K102542 · Atom Medical Corporation · BTL · Apr 6, 2011 · Anesthesiology
Device Facts
Record ID
K102542
Device Name
RESUSCI FLOW; RESUSCI FLOW 104
Applicant
Atom Medical Corporation
Product Code
BTL · Anesthesiology
Decision Date
Apr 6, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5925
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Resusci Flow series (Resusci Flow and Resusci Flow with blender) is a manually operated ventilator intended to assist a doctor in resuscitating an infant who has spontaneous breathing by providing the positive end-expiratory pressure (PEEP) and the peak inspiratory pressure (PIP). The Resusci Flow series (Resusci Flow 104 PV and Resusci Flow 104EV) is intended for resuscitation of neonates in a delivery room and a NICU. It is used to remove secretions from the neonatal mouth and nostrils or to ventilate or assist in ventilating neonates suffering from apnea or other respiratory failure. All are intended for neonates and infants (≤ 10kg or 22 lbs) in a delivery room and a NICU.
Device Story
Resusci Flow series are manually operated gas-powered resuscitators for neonates/infants. Inputs: oxygen and air compression gas sources. Operation: clinician-controlled T-piece system provides PEEP and PIP; models 104PV/104EV include suction capabilities (piping or Venturi-based). Outputs: controlled airway pressure for ventilation; suction for secretion removal. Used in delivery rooms and NICUs by physicians/clinicians. Device provides pressure monitoring via integrated gauges and pressure relief valves to ensure safe delivery of respiratory support. Benefits: assists breathing in infants with respiratory failure or apnea; facilitates airway clearance.
Clinical Evidence
Bench testing only. Performance verified against ISO 10651-5 and ISO 10079-3 standards. Tests included display accuracy of airway pressure gauge, gas discharge flow rate, oxygen concentration accuracy, inspiratory/expiratory resistance, pressure limitation, PEEP/PIP characteristics, vibration, drop testing, and alarm functionality (pressure loss/gas mixer). All models met pass/fail criteria.
Technological Characteristics
Gas-powered manual resuscitator; T-piece design. Features include pressure relief valves, pressure gauges, and optional oxygen/air blenders. Models 104PV/104EV include suction (piping or Venturi). Connectivity: gas inlet sources. Standards: ISO 10651-5 (resuscitators), ISO 10079-3 (suction). No software/electronics.
Indications for Use
Indicated for neonates and infants ≤ 10kg (22 lbs) in delivery rooms and NICUs requiring resuscitation, secretion removal, or ventilation support for apnea or respiratory failure.
Regulatory Classification
Identification
A powered emergency ventilator is a demand valve or inhalator intended to provide emergency respiratory support by means of a face mask or a tube inserted into a patient's airway.
{0}------------------------------------------------
K102542
510(k) Summary Page 1 of 5 27-Mar-11
APR -- 6 2011
Atom Medical Corporation 3-18-15, Hongo, Bunkyo-ku
| 3-18-15, Hongo, Bunkyo-ku<br>Tokyo, Japan 113-0033 | Tel - 011-81-3-3815-3632<br>Fax - 011-81-3-3812-3199 |
|----------------------------------------------------|----------------------------------------------------------------------------------------|
| Official Contact: | Tsuyoshi Sugino - Regulatory Affairs Manager |
| Proprietary or Trade Name: | Resusci Flow series resuscitator |
| Common/Usual Name: | Infant Resuscitator |
| Classification Name/Code: | BTL - powered emergency ventilator<br>CFR 868.5925 |
| Device: | Resusci Flow<br>Resusci Flow and blender<br>Resusci Flow 104 EV<br>Resusci Flow 104 PV |
| Predicate Devices: | GE - Giraffe and Panda – K070210<br>Fisher & Paykel - NeoPuff – K892885 |
## Device Description:
There are four (4) models of the Resusci Flow resuscitator.
- Resusci Flow ●
- Resusci Flow with blender ●
- Resusci Flow 104PV .
- Resusci Flow 104EV .
They all have the same basic design. Table 1 outlines the basic features and differences of each model.
- -
- )
{1}------------------------------------------------
## 510(k) Summary Page 2 of 5 27-Mar-11
#### Table 1 Resusci Flow Series
| Feature | Resusci Flow | Resusci Flow<br>with Blender | Resusci Flow<br>104PV | Resusci Flow<br>104EV |
|-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------|
| PIP (Peak Inspiratory.<br>Pressure) | Yes | Yes | Yes | Yes |
| Suction(Piping Type) | No | No | Yes | No |
| Suction(Venturi Type) | No | No | No | Yes |
| Oxygen / Air Mixture | No | Yes | Yes | Yes |
| Inlet Gas Sources | Mixture of oxygen<br>and air | Oxygen<br>compression<br>gas<br>Air<br>compression<br>gas | Oxygen compression<br>gas<br>Air compression gas<br>Suction compression<br>gas | Oxygen<br>compression gas<br>Air compression<br>gas |
| Pressure Relief Valve | Yes | Yes | Yes | Yes |
| Pressure Gauge | Yes | Yes | Yes | Yes |
| Suction Pressure Gauge | No | No | Yes | Yes |
| Outlet | Single outlet | Single outlet | Double outlets<br>(Outlet A/B) | Double outlets<br>(Outlet A/B) |
| | This outlet is for resuscitation using<br>aerosol tube and "T" Flow Valve.<br>It may be connected a face mask or<br>tracheal tube. | | Outlet A is for resuscitation using<br>aerosol tube and "T" Flow Valve.<br>It may be connected a face mask or<br>tracheal tube.<br>Outlet B can be connected to oxygen<br>tent, oxygen mask, oxygen cannula and<br>humidifier jar etc. | |
| Predicates | Fisher & Paykel<br>RD1000 NEOPUFF<br>INFANT<br>RESUSCITATOR | Giraffe and Panda T-piece Resuscitation System<br>(K070210) | | |
#### Indications for Use:
The Resusci Flow series (Resusci Flow and Resusci Flow with blender) is a manually operated ventilator intended to assist a doctor in resuscitating an infant who has spontaneous breathing by providing the positive end-expiratory pressure (PEEP) and the peak inspiratory pressure (PIP).
The Resusci Flow series (Resusci Flow 104 PV and Resusci Flow 104EV) is intended for resuscitation of neonates in a delivery room and a NICU. It is used to remove secretions from the neonatal mouth and nostrils or to ventilate or assist in ventilating neonates suffering from apnea or other respiratory failure.
All are intended for neonates and infants (< 10kg or 22 lbs) in a delivery room and a NICU.
Environment of Use: Hospitals, delivery suites, NICU
{2}------------------------------------------------
## 510(k) Summary Page 3 of 5 27-Mar-11
### Summary of substantial equivalence
### Resusci Flow
The Resusci Flow model was compared to the predicate Fisher & Paykel NeoPuff (K892885) and found to be substantially equivalent based upon:
Indications for Use - The Resusci Flow has equivalent indications for use to the predicate.
The details of the NeoPuff indications are not available, except in their labeling due to the age of the clearance, 1989. Nonetheless the indications for use are equivalent.
Patient Population - The Resusci Flow has the equivalent patient population neonates and infants. We have been more specific than the predicate NeoPuff and included a limitation of patient weight.
Environment for use - The Resusci Flow has the identical environments for use as the predicate.
Prescriptive - The Resusci Flow is identical to the predicate.
Design and Technology - The Resusci Flow has an equivalent design and features as the predicate and has the identical technology as the predicate.
Performance and Specifications - The Resusci Flow has equivalent specifications of performance as the predicate.
Compliance with standards - Both devices declare compliance with the internal standard ISO 10651-5 for powered resuscitators.
#### Conclusion
The Resusci Flow is substantially equivalent to the predicate NeoPuff (K892885) in indications for use, patient population, and environment for use, technology characteristics, specifications / performance and compliance with international standards
{3}------------------------------------------------
# 510(k) Summary Page 4 of 5 27-Mar-11
# Resusci Flow with Blender, Resusci Flow 104 PV and Resusci Flow 104 EV
The Resusci Flow with Blender, Resusci Flow 104PV, and Resusci Flow 104EV were compared to the predicate GE Giraffe and Panda T-piece resuscitator (K070210). As well as the Resusci Flow with Blender, Resusci Flow 104PV, and Resusci Flow 104EV were compared to the predicate GE Giraffe and Panda (K070210) for equivalence.
Indications for Use - The Resusci Flow with Blender, Resusci Flow 104PV, and Resusci Flow 104EV have equivalent indications for use to the predicate.
Patient Population - The Resusci Flow with Blender, Resusci Flow 104PV, and Resusci Flow 104EV have the equivalent patient population neonates and infants.
Environment for use - The Resusci Flow with Blender, Resusci Flow 104PV, and Resusci Flow 104EV have the identical environments for use as the predicate.
Prescriptive - The Resusci Flow with Blender, Resusci Flow 104PV, and Resusci Flow 104EV are identical to the predicate.
Design and Technology - The Resusci Flow with Blender, Resusci Flow 104PV, and Resusci Flow 104EV have an equivalent design and features as the predicate and has the identical technology as the predicate.
Performance and Specifications - The Resusci Flow with Blender, Resusci Flow 104PV, and Resusci Flow 104EV have equivalent specifications of performance as the predicate.
Compliance with standards - Both devices declare compliance with the internal standard ISO 10651-5 for gas powered resuscitators and ISO 10079-3 for suction equipment.
#### Conclusion
The Resusci Flow with Blender. Resusci Flow 104PV, and Resusci Flow 104EV is substantially equivalent to the predicate GE Giraffe and Panda (K070210) in indications for use, patient population, and environment for use, technology characteristics, specifications / performance and compliance with international standards.
{4}------------------------------------------------
## 510(k) Summary Page 5 of 5 27-Mar-11
## Performance Testing
We have performed bench tests which included the list below and found that the Resusci Flow series (Resusci Flow, Resusci Flow with blender, Resusci Flow 104 PV, and Resusci Flow 104 EV) resuscitators met all pass /fail criteria, cited standards requirements or were found to be equivalent in comparison to the predicates.
- ISO 10651-5 clause 5.1.10 Display accuracy of the airway pressure gauge .
- . Gas discharge flow rate from the patient circuit
- ISO 10651-5 Clause 7.7.1 Delivered concentration accuracy of oxygen .
- ISO 10651-5 clause 7.1.2.2 Inspiratory resistance .
- ISO 10651-5 clause 7.2.4 pressure limitation .
- ISO 10651-5 clause 7.2.4 and 5.1.10 Pressure characteristics of peak . inspiratory pressure (PIP inspiratory resistance)
- . Confirm the maximum open pressure (Pmax)
- ISO 10651-5 clause 7.2.4 and 5.1.10 positive end-expiratory pressure (PEEP) .
- . Environment Performance
- . Vibration Testing
- . Drop Testing
- Test pressure loss alarm .
- ISO 11195 Gas mixers alarm for loss of pressure .
- . ISO 10079-3 for suction equipment
- . Safety requirements
- Vomitus Resistance .
- Expiratory Resistance .
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, with three stylized wing segments and a flowing tail.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Atom Medical Corporation C/O Mr. Paul Dryden ProMedic, Incorporated 24301 Woodsage Drive Bonita Springs, Florida 34134
APR -- 6 2011
Re: K102542
Trade/Device Name: Resusci Flow Series, Resusci Flow, Resusci Flow with Blender, Resusci Flow 104 PV and Resusci Flow 104 EV Regulation Number: 21 CFR 868.5925 Regulation Name: Powered Emergency Ventilator Regulatory Class: II Product Code: BTL Dated: March 27, 2011 Received: March 29, 2011
Dear Mr. Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Mr. Dryden
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal requirem. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its offective number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
in fer
Anthony Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
## Indications for Use Statement
510(k) Number:
K102542
Device Name:
Resusci Flow series Resusci Flow Resusci Flow with blender Resusci Flow 104 PV Resusci Flow 104 EV
Indications for Use:
The Resusci Flow series (Resusci Flow and Resusci Flow with blender) is a manually operated ventilator intended to assist a doctor in resuscitating an infant who has spontaneous breathing by providing the positive end-expiratory pressure (PEEP) and the peak inspiratory pressure (PIP).
The Resusci Flow series (Resusci Flow 104 PV and Resusci Flow 104EV) is intended for resuscitation of neonates in a delivery room and a NICU. It is used to remove secretions from the neonatal mouth and nostrils or to ventilate or assist in ventilating neonates suffering from apnea or other respiratory failure.
All are intended for neonates and infants (≤ 10kg or 22 lbs) in a delivery room and a NICU.
Prescription Use XX (Part 21 CFR 801 Subpart D) or '
Over-the-counter use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
f. Schulthess
(Division Sign-Off)
Division of Anesthesiology, General Hospital
Infection Control and Dental Devices
510(k) Number: K.V. 542
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.