The IntubaidFlex™ is intended for use as an aid in the placement of an Endotracheal Tube (ETT) directly or through an Intubating Laryngeal Mask Airway (LMA) during non difficult and difficult intubation procedures. The IntubaidFlex achieves its purpose by providing the user with a visual confirmation of where the tip of the device is in the human anatomy. The flexible tip of the IntubaidFlex allows the user to guide the ETT in the desired direction.
Device Story
IntubaidFlex is a disposable, sterile, handheld, portable stylet with an integrated camera and light source. It functions as an accessory to an Endotracheal Tube (ETT) or Intubating Laryngeal Mask Airway (LMA). The device provides real-time video visualization of the laryngeal inlet to assist clinicians during intubation. The stylet features an articulating flexible tip, allowing the user to steer the ETT into the desired position. The device connects via an integrated cable to a TV monitor, handheld monitor, or PC (via USB) for live video display. By providing visual confirmation of the device tip location within the anatomy, it aids in successful intubation. It is intended for use by healthcare professionals in clinical settings.
Clinical Evidence
Bench testing only. Compliance with ISO 10993 (biocompatibility), IEC 60601-1 and IEC 60601-1-2 (electrical safety/EMC), and ISO 8600-3 and ISO 8600-5 (endoscope performance). Mechanical testing included tensile strength (5.0 lb load), 36-inch drop test, and fatigue testing of the plunger mechanism.
Technological Characteristics
Disposable, sterile, handheld stylet with embedded video camera and light source. Features an articulating flexible tip. Connectivity via integrated cable to external monitors or PC (USB). Compliant with ISO 10993, IEC 60601-1, IEC 60601-1-2, ISO 8600-3, and ISO 8600-5.
Indications for Use
Indicated for use as an aid in endotracheal tube (ETT) placement, either directly or via intubating laryngeal mask airway (LMA), during both non-difficult and difficult intubation procedures in patients requiring airway management.
Regulatory Classification
Identification
A tracheal tube is a device inserted into a patient's trachea via the nose or mouth and used to maintain an open airway.
{0}------------------------------------------------
# Section V-510(k) Summary
## AUG 0 4 2009
#### 5.1. Device Information
| Category | Comments |
|-------------------------------------|--------------------------------------------------|
| Sponsor: | Company: EZC Medical, LLC |
| | Address: The Presidio |
| | 16A Funston Avenue |
| | San Francisco, Ca, 94129, USA |
| | Tel: (415) 561-2550 |
| | |
| Correspondence Contact Information: | Pavan Sethi, Ph.D. |
| | Vice President, Regulatory, Quality & Scientific |
| | Affairs |
| | Tel: (408) 513-7529 |
| | Pavan@ezcmedical.com |
| | |
| | |
| Device Common Name: | Stylet / Videoscope |
| Device Classification and Code: | Class II, BTR |
| Device Classification Name: | Tracheal Tube (Accessory) |
| Device Proprietary Name: | IntubaidFlexTM |
#### Predicate Device Information 5.2.
| Predicate Device: | ETView Tracheoscopic Ventilation Tube (TVTTM) |
|-------------------------------------------|-----------------------------------------------|
| Predicate Device Manufacturer: | ETView Ltd |
| Predicate Device Common Name | Tracheal Tube |
| Predicate Device Classification and Code: | Class II, BTR |
| Predicate Device Classification: | 21CFR868.5730 |
#### 5.3. Date Summary Prepared
Date Submitted: March 20, 2009
#### Description of Device: 5.4.
The IntubaidFlex" is a simple, disposable, single-patient use, sterile, handheld, portable stylet with a built in camera, that fits within an Endotracheal tube (ETT) directly or through an Intubating Laryngeal Mask Airway(LMA) and provides for video-assisted visualization of the laryngeal inlet. The IntubaidFlex" can plug into any TV monitor, hand held monitor or PC (via USB) for live video visualization.
#### 5.5. Intended Use
The IntubaidFlex™ is intended for use as an aid in the placement of an Endotracheal Tube (ETT) directly or through an Intubating Laryngeal Mask Airway (LMA) during non difficult and difficult intubation procedures. The IntubaidFlex achieves its purpose by providing the user with a visual confirmation of where the
V-1
{1}------------------------------------------------
tip of the device is in the human anatomy. The flexible tip of the IntubaidFlex allows the user to guide the ETT in the desired direction.
#### 5.6. Comparison to Predicate Device:
The EZC Medical's IntubaidFlex™ is substantially equivalent to ETView Tracheoscopic Ventilation Tube (TVTTM) in intended use and operation (K082420).
Both the devices are endoscopic, and transmit images through a video camera to a video monitor and use a light source for illumination. Both the devices are disposable, single-patient use, sterile, have an embedded video camera, an embedded light source, and an integrated cable which connects to the video monitor.
The devices are different in design. The IntubaidFlex™ is an accessory to the ETT, where as TVT" is an Endotracheal Tube by itself. The IntubaidFlex " has an articulation capability, which TVT" does not have.
However, these differences do not impact the intended use or optical performance of the IntubaidFlex. The testing described below demonstrates that the differences in the devices do not raise any unresolved issues of safety or efficacy.
EZC Medical concludes that the devices are substantially equivalent.
#### 5.7. Summary of Supporting Data:
Biocompatibility data demonstrates that the device is in compliance with ISO 10993.
Electrical Safety and Electromagnetic Compatibility data demonstrates that the device is in compliance with IEC 60601-1 and IEC 60601-1-2.
Bench testing has demonstrated that the device is in compliance with ISO 8600-3 and ISO 8600-5. In addition the bench testing included the Tensile Strength test, Drop test and Fatigue test. The Tensile Strength Test demonstrated that joints can withstand a 5.0 lb tensile load, the Drop Test demonstrated that the device can withstand a 36 inch fall and the Fatigue test demonstrated that the plunger can withstand multiple actuations without affecting the image functionality of the device.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows a logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the seal is a stylized symbol featuring three curved lines, which may represent people or services.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-0609 Silver Spring, MD 20993-0002
# AUG 0 4 2009
Ms. Pavan Sethi Vice President Regulatory, Quality & Scientific Affairs EZC Medical, L.L.C. 16 A Funston Avenue San Francisco, California 94129
Re: K090777
Trade/Device Name: EZC Medical, LLC, IntubaidFlexTM Regulation Number: 21 CFR 868.5730 Regulation Name: Tracheal Tube Regulatory Class: II Product Code: BTR Dated: July 27, 2009 Received: July 29, 2009
Dear Ms. Sethi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
## Page 2- Ms. Sethi
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours,
signature
Susan Runner, D.D.S., M.A. Acting Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Section IV-Indications for Use Statement
510(k) Number: K090777
EZC Medical, LLC, IntubaidFlex™ Device Name:
Indications for Use:
The IntubaidFlex" is intended for use as an aid in the placement of an Endotracheal Tube (ETT) directly or through an Intubating Laryngeal Mask Airway (LMA) during non difficult and difficult intubation procedures. The IntubaidFlex achieves its purpose by providing the user with a visual confirmation of where the tip of the device is in the human anatomy. The flexible tip of the IntubaidFlex allows the user to guide the ETT in the desired direction.
Prescription Use X (Per 21 CFR §801 subpart D)
Over-The-Counter Use (Per 21 CFR §801 subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
OR
L. Schutte
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K090777
Section IV-Indications for Use Statement
Company Confidential
Page IV-1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.