The GE Datex-Ohmeda Aisys Anesthesia System is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (neonatal, pediatric, adult). The device is intended for volume or pressure control ventilation. The Aisys is not suitable for use in a MRI environment.
Device Story
Aisys Anesthesia System provides inhalation anesthesia and ventilatory support; used by trained medical professionals in clinical settings. Inputs include medical gases (O2, air, N2O) and user settings via keypad/rotary controller. System uses electronic gas mixing and microprocessor-controlled 7900 Anesthesia Ventilator to deliver gas/anesthetic agent mixtures to patient breathing circuit. Anesthesia computer controls agent delivery via Aladin/Aladin 2 cassettes. Ventilator monitors inspired O2, airway pressure, and exhaled volume; compensates for compression losses and leaks. Output displayed on system monitor; includes ventilation parameters, gas concentrations, and alarm notifications. System supports various ventilation modes (VCV, PCV, SIMV/PSV, etc.). Integrated respiratory gas monitoring available via M-Gas or E-Gas modules. Device benefits patients by providing precise, automated anesthesia delivery and respiratory support, reducing risk of hypoxic mixtures and system failures.
Clinical Evidence
No clinical data. Functionality of modified software features evaluated via nonclinical design verification and validation testing.
Technological Characteristics
Microprocessor-controlled, pneumatically driven anesthesia system. Electronic gas mixing; electronic vaporizer technology. Integrated respiratory gas monitoring modules. RS-232 serial communication. Ventilator modes: VCV, PCV, SIMV/PSV, PSVPro, SIMV-PC, PCV-VG, CPAP/PSV. Modular frame configuration with integrated display and cable management.
Indications for Use
Indicated for general inhalation anesthesia and ventilatory support (volume or pressure control) in neonatal, pediatric, and adult patients. Not for use in MRI environments.
Regulatory Classification
Identification
A gas machine for anesthesia is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. The device may include a gas flowmeter, vaporizer, ventilator, breathing circuit with bag, and emergency air supply. A gas machine for analgesia is a device used to administer to a patient an analgesic agent, such as a nitrous oxide-oxygen mixture (maximum concentration of 70 percent nitrous oxide).
{0}------------------------------------------------
Page 1 of 4
## JUL 20 2011
## Premarket Notification 510(k) Summary As required by section 807.92 GE Datex-Ohmeda Aisys
## GENERAL COMPANY INFORMATION as required by 807.92(a)(1)
## COMPANY NAME/ADDRESS/PHONE/FAX:
Datex-Ohmeda, Inc. PO Box 7550 Madison, WI 53707 USA Tel: 608-221-1551 Fax: 608-299-2132
## NAME OF CONTACT:
Mr. James P. Raskob Ms. Monica Morrison (alternate)
## DATE:
June 6, 2011
#### DEVICE NAME as required by 807.92(a)(2)
## TRADE NAME:
GE Datex-Ohmeda Aisys Anesthesia System
## COMMON NAME:
Gas Machine, Anesthesia
#### CLASSIFICATION NAME:
Anesthesiology, 73 BSZ, 21 CFR 868.5160 Gas Machine, Anesthesia
## NAME OF LEGALLY MARKETED DEVICE FOR WHICH A CLAIM OF SUBSTANTIAL EQUIVALENCE IS MADE as required by 807.92(a)(3)
The GE Datex-Ohmeda Aisys is substantially equivalent in safety and effectiveness to the legally marketed (predicate) GE Datex-Ohmeda Aisys (K090233) and GE Datex-Ohmeda Engstrom ventilator (K093886).
{1}------------------------------------------------
## DEVICE DESCRIPTION as required by 807.92(a)(4)
The GE Datex-Ohmeda Aisys is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (neonatal, pediatric, adult). It represents one of the systems in a long line of products based on the Datex-Ohmeda Excel. Aestiva, Aespire, and Avance Anesthesia Systems. It is to be used only by trained and qualified medical professionals.
The GE Datex-Ohmeda Aisys Carestation supplies set flows of medical gases to the breathing system using electronic gas mixing. Gas flows are selected by the user using the keypad and rotary controller on the main display unit and then displayed as electronic flow indicators on the svstem display unit. The Aisvs is equipped with a pneumatic back-up O2 delivery system and traditional flow tube, as well. A large selection of frames, gases, and vaporizers are available to give the user control of the system configuration. The Aisys is also available in a pendant model. It is available with two or three gases, and up to three cylinder connections. All models have 02. The Aisys comes with up to two optional gases (air, N2O). Safety features and devices within the Aisys are designed to decrease the risk of hypoxic mixtures and complete power or sudden gas supply failures. The Aisys system is available with optional integrated respiratory gas monitoring. When supplied as an option, the integrated respiratory gas monitoring is provided via the Datex-Ohmeda M-Gas Module (M-CAiO and M-CAiOV software revision 3.2 and above cleared via K001814) and E-Gas Module (E-CAiOVX software revision 4.5 and above cleared via K051092) which can be physically integrated into the Aisys, receive electronic power from the Aisys and communicate measured values to the Aisys for display on the system display unit.
The anesthetic agent delivery for the Aisys is controlled via an anesthesia computer through user input from the central display. The vaporization technology is based upon the electronic vaporizer cleared as part of the Datex-Ohmeda Anesthesia Delivery Unit (ADU) cleared via K973985. An Aladin cassette (also cleared as part of K973895) or Aladin 2 is inserted into the active cassette bay. The cassette holds the agent to be delivered - Halothane, Enflurane, Isoflurane, Desflurane or Sevoflurane. Agent is delivered as a percent volume/volume. The Aisvs is designed to allow only one active cassette at a time. Per the user input into the main display, valves within the active cassette bay will open and allow agent to be delivered. The agent is mixed with gas from the FGC unit. After mixing, the combination of gases and agent is delivered to the breathing system and then on to the patient.
The Datex-Ohmeda 7900 Anesthesia Ventilator is used in the Aisys Anesthesia System. It is a microprocessor based, electronically controlled, pneumatically driven ventilator that provides patient ventilation during surgical procedures. The 7900 ventilator is equipped with a built-in monitoring system for inspired oxygen, airway pressure and exhaled volume. Sensors in the breathing circuit are used to control and monitor patient ventilation as well as measure inspired oxygen concentration. This allows for the compensation of compression losses, fresh gas contribution and small leakage in the breathing absorber, bellows and system. User setting and microprocessor calculations control breathing patterns. The user interface keeps settings in memory. The user may change settings with a simple and secure setting sequence. A bellows contains breathing gasses to be delivered to the patient. Positive End Expiratory Pressure (PEEP) is requlated electronically. Positive pressure is maintained in the breathing system so that any leakage that occurs is outward. An RS-232 serial digital communications port connects to and communicates with external devices. Ventilator modes for the device include Volume Control Ventilation (VCV), Pressure Control Ventilation (PCV) (optional), Synchronized Intermittent Mandatory Ventilation/Pressure Support (SIMV/PSV) (optional), Pressure Support Ventilation (PSVPro) (optional), Synchronized Intermittent Mandatory Ventilation-Pressure Control (SIMV-PC) (optional), Pressure Control Ventilation-Volume Guaranteed (PCV-VG) (optional) and Constant Positive Airway Pressure/Pressure Support Ventilation (CPAP/PSV). Ventilator parameters and measurements are displayed on the system display unit.
{2}------------------------------------------------
The system display unit is mounted to an arm on the top shelf of the Aisys. The arm is counter balanced and capable of moving vertically and/or horizontally, and also tilting the display, enabling the user to position the display to the most advantageous viewing position. The arm length is limited such that the display position is always within the footprint of the Aisys frame. The arm also supports the mounting of additional display units for a variety of patient monitors.
Several frame configurations are available, including one that allows for the physical integration of the GE Monitors (most recently cleared Carescape B850 via K092027 and Carescape B650 via K102239). This configuration also provides cable management solutions such that the necessary connections from the monitor display unit to the monitor are hidden within the Aisys frame. An additional option allows the monitor to be linked to the power supply of the Aisys such that when the Aisys is turned on, the monitor is also turned on. Additional configurations allow for the mounting of various patient monitors on the top shelf of the Aisys.
## INTENDED USE as required by 807.92(a)(5)
The GE Datex-Ohmeda Aisys Anesthesia System is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (neonatal, pediatric, adult). The device is intended for volume or pressure control ventilation. The Aisys is not suitable for use in a MRI environment.
## SUMMARY OF TECHNOLOGICAL CHARACTERISTICS OF DEVICE COMPARED TO THE PREDICATE DEVICE as required by 807.92(a)(6)
The GE Datex-Ohmeda Aisys has been updated from the predicate version (K090233). There have been no changes to the intended use or fundamental scientific technology.
The software for the Aisys has been updated to introduce several new features. A summary of the changes to the software and display monitor include:
- . Pediatric Improvements: 02 concentration range improvement at low flows, reduction of spurious alarms when patient is adequately ventilated, improved control of the agent quick key, gas monitor ventilator Tidal Volume compensation, Breathing Tube Compliance On/Off option
- Highlight the border of the menu button for increased visibility .
- Checkout improvement: Remove "Leak < 250" question .
- Temporary gas flow suspension capability .
- Improved "unconfirmed" setting notification .
- Remind clinician to set the agent .
- . Ventilator defaults for each individual vent mode
- New ventilation mode: CPAP/PSV with backup breaths .
- Lung mechanics procedure options of Vital Capacity and Cycling .
- Procedures button added to the existing display monitor. Quick Keys .
{3}------------------------------------------------
## SUMMARY OF NONCLINICAL TESTING FOR THE DEVICE and CONCLUSIONS as required by 807.92(b)(1)(3)
The GE Datex-Ohmeda Aisys has been thoroughly tested through verification of specifications and validation, including software validation. Verification of compliance with applicable voluntary standards has also been made to support safe use of the device in its intended environment. The following quality assurance measures were applied to the development of the system:
- Risk Analysis .
- Requirements Reviews .
- Desian Reviews .
- . Testing on unit level (Module verification)
- Integration testing (System verification) .
- Performance testing (Verification) ●
- Safety testing (Verification) .
- Simulated use testing (Validation) .
## SUMMARY OF CLINICAL TESTING FOR THE DEVICE and CONCLUSIONS as required by 807.92(b)(2)
The modifications made to the GE Datex-Ohmeda Aisys did not require clinical testing. The functionality of the modified software features were completely evaluated by performing nonclinical tests of design verification and validation testing.
#### CONCLUSION:
The summary above shows that there are no new questions of safety and effectiveness for the GE Datex-Ohmeda Aisys as compared to the predicate devices. The performance data demonstrates the device is substantially equivalent to the predicates.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Datex-Ohmeda, Incorporated Mr. Jim Raskob Care Business Area PO Box 7550 3030 Ohmeda Drive Madison, Wisconsin 53707-7550
JUL 20 2011
Re: K110213
Trade/Device Name: GE Datex -- Ohmeda Aisys Anesthesia System Regulation Number: 21 CFR 868.5160 Regulation Name: Gas Machine for Anesthesia or Analgesia Regulatory Class: II Product Code: BSZ Dated: July 13, 2011 Received: July 14, 2011
Dear Mr. Raskob:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provice of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements, rits s r, uns ooo to over in the Federal Register.
{5}------------------------------------------------
Page 2- Mr. Raskob
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal gencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803), good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm 115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address. http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Antnong, D. Watson
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): K
Device Name: GE Datex-Ohmeda Aisys Anesthesia System
Indications For Use:
The GE Datex-Ohmeda Aisys Anesthesia System is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (neonatal, pediatric, adult). The device is intended for volume or pressure control ventilation. The Aisys is not suitable for use in a MRI environment.
Prescription Use XXX (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
Page 1 of __ 1________________________________________________________________________________________________________________________________________________________________
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
E. Shultten
Division Sign-Off) **vision of Anesthesiology, General Hospital ·ection Control, Dental Devices
K110213 . 10(k) Number:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.