CODONICS SAFE LABELING SYSTEM (SLS), MODEL 500I/500P
K101439 · Codonics, Inc. · BSZ · Feb 7, 2011 · Anesthesiology
Device Facts
Record ID
K101439
Device Name
CODONICS SAFE LABELING SYSTEM (SLS), MODEL 500I/500P
Applicant
Codonics, Inc.
Product Code
BSZ · Anesthesiology
Decision Date
Feb 7, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5160
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K101439 · Feb 7, 2011
CODONICS SAFE LABELING SYSTEM (SLS), MODEL 500I/500P
Codonics, Inc.
Clinical application of prototype device at Massachusetts General Hospital (MGH)
The sponsor utilized clinical application data from a prototype system at MGH to verify the device's operational performance and clinical utility in a real-world surgical environment.
Clinical application of the prototype Safe Label system at MGH; Clinical application/observational
Patients undergoing surgical procedures at MGH; Number of Sites: 1 (MGH)
Not applicable for this study
Operational performance and clinical operations
Indications for Use
The Codonics Safe Label System (SLS) and SLS Software provides a simple computerbased bar code scanning & printing system to automatically verify drug identity from NDC and other drug vial UDI Barcodes, and to print labels for prepared drugs and other items in use on patients during surgical procedures. Codonics Safe Label System (SLS) is generally placed in, however not limited to, the perioperative environment to identify syringes prepared for anesthesiology use during surgery. Additional uses include producing labels for IVs and other artifacts used during a surgical procedure. SLS can also be used to print "non-surgical environment" color & text labels as required. Typical users of this system are trained professionals, including but not limited to physicians, nurses, and technicians
Device Story
System uses barcode scanner to read NDC/UDI from drug vials; compares input against hospital-managed formulary database (containing drug name, concentration, expiration, site-specific rules). System provides audible/visual confirmation of drug identity; prints waterproof, color-coded, ISO/ASTM-compliant labels with 2D barcodes. Used in perioperative/clinical settings by trained professionals (physicians, nurses, technicians). Integration with Anesthesia Information Management Systems (AIMS) allows real-time documentation of drug administration via barcode scan. Benefits include improved medication safety, reduced labeling errors, and standardized documentation.
Clinical Evidence
Bench testing and clinical operations verification performed. Non-clinical lab tests confirmed error-free NDC/UDI vial reading and labeling compliance with ASA/ISO standards. Clinical application of prototype system at MGH documented in original submission. No adverse clinical outcomes reported; device designed to meet EN60601-1/2 safety standards.
Technological Characteristics
PC-based (x86 Pico-ITX) system with SSD/Flash storage, touch screen interface, barcode scanner, and ink-jet label printer. Medical grade power supply (EN60601-1). Connectivity via Ethernet/WIFI. Software runs on Linux OS. Complies with ASTM D4774, ISO 26825, ASTM D4267, ASTM D6398, and FCC Part 15 Class B.
Indications for Use
Indicated for trained professionals (physicians, nurses, technicians) to identify prepared syringes, IVs, and other items in perioperative and non-surgical environments by verifying drug identity via barcode scanning and printing compliant labels.
Regulatory Classification
Identification
A gas machine for anesthesia is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. The device may include a gas flowmeter, vaporizer, ventilator, breathing circuit with bag, and emergency air supply. A gas machine for analgesia is a device used to administer to a patient an analgesic agent, such as a nitrous oxide-oxygen mixture (maximum concentration of 70 percent nitrous oxide).
Philips CompuRecord Perioperative System (K030939)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains a logo with a geometric shape on the left and the word "CODONICS" in bold, sans-serif font on the right. Below the word "CODONICS" is the tagline "We bring the future into focus" in a smaller, italicized font. The geometric shape resembles a stylized hexagon with a line extending from the top left corner.
# FEB - 7 20il
Submitter Information:
- 1.1 Submitter:
510(k) Summary
1
Codonics, Inc. 17991 Englewood Drive Middleburg Heights, Ohio 44130
# 1.2 Manufacturing Facility:
Same as above
### 1.3 Representative:
Not applicable at this time
#### 1.4 Contact:
Gary W. Enos, Phone: (440) 243-1198 / Fax : (440) 243-1334 17991 Englewood Drive Middleburg Heights, Ohio 44130
- 1.5 Date: February 1, 2011
#### 2 Device Name
807.92(a){2) & 807.92(a)(3)
- 2.1 Anesthesia Devices: Therapeutic Devices: Accessories
- Gas machine for anesthesia or analgesia, Accessory 2.2 Classification Name:
807.92(c)
807.92(a)(1)
- Classification Number: 868.5160 2.3 Classification Code: BSZ
- Codonics Safe Labeling System 2.4 Trade/Proprietary Name: Codonics Safe Labeling Software & Standalone Center Included Products: Codonics Safe Labeling System Family (Integrated) Codonics Safe Labeling (SLS XXX) Multiple Models Pending Configuration
- 2.5 Predicate/Comparitive Devices: 807.92(a)(3)
Codonics has identified Medley System with Bar Code Module (Premarket notification K041241) and the SAFERsleep System (Pre-market notification K050883) and DocuSys Digital Drug Management (Premarket notification K062388) and Philips CompuRecord Perioperative System (Pre-market notification K030939)
{1}------------------------------------------------
#### 3 Device Description 807.92(a)(4)
### 3.1 Function
Drug preparation and administration in the perioperative environment are integral aspects of anesthesiologist's patient care responsibilities.
Codonics Safe Labeling System is a simple, integrated system utilizing a bar code scanner to read and confirm drug identity from NDC and other drug ID Barcodes from vials automatically print labels for prepared drugs and other items in use on patients during surgical procedures. The labels are compliant with national regulations focused on improving medication safety in the perioperative environment.
The software components provide functions for scanning vials, creating, indexing, and approved hospital managed promotion of a formulary database, displaying on screen and audibly confirming drug type, printing color The Joint Commission ISO and ASTM compliant labels with 2-D barcodes. The system reads drug vial barcodes and produces waterproof, color labels.
The system can be integrated to function with AIMS system workflow to provide real-time documentation of drug administration when the syringe "2D Barcode" is read.
- 3.2 Scientific Concepts: NDC or UDI barcodes are read from drug vials and compared to an formulary of drugs approved for use by the hospital pharmacy. The formulary contains drug name, concentration, expiration, site specific warnings, dilutions and class of drug templates (for color and patter label production). Once identified and confirmed, a vial label is displayed along with audible indication of the drug scanned. Specific site directed rules for approved conditions of use are promoted to each SLS. When vials scanned are not found in the database, a "Drug not found" indication is made and the user is warned and prompted to produce a manual label.
# 3.3 Physical And Performance Characteristics:
Codonics Safe Labeling System consists of :
- PC (x86 Pico-ITX based w/on-board RAM) with Ethernet interface .
- . SSD/Flash disk
- Touch screen end-user interface (for configuring and controlling authorized users, . barcode scan and print jobs, selection of label type and manual/automatic label production % dilution factor indication)
- Barcode scanner/decoder .
- Label printing engine .
- Medical grade (EN60601-1) compliant power supply .
Software to support primary functions:
- Linux OS for CPU, I/O (USB, IDE or SATA, AUDIO, Network (Ethernet/WIFI), SSD . Disk access, and Touch Screen LCD display
- Specific Drivers for Touch Screen LCD, Ink Jet label printer, Barcode scanner, . Ethernet/WIFI communications
- NDC or other UDI (Unique Drug Identifier, commonly known in the US as an "NDC") . drug formulary persistent repository (or database) via SSD/USB Flash
{2}------------------------------------------------
In addition to these primary functions, there are other functions that are provided by the software, to include:
- System configuration (network, security, profiles, etc.) .
- Security management to perform accounting and authentication of user data; add . users to the User Database when requested and verify that any authenticating user credentials are correct
- Settings management global component available to all other components in the . application; lookup configuration values given the appropriate configuration key
- User feedback (job and device status, errors, etc.) various events that describe the . current state of the system are generated
- Job function application managers. Each manager is responsible for a specific set of . functionality grouped by a logical theme. For example, the Label Manager handles creating labels based on a set of Label Parameters, the Security Manager handles all issues relating to user management and authentication, and the Print Manager sends commands to and receives responses from the printer hardware
Software Application Description: The software scans user Identification (authentication) and drugs by means of a bar code, and identifies connected devices (printers, computer, and barcode scanner, etc.). The software allows a label to be printed which has a color or colors identifying the type of drug contained in the syringe, the name of the drug or drugs, the units of the drug or drugs, the amount of the drug or drugs, total volume of the syringe, the user ID of the preparer, the time and date of the preparation, and a bar code identifying the contents of the syringe. The software also enables the clinician to document the administration of the drug or drugs by reading the bar code printed on the label and transmitting the drug identification to a third party application (i.e. AIMS system). The system also allows for the labeling of drugs which are combined with other drugs or are diluted. Finally, the software also allows the printing of blank labels, with just the clinician's ID and the data/time.
The major characteristics and functions of the family of devices include:
- Scanning the FDA required drug vial barcode directly from the vial .
- Decoding the manufacturer issued barcode into the required FDA national drug code o (NDC) or Unique Drug Identifier (UDI) number
- Referring the NDC/UDI number to a site managed formulary lookup database .
- Providing audio and ISO-compliant visual "readback" of the drug name .
- Providing a clinical alert if the drug vial is listed as recalled in the site formulary .
- Printing an easy to read, waterproof ISO 26825 compliant color label meeting The . Joint Commission medication management standards and the American Society of Anesthesiologists guidelines for labeling
- Including on same label a printed barcode compliant with national standards for . machine readability allowing integration with an anesthesia information management system (AIMS)
- Providing the basic information by which the printed label barcode can be read to . document medication administration in an AIMS system
- Providing a 2D barcode of information to permit integration with AIMS systems with . potential for alerts if the prepared drug has expired based on preparation time and site specific drug use criteria
- Printing labels with insertion and expiration date and time for IV lines .
{3}------------------------------------------------
#### 807.92(a){5} Device Intended Use: 4
- 4.1 The Codonics Safe Label System (SLS) and SLS Software provides a simple computerbased bar code scanning & printing system to automatically verify drug identity from NDC and other drug vial UDI Barcodes, and to print labels for prepared drugs and other items in use on patients during surgical procedures. An integrated formulary database reliably confirms vial barcodes with both audible and visual display of the drug name and concentration. The SLS system produces waterproof, ASA compliant color labels per ISO and ASTM standards. The software components provide the functions for scanning vials, indexing against a hospital managed formulary database, displaying on screen and audibly confirming drug type, and printing color The Joint Commission compliant labels with 2-D barcodes.
- 4.2 The use of drug class specific pattern and color per ASTM D4774 and ISO 28625 Specifications for User Applied Drug Labels in Anesthesiology is configurable by site and "Data-set". Data-sets are uniquely named configurations that may differ in drugs, colors, dilutions and comments to accommodate different practices within a single site or hospital (pediatric versus cardiac for example).
- 4.3 Additional uses include producing labels for IVs and other artifacts used during a surgical procedure.
- 4.4 Codonics Safe Label System (SLS) is generally placed in, however not limited to, the perioperative environment to identify syringes prepared for anesthesiology use during surgery..
- 4.5 Typical users of this system are trained professionals, including but not limited to physicians, nurses, and technicians
- 4.6 Barcode scanning with SLS formulary confirmation of drug, expiration, and suitability is equivalent to the predicate devices employing similar technology. Labeling or identification of the syringe with input to AIMS system in the process of patient surgical care is equivalent in technology to previously cleared devices employing similar technology.
# 4.7 510(k) Indications for Use Statement: Prescription Use Device
The Codonics Safe Label System (SLS) and SLS Software provides a simple computerbased bar code scanning & printing system to automatically verify drug identity from NDC and other drug vial UDI Barcodes, and to print labels for prepared drugs and other items in use on patients during surgical procedures.
Codonics Safe Label System (SLS) is generally placed in, however not limited to, the perioperative environment to identify syringes prepared for anesthesiology use during surgery. Additional uses include producing labels for IVs and other artifacts used during a surgical procedure. SLS can also be used to print "non-surgical environment" color & text labels as required. Typical users of this system are trained professionals, including but not limited to physicians, nurses, and technicians
#### Device Technological Characteristics: 807.92(a)(6) ട്
- 5.1 The Codonics Safe Label System is a new device in terms of a product demonstrating the same completely integrated functionality. The SLS device does employ substantially equivalent technology with respect to reading bar codes as the Medley System with Bar Code Module (Pre-market notification K041241) and the SAFERsleep System (Premarket notification K050883) with visual and auditory feed back confirmation. The SLS system also employs equivalent technology as DocuSys Digital Drug Management (Pre-
{4}------------------------------------------------
market notification K062388) and Philips CompuRecord Perioperative System (Premarket notification K030939) in terms of barcode read information reference with a drug formulary list/database and claims of similar system function and intended uses.
- 5.2 The technology and applications are substantially equivalent to these products already cleared to market by the FDA with respect to drug identification, alert management and in the ability to record/input the time and type of drug administration during surgery via labeling barcodes read by AIMS (Anesthesia Management Systems).
- 5.3 The software components provide functions for scanning vials, indexing against a hospital managed formulary database, displaying on screen and audibly confirming drug type with printing color compliant labels with 2-D barcodes. End-user feedback and confirmation improves the safety and effectiveness of the device when used as labeled. .
- 5.4 The system integration with AIMS can provide real-time documentation of drug administration.
#### 6 Testing and Equivalence: 807.92(b)(1), 807.92(b)(2) & 807.92(b)(3)
- 6.1 In the code implementation, electrical compliance tests, simulation, Bar code reading, UDI/NDC confirmation, NPSG.03.04.01 Labeling, and clinical operations, results and outcomes have been thoroughly reviewed with proper operation and intended functions verified. Both laboratory (non-clinical environment) and surgical (clinical) tests have shown error free NDC/UDI vial reading and labeling of prepared drugs to ASA/ISO standards.
- 6.2 The non-clinical lab tests were conducted utilizing NDC/UDI codes vials scanned and compared to the output of the MGH original SafeLabels system is referenced in the original K101439 submission Attachment 6, Tab 11 and documented in Attachment 5, Tab 8..
- 6.3 The clinical application of the prototype Safe Label system at MGH is referenced in the original K101439 submission Attachment 6, Tab 11 and documented in Attachment 5, Tab 8.
- 6.4 The production ready device has been designed under ISO 13485 certified controls and has passed the series of electrical safety tests including:
Additional Information Reasonably Deemed Necessary to access safe and effective use 807.92(d)
#### Emissions
FCC Part 15.109 Class B Radiated Emissions FCC Part 15.107 Class B Conducted Emissions
#### Safety
ETL/UL 60601-1 UL or associated lab to test for USA, Canada and various international regional requirements
The subject device has no patient contact however has been designed to meet patient contact (EN60601-1) and anesthesia environment (EN60601-2) electrical safety. The device family does not control, monitor or otherwise affect any devices directly connected to or affecting the patient. Medical personnel review the results and inputs processed by the Codonics SLS and offers ample opportunity for competent human intervention in the case of a malfunction or other failure.
{5}------------------------------------------------
#### Quality System
QSR compliance to 21 CFR Part 820
Laboratory and preliminary tests have documented effective application, throughput, reliability, and expected results consistent with predicate devices currently in commercial distribution and additional verification and validation testing is planned prior to release.
The SLS device employs substantially equivalent technology with respect to reading bar codes as the Medley System with Bar Code Module (Pre-market notification K041241) and the SAFERsleep System (Pre-market notification K050883) with visual and auditory feed back confirmation. The SLS system also employs equivalent technology as DocuSys Digital Drug Management (Pre-market notification K062388) and Philips CompuRecord Perioperative System (Pre-market notification K030939) in terms of barcode read information reference with a drug formulary list/database and claims of similar system function and intended uses.
The technology and applications are substantially equivalent to these products already cleared to market by the FDA with respect to drug identification, alert management and in the ability to record/input the time and type of drug administration during surgery via labeling barcodes read by AIMS (Anesthesia Management Systems). Attachment 8 of the original K101439 submission contains information describing these predicate devices and provides a comparison of the Codonics Safe Label System (SLS) technology to the predicate device(s) and describes how any differences of note are substantially equivalent.
# APPLICABLE PERFORMANCE STANDARDS: Codonics has designed the Safe Labeling System to comply with the following standards:
6.5 Recognized Consensus Standards With respect to prepared drug labeling & devices
- ASTM D4774 06 Standard Specification for User Applied Drug Labels in . Anesthesiology: size, color and pattern, and type used on labels applied to unlabeled syringes filled by the users or their agents to identify the drug content. Not intended to cover labels applied by the drug manufacturer.
- ISO 26825:2008 Anesthetic and respiratory equipment User-applied labels for . syringes containing drugs used during anesthesia - Color, design and performance
- ASTM D4267, Standard Specification for Labels for Small-Volume (100 ml or less) . Parenteral Drug Containers: contrast; label content; legibility; legibility; legibility test; parenteral drug; type size applied to unlabeled volumes filled by the users or their agents to identify the drug content
- ASTM D6398, Standard Practice to Enhance Identification of Drug Names on . Labels: covers the shape, size, color, layout, typeface, and bar-coding practices, use of TALLMAN lettering
- National Drug Code (NDC) number (21 CFR 201.25), Bar Code Recognition .
- Joint Commission on Accreditation of Healthcare Organizations NPSG. . 03.03.01, 03.04.01, and 03.05.01 Re: Enhanced ID of Look Alike/Sound Alike Drugs, Labeling all medications, syringes, medication containers, and other solutions on and off the sterile field in perioperative and other procedural settings, ID for increased safety of anticoagulant preparations
- ISMP -The Institute for Safe Medication Practices and the FDA-U.S. Food and Drug . iomistration Look-Alike Drug Name Sets With Recommended Tall Man Letters
- ASA GUIDE ON LABELING OF PHARMACEUTICALS FOR USE IN . ANESTHESIOLOGY
{6}------------------------------------------------
- IEC 60601-1, 60601-1-2 and 61000: Medical electrical equipment standards . (electrical safety, emissions).
#### Hazard Analysis and Safety Concerns 7
- 7.1 ) Codonics initially determined the Software Level of Concern of this device to be Major based on: the software functions and intended use in combination with a drug or biologic.
- 7.2 ) Codonics considered the risks to be mitigated that could result in potential human injury if wrong drug as labeled is administered:
- a- misreading of barcode
- b- failure to database index proper drug
- c- failure to visibly or audibly validate the drug
- d- failure to print the correct color label and/or drug identification on label
- e- failure to present the correct barcode.
- 7.3) Hazard analysis on this product has been performed throughout the product concept and testing phases of the product development and implementation. This process has emphasized:
- Identification of potential hazards, their causes and their effects .
- Development of methodologies to control the occurrence of hazards and to constrain . their effects:
- Determine any effect on patient safety and system effectiveness .
- The FMEA Electrical, Software and Supplemental System Actions to mitigate potential . risks have been completed with action taken described as part of the FMEA forms submitted
The potential hazards associated with this product are not different than those of other devices in this category. These are primarily related to the failure of computer system components, and may be variously obviated by decisions taken by the end users of the product. None of the failures are expected to materially contribute to patient death or injury.
It is our conclusion that there is no hardware or software component, operating in a properly configured environment, whose latent design defect would be expected to result in death or injury of the patient. Thus the "level of Concern" is reduced to "Moderate" in the production ready 1.0.0 version device.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows a seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol consisting of three curved lines, possibly representing people or services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Garv W. Enos VP, Business and Technology Integration Codonics, Incorporated 17991 Englewood Drive Middleburg Heights, Ohio 44130
FEB - 7 2011
Re: K101439
Trade/Device Name: Codonics Safe Label System (SLS) Regulation Number: 21 CFR 868.5160 Regulation Name: Gas Machine for Anesthesia or Analgesia Regulatory Class: II Product Code: BSZ Dated: February 1, 2011 Received: February 3, 2011
Dear Mr. Enos:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{8}------------------------------------------------
Page 2- Mr. Enos
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
hn for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{9}------------------------------------------------
Indications for Use 510(k) Number (if known): K101439
-
Device Name: Codonics Safe Label System (SLS)
Included Products: Codonics Safe Label System Models: Codonics Safe Labeling System (Integrated) Codonics Safe Labeling Software & Standalone Center Codonics Safe Labeling (SLS XXX) Multiple Models Pending Configuration (exact model number to be determined)
Indications for Use:
The Codonics Safe Label System (SLS) and SLS Software provides a simple computerbased bar code scanning & printing system to automatically verify drug identity from NDC and other drug vial UDI Barcodes, and to print labels for prepared drugs and other items in use on patients during surgical procedures.
Codonics Safe Label System (SLS) is generally placed in, however not limited to, the perioperative environment to identify syringes prepared for anesthesiology use during surgery. Additional uses include producing labels for IVs and other artifacts used during a surgical procedure. SLS can also be used to print "non-surgical environment" color & text labels as required. Typical users of this system are trained professionals, including but not limited to physicians, nurses, and technicians
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
L. Shultz
Division Sign-Off) امارتونا Division of Anesthesiology, General ( Intection Control, Dental Devices
510(k) Number:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.