K072884 · Draeger Medical AG & Co. KG · BSZ · Jul 30, 2008 · Anesthesiology
Device Facts
Record ID
K072884
Device Name
FABIUS MRI
Applicant
Draeger Medical AG & Co. KG
Product Code
BSZ · Anesthesiology
Decision Date
Jul 30, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5160
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Fabius MRI is indicated as a continuous flow anaesthesia system. The Fabius MRI may be used for spontaneous, manually assisted or automatic ventilation, delivery of gases and anaesthetic vapor, and monitoring of oxygen concentration, breathing pressure and respiratory volume. The Fabius MRI is indicated for use in MRI scanner rooms with 1.5 and 3.0 tesla magnets at a location of 40 m tesla (400 gauss) or less. Federal law restricts this device to sale by, or on the order of a physician.
Device Story
Continuous flow anesthesia system for MRI environments; utilizes O2, N2O, and AIR from pipelines or cylinders. Electrically driven, electronically controlled ventilator provides volume-controlled, pressure-controlled, pressure-support, SIMV/PS, manual, and spontaneous ventilation. Monitors airway pressure, volume, and FiO2; requires external CO2 and anesthetic agent monitoring per IEC 60601-2-13. Operated by clinicians in ORs, induction/recovery rooms, and MRI suites. Features MRI-specific modifications including reduced ferromagnetic materials, tesla sensor, and integral gauss alarm for safety within 400 gauss limit. Benefits patient by enabling safe anesthesia delivery during MRI procedures.
Clinical Evidence
Bench testing only. Evidence includes risk management reports, system-level qualification, and verification testing according to applicable standards, including testing in an MRI environment.
Technological Characteristics
Continuous flow anesthesia system; electrically driven/electronically controlled ventilator. Materials modified for MRI compatibility (reduced ferromagnetic content). Connectivity includes medical gas pipeline/cylinder inputs. Monitors: airway pressure, volume, FiO2. Standards: IEC 60601-2-13. Safety features: integral gauss alarm for 400 gauss limit.
Indications for Use
Indicated for patients requiring anesthesia delivery and ventilation support in MRI environments (1.5T or 3.0T magnets, <400 gauss). Used for spontaneous, manual, or automatic ventilation, gas/anesthetic delivery, and monitoring of O2, pressure, and volume.
Regulatory Classification
Identification
A gas machine for anesthesia is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. The device may include a gas flowmeter, vaporizer, ventilator, breathing circuit with bag, and emergency air supply. A gas machine for analgesia is a device used to administer to a patient an analgesic agent, such as a nitrous oxide-oxygen mixture (maximum concentration of 70 percent nitrous oxide).
GE Datex Ohmeda Aestiva /5 MRI Anesthesia System (K050055)
Submission Summary (Full Text)
{1}------------------------------------------------
510(k) Summary Fabius MRI
Image /page/0/Picture/1 description: The image shows the logo for Dräger medical, which is a company that specializes in medical technology. Above the logo is the number 1072884. Below the logo, it states that Dräger is a Dräger and Siemens Company. The logo is in black and white.
# 510(k) Summary
21 CFR part 807.92
JUL 3 0 2008
| Applicants Name and Address: | Dräger Medical AG & Co. KG<br>Moislinger Allee 53-55<br>23542 Lübeck<br>Germany |
|----------------------------------|----------------------------------------------------------------------------------|
| Manufacturer Name and Address: | Dräger Medical AG & Co. KG<br>Moislinger Allee 53-55<br>23542 Lübeck<br>Germany |
| Establishment Regulation Number: | 9611500 |
| Contact Person: | Dr. Karin Luebbers<br>Senior Manager Regulatory Affairs |
| | Phone: + 49 (451) 882-5367<br>Fax: + 49 (451) 882-4351 |
| Applicants US Contact Person: | Joyce Kilroy<br>Vice President, PQR<br>(Processes, Quality & Regulatory) |
| | Phone: +1 (215) 660-2626<br>Fax: +1 (215) 721-5424 |
| Date submission was prepared: | 2007-10-05 |
| Device Name: | Fabius MRI |
| Common Name: | Gas-machine, anesthesia (73 BSZ) |
| Classification Name: | 21 CRF 868.5160 |
| Regulation Number: | II |
| Class: | |
| Predicate Device Identification: | Fabius GS K042419<br>GE Datex Ohmeda Aestiva /5 MRI Anesthesia<br>System K050055 |
、
Section 3 - Page 1 of 3
:
{2}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for Dräger Medical. The logo is in black and white, with the word "Dräger" in a bold, sans-serif font. Below the logo is the text "A Dräger and Siemens Company" in a smaller font. The logo is simple and professional, and it is likely used on the company's website, marketing materials, and products.
### Device Description:
The Fabius MRI is a continuous flow anesthesia system usable in an MRI environment.
#### Intended Use:
Fabius MRI is an inhalation anesthesia machine for use in MRI environments and standard ORs, induction and recovery rooms. It can be used in MR/ environments with 1.5 and 3 tesla magnets at a location of < 40 millitesta (400 gauss).
It may be used with O2, N2O, and AIR supplied by a medical gas pipeline system or by externally mounted gas cylinders.
Fabius MRI is equipped with a compact breathing system, providing fresh gas decoupling, PEEP, and pressure limitation.
The following ventilation options are available:
- Volume Controlled Ventilation �
- ◆ Pressure Controlled Ventilation
- . Pressure Support (Optional)
- SIMV/PS (Optional) ◆
- Manual Ventilation ◆
- . Spontaneous Breathing
Fabius MRI is equipped with an electrically driven and electronically controlled ventilator and monitors for airway pressure (P), volume (V), and inspiratory oxygen concentration (FiO2),
As per IEC 60601-2-13 (Anesthetic Workstations and their Modules - Particular Reguirements), additional monitoring of the concentrations of CO2 and anesthetic agent is required when the machine is in use.
#### Predicate Devices:
| 510(k) Number | Device Name | Manufacturer |
|---------------|--------------------------------|-----------------------------|
| K042419 | Fabius GS | Draeger Medical AG & Co. KG |
| K050055 | GE Datex Ohmeda Aestiva /5 MRI | GE Healthcare |
#### Substantial Equivalence:
The Fabius MRI is substantially equivalent to the Fabius GS (K042419) and the GE Datex Ohmeda Aestiva /5 MRI (K050055).
The Fabius GS was modified into the Fabius MRI. The two devices are identical with exception of modifications for the MRI environment (e.g. testa sensor, alarm indicators, MRI specific labeling reduction of ferromagnetic materials) and a few other minor differences, (e.g. auxiliary outlet standard, 2 vapors instead of 3).
{3}------------------------------------------------
# 510(k) Summary Fabius MRI
Image /page/2/Picture/1 description: The image shows the word "Drägermedical". The word is written in a sans-serif font. The first letter "D" is capitalized, and the rest of the letters are in lowercase. The word is written in black color.
A Dräger and Siemens Company
The Fabius MRI is similar to the Aestival5MRI in that both devices can be used in MRI environments with shielded magnets of 1.5 and 3.0 Tesla. Both devices have limitations to the location of use. The Fabius MRI may be used at a location of <400 gauss, while the Aestival5MRI may be used at locations of 300 gauss. Both devices have integral gauss alarms that notify the user when the device in excess of their operating parameters (> 400 gauss, 300 gauss).
Qualification of the Fabius MRI includes a risk management report, system level qualification and verification testing according to applicable standards and testing in an MRI environment.
Section 3 - Page 3 of 3
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The text is in all caps and is in a sans-serif font.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 30 2008
Draeger Medical AG & CO. KG Ms. Joyce Kilroy Vice President of Processes, Quality, and Regulatory Draeger Medical System, Incorporated 3135 Quarry Road Telford, Pennsylvania 18969
Re: K072884
Trade/Device Name: Fabius MRI Regulation Number: 21 CFR 868.5160 Regulation Name: Gas Machine for Anesthesia or Analgesia Regulatory Class: II Product Code: BSZ Dated: July 22, 2008 Received: July 25, 2008
Dear Ms. Kilroy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Ms. Kilroy
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (QSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
TK Armele-Lund, m/frell
Chiu Lin. Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): K
Device Name: Fabius MRI
# Indications For Use:
The Fabius MRI is indicated as a continuous flow anaesthesia system. The Fabius MRI may be used for spontaneous, manually assisted or automatic ventilation, delivery of gases and anaesthetic vapor, and monitoring of oxygen concentration, breathing pressure and respiratory volume.
The Fabius MRI is indicated for use in MRI scanner rooms with 1.5 and 3.0 tesla magnets at a location of 40 m tesla (400 gauss) or less.
Federal law restricts this device to sale by, or on the order of a physician.
AND/OR Prescription Use X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use_ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
shl
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: Ko 72884
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.