← Product Code [BSZ](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BSZ) · K122445

# AESPIRE 7900, AESPIRE VIEW (K122445)

_Datex-Ohmeda · BSZ · Oct 12, 2012 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BSZ/K122445

## Device Facts

- **Applicant:** Datex-Ohmeda
- **Product Code:** [BSZ](/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BSZ.md)
- **Decision Date:** Oct 12, 2012
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 868.5160
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic

## Indications for Use

The family of GE Datex-Ohmeda Aespire anesthesia systems with 7900 ventilator (Aespire 7900 and Aespire View) is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients. The devices are intended for volume or pressure control ventilation.

## Device Story

Anesthesia system providing inhalation anesthesia and ventilatory support; utilizes 7900 ventilator for patient ventilation during surgical procedures. Inputs: medical gases (O2, air, N2O), vaporizer-delivered anesthetic agents, and sensor data from breathing circuit (inspired O2, airway pressure, exhaled volume). Operation: microprocessor-controlled, pneumatically driven; regulates breathing patterns and PEEP; compensates for compression losses, fresh gas contribution, and system leaks. Output: controlled gas delivery and ventilation parameters displayed on monitor (monochromatic for Aespire 7900, color for Aespire View). Used in clinical settings (OR) by trained medical professionals. Healthcare providers use output to monitor patient status and adjust anesthesia/ventilation settings. Benefits: provides controlled anesthesia and respiratory support during surgery; safety features mitigate hypoxic mixtures and power/gas supply failures.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including risk analysis, design reviews, performance verification, and safety testing (environmental and electromagnetic compatibility).

## Technological Characteristics

Microprocessor-based, pneumatically driven anesthesia system. Features Selectatec manifold for Tec 4, 5, 6, 6+, and 7 vaporizers. Includes electronic sensor interface for airway overpressure detection and mechanical overpressure relief valve. Ventilator modes: VCV, PCV, SIMV/PSV, PSVPro, SIMV-PC, and PCV-VG. Connectivity: RS-232 serial digital communications port. User interface: monochromatic (Aespire 7900) or color (Aespire View) display.

## Regulatory Identification

A gas machine for anesthesia is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. The device may include a gas flowmeter, vaporizer, ventilator, breathing circuit with bag, and emergency air supply. A gas machine for analgesia is a device used to administer to a patient an analgesic agent, such as a nitrous oxide-oxygen mixture (maximum concentration of 70 percent nitrous oxide).

## Predicate Devices

- GE Datex-Ohmeda Aespire 7900, Aespire View ([K092864](/device/K092864.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K122445

Image /page/0/Picture/1 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circle. There are decorative flourishes around the circle, adding to the logo's classic design.

# GE Healthcare

Datex-Ohmeda Inc. 3030 Ohmeda Drive P.O. Box 7550 Madison, WI 53707-7550 USA

OCT 12 2012

## Premarket Notification 510(k) Summary - As required by section 807.92

GE Datex-Ohmeda Aespire Anesthesia System

## GENERAL COMPANY INFORMATION as required by 807.92(a)(1) COMPANY NAME/ADDRESS/PHONE/FAX:

GE Healthcare Datex-Ohmeda, Inc. . 3030 Ohmeda Drive PO Box 7550 Madison, WI 53707 USA Tel: 608-221-1551 Fax: 608-223-2132

#### NAME OF CONTACT:

Mr. James P. Raskob Ms. Monica Morrison (alternate)

#### DATE:

October 12, 2012

DEVICE NAME as required by 807.92(a)(2) TRADE NAME:

GE Datex-Ohmeda Aespire Anesthesia System

### COMMON NAME:

Gas Machine, Anesthesia

## CLASSIFICATION NAME:

Anesthesiology, 73 BSZ, 21 CFR 868.5160 Gas Machine, Anesthesia

{1}------------------------------------------------

## NAME OF LEGALLY MARKETED DEVICE FOR WHICH A CLAIM OF SUBSTANTIAL EQUIVALENCE IS MADE as required by 807.92(a)(3)

The GE Datex-Ohmeda Aespire 7900 and Aespire View products are considered substantially equivalent to, and as safe and effective as, the legally marketed (predicate) GE Datex-Ohmeda Aespire 7900, Aespire View (K092864).

#### DEVICE DESCRIPTION as reguired by 807.92[a)(4)

The family of GE Datex-Ohmeda Aespire anesthesia systems with 7900 ventilator (Aespire 7900 and Aespire View) is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients. The systems are to be used only by trained and qualified medical professionals.

The Aespire 7900 and Aespire View supply set flows of medical gases to the breathing system. A large selection of frames, gases, and vaporizers are available to aive the user control of the system configuration. They are available in trolley and pendant models, with two or three gases, two vaporizer positions and up to three cylinder connections. All models connect to oxygen and can additionally connect with up to two optional gases (air, N2O). The Aespire systems accept Tec 4. Tec 5. Tec 6. Tec 6+ and Tec 7 vaporizers on a Selectatec manifold. Safety features are designed to decrease the risk of hypoxic mixtures and complete power or sudden aas supply failures. The Aespire View product provides optional electronic Total Fresh Gas Flow (TFS) monitoring. The Aespire View also features a color display, while the Aespire 7900 uses a monochromatic display.

The Datex-Ohmeda 7900 Anesthesia Ventilator is used in this family of Anesthesia Systems. It is a microprocessor based, electronically controlled, pneumatically driven ventilator that provides patient ventilation during suraical procedures. The 7900 ventilator is equipped with a built-in monitoring system for inspired oxygen, airway pressure and exhaled volume. Sensors in the breathing circuit are used to control and monitor patient ventilation as well as measure inspired oxygen concentration. This allows for the compensation of compression losses, fresh aas contribution and small leakage in the breathing absorber, bellows and system. User setting and microprocessor calculations control breathing patterns. The user interface keeps settings in memory. The user may change settings with a simple and secure setting sequence. A bellows contains breathing gasses to be delivered to the patient. Positive End Expiratory Pressure (PEEP) is requlated electronically. Positive pressure is maintained in the breathing system so that any leakage that occurs is outward. An RS-232 serial digital communications port connects to and communicates with external devices. Ventilator modes for the device include Volume Control (VCV) Mode, Pressure Control (PCV) Mode (Optiongl), Svnchronized Intermittent Mandatory Ventilation with Pressure Support Ventilation (SIMV/PSV) Mode, Pressure Support with Apnea Backup.(PSVPro) Mode (Optional), Synchronized Intermittent Mandatory Ventilation with Pressure Control (SIMV-PC) Mode (Optional), and Pressure Control Ventilation – Volume Guaranteed (PCV-VG) mode (Optional on Aespire View variant only).

{2}------------------------------------------------

### INTENDED USE as required by 807.92(a){5)

The family of GE Datex-Ohmeda Aespire anesthesia systems with 7900 ventilator (Aespire 7900 and Aespire View) is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients. The devices are intended for volume or pressure control ventilation.

## SUMMARY OF TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED TO THE PREDICATE DEVICE as reguired by 807.92(a)(6)

The GE Datex-Ohmeda Aespire 7900 and Aespire View have been updated from the predicate version (K092864). There have been no changes to the intended use or fundamental scientific technology.

The electronic sensor interface circuit board used within the Aespire 7900 and Aespire View has been updated. The primary change included replacement of a mechanical switch for airway overpressure detection with an electronic monitoring circuit for airway overpressure detection. The existing mechanical overpressure relief valve that serves as an additional safety mitigation remains unchanged.

## SUMMARY OF NONCLINICAL TESTING FOR THE DEVICE and CONCLUSIONS as required by 807.92(b)(1)(3)

The GE Datex-Ohmeda Aespire 7900 and Aespire View Anesthesia Systems have been thoroughly tested through verification of specifications. The following quality assurance measures were applied to the development of the revised circuit board:

- Risk Analysis .
- Design Reviews .
- Performance testing (Verification) .
- . Safety testing (Verification)

Verification tests completed during the development process included environmental verification testing and electromagnetic compatibility verification testing.

## SUMMARY OF CLINICAL TESTING FOR THE DEVICE and CONCLUSIONS as required by 807.92(b)(2)

The modification to the family of GE Datex-Ohmeda Aespire anesthesia systems with 7900 ventilator did not require clinical testing.

{3}------------------------------------------------

## CONCLUSION:

GE Healthcare considers the GE Datex-Ohmeda Aespire 7900 and Aespire View anesthesia machines to be as safe, as effective, and have performance substantially equivalent to the predicate device.

.

.

1999 - 1999 - 1999

:

ー

. .

{4}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" arranged around the perimeter. Inside the circle is a stylized symbol resembling an abstract caduceus or a person embracing another, rendered in black.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

OCT
12 2012

Mr. James P. Raskob Regulatory Affairs Leader GE Healthcare Datex-Ohmeda, Incorporated 3030 Ohmeda Drive P.O. Box 7550 Madison, Wisconsin 53707-7550

Re: K122445

Trade/Device Name: GE Datex-Ohmeda Aespire Anesthesia System Regulation Number: 21 CFR 868.5160 Regulation Name: Gas Machine for Anesthesia or Analgesia Regulatory Class: II Product Code: BSZ Dated: September 12, 2012 Received: September 13, 2012

Dear Mr. Raskob:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

### Page 2 - Mr. Raskob

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

For S.D. Shuckett & Co.

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Infection Control and Dental Devices

Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

# Indications for Use

510(k) Number (if known): K

Device Name: GE Datex-Ohmeda Aespire Anesthesia System

Indications For Use:

The family of GE Datex-Ohmeda Aespire anesthesia systems with 7900 ventilator (Aespire 7900 and Aespire View) is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients. The devices are intended for volume or pressure control ventilation.

Prescription Use XXX (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of __ 1

(Division Sign-Un) Division of/Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K122445

---

**Source:** [https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BSZ/K122445](https://fda.innolitics.com/submissions/AN/subpart-f%E2%80%94therapeutic-devices/BSZ/K122445)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
