IST RESPONSE MANUAL RESUSCITATOR, MODELS, 008000, 008003,008006
Applicant
Sims Portex, Inc.
Product Code
BTM · Anesthesiology
Decision Date
Sep 13, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5915
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The 1st Response Manual Resuscitator is a pulmonary-assist device intended to provide respiratory support to patients suffering from respiratory distress. It is intended for use on patients with a body mass of 25 kg (55 lbs) or more.
Device Story
1st Response Manual Resuscitator is a disposable, single-use emergency manual ventilator. Device consists of a compressible plastic bag with inlet and patient-end valves. Inlet valve connects to bag or tube reservoirs for supplemental oxygen. Patient-end valve features 15 mm ID x 22 mm OD connector and exhalation port; optional manometer port, PEEP valve, or exhalation filter available. Operated by clinicians to provide manual respiratory support to patients in respiratory distress. Device functions by manual compression of the bag to deliver gas to the patient; valves ensure unidirectional flow and prevent gas escape during compression. Provides immediate respiratory assistance in emergency settings.
Clinical Evidence
Bench testing only. Performance and specifications verified against ISO 8382:1988 (Resuscitators intended for use with humans) and ISO 5356-1:1987 (Anaesthetic and respiratory equipment - Conical connectors).
Technological Characteristics
Manual, disposable, single-use resuscitator. Materials: plastic bag, valve assemblies. Connectivity: none. Energy source: manual compression. Standards: ISO 8382:1988, ISO 5356-1:1987. Form factor: compressible bag with 15mm/22mm connectors.
Indications for Use
Indicated for patients with body mass of 25 kg (55 lbs) or more suffering from respiratory distress requiring pulmonary-assist respiratory support.
Regulatory Classification
Identification
A manual emergency ventilator is a device, usually incorporating a bag and valve, intended to provide emergency respiratory support by means of a face mask or a tube inserted into a patient's airway.
Laerdal Medical Corporation, Laerdal Silicone Resuscitator, Adult
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Smiths Industries Medical Systems. The logo includes the text "SMITHS INDUSTRIES" in bold, uppercase letters, with "Medical Systems" written in a smaller, italicized font below. To the left of the text is a stylized graphic. Above the logo, the text "SEP | 3 | 1999" is visible.
KGG2057
#### SIMS Portex Inc.
10 Bowman Drive PO Box 0724 Keene NH 03431 USA Telephone: 603-352-3812 Fax: 603-352-3703
# J: 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
### 510(K) SUMMARY:
### COMPANY INFORMATION:
SIMS Portex Inc 10 Bowman Drive Keene, NH 03431 (603) 352-3812 Contact: Timothy J. Talcott Manager, Regulatory Affairs
### PREPARATION DATE OF SUMMARY:
June 16, 1999
### TRADE NAME:
1st Response Manual Resuscitator
#### COMMON NAME:
Manual Resuscitator
### PRODUCT CLASS/CLASSIFICATION:
Class II, 73 BTM, 21 CFR 868.5915
## PREDICATE DEVICE(S):
- SIMS Portex Inc., Ft. Myers Florida, 1ª Response Adult Manual Resuscitators, Cat. No. 008000, 008003, and 008006. Laerdal Medical Corporation, Wappingers Falls, New York, Laerdal Silicone
Resuscitator, Adult.
{1}------------------------------------------------
#### DESCRIPTION:
The 1st Response manual resuscitator is a disposable, single use emergency manual ventilator. Each device consists of a plastic compressible ventilator bag fitted with control valves at each of the two ends.
The inlet valve, opposite the patient end, allows entry of fresh gas into the compressible ventilator bag. The valve blocks escape of fresh gas from the ventilator bag during its compression. Attached to this valve are one of two types of reservoirs; bag reservoir or tube reservoir. These reservoirs serve to collect an overflow of oxygen when a supplemental oxygen supply is used.
The patient end of the ventilator bag is fitted with a second valve assembly. This valve consists of a 15 mm ID x 22 mm OD patient connector and exhalation port. A manometer port is optional.
The device is provided with or without a facemask. Special configurations are available which include a disposable manometer, PEEP valve with adapter, or exhalation filter.
#### INDICATIONS FOR USE:
The 18 Response Manual Resuscitator is a pulmonary-assist device intended to provide respiratory support to patients suffering from respiratory distress. It is intended for use on patients with a body mass of 20 kg (44 lbs) or more.
### TECHNICAL CHARACTERISTICS:
The device has the same technical characteristics as the device we have authorization to market under premarket notification K972283.
#### NON-CLINICAL DATA:
Performance and specifications of the modified device are consistent with all requirements for this device type specified by ISO 8382:1988 (E) Resuscitators intended for use with humans and ISO 5356-1: 1987 - Anaesthetic and respiratory equipment -Conical connectors - Part 1: Cones and sockets.
{2}------------------------------------------------
## CONCLUSION:
The comparison to the predicate devices demonstrate that the proposed device is safe and effective and is substantially equivalent to the predicate devices.
Very truly yours,
SIMS PORTEX INC.
Timothy Halcon
Timothy J. Talcott Manager, Regulatory Affairs
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, consisting of a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 13 1999
Mr. Timothy J. Talcott SIMS Portex Inc. 10 Bowman Drive Keene, NH 03431
Re: K992057 15t Response Adult Manual Resuscitator Regulatory Class: II (two) Product Code: 73 BTM Dated: June 16, 1999 Received: June 18, 1999
Dear Mr. Talcott:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to leqally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 A substantially equivalent determination assumes compliance to 895. with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
#### Page 2 - Mr. Timothy J. Talcott
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its jnternet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Thomas J. Callahan
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
TO 13014804204
# B: INTENDED USE OF DEVICE
### PROPOSED INDICATIONS FOR USE:
Page 1 of 1
510(k) Number (if known): K992057
Device Name: 1st Response Manual Resuscitator
Indications For Use:
The 14 Response Manual Resuscitator is a pulmonary-assist device intended to provide respiratory support to patients suffering from respiratory distress. It is intended for use on patients with a body mass of 25 kg (55 lbs) or more.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) Aus extration OR Over-The-Counter Use Prescription Use (Division Sign-Off) Division of Cardiovascular. Respiratory,
and Neurological Devices 510(k) Number _