------------------------------------------
## Substantial Equivalence Comparative Table: Safety Clear Pediatric Endotracheal Tube (Subject Device 3)
| Features | Teleflex Medical<br>Rusch Safety Clear<br>Endotracheal Tube<br>(Subject Device 2) | Rusch/Teleflex Medical<br>Oral/Nasal Tracheal<br>Tube/Safety Clear Plus<br>Tracheal Tube<br>(Predicate K961837) | Comments |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification<br>Name | Tube, Tracheal (w/wo<br>connector) | Same | Same |
| Product Code | BTR | Same | Same |
| Regulation<br>Number | 868.5730 | Same | Same |
| Indications for Use | Rusch Endotracheal Tubes,<br>Cuffed and Uncuffed, with<br>or without Murphy Eye are<br>for oral or nasal intubation.<br>They are indicated for<br>airway management. The<br>Rusch Endotracheal Tubes<br>are intended to be used on<br>pediatric. | Rusch Oral/Nasal Tracheal<br>Tube, Cuffed.<br>Magill/Murphy is a device<br>inserted into a patient's<br>trachea via the nose or<br>mouth and used to<br>maintain an open airway. | Similar,<br>updated to align<br>with the<br>indication for<br>the rest of the<br>Rusch Safety<br>Clear<br>Endotracheal<br>Tube product<br>family. |
| Environment of<br>Use | Locations where ET<br>intubation may be<br>performed | Not stated | Same |
| Patient Population | Pediatric | Not stated | Same |
| Contraindications | None | Same | Same |
| Single Use | Yes | Same | Yes |
| Size Range | 2.5 mm - 4.0 mm | 2.0 mm - 4.0 mm | Within cleared<br>range |
| Cuffed | Yes | Same | Same |
{9}------------------------------------------------
| Radiopaque | Yes | Same | Same |
|-------------------------------------|-----------------------------------------------|------|-------------------------------------------------------------------------------------------------------------|
| Connection to<br>ventilation source | 15 mm connector | Same | Same |
| Method of<br>Sterilization | Ethylene Oxide 10-6 SAL | Same | Same |
| Biocompatibility | Materials have been tested<br>per ISO 10993-1 | Same | Same |
| Made without<br>DEHP | Yes | No | While slightly<br>different, the<br>prosed device is<br>made without<br>the known<br>hazardous<br>substance |
| Sterile | Yes | Same | Same |
| Eye | Murphy | Same | Same |
| Graduations | Multiple cm markings | Same | Same |
| Shaft (main tube) | PVC | Same | Same |
| Cuff | PVC | Same | Same |
## J. Comparison to the Predicate
The table above illustrates the similarities and differences between the subject and predicate devices. The basic technological and operating principles are the same for both devices. Although the indications for use wording is not identical, both the subject and predicate devices are for airway management. Both the subject and predicate devices are disposable, sterile, single patient use devices. The subject devices are substantially equivalent to the predicate device. There are no significant differences that would affect safety and efficacy.
#### K. Performance Data
The bench testing performed verifies that the performance of the subject device is substantially equivalent in terms of critical performance characteristics to the predicate device. These tests include:
- Testing performed included visual inspection, Dimensional testing, Bonding strength (main tube to the side arm, inflation tube and connector), Tube curvature, cuff restrained burst, Bevel angle Cuff inflation, Kink resistance, Cuff herniation, Cuff diameter, Tube collapse and Tracheal seal testing, as well as testing required by ISO 10993-1 Fifth Edition 2018-08 - Biological Evaluation of Medical Devices -- Part 1: Evaluation and testing within a risk management process.
The testing performed verifies that the performance of the subject device is substantially equivalent in terms of critical performance characteristics to the predicate device.
Biological testing was performed according to ISO 10993:
- Cytotoxicity ●
#### 510(k) Summary - Page 6 of 7
{10}------------------------------------------------
- Sensitization ●
- Irritation
- Acute Systemic Toxicity ●
- Material Mediated Pyrogenicity ●
- Implantation
- Subacute Systemic Toxicity
- Genotoxicity
- Chemical Characterization-Exhaustive and Simulated Use. Toxicological Risk ● Assessment Related to Extractable and Leachables
- Particulates and VOC. Toxicological Risks Related to Inhalation of VOCs and ● Particulates
#### L. Conclusion
The subject device has the same intended use and technological characteristics and construction as the predicate. Test results demonstrate that the subject device meets it's intended use. It is for these reasons that the subject device can be found substantially equivalent to the predicate devices.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.