K133958 · Maquet Critical Care AB · BSZ · Mar 20, 2014 · Anesthesiology
Device Facts
Record ID
K133958
Device Name
FLOW-I ANESTHESIA SYSTEM C20, C30 & C40
Applicant
Maquet Critical Care AB
Product Code
BSZ · Anesthesiology
Decision Date
Mar 20, 2014
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 868.5160
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The indication for FLOW-i Anesthesia System is administering inhalation Anesthesia while controlling the entire ventilation of patients with no ability to breathe, as well as in supporting patients with a limited ability to breathe. The system is intended for use on neonatal to adult patient populations. The system is intended for use in hospital environments, except MRI environment, by healthcare professionals trained in inhalation Anesthesia administration.
Device Story
FLOW-i Anesthesia System is a software-controlled, semi-closed inhalation anesthesia delivery system. It utilizes ICU-ventilator technology, volume reflector technology, and electronically controlled injector vaporizers. Input includes anesthesia agents (Sevoflurane, Desflurane, Isoflurane) and gases (O2, Air, N2O). The system provides controlled or supported ventilation and monitors patient gas concentrations via a Patient Gas Analyzer (PGA). Operated by trained healthcare professionals in hospital settings, the system features O2Guard (hypoxic mixture prevention), HLM mode for cardiopulmonary bypass, and apnea mute functions. Output is displayed on a GUI, providing clinicians with real-time ventilation parameters, trend data, and gas concentration metrics to guide anesthesia administration. Benefits include precise agent delivery, integrated lung mechanics measurement, and enhanced safety features for diverse patient populations.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical design verification and validation, including requirement verification, regression testing, free user testing (FUT), code review, static code analysis, and adherence to applicable product standards (e.g., IEC 60601-1, ISO 5356-1, ASTM F1101-90).
Technological Characteristics
Software-controlled semi-closed anesthesia system. Features volume reflector technology and electronically controlled injector vaporizers. Connectivity includes remote service functionality for log transfer. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-2-13, ISO 5356-1, ISO 5360, and ASTM F1101-90. Hardware includes core unit, backup gas system, and various mounting accessories.
Indications for Use
Indicated for neonatal to adult patients requiring inhalation anesthesia and mechanical ventilation support (full or partial). Contraindicated for use in MRI environments.
Regulatory Classification
Identification
A gas machine for anesthesia is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. The device may include a gas flowmeter, vaporizer, ventilator, breathing circuit with bag, and emergency air supply. A gas machine for analgesia is a device used to administer to a patient an analgesic agent, such as a nitrous oxide-oxygen mixture (maximum concentration of 70 percent nitrous oxide).
{0}------------------------------------------------
## 510(k) SUMMARY
SUMMARY as required by section 807.92
# Submitter Name & Address
Maquet Critical Care AB Röntgenvägen 2 SE-171 54 Solna, Sweden Tel: (011) 46 8 730 7300 Fax: (011) 46 8 730 7838
Contact Persons for this submission:
Carina Lundberg Regulatory Affairs Manager Phone: direct: (011) 46 8 730 7786 Email: carina.lundberg@maquet.com
Application Correspondent:
Lia Liebgold Regulatory Affairs Specialist Maquet Medical Systems USA 45 Barbour Pond Drive Wayne, NJ 07470 Email: lia.liebgold(@;maquet.com Phone: 973-709-7340 Fax: 973-807-9210
Date prepared: 12/20/2013
| Trade Name : | Model: | Model no: |
|--------------------------|------------|-----------|
| FLOW-i Anesthesia System | FLOW-i C20 | 66 77 200 |
| | FLOW-i C30 | 66 77 300 |
| | FLOW-i C40 | 66 77 400 |
# MAR 2 0 2014
{1}------------------------------------------------
#### Device Classification
| Common Name | Classification Number | Class | Regulation Number |
|-------------------------|-----------------------|-------|-------------------|
| Gas-Machine, Anesthesia | BSZ | II | 21 CFR 868.5160 |
#### Predicate Device Identification
| Legally marketed devices to which equivalence is being claimed | 510(k) # |
|----------------------------------------------------------------|----------|
| MAQUET FLOW-i Anesthesia System | K112114 |
#### Indications for Use
The indication for FLOW-i Anesthesia System is administering mhalation Anesthesia while controlling the entire ventilation of patients with no ability to breathe, as well as in supporting patients with a limited ability to breathe. The system is intended for use on neonatal to adult patient populations. The system is intended for use in hospital environments, except MRI environment, by healthcare professionals trained in inhalation Anesthesia administration.
#### Intended use of the Device
The system is intended for use in administering Anesthesia while controlling the entire ventilation of patients with no ability to breathe, as well as in supporting patients with a limited ability to breathe.
The system is intended for use by healthcare professionals, trained in the administration of Anesthesia.
The system is intended for use on neonatal to adult patient populations.
The system is intended for use in hospital environments, except MRI environment,
When not in operation, the system is designed for in-hospital transport.
#### Device Description
The modified FLOW-i Anesthesia System version 3.0 is an Anesthesia System designed to meet the many ventilatory challenges within Anesthesia, as well as to provide inhalation Anesthesia. It is intended to serve a wide range of patients from neonatal to adult.
FLOW-i Anesthesia System is a software-controlled semi-closed system for inhalation Anesthesia (Sevoffurane, Desflurane, Isofturane and/or nitrous oxide).
The most important performance features of the FLOW-i Anesthesia System are:
- a ventilator whose functionality is based on ICU-ventilator technology, o
- the volume reflector technology, o
- the electronically controlled injector vaporizers and o
- the ergonomic design. o
{2}------------------------------------------------
This 510(k) submission for the FLOW-i Anesthesia System is based on the following modifications :
- ♥ Removal of the Control Gas Analyzer (CGA)
- Measured Tidal Volume lower range in Pressure Control extended .
- Inspiratory and Expiratory Hold (Software functionality) .
- . Apnea Mute (Software functionality)
- HLM Mode (Software functionality) ●
- O2Guard Safety function (Software functionality) .
- Volume Reflector Indicator (Software functionality) .
## Non-clinical Testing and Performance
Design verification and validation has demonstrated that the FLOW-i Anesthesia System performs within its specifications and within the limits of the applied performance standards. The following characteristics of FLOW-i Anesthesia System were thoroughly tested: technical data. measurement ranges and measurement accuracy, delivery accuracy, construction, features, interfaces, handling, critical situations and interventions.
The design verification activities for the modified FLOW-i Anesthesia System version 3.0 consist of:
- Requirement verification of affected requirements ●
- ↓ Regression testing
- . Free User Testing (FUT)
- Code review and static code analysis .
- Verification of applicable product standards .
- IEC 60601-1
- IEC 60601-1-1 0
- IEC 60601-1-2 0
- IEC 60601-1-8 O
- IEC 60601-2-13 0
- ISO 5356-1 O
- ISO5360 O
- ASTM F1101-90 O
- o CGA V-5
Design validation has been performed for each market requirement for FLOW-i Anesthesia System in order to assure the product meets its intended use and user needs, including usability.
#### Clinical Investigation
The functionality added in the proposed FLOW-i Anesthesia System version 3.0 (K133958) does not add any new functions that need to be validated by clinical investigation.
{3}------------------------------------------------
## CHANGES AND SUBSTANTIAL EQUIVALENCE DISCUSSION
#### Comparison of Intended Use
The Intended Use for the modified FLOW-i Anesthesia System version 3.0 (K133958) is identical to the predicate device, FLOW-i Anesthesia System version 2.1 (K112114).
## Comparison of Technology Used
The technological characteristics for the modified FLOW-i Anesthesia System version 3.0 (K133958) is identical to the predicate device, FLOW-i Anesthesia System version 2.1 (K112114).
## Similarities and differences
The main improvements introduced in the modified FLOW-i Anesthesia System version 3.0 (K133958) from the predicate device FLOW-i Anesthesia System version 2.1 (K112114) are:
## Removal of the Control Gas Analyzer (CGA)
The Control Gas Analyzer (CGA) in the fresh gas channel was removed in the modified FLOW-i Anesthesia System version 3.0 (K133958). In the predicated device FLOW-i Anesthesia System version 2.1 (K112114) the CGA was used to supervise the vaporizer.
In system version 3.0, the vaporizer has been improved to secure the agent delivery. This, together with some minor and complimentary functions of the Patient Gas Analyzer (PGA) in supervising the agent concentration, will ensure the system's safety and reliability and in turn makes the CGA redundant.
The FLOW-i Anesthesia System uses safety measures at different levels. After removal of the CGA (previously the third of four layers) the safety system consists of three layers with varying responsiveness and accuracy:
- . The first layer (Vaporizer) supervises the functions in the vaporizer and detects major errors in the Anesthesia Agent (AA) delivery, such as a continuous spray of AA from the injector. The fastest detection of errors occurs in the vaporizer.
- The second layer (Monitor) supervises the communication between the control node and . the vaporizer. This layer is also very fast.
- . The third layer (Gas analyzer) samples patient gas at the v-piece and supervises the delivered AA-, and oxygen concentration to the patient using the Patient Gas Analyzer (PGA). This is the most accurate and direct part of the safety system, but also the slowest.
In system version 3.0, the responsibility of the CGA to supervise the gas concentrations was transferred to the Vaporizers and the Patient Gas Analyzer (PGA). This change will not be noticed by the user.
This change will not affect the performance of the FLOW-i Anesthesia System.
{4}------------------------------------------------
# Measured Tidal Volume lower range in Pressure Control extended
In the modified FLOW-i Anesthesia System version 3.0 (K133958) the lower range of the measured Tidal Volume in ventilation mode Pressure Control is extended from 20 ml to 5 ml which will allow the user to view a measured value on the control panel for Tidal Volumes between 5-2000 ml. In the predicated device FLOW-i Anesthesia System version 2.1 (K112114) the user can view a measured value on the control panel for Tidal Volumes between 20-2000 ml. This change is for measuring of the Tidal Volume.
# Inspiratory and Expiratory Hold (Software functionality)
In the modified FLOW-i Anesthesia System version 3.0 (K133958) a new feature was added which allows the user to hold the patient's breathing cvcle at inspiration and gives the user a lung mechanics measurement tool for static compliance as well as total PEEP.
# Apnea Mute (Software functionality)
Induction and some other procedures might warrant the possibility to turn off the audio signal of alarms associated with breathing and breathing parameters. In the modified FLOW-i Anesthesia System version 3.0 (K133958) a new Apnea Mute function which enables the possibility to mute the Apnea related alarms to reduce the stress from annoying audible alarms was added.
# HLM Mode (Software functionality)
During open-heart surgery the Heart-Lung Machine (HLM) takes over the patient blood perfusion. During HLM (CardioPulmonary Bypass - CPB) patient related audio alarms are often activated by the anesthesia machine which unintentionally disturbs the surgeon and the operation staff. To avoid the audio disturbance the HLM mode which enable the possibility to set applicable alarms to audio off was added in the modified FLOW-i Anesthesia System version 3.0 (K133958).
# O2Guard Safety function (Software functionality)
In the modified FLOW-i Anesthesia System version 3.0 (K133958) a new hypoxic function was added to support the user to avoid hypoxic gas mixtures from being supplied to the patient. When Oxygen/AIR gas mixture is used and the system detects that the fraction of inspired oxygen (Fi(2) is below a certain threshold the system will automatically adjust the fresh gas flow and the oxygen concentration to predefined fixed levels. A dialog window appears informing of the alterations.
{5}------------------------------------------------
Volume Reflector Indicator (Software functionality)
In the modified FLOW-i Anesthesia System version 3.0 (K133958) the Volume Reflector Indicator (VRI) option was added. The VRI is a graphic representation which visualizes the volume reflector activity and re-breathing fraction.
- The VRI is a graphic representation of the ratio of reflector gas to exhaled gas in the । volume reflector.
n । ७७५७०
- The VR balance is a numerical value that describes the net flow of gas in the volume reflector.
- -The RBF (Rebreathing Fraction) is a metric that describes how much of the exhaled gas is being re-used during each successive breath.
Other minor modifications to the FLOW-i Anesthesia System 3.0 (K133958) consist of:
- Software upgrades: .
- Graphical User Interface (GUI) improvements. .
- Addition of short trend tab . The short trend tab depicts waveform data going back 10, 30 or 60 minutes.
- Inlet pressure on the main display . Addition of possibility to display the inlet pressure on the main display.
- Remote Services functionality . Addition of ability to transfer functionality status information and technical logs to a service remote server.
- . Software changes to handle component replacement due to Last Time Buy and implementation of the RoHS (Restriction of Hazardous Substances) EU directive as well as correction of software anomalies.
None of the modifications above implemented in the modified FLOW-i Anesthesia System version 3.0 changes the technology or performance of the FLOW-i Anesthesia System.
- . Labeling changes:
- Specification change for lower measured Tidal Volume limit from 20 ml to 5 ml. .
- Product label updated with symbol for CSA and IPx1. .
- . Addition of Sevofturane of brand Piramal. The Sevofiturane Vaporizer is also approved for use with the brand Piramal.
None of the labeling changes above implemented in the modified FLOW-i Anesthesia System version 3.0 changes the technology or performance of the FLOW-i Anesthesia System.
{6}------------------------------------------------
- Hardware/Accessories changes: .
- System Checkout valve inserted. .
- Addition of FLOW-i Remote services adaptor. .
- New accessories and modification to existing accessories: t
- . Addition of EVAC restrictor
The EVAC restrictor reduces the flow between high evacuation systems and the FLOW-i AGS outlet (Anesthesia Gas Scavenging outlet) to ensure proper gas evacuation.
- . Addition of Vaporizer Holder A holder for a third Vaporizer for easy access to an additional vaporizer during surgical procedures.
- . Addition of Vaporizer slot cover The vaporizer slot cover is intended to protect the gas and electrical comnections inside the vaporizer slot when no vaporizer is connected and to make the cleaning easier.
- I Addition of Universal bracket C20 for right side The Universal bracket is an accessory which supports additional space for mounting of accessory equipment, e.g. a desktop PC or patient monitor power transformer.
The FLOW-i Anesthesia System C20 already includes a Universal bracket for mounting on the rear side. This version will be for mounting on the right side of the FLOW-i Anesthesia System
- L Addition of Manual breathing bag rigid support arm The support arm is used to provide a static and secure support for the manual breathing bag and associated tubings. The support arm is fastened to the vertical railing on the core unit of the FLOW-i
- . Redesign of the Gas Backup system for oxygen/air and oxygen/nitrous oxide The backup gas system acts as a reserve system if the central gas supply would be unavailable. The FLOW-i C20 and C30 can be equipped with a backup system configured to deliver either oxygen/air or oxygen/nitrous oxide. The redesign of the current backup system includes removal of the wheels and a smother holder. The backup system, gas hoses and cable is as the current system permanently attached to the Core unit of the FLOW-i Anesthesia System.
None of the hardware/accessories changes above implemented in the modified FLOW-i Aresthesia System version 3.0 change the technology or performance of the FLOW-i Anesthesia System.
{7}------------------------------------------------
The modified FLOW-i Anesthesia System version 3.0 (K133958) is an enhancement of the
predicate device FLOW-i Anesthesia System version 2.1 (K112114) with the above listed changes/modifications included.
# Conclusion for Substantial Equivalence
MAQUET believes these modifications do not affect the intended use of the device, the indications for use nor alter the fundamental scientific technology of the device.
MAQUET has conducted the risk analysis and performed the necessary verification and validation activities to demonstrate that the design outputs of the modified device meet the design input requirements. The proposed changes do not affect the safety and effectiveness of the FLOW-i Anesthesia System. MAQUET has concluded that the modified FLOW-i Anesthesia System version 3.0 (K133958) is substantially equivalent to the predicate device, FLOW-i Anesthesia System version 2.1 (K112114).
# SUBSTANTIAL EQUIVALENCE TABLE
| | Predicate Device | SUBJECT DEVICE |
|-----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| Device | FLOW-i Anesthesia System version 2.1 | FLOW-i Anesthesia System version 3.0 |
| Manufacturer | Maquet Critical Care AB | Maquet Critical Care AB |
| Device<br>Classification Name | Gas-Machine, Anesthesia | Gas-Machine, Anesthesia |
| 510(k) Number | K112114 | K133958 |
| Indications for Use<br>according to the<br>510(k) Summary | The indication for FLOW-i Anesthesia System is<br>administering inhalation Anesthesia while controlling<br>the entire ventilation of patients with no ability to<br>breathe, as well as in supporting patients with a limited<br>ability to breathe. The system is intended for use on<br>neonatal to adult patient populations. The system is<br>intended for use in hospital environments, except MRI<br>environment, by healthcare professionals trained in<br>inhalation Anesthesia administration. | Same |
| Patient range | Neonatal to adult patient populations | Same |
| Settings (adjustable<br>parameters): | | |
| Ventilation Modes | • Manual Ventilation<br>• Volume Control<br>• Pressure Control<br>• Pressure Support + backup<br>• Pressure Regulated Volume Control (PRVC)<br>• SIMV | Same |
| SIMV rate (b/min) | 1-60 | Same |
| Backup rate (b/min) | 2-60 | Same |
| Fresh gas oxygen conc | 21 - 100 % when Oxygen/Air is selected<br>28- 100 % when Oxygen/Nitrous Oxide is selected.<br>Accuracy:<br>$21-59%: ±3 % v/v$<br>$60-100%: ±5 % v/v$ | Same |
| Isoflurane conc | 0-5%.<br>Accuracy: ±15% of set value or ±5% of maximum<br>possible setting (whichever is greater) | Same |
| | Predicate Device | SUBJECT DEVICE |
| Device | FLOW-i Anesthesia System version 2.1 | FLOW-i Anesthesia System version 3.0 |
| Manufacturer | Maquet Critical Care AB | Maquet Critical Care AB |
| Sevoflurane conc | 0-8%.<br>Accuracy: ±15% of set value or ±5% of maximum possible setting (whichever is greater) | Same |
| Desflurane conc | 0-10%: for set Fresh Gas Flowless than 20 l/min and maximal Fresh Gas Flowless than 75 l/min<br><br>0-18%: for set Fresh Gas Flowless than 10 l/min and maximal Fresh Gas Flowless than 40 l/min<br><br>Accuracy: ±15% of set value or ±5% of maximum possible setting (whichever is greater) | Same |
| Fresh gas flow | 0.3-20 l/min<br>Accuracy:<br>± 10 % or ± 50 ml/min (whichever is greater) | Same |
| Tidal volume | 100 - 2000 ml in the adult patient category.<br><br>20 - 350 ml in the infant patient category.<br><br>Accuracy: ±10% or 10 ml, whichever is greater | Same |
| Minute volume | 0.5 - 60 l/min in the adult patient category.<br><br>0.3 - 20 l/min in the infant patient category.<br><br>Accuracy: ±10% or 0.3 l/min (whichever is greater) | Same |
| PEEP (positive end<br>expiratory pressure) | 0-50 cmH2O<br>Accuracy: ±2 cmH2O or ±10% (whichever is greater) | Same |
| Pressure level above PEEP | 0 to 120 cmH2O- PEEP in the adult patient category.<br><br>0 to 80 cmH2O-PEEP in the infant patient category.<br><br>Accuracy: ±15% or ±2 cmH2O (whichever is greater) | Same |
| Breathing frequency | 4 - 100 breaths per minute.<br>Accuracy:<br>±5 % or ± 1 bpm (whichever is greater) | Same |
| Inspiration and Expiration<br>ratio; I:E | 1:10-4:1 in automatic modes | Same |
| Monitoring<br>Measurement range<br>and accuracy: | | |
| Pressure | -30 cmH2O - 140 cmH2O<br>Accuracy:<br>±5% or ±2 cm H2O (whichever is greater) | Same |
| Oxygen conc | 0 to 100%<br>Accuracy:<br>±1 vol% @ (0-25%)<br>±2 vol% @(25-80%)<br>±3 vol% @(80-100%) | Same |
| Tidal volume | Exp. Tidal volume:<br>Measurement range: 20-2000 ml<br>Accuracy :<br>±10% or 10 ml, whichever is greater | Exp. Tidal volume:<br>Measurement range: 5-2000 ml<br>Accuracy:<br>±4ml (5-20 ml)<br>±10% or 10 ml, whichever is greater (20-2000ml) |
| | Predicate Device | SUBJECT DEVICE |
| Device | FLOW-i Anesthesia System version 2.1 | FLOW-i Anesthesia System version 3.0 |
| Manufacturer | Maquet Critical Care AB | Maquet Critical Care AB |
| Carbon dioxide conc | 0 to 10%<br>Accuracy:<br>$\pm$ 0.1 vol%@ (0-1%)<br>$\pm$ 0.2 vol%@ (1-5%)<br>$\pm$ 0.3 vol%@ (5-7%)<br>$\pm$ 0.5 vol%@ (7-10%) | Same |
| Isoflurane conc | 0 to 5%<br>Accuracy:<br>$\pm$ 0.15 vol%@ (0-1%)<br>$\pm$ 0.2 vol%@ (1-5%) | Same |
| Sevoflurane conc | 0 to 8%<br>Accuracy:<br>$\pm$ 0.15 vol%@ (0-1%)<br>$\pm$ 0.2 vol%@ (1-5%)<br>$\pm$ 0.4 vol%@ (5-8%) | Same |
| Desflurane conc | 0 to 18%<br>Accuracy:<br>$\pm$ 0.15 vol%@ (0-1%)<br>$\pm$ 0.2 vol%@ (1-5%)<br>$\pm$ 0.4 vol%@ (5-10%)<br>$\pm$ 0.6 vol%@ (10-15%)<br>$\pm$ 1.0 vol%@ (15-18%) | Same |
Table 1: Specifications from FLOW-i User's Manual, section 14 Technical Specifications.
{8}------------------------------------------------
·
{9}------------------------------------------------
Table 2, software and hardware changes
| | Predicate Device | SUBJECT DEVICE |
|------------------------------------------|--------------------------------------|--------------------------------------|
| Device | FLOW-i Anesthesia System version 2.1 | FLOW-i Anesthesia System version 3.0 |
| Manufacturer | Maquet Critical Care AB | Maquet Critical Care AB |
| Device<br>Classification Name | Gas-Machine, Anesthesia | Gas-Machine, Anesthesia |
| 510(k) Number | K112114 | K133958 |
| Control Gas Analyzer<br>(CGA) | Available | Removed |
| Inspiratory and Expiratory<br>Holdoption | Not available | New software functionality |
| Apnea Mute | Not available | New software functionality |
| HLM (CPB) Mode | Not available | New software functionality |
| Volume Reflector<br>Indicator (VRI) | Not available | New software functionality |
| 02 Guard Safety Function | Not available | New software functionality |
| EVAC restrictor | Not available | New accessory |
| Vaporizer holder | Not available | New accessory |
| Vaporizer cover | Not available | New accessory |
| Universal holder C20 for<br>right side | Not available | New accessory |
| Manual Breathing bag<br>rigidsupport arm | Available | New shape and fastening |
| Gas Backup system O2/Air<br>or O2/N2O | Available | Redesigned |
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle or bird-like figure, with three curved lines forming its body and wings. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 20, 2014
Maquet Critical Care AB C/O Lia Liebgold Regulatory Affairs Specialist Maquet Medical Systems USA 45 Barbour Pond Drive Wayne, NJ 07470
Re: K133958
Trade Name: Maquet FLOW-i Anesthesia System, Models C20, C30 and C40 Regulation Number: 21 CFR 868.5160 Regulation Name: Gas Machine for Anesthesia or Analgesia Regulatory Class: Class II Product Code: BSZ Dated: January 10, 2014 Received: January 13, 2014
Dear Ms. Liebgold:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{11}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industrv/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safetv/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Image /page/11/Picture/7 description: The image shows a signature and some text. The text includes "Sincerely yours," followed by the name "Tejashri Purohit-Sheth, M.D." and the title "Clinical Deputy Director." The acronym "DAGRID" is also present. The word "FOR" is at the bottom right.
Erin Keith. M.S. Acting Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{12}------------------------------------------------
# Indications for Use
510(k) Number (if known) K133958
Device Name FLOW-i Anesthesia System
## Indications for Use (Describe)
The indication for FLOW-i Anesthesia System is administering inhalation Anesthesia while controlling the entire ventilation of patients with no ability to breathe, as well as in supporting with a limited ability to breathe. The system is intended for use on neonatal to adult patient populations. The system is intended for use in hospital environments, except MRI environment, by healthcare professionals trained in inhalation Anesthesia administration.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
CALL LA LI . . . . . . . . . . . . FOR FOR FOR FOR FOR . . . . . .
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Todd D. Courtney -S
2014.03.18 10:30:32 -04:00
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.